Title: NonInvasive Hemodynamics with a Small Overview of Invasive Hemodynamic Monitoring
1Non-Invasive Hemodynamicswith a Small Overview
of Invasive Hemodynamic Monitoring
- SVCCs Cardiopulmonary and Neurodiagnostic
Sophomore Class
2Review of the Heart and Blood Circulation
- The cardiovascular system performs two major
tasks - Delivers oxygen and nutrients to the body organs
- Removes waste products of metabolism from tissue
cells
3Major components of the cardiovascular system
- The heart- a hollow muscular pump with two sides
pulmonary and systemic - Circulatory system-consists of large and small
elastic vessels that transport blood throughout
the body - The amount of blood pumped through the heart of a
normal healthy adult could reach 2100 gallons per
day!
4The Human Heart
- Central organ of the cardiovascular system,
located between the two lungs in the middle of
the chest - The adult heart is about the size of two clinched
fists. It is shaped like a cone and weighs about
7 to 15 ounces, depending on the individual - The heart pumps one-million barrels of blood in
an average lifetime, with 3 million beats per
year, for a healthy adult
5Chambers of the Heart
- The human heart is divided into four chambers
the right atrium, right ventricle, left atrium
and left ventricle - There are special muscles that create the walls
of the chambers, hence the myocardium, that
contracts rhythmically under the stimulation of
electrical currents
6Chambers of the Heart Cont.
- Both the left and right sides of the heart are
separated medially by a muscle called the septum.
(The septum may be termed as the atrial or
ventricular septum, depending on its location.) - Contained within the chambers are four one-way
valves, closing off at appropriate times to keep
a forward flow of blood in the cardiac cycle
7Valves of the Heart
- The heart contains four vital valves that allow
the blood to flow through, then collapse
preventing backflow - The two valves that separate the ventricles to
the circulatory system are the pulmonic and
aortic valves, also known as the semilunar valves
due to there crescent shape
8Valves of the Heart Cont.
- The two valves consist of three cusps. The
pulmonic valve opens freely when the right
ventricle contracts and then falls back as it
relaxes. Where the aortic valve opens as the
left ventricle pushes the blood into circulation,
then shuts from the pressure of the aorta
9Valves of the Heart Cont.
- The two valves that separate the ventricle from
the atria are the mitral valve and the tricuspid
valve. The two valves consist of cusps, but in
addition they have a strong fibrous cords called,
chordae tendinae, which help keep the valves
attached to the ventricular walls - The mitral valve separates the left ventricle and
left atrium, where the tricuspid valves separate
the right ventricle and atrium. Both valves
prevent backflow into the ventricles, in normal
functioning hearts
10Circulation of Blood
- The poor oxygenated blood returns from the body,
which is rich with carbon dioxide - The blood first travels through the Superior and
Inferior Vena Cava, along with the coronary
sinuses into the right atrium - The blood flows from the right atrium through the
tricuspid valve into the right ventricle
11Circulation of Blood Cont.
- When the right ventricle contracts the tricuspid
valve collapse, the right ventricle will then
expel the blood through the pulmonic valve into
the pulmonary trunk, which divides into a right
and left pulmonary artery. Each of which carries
blood to one lung. (The pulmonary circuit)
12Circulation of Blood Cont.
- The blood then flows through the pulmonary
arteries into the lungs (where oxygen and carbon
dioxide are exchanged in the pulmonary
capillaries) and then to the pulmonary veins.
The carbon dioxide is now exhaled, while the left
atrium receives oxygenated blood from the lungs,
via four pulmonary veins, two from each lung
13Circulation of Blood Cont.
- The blood will flow from the left atrium through
the mitral valve, also known as the bicuspid
valve, into the left ventricle, causing the
bicuspid valve to close off - As the blood leaves the left ventricle it travels
through aortic valve into the aorta, which is a
network of small branches, circulating the blood
to the systemic circuit (the rest of the body)
14Circulation of Blood Cont.
- After the oxygenated blood travels through the
whole body, the blood returning from the lower
body empties into the inferior vena cava and the
blood from the tissues of the head and neck
region will empty into the superior vena cava.
Back into the atrium
15What is Non-Invasive Hemodynamic Monitoring How
Does It Work?
- A device that provides hemodynamic parameters
based on the measurement of thoracic electrical
bioimpedance. This device works by transmitting
an alternating current through the chest that
seeks the path of least resistance and ends up in
the blood filled aorta. It also measures the
baseline impedance to this current and then with
each heartbeat, as the blood volume and velocity
in the aorta change, it measures the
corresponding change in impedance. This
information is then processed with statistics on
blood pressure and heart rate to determine fluid
levels and cardiac function.
16Indications for Non-Invasive Hemodynamic
Monitoring
- Suspected or known cardiovascular disease
- Patients with the need for fluid management
- Differentiation of cardiogenic from pulmonary
causes of acute dyspnea - Optimization of atrioventricular interval for a
patient with A/V sequential cardiac pacemakers - Patients with need of determination for
intravenous inotropic therapy - Post heart transplant myocardial biopsy patients
17It is used on patients with
- Heart Failure
- Hypertension/Hypotension
- Pacemaker Syndromes
- Coronary Artery Disease
- Pericardial Disease with Effusion
- Critical Multi-System Illness
- Acute/Chronic Renal Failure
18Warnings and Precautions of Hemodynamic
Monitoring, General Warnings
- The BioZ is not specifically intended to monitor
variations in cardiac performance that could
result in immediate danger to the patient - Explosive Hazard Do not use in the presence of
flammable anesthetics or gases. - Sensors are to be placed externally on the skin
only and are not for direct cardiac application - The conductive gel of the sensors should not
contact any other conductive materials during
patient monitoring - This device is to be connected to a grounded
receptacle
19Warnings and Precautions of Hemodynamic
Monitoring, General Warnings Cont.
- The external Ferrite Core must remain attached to
the patient cable at all times - The patient cables specified and included with
the BioZ are designed specifically for protection
against the effects of cardiac defibrillators and
radio-surgery equipment - Do not use any other type of patient cable with
this device - Disposal of this product and/or any of its
accessories shall be in accordance with any and
all local regulations - The BioZ should only be used with a
Hospital/Medical grade power supply
20Conditions that May Limit the Accuracy of the Data
- Septic Shock
- Aortic valve regurgitation
- Severe hypertension (MAP gt 130 mmHg)
- Patient heights measuring below 48 (120 cm) or
above 90 (230 cm)
21Conditions that May Limit the Accuracy of the
Data Cont.
- Patient weights measuring less that 67 lbs. (30
kg) or greater than 341 lbs. (155 kg) - Patient Movement
- Aortic Balloon Pump
- The BioZ should not be used concurrently on
patients with Minute Ventilation pacemakers when
the MV sensor is activated
22Procedure
- Skin preparation
- 1. Shave the hair over the sensor site if
necessary. - 2. Dry prep the sensor sites by mildly
abrading the skin with the perforation on the
sensor backing
23Sensor Application
- Check expiration date on sensor package before
opening, after opening the sealed pouch and
remove the four dual sensor patches - Remove the four dual sensor patches from the
backing material and apply each sensor, adhesive
side down to the proper sites as in the diagram
on the next page
24Diagram
25Sensor Application Cont.
- The rectangular shaped end should be positioned
closest to the heart - Depress the tab at the end of the connector,
place the connector over the sensor stud and
release the tab - The blue left and right lead wire connectors to
the respective circular shaped transmitting
sensors on the neck - The violet left and right lead wire connectors to
the respective rectangular shaped detecting
sensors on the neck
26Sensor Application Cont.
- The green left and right lead wire connectors to
the respective rectangular shaped detecting
sensors on the thorax - The orange left and right lead wire connectors to
the respective circular shaped transmitting
sensors on the thorax - The sensors must be positioned so that they are
180 degrees opposite each other
27Sensor Application Cont.
28How Impedance Cardiography Works
- An alternating current is transmitted through the
chest - The current seeks the path of least resistance
the blood filled aorta - Baseline impedance to current is measured
- Blood volume and velocity in aorta change with
each heartbeat - Corresponding changes in impedance are used with
ECG to provide hemodynamic parameters
29Non-Invasive Hemodynamic monitors the following
- Cardiac Output
- Stroke Volume
- Systemic Vascular Resistance
- Acceleration Index
- Thoracic Fluid Content
- Velocity Index
- Systolic Time Ratio
- Left Ventricular Ejection Time
- Pre-Ejection Period
- Left Cardiac Work/Index
- Heart Rate
30Brief Overview of Invasive Hemodynamic Monitoring
31Basic Indications of Invasive Hemodynamic
Monitoring
- Pulmonary arterial pressure monitoring
- For diagnosis, management, and treatment of
cardiopulmonary insufficiency - For management and treatment of cardiac shock
- For assessment of pulmonary vascular function
32Basic Indications of Invasive Hemodynamic
Monitoring
- For assessment of cardiac function
- For cardiac pacing
- In application of mechanical ventilation to
assess patient status
33Indications of Invasive Hemodynamic Monitoring
Related to Arterial Pressure Monitoring
-
- During manifestations of hemodynamic instability
- For measurement of the hemodynamic response to
therapeutic intervention - Example Administration of vasoactive
pharmacologic agents - For repeated arterial blood gas sampling
34Indications of Invasive Hemodynamic Monitoring
Related to Central Venous Pressure Monitoring
- For assessment of intravascular volume status and
venous return - For administration of fluids and/or drugs
- For assessment of cardiac function
- Mixed venous saturation and sampling
35Application of Invasive Hemodynamic Monitoring
Related Disease Processes
- Hypotension/Hypertension
- Acute MI
- Mitral regurgitation
- Ventricular septal perforation
- Cardiac tamponade
- Congestive Heart Failure
36Application of Invasive Hemodynamic Monitoring
Related Disease Processes
- Pulmonary Edema
- Sepsis
- Respiratory failure
- Renal failure
- Dialysis with complications
- Drug overdose
37Precautions In Application of Invasive
Hemodynamic Monitoring
- Read patients chart carefully before applying
Invasive Hemodynamic Monitoring - Check documented vitals for any abnormalities
such as hypotension, hypertension, tachypnea,
bradypnea, etc. - Check for recent arterial blood gas results
- Assess the patient
- Assess mechanical ventilation settings
38Contraindications of Invasive Hemodynamic
Monitoring
- Presence of Left Bundle Branch Block on EKG
- Placement of right heart catheter
- Severe hypothermia
- Inadequate monitoring equipment
- Patient refusal
- Poor collateral circulation
- Coagulopathies, systemic anticoagulation, and
interventional thrombolysis
39Complications of Invasive Hemodynamic Monitoring
- Bleeding or bruising at the catheter insertion
site - Hemorrhage
- Infection
- Pneumothorax or hemothorax darning venipuncture
for insertion of central venous catheters
40Complications of Invasive Hemodynamic Monitoring
- Dysrhythmias attributed to central venous or
pulmonary artery catheter migration or irritation
of the myocardium - Pulmonary artery rupture with inflation of balloon
41Complications of Invasive Hemodynamic Monitoring
- Pulmonary infarction or ischemia from prolonged
wedging of pulmonary artery catheter balloon or
embolization of a thrombus from the tip of a
catheter - Obtaining inadequate or poor quality data or
inappropriately interpreting data
42Please see your handout for an overview of
Swan-Ganz catherization
43References
- Online site www.cardiodynamics.com BioZ tect,
ICG Sensor and Cable System - BioZ.com Operators Manual
- Online site www.cardiodynamics.com BioZ CG
Monitor Specifications - Online site www.medobserver.com/may2002/hemodyna
mics.html Noninvasive Hemodynamics - Online site www.inpedancecardiography.com/icgove
r10.html Overview of Impedance Cardiography
44References
- Matthew Acampora, M.D. Internal Medicine of
Charlotte, P.A. Protocol for Use of BioZICG
(Impedence Cariography). - Non-Invasive Hemodynamics Monitoring
www.fciheart.com/NonInvasive20CHF20Monitor.html
accessed on 9/1/04. - Gary A. Thibodequ and Kevin T. Patton. Anatomy
Physiology, 4th Edition. - St. Louis Mosby, 1999.
- Braunwald, Eugene and others. Harrisons
Principles of Internal Medicine, 11th Edition.
45References
- New York McGraw-Hill Book Company, 1987
- Baum, Gerald L., J.D. Crapo, B.R. Celli, and J.B.
Karlinsky. - Pulmonary Diseases 6th Edition.
- Philadelphia Lippincott Raven, 1998
46References
- Pilbeam, Susan. Mechanical Ventilation
Physiological and Clinical Applications, 3rd ed. - White, Gary C. Basic Clinical Lab Competencies
for Respiratory Care An Integrated Approach, 4th
ed. - Wilkens, Robert L. Egans Fundamentals of
Respiratory Care, 8th ed. - Miller-Keane Encyclopedia Dictionary of Medicine,
Nursing, Allied Health, 6th ed. - http//www.medscape.com/viewarticle/463474
- http//www.cyber-nurse.com/veetac.cham2.htm
- http//hemodynamicsociety.org/hemodyn.html