Title: The iron content of runoff from a banana ranch is a necessary analytical parameter to analyze. A 25.0mL sample of the runoff was acidified with HNO3 and treated with excess KSCN to form a red complex. (KSCN itself is colorless.) The solution then was
1- The iron content of runoff from a banana ranch is
a necessary analytical parameter to analyze. A
25.0mL sample of the runoff was acidified with
HNO3 and treated with excess KSCN to form a red
complex. (KSCN itself is colorless.) The
solution then was diluted to 100.0mL and put in a
variable pathlength cell. For comparison, a
10.0mL reference sample of 6.80x10-4 M Fe3 was
treated with HNO3 and KSCN and diluted to 50.0mL.
The reference was placed in a cell with a 1.00cm
light path. The runoff sample exhibited the same
absorbance as the reference when the pathlength
of the runoff cell was 2.48cm. What was the
concentration of iron in the banana runoff?
2UV Spectroscopy and Qualitative Analysis
- UV-vis spectroscopy is usually not very useful
for qualitative analysis because there are few
absorption maxima and minima - Solvents
- Must be transparent in region of interest
- Should not interfere with absorbing species (but
usually it does). Polar solvents tend to
obliterate fine structural detail in molecular
spectra
3UV Spectroscopy and Qualitative Analysis
- 3) UV-vis spectroscopy does provide some
information on functional groups
4UV-vis spectroscopy and Quantitative Analysis
- 1)Scope is huge
- a.)95 of all quantitative analyses in health
care field are done by UV-vis spectroscopy - b.) wide applicability to organic and inorganic
analyses - c.) even non-absorbing species can be used by
doing colorimetric reactions (reactions must go
near to completion)
5UV-vis spectroscopy and Quantitative Analysis
- 2. L.O.D. is low, typically 10-4 ? 10-5M but can
be as low as 10-6 ? 10-7M - 3. Moderate to high selectivity
- 4. Accuracy to within 1-3 with minimal training
- 5. Easy and accurate data acquisition
6Construction of Calibration Curves is often done
on the absorption max. Why?
Sample and Reference cells should be matched
7Example
- The ultraviolet absorbances of a series of 9
standards having different nitrate concentrations
were determined at 220nm using a 1.0cm cell 8
samples of river water were taken downstream from
a chemical plant, avg. absorbance 0.642. What is
the nitrate content of the river in mg/mL?
NO3 (mg/mL 0 .004 .015 .025 .035 .04 .05 .06 .07
Abs. .003 .10 .211 .350 .453 .556 .623 .671 .691
8(No Transcript)
9The n ? p (T1) transition occurs at 397nm, the n
? p (S1) transition occurs at 355nm. What is
the difference in energy between the n ? p (T1)
state?
10Relaxation Processes
- Radiationless ? loss of energy in small steps
excitation energy converted into kinetic energy
by small collisions with other molecules, small
increase in temperature - Fluorescence radiative form of relaxation
- Resonance fluorescence no change in wavelength
from excitation to emission Lowest e- state,
vibrational, rotational state of each excited
state produces resonance ? mostly in atoms
11Stokes Shift
Results from thermal energy losses
- Molecules see more non-resonance fluorescence
- Not all excited energy is transmitted as
radiation (some is non-radiative) - Makes emission spectrum look like mirror image of
excitation spectrum - This shift in the spectrum toward longer
wavelengths is called Stokes Shift
12More about fluorescence
- Why do some molecules fluoresce and others dont?
- Want as fast of way as possible to get down to
the ground state, generally non-radiative
internal conversion is fastest but sometimes due
to molecules configuration, fluorescence may be
faster
13More about excess energy loss
- 1. Emission of radiation
- Excited particles (ions, atoms, or molecules)
relax to lower energy level by giving up excess
energy as photons - Excitation brought about by bombardment with e-,
exposure to high potential current, or heat
treatment by arc or flame
14More about excess energy loss
- 2.Thermal radiation a.k.a blackbody radition
- 1. radiation emission lmax prop. To 1/T
- 2. Energy emitted varies as the 4th power of
temperature - 3. Emissive power varies 1/l5
- Heated solids produce IR, vis, and longer l UV
153. Phosphorescence (another way to lose excess
energy)
- Radiative form of relaxation, involves
inter-system crossing where an e- flips spin.
Long lifetime!
16Riboflavin Demo
17- You are in graduate school and just synthesized a
new organic complex that absorbs UV light. Your
advisor is really excited and tells you that you
can write a paper about it for submission to the
Journal of Organic Chemistry but first wants more
information about the new compound for inclusion
in the paper. One piece of useful information
would be the e. Design an experiment to do this.