INTRAOCULAR LENS POWER CALCULATION BY IMMERSION A-SCAN BIOMETRY VERSUS CONTACT A-SCAN BIOMETRY MEASUREMENTS BEFORE CATARACT SURGERY - PowerPoint PPT Presentation

About This Presentation
Title:

INTRAOCULAR LENS POWER CALCULATION BY IMMERSION A-SCAN BIOMETRY VERSUS CONTACT A-SCAN BIOMETRY MEASUREMENTS BEFORE CATARACT SURGERY

Description:

intraocular lens power calculation by immersion a-scan biometry versus contact a-scan biometry measurements before cataract surgery burak bilgin**, m.d., kadir ... – PowerPoint PPT presentation

Number of Views:89
Avg rating:3.0/5.0
Slides: 8
Provided by: ascrs2009A
Category:

less

Transcript and Presenter's Notes

Title: INTRAOCULAR LENS POWER CALCULATION BY IMMERSION A-SCAN BIOMETRY VERSUS CONTACT A-SCAN BIOMETRY MEASUREMENTS BEFORE CATARACT SURGERY


1
INTRAOCULAR LENS POWER CALCULATION BY IMMERSION
A-SCAN BIOMETRY VERSUS CONTACT A-SCAN BIOMETRY
MEASUREMENTS BEFORE CATARACT SURGERY
  • Burak Bilgin, M.D., Kadir Eltutar, M.D., N.
    Demet Ozcelik, M.D.

Istanbul Education and Research Hospital
Opthalmology Department, Turkiye. Almangoz
Eye Center, Nisantasi, Istanbul, Turkiye.
The authors state that they have no proprietary
interest and they have not received any grants or
funds in support of the study.
2
PURPOSE
  • The study was designed to evaluate the difference
    in contact A - scan biometry versus immersion A -
    scan biometry measurements for intraocular lens
    (IOL) power calculations before cataract surgery.
    .

3
METHODS
  • Foldable acrylic hydrophobic IOLs ( AcrySof
    SA60AT, SN60WF, SA30AT, SN60AT ) were implanted.
  • Axial length (AxL) measured by contact A - scan
    and immersion biometry were compared.
  • The estimated refraction ( -0,50 Diopters ) was
    compared with postoperative refraction achieved
    on the days 1, 7,15 and 30.
  • Highly myopic eyes with posterior staphyloma were
    excluded from the study to avoid the
    miscalculation of IOL power.

4
RESULTS
  • Immersion A-scan measured longer axial length
    than contact A-scan biometry ( p lt 0,05 ).

5
RESULTS
  • Compared to contact A-Scan biometry, estimated
    refraction measurements with immersion A-Scan
    biometry were closer to refraction achieved
    postoperatively.

SE Spherical Equivalent
6
RESULTS
  • Lens power prediction error for the contact
    A-scan group was 0,93 0,67 D at the visit on
    day 30, and 0,47 0,46 for the immersion A-scan
    group respectively. The difference was
    statistically significant (plt0,05 ).

SE Spherical Equivalent
7
CONCLUSIONS
  • IOL power calculation to achieve postoperative
    emmetropia, is an important factor in
    phacoemulsification surgery.
  • Our results show that lens power prediction
    error with immersion A-Scan biometry is lower
    compared to contact A-Scan biometry ( p lt 0,05 ).
  • Immersion biometry gives longer axial length
    measurements compared to contact A-Scan biometry
    ( p lt 0,05 ).
  • Choosing the appropriate formula for IOL power
    calculation according to axial length measured by
    ultrasound biometry plays a major role for
    achieving emmetropia after uncomplicated
    phacoemulsification surgery.