Current UK legislation and guidelines for radiation protection of patients and staff - PowerPoint PPT Presentation

1 / 31
About This Presentation
Title:

Current UK legislation and guidelines for radiation protection of patients and staff

Description:

Current UK legislation and guidelines for radiation protection of patients and staff Prepared by: Dr D. Mladenova General principles of the International Commission ... – PowerPoint PPT presentation

Number of Views:41
Avg rating:3.0/5.0
Slides: 32
Provided by: silverswo
Category:

less

Transcript and Presenter's Notes

Title: Current UK legislation and guidelines for radiation protection of patients and staff


1
Current UK legislation and guidelines for
radiation protection of patients and staff
  • Prepared by Dr D. Mladenova

2
General principles of the International
Commission on Radiological Protection (ICRP)
1.No practice shall be adopted unless its
introduction produces positive net benefit
(Justification)2.All exposures shall be kept
as low as reasonably practicable (ALARP)

3
Principles of ICRP- contd
  • Taking economic and social factors
  • into account( Optimization)
  • The dose equivalent to individuals shall
  • not exceed the limits recommended by the
  • ICRP (Limitation)

4
Regulations for intraoral radiography
  • Tube voltage should not be lower than 50
  • kV preferably 70-90 kV
  • Beam diameter should nor exceed 60 mm
  • Rectangular collimation should be used

5
Intraoral radiography-contd
  • Total beam filtration ( inherent and added )
  • -1.5 mm aluminium disc for sets operating
    bellow 70 kV
  • - 2.5 mm aluminium for sets operating above
    70 kV

6
Intraoral radiography-contd
  • The focal spot should be marked on
  • the outer casting of the tubehead
  • Focal spot to skin distance ( FSD) should be at
    least 100 mm for sets operating below 60 kV and
    200 mm for sets operating above 60 kV.

7
Intraoral radiography-contd
  • Film speed controls and finely
  • adjustable exposure time settings should be
    provided
  • The fastest film available( E or F speed) that
    will produce satisfactory diagnostic images
    should be used

8
Panoramic radiography
  • Equipment should have a range of tube potential
    settings, preferably 60-90 kV.
  • The beam height at the receiving slit of cassette
    holder should not be greater than

9
Panoramic radiography-contd
  • greater than the film in use (normally 125 mm or
    150 mm).
  • The width of the beam should not be greater than
    5 mm

10
Panoramic radiography-contd
  • Equipment should be provided with adequate
    patient-positioning aids incorporating light beam
    markers
  • New equipment should provide facilities for
    field-limitation techniques

11
Cephalometric radiography
  • Equipment must be able to ensure the precise
    alignment of X-ray beam, cassette and patient
  • The beam should be collimated to include only the
    diagnostically relevant area

12
Cephalometric radiography-contd
  • To facilitate the imaging of the soft
  • tissues, an aluminium wedge filter
  • should be provided at the X-ray tube

13
All equipment
  • Should have a light on the control panel to show
    that the mains supply is switched on
  • Should be fitted with a light and audible
    warnings that gives a clear and visible
    indication to the operator that an exposure is
    taking place

14
All equipment-contd
  • Exposure switches (timers) should only function
    while continuous pressure is maintained on the
    switch and terminate if pressure is released
  • Exposure switches should be positioned so that
    the operator can

15
All equipment-contd
  • remain outside the controlled area
  • and at least 2 m from the X-ray tube and patient

16
Justification
  • -the availability and/or findings of previous
    radiographs
  • - the specific objectives of the exposure in
    relation to the history and examination of the
    patient

17
Justification-contd
  • the total potential diagnostic benefit to the
    patient
  • - the radiation risk associated with the
    radiographic examination

18
Justification-contd
  • the efficacy, benefits and risks of
  • alternative techniques having the
  • same objective but no or less
  • exposure to ionizing radiation

19
Lead protection
  • There is no justification for the routine use of
    lead aprons for the routine use of lead aprons
    for patients in dental radiography
  • Thyroid collar to be used only in maxillary
    occlusal radiography

20
Lead protection-contd
  • Lead aprons do not protect against scattered
    radiation for adult who support a patient during
    exposure
  • Lead aprons should not be folded

21
Specific requirements for pregnant women
  • Only radiographs that are absolutely necessary
    are taken
  • ALARP and the patient is given the option to
    delay the radiography

22
Selection criteria in Dental Radiography 1998
  • No radiographs should be taken without a history
    and clinical examination
  • New patient
  • Child (primary and mixed dentition)- panoramic
    and bitewings radiographs

23
Criteria for radiographs-contd
  • Adult (dentate patient)-patient-specific
    radiographs depending on clinical examination
  • Edentulous patients- panoramic and/or periapical
    radiographs in selected areas

24
Dose limitations and annual dose
  • Patients
  • Radiation workers ( classified and non-
    classified)
  • General public

25
Radiographic investigations for patients
  • Examinations directly associated with illness
  • Systematic examinations
  • Examinations for occupational, medico-legal or
    insurance purposes
  • Examinations for medical research

26
Annual dose limits-contd
  • Non-classified workers 6 mSv
  • General public 1 mSv
  • Dose Constraints
  • non-classified workers 1 mSv

27
Annual dose limits-contd
  • For employee not directly
  • Involved with radiography and
  • General public 0.3 mSv
  • Pregnant staff member 1 mSv

28
Source of radiation to dentist and their staff
  • The primary beam, if they stands in its path
  • Scattered radiation from the patients
  • Radiation leakage form the tubehead

29
Protective measures
  • Personal monitoring is recommended
  • for practice exceeding 100 intraoral or
  • 50 panoramic film
  • Staff should stand outside the controlled area
    and not in the line of the primary beam

30
Protective measures-contd
  • Safe use of equipment
  • Safe use of radiographic techniques

31
Main methods of monitoring and measuring
radiation dose
  • Film badges
  • Thermoluminescent dosemeters
  • Ionizing chambers
Write a Comment
User Comments (0)
About PowerShow.com