Title: Laboratory Evaluation of a Next Generation Transversal Ultrasound System
1Laboratory Evaluation of a Next Generation
Transversal Ultrasound System
- Mark E. Schafer, Ph.D.
- President and Principal Scientist, Sonic Tech,
Inc. - Chief Technology Officer and VP, RD,
- Sound Surgical Technologies, LLC
I am a consultant to several phaco companies, and
have received research funding for this work
Contact email marks_at_sonictech.com
2Purpose Outline
- Purpose
- To provide baseline comparative data for
different phaco devices, specifically for a newly
introduced transversal system AMO Whitestar
Signature with Ellips FX - Testing focused on Clinically Relevant settings
- Outline
- Measurement results using several previously
studied approaches - Needle Shaft Temperature Rise (non-frictional)
- Cutting Effectiveness into artificial lens
material - Acoustic Output
- Compare results with another non-longitudinal
phaco device Infiniti Torsional Ozil - Discussion and Conclusion
3What are the technologies/modes?
- Standard Longitudinal phaco is an in/out motion
- Torsional or T-phaco uses a twisting motion
of the tip rather than a longitudinal motion
(only one of the two modes can operate at a time,
therefore the need to switch back and forth) - Transverse or Ellips uses a combination of
side to side and in/out simultaneously may be
used with either straight or bent tips
4Thermal Measurements Materials Methods
- High resolution thermal imaging (FLIR) camera
records thermal image of hub, needle, and tip - An open fluid chamber allows the entire length of
the needle to be examined - Cup designed so that fluid would only touch tip,
to allow aspiration, but not interfere with
thermal imaging - Inner cutaway cup maintained fluid at desired
level outer cup held overflow - Hub, shaft, and tip tracked and separately
recorded
- Full computer control and acquisition of thermal
images for analysis - Irrigation and aspiration flows balanced to mimic
closed system in eye Fluid rate precisely
measured at 30ml/min, at temperature of 220.5C - Systems run for 5 second foot pedal time results
scaled to 0.5 or 1 sec
5Thermal Testing Results
- Thermal imaging showed distributed heat sources
for transversal mode included both the hub region
and the shaft - In longitudinal mode, the heat source was
predominantly in the hub transition region - Hyperpulse (WhiteStar) modes reduced the thermal
rise - Analysis was scaled to the one second on-time
point, as shown in the graph by the vertical line - Multiple experiments were conducted under the
same conditions and the results averaged
6Thermal Testing Results
- The one-second temperature rise for Elllips at
typical clinical settings of 25-50 amplitude, in
a WhiteStar 6-12 mode, ranged from 2.1 to 3.5C - In Longitudinal mode, for the same WS mode, the
temperature rise was lower, on the order of 0.5
to 1.0C - The difference is due to the location of the heat
source - Note that this does not account for friction
between the shaft and the sleeve, which would be
significantly higher in longitudinal mode than in
transversal or torsional
7Measuring Cutting Materials Methods
- Custom cutting force system
- Full computer control of motor and high
resolution acquisition of position and force data
for analysis - Simulated lens target material
- Constant force (60g weight), measure displacement
as a function of time
- Fluid rates maintained at 30cc/min
- 10 second experiment initial 2 seconds to
establish position baseline systems then run for
8 second foot pedal time - Systems operated over a range of clinically
relevant/recommended settings
8Cutting Effectiveness Results
- Multiple test runs conducted for each measured
condition - Penetration measured and analyzed to find a
cutting rate in terms of millimeters per second
- Ellips cutting rates varied from 0.35mm/sec at
25 setting to 1.7mm/sec at the 50 setting
9Acoustic Output Materials Methods
- Acoustic measurement system captures both the low
frequency (handpiece drive) energy as well as the
cavitational energy - Rotational fixture allows mapping the
distribution of energy, which relates to the
motional direction of the tip - Data can correlate to cutting efficiency
depending upon cavitation readings
10Acoustic Output Results
- Drive energy levels match theory for
longitudinal for transversal, less acoustic
energy is generated because of the acoustic
dipole pattern, as confirmed by angular data - Difference between theory and measurement
represents low frequency ultrasonic energy
converted into Cavitation - In Transveral mode, Cavitation is generated in
region around the tip including the sides, rather
than just at the front, as is the case for
Longitudinal - The front panel setting which denotes the start
of Cavitation action matches the setting which
demonstrates increased cutting effectiveness
11Comparison to Torsional
Test Metric Ellips FX (50 amp, WS 6-12) Torsional (100)
Thermal Rise at 1 second 3.5C 9.9C
Cutting Rate into lens material 1.7mm/sec 0.6mm/sec
Acoustic Output Comparable, and significantly less than longitudinal mode Comparable, and significantly less than longitudinal mode
- Transversal mode produces less internal heating,
with improved cutting performance, in comparison
with Torsional, at clinically relevant/recommended
amplitude settings - This is due in part because Transversal motion
preserves some longitudinal component - Transversal mode can also be used with straight
tips, which permits more choices for the surgeon
bent tips would be expected to have slightly
better cutting performance, and higher acoustic
output, without significantly higher thermal rise
12Discussion / Conclusion
- This study provided an initial investigation into
this new phaco modality, which was only
introduced into the market this year - Additional work will be required to examine all
the interdependent configuration possibilities,
such as drive setting/mode and tip size and
configuration - Preliminary data suggest that Transversal mode
demonstrates superior characteristics in terms of
the balance of safety and efficacy, relative to
other non-longitudinal modalities - Ultimate goal is to provide users with a
consistent set of metrics by which to compare
different systems, permitting a better
understanding of energy input and clinical
outcome