Comparison of Cone Beam Computed Tomography With MV Imaging for Paraspinal Radiosurgery Patient Alig - PowerPoint PPT Presentation

1 / 26
About This Presentation
Title:

Comparison of Cone Beam Computed Tomography With MV Imaging for Paraspinal Radiosurgery Patient Alig

Description:

A pair of MV localization images has traditionally been used for setup. ... Systematic errors between the radiation and CBCT isocenters were measured ... – PowerPoint PPT presentation

Number of Views:335
Avg rating:3.0/5.0
Slides: 27
Provided by: ser8187
Category:

less

Transcript and Presenter's Notes

Title: Comparison of Cone Beam Computed Tomography With MV Imaging for Paraspinal Radiosurgery Patient Alig


1
Comparison of Cone Beam Computed Tomography With
MV Imaging for Paraspinal Radiosurgery Patient
Alignment and Position VerificationS.
Kriminski, D.M. Lovelock, H. Amols, I. Ali, Y.
YamadaMemorial Sloan-Kettering Cancer Center,
New York, NYASTRO 2006
2
  • Introduction
  • Patients with oligo-metastatic paraspinal disease
    are treated at MSKCC to a high dose (18 24 Gy)
    image guided single fraction treatment
  • Patients are setup to within 2 mm in an
    immobilization cradle
  • A pair of MV localization images has
    traditionally been used for setup.

3
  • MV images for patient alignment
  • Quality of the MV images for clinically
    acceptable doses of few MU can be poor
  • Only in 1/3 of all cases registration is
    possible based on patient bony anatomy
  • It is difficult to say in advance whether bony
    registration would be feasible for a given
    patient
  • All patients were required to undergo
    implantation of 6 gold fiducial markers into the
    spine.

4
  • kV Cone Beam Computed Tomography
  • MSKCC recently acquired linac with gantry mounted
    kV source and 2D flat panel detector
  • Permits the acquisition of a volumetric image
    (cone beam computed tomography CBCT) at the
    time of treatment
  • Improved bone contrast is expected to permit
    precise registration of the cone beam image with
    the planning CT scan without the use of implanted
    fiducial markers

5
  • Motivation
  • Disadvantages of MV localization
  • Due to poor quality of images patient alignment
    using MV localization images requires
    implantation of gold markers into the spine
  • Automatic registration of MV images is not
    feasible
  • Advantages of kV CBCT
  • Improved bone contrast (due to use of kV X-ray)
    together with 3D imaging makes registration more
    robust
  • Automatic 3D registration is feasible
  • 3D image gives clear indication of possible
    rotational setup errors

6
  • Objective
  • To investigate whether the setup corrections from
    kV CBCT are consistent with that from MV
    localization images
  • If corrections are consistent
  • The need for implantation of markers would be
    eliminated

7
  • Materials and Methods
  • Patients
  • 16 patients with paraspinal disease eligible for
    single fraction treatment to spinal metastasis
    were accrued to an IRB approved protocol.
  • Each patient underwent implantation of 6 gold
    fiducial markers, generally into the left and
    right lamina of three different vertebrae
  • Three MV image pairs (AP LL or LAO) were
    acquired for each of 16 patients
  • Initial, to determine couch position correction
  • Verification, after corrections, prior to
    treatment
  • Final, after treatment
  • Each MV pair was accompanied by kV CBCT scan

8
  • Positioning with MV images
  • Digitally composited radiographs (DCRs) created
    using AcqSim software (Philips Medical Systems)
    were used as reference
  • Similar to DRRs but voxels with HU corresponding
    to implanted metal have their opacity increased
  • Results in images in which the fiducial markers
    are sharply demarcated
  • Every MV image pair was manually registered to
    DCRs from plan CT to determine setup corrections
  • All patient setup corrections were based on the
    corrections from the MV technique.

9
  • Examples of MV images and DCR
  • Outlines (yellow) are drawn manually on an
    overlay layer DCR using vendors software (Varian
    Medical Systems)
  • The MV image is processed to enhance markers
  • The overlay layer with the yellow marker outlines
    is manually registered with the markers on MV
    image

10
  • Systematic Uncertainties in localization by MV
    images
  • Anatomic features seen in the MV images are
    referenced to a grid projected onto all images
  • The grid is projected by a physical device
    inserted into the head of the Linac
  • Monthly quality assurance procedures keep the
    grid positioning error to less than 1 mm

11
  • Positioning with kV CBCT
  • Three 3D CBCT scans were acquired for each
    patient following MV images using gantry-mounted
    kV on-board imaging system (Varian Medical
    Systems, Palo Alto, CA)
  • 3D image registration of CBCT to plan CT was
    performed automatically by maximizing mutual
    information using a region of interest excluding
    markers
  • Corrections from MV images were used for
    comparison with that from CBCT

12
  • 3D Registration of CBCT to plan CT
  • Cone beam scan (cyan) is overlaid with the
    planning CT scan (red)
  • In the superior/inferior direction, the ROI
    extends 2 or 3 vertebrae

Selection of ROI excluding implanted markers
prior to automatic registration
Result of maximizing mutual information
13
  • Systematic Uncertainties in localization by kV
    CBCT
  • Systematic errors between the radiation and CBCT
    isocenters were measured frequently over a 18
    month period and found to be very stable
  • Position of radiation isocenter (for HFS
    coordinates)
  • These offsets were corrected for in all
    registrations
  • Residual systematic errors are estimated to be
    0.3mm

14
  • Results
  • Random component of MV localization
  • Histograms of the difference between SI setup
    corrections determined from AP and LAO or LL MV
    localization images
  • Mean and std of absolute difference are given
  • If and AP LAO are used for alignment, LR and AP
    random errors have additional factors of 1.7 and
    1.4

0.50.4 mm
15
  • Is this precision acceptable?
  • Goal of image guidance is to reduce setup errors
    to lt 2 mm
  • With MV systematic error lt 1 mm, this precision
    is sufficient for SRS
  • This leaves out the issue that MV registration
    requires implantation of seeds
  • Also, leaves out the question whether seeds
    represent tumor position correctly

16
  • An example of CBCT to plan CT registration
  • Red plan CT
  • Cyan - CBCT
  • Contour - PTV
  • Rectangle ROI for registration
  • It is seen that with good bony registration seeds
    are slightly misaligned
  • Registrations based on bony structures and based
    on BB are slightly different

17
  • Comparison of MV and CBCT setup corrections
  • Histograms of the differences between setup
    corrections determined from the CBCT scans and
    corresponding MV localization image
  • Mean and std of absolute difference are given

1.00.7 mm 1.00.6 mm
0.80.6 mm
18
  • Are these differences acceptable?
  • Goal of image guidance is to reduce setup errors
    to lt 2 mm
  • kV CBCT provides sufficient accuracy and
    precision for high dose single fraction treatment

19
  • Source of MV and CBCT differences
  • Differences between CBCT and MV corrections,
    which are not due to CBCT errors, are caused by
  • Uncertainty in MV / DCR registration
  • Systematic errors associated with the field
    localization in MV images
  • Patient motion within the immobilization cradle
  • Couch re-positioning errors between the MV images
    and CBCT scans
  • The random component in these sources can be
    estimated by comparing the MV corrections before
    and after treatment

20
  • Pre-post change in patient position from MV
  • Histograms of the differences between setup
    corrections determined from the MV localization
    image pairs before and after treatment
  • Mean and std of absolute of the difference are
    given

1.11.0 mm 1.20.8 mm
1.11.0 mm
21
  • Are the differences between CBCT and MV
    corrections reasonable?
  • Differences between setup corrections determined
    from MV images before and after treatment
    1.11.0, 1.20.8, and 1.11.0 mm (meanstd) for
    LR, AP and SI
  • Differences between the CBCT and MV setup
    corrections are 1.00.7, 1.10.6, and 0.80.6 mm
  • These differences are similar
  • We speculate that the contribution to these
    differences by errors in setup correction from
    CBCT is comparable or smaller than by that from MV

22
  • Shifts during treatment
  • Histograms of the differences between the setup
    corrections determined from kV CBCT before and
    after treatment
  • Mean and std of absolute of the difference are
    given

0.50.5 mm 0.70.6 mm
1.00.9 mm
23
  • Are shifts during treatment significant?
  • Differences between setup corrections determined
    from kV CBCT before and after treatment are
    0.50.5, 0.70.6, and 1.00.9 mm (meanstd) for
    LR, AP and SI
  • Differences between setup corrections determined
    from MV image pairs before and after treatment
    are 1.11.0, 1.20.8, and 1.11.0 mm (meanstd)
  • Both MV imaging and kV CBCT indicate that patient
    displacements during treatment are small

24
  • Shifts from CBCT vs. shift from MV
  • Differences between the setup corrections
    determined from kV CBCT before and after
    treatment vs differences between the setup
    corrections determined from MV image pair before
    and after treatment
  • Lines show ideal correlations between the
    differences

25
  • Some shifts are not correlated why?
  • Patient displacements marked by red circles are
    inconsistent between CBCT and MV
  • For one outlier a change in posture (bending of
    the spinal cord) was observed, as clearly seen on
    CBCT
  • Because MV and CBCT use different ROIs/BBs for
    patient registrations, change in the shape of
    spinal cord leads to differences in patient
    alignments
  • Other discrepancies shown may be attributed to
    patient displacements between CBCT scans and MV
    imaging

26
  • Conclusions
  • Setup corrections determined using CBCT without
    the use of implanted markers were consistent with
    that from marker registration in MV localization
    images and DCRs
  • This and other advantages of CBCT such as robust
    automatic 3D registration and clear indication of
    possible rotational setup errors have led our
    clinic to adopt CBCT for the setup of all single
    fraction paraspinal patients
  • Both MV imaging and kV CBCT indicate that patient
    displacements during treatment are small
Write a Comment
User Comments (0)
About PowerShow.com