Chapter%2074:%20Biopotentials%20and%20Electrophysiology%20Measurement - PowerPoint PPT Presentation

About This Presentation
Title:

Chapter%2074:%20Biopotentials%20and%20Electrophysiology%20Measurement

Description:

Electrocardiography (ECG) Measures galvanically the electric activity of the heart ... A normal, scalar ECG curve can be formed from this vectro representation, ... – PowerPoint PPT presentation

Number of Views:546
Avg rating:3.0/5.0
Slides: 31
Provided by: ramot
Category:

less

Transcript and Presenter's Notes

Title: Chapter%2074:%20Biopotentials%20and%20Electrophysiology%20Measurement


1
Chapter 74 Biopotentials and Electrophysiology
Measurement
  • Teemu Rämö
  • teemu.ramo_at_nokia.com

butler.cc.tut.fi/malmivuo/bem/bembook/
2
Agenda
  • 1st half
  • Introduction to biopotentials
  • Measurement methods
  • Traditional ECG, EEG, EMG, EOG
  • Novell VCG
  • 2nd half
  • Measurement considerations
  • Electronics
  • Electrodes
  • Practices
  • QA

3
What are biopotentials
  • Biopotential An electric potential that is
    measured between points in living cells, tissues,
    and organisms, and which accompanies all
    biochemical processes.
  • Also describes the transfer of information
    between and within cells
  • This book focuses strictly on the measurement of
    potentials

4
Mechanism behind biopotentials 1/2
  • Concentration of potassium (K) ions is 30-50
    times higher inside as compared to outside
  • Sodium ion (Na) concentration is 10 times higher
    outside the membrane than inside
  • In resting state the member is permeable only for
    potassium ions
  • Potassium flows outwards leaving an equal number
    of negative ions inside
  • Electrostatic attraction pulls potassium and
    chloride ions close to the membrane
  • Electric field directed inward forms
  • Electrostatic force vs. diffusional force
  • Nernst equation
  • Goldman-Hodgkin-Katz equation

5
Mechanism behind biopotentials 2/2
  • When membrane stimulation exceeds a threshold
    level of about 20 mV, so called action potential
    occurs
  • Sodium and potassium ionic permeabilities of the
    membrane change
  • Sodium ion permeability increases very rapidly at
    first, allowing sodium ions to flow from outside
    to inside, making the inside more positive
  • The more slowly increasing potassium ion
    permeability allows potassium ions to flow from
    inside to outside, thus returning membrane
    potential to its resting value
  • While at rest, the Na-K pump restores the ion
    concentrations to their original values
  • The number of ions flowing through an open
    channel gt106/sec
  • Body is an inhomogeneous volume conductor and
    these ion fluxes create measurable potentials on
    body surface

6
Electrocardiography (ECG)
  • Measures galvanically the electric activity of
    the heart
  • Well known and traditional, first measurements
    byAugustus Waller using capillary electrometer
    (year 1887)
  • Very widely used method in clinical environment
  • Very high diagnostic value

7
ECG basics
  • Amplitude 1-5 mV
  • Bandwidth 0.05-100 Hz
  • Largest measurement error sources
  • Motion artifacts
  • 50/60 Hz powerline interference
  • Typical applications
  • Diagnosis of ischemia
  • Arrhythmia
  • Conduction defects

8
12-Lead ECG measurement
  • Most widely used ECG measurement setup in
    clinical environment
  • Signal is measured non-invasively with 9
    electrodes
  • Lots of measurement data and international
    reference databases
  • Well-known measurement and diagnosis practices
  • This particular method was adopted due to
    historical reasons, now it is already rather
    obsolete

Einthoven leads I, II III
Goldberger augmented leads VR, VL VF
Precordial leads V1-V6
9
Why is 12-lead system obsolete?
  • Over 90 of the hearts electric activity can be
    explained with a dipole source model
  • Only 3 orthogonal components need to be measured,
    which makes 9 of the leads redundant
  • The remaining percentage, i.e. nondipolar
    components, may have some clinical value
  • This makes 8 truly independent and 4 redundant
    leads
  • 12-lead system does, to some extend, enhance
    pattern recognition and gives the clinician a
    few more projections to choose from
  • but.
  • If there was no legacy problem with current
    systems, 12-lead system wouldve been discarded
    ages ago

10
Electroencephalography (EEG)
  • Measures the brains electric activity from the
    scalp
  • Measured signal results from the activity of
    billions of neurons
  • Amplitude 0.001-0.01 mV
  • Bandwidth 0.5-40 Hz
  • Errors
  • Thermal RF noise
  • 50/60 Hz power lines
  • Blink artifacts and similar
  • Typical applications
  • Sleep studies
  • Seizure detection
  • Cortical mapping

11
EEG measurement setup
  • 10-20 Lead system is most widely clinically
    accepted
  • Certain physiological featuresare used as
    reference points
  • Allow localization of diagnostic features in the
    vicinity of the electrode
  • Often a readily available wire or rubber mesh is
    used
  • Brain research utilizes even 256 or 512 channel
    EEG hats

12
Electromyography (EMG)
  • Measures the electric activity of active muscle
    fibers
  • Electrodes are always connected very close to the
    muscle group being measured
  • Rectified and integrated EMG signal gives rough
    indication of the muscle activity
  • Needle electrodes can be used to measure
    individual muscle fibers
  • Amplitude 1-10 mV
  • Bandwidth 20-2000 Hz
  • Main sources of errors are 50/60 Hz and RF
    interference
  • Applications muscle function, neuromuscular
    disease, prosthesis

13
Electrooculography (EOG)
  • Electric potentials are created as a result of
    the movement of the eyeballs
  • Potential varies in proportion to the amplitude
    of the movement
  • In many ways a challenging measurement with some
    clinical value
  • Amplitude 0.01-0.1 mV
  • Bandwidth DC-10 Hz
  • Primary sources of error include skin potential
    and motion
  • Applications eye position, sleep state,
    vestibulo-ocular reflex

14
Vectorcardiogram (VCG or EVCG)
  • Instead of displaying the scalar amplitude (ECG
    curve) the electric activation front is measured
    and displayed as a vector (dipole model,
    remember?)
  • ? It has amplitude and direction
  • Diagnosis is based on the curve that the point of
    this vector draws in 2 or 3 dimensions
  • The information content of the VCG signal is
    roughly the same as 12-lead ECG system. The
    advantage comes from the way how this information
    is displayed
  • A normal, scalar ECG curve can be formed from
    this vectro representation, although (for
    practical reasons) transformation can be quite
    complicated
  • Plenty of different types of VCG systems are in
    use
  • ? No legacy problem as such

15
  • Short break,
  • Kahvia ja pullaa!

16
The biopotential amplifier
  • Small amplitudes, low frequencies, environmental
    and biological sources of interference etc.
  • Essential requirements for measurement equipment
  • High amplification
  • High differential gain, low common mode gain ?
    high CMRR
  • High input impedance
  • Low Noise
  • Stability against temperature and voltage
    fluctuations
  • Electrical safety, isolation and defibrillation
    protection

17
The Instrumentation Amplifier
  • Potentially combines the best features desirable
    for biopotential measurements
  • High differential gain, low common mode gain,
    high CMRR, high input resistance
  • A key design component to almost all biopotential
    measurements!
  • Simple and cheap, although high-quality OpAmps
    with high CMRR should be used

18
Application-specific requirements
  • ECG amplifier
  • Lower corner frequency 0.05 Hz, upper 100Hz
  • Safety and protection leakage current below
    safety standard limit of 10 uA
  • Electrical isolation from the power line and the
    earth ground
  • Protection against high defibrillation voltages
  • EEG amplifier
  • Gain must deal with microvolt or lower levels of
    signals
  • Components must have low thermal and electronic
    noise _at_ the front end
  • Otherwise similar to ECG
  • EMG amplifier
  • Slightly enhanced amplifier BW suffices
  • Post-processing circuits are almost always needed
    (e.g. rectifier integrator)
  • EOG amplifier
  • High gain with very good low frequency (or even
    DC) response
  • DC-drifting ? electrodes should be selected with
    great care
  • Often active DC or drift cancellation or
    correction circuit may be necessary

19
Electrical Interference Reduction
  • Power line interference (50 or 60 Hz) is always
    around us
  • Connects capacitively and causes common mode
    interference
  • The common mode interference would be completely
    rejected by the instrumentation amplifier if the
    matching would be ideal
  • Often a clever driven right leg circuit is used
    to further enhance CMRR
  • ? Average of the VCM is inverted and driven back
    to the body via reference electrode

20
Filtering
  • Filtering should be included in the front end of
    the InstrAmp
  • Transmitters, motors etc. cause also RF
    interference

Small inductorsor ferrite beadsin the lead
wiresblock HF frequencyEM interference
RF filtering withsmall capacitors
21
50 or 60 Hz notch filter
  • Sometimes it may be desirable to remove the power
    line interference
  • Overlaps with the measurement bandwidth
  • May distort the measurement result and have an
    affect on the diagnosis!
  • Option often available with EEG EOG measuring
    instruments

Twin Tnotch filter
22
Artifact reduction
  • Electrode-skin interface is a major source of
    artifact
  • Changes in the junction potential causes slow
    changes in the baseline
  • Movement artifacts cause more sudden changes and
    artifacts
  • Drifting in the baseline can be detected by
    discharging the high-pass capacitor in the
    amplifier to restore the baseline

23
Electrical isolation
  • Electrical isolation limits the possibility of
    passage of any leakage current from the
    instrument in use to the patient
  • Such passage would be harmful if not fatal!
  • Transformer
  • Transformers are inherently high frequency AC
    devices
  • Modulation and demodulation needed
  • Optical isolation
  • Optical signal is modulated in proportion to the
    electric signal and transmitted to the detector
  • Typically pulse code modulated to circumvent the
    inherent nonlinearity of the LED-phototransistor
    combination

24
Defibrillation Protection
  • Measuring instruments can encounter very high
    voltages
  • E.g. 15005000V shocks from defibrillator
  • Front-end must be designed to withstand these
    high voltages

25
Electrodes Basics
  • High-quality biopotential measurements require
  • Good amplifier design
  • Use of good electrodes and their proper placement
    on the patient
  • Good laboratory and clinical practices
  • Electrodes should be chosen according to the
    application
  • Basic electrode structure includes
  • The body and casing
  • Electrode made of high-conductivity material
  • Wire connector
  • Cavity or similar for electrolytic gel
  • Adhesive rim
  • The complexity of electrode design often neglected

26
Electrodes - Basics
  • Skin preparation by abrasion or cleansing
  • Placement close to the source being measured
  • Placement above bony structures where there is
    less muscle mass
  • Distinguishing features of different electrodes
  • How secure? The structure and the use of strong
    but less irritant adhesives
  • How conductive? Use of noble metals vs. cheaper
    materials
  • How prone to artifact? Use of low-junction-potenti
    al materials such as Ag-AgCl
  • If electrolytic gel is used, how is it applied?
    High conductivity gels can help reduce the
    junction potentials and resistance but tend to be
    more allergenic or irritating

Baseline drift due to thechanges in
junctionpotential or motion artifacts?Choice of
electrodes
Muscle signalinterference? Placement
Electromagneticinterference? Shielding
27
Ag-AgCl, Silver-Silver Chloride Electrodes
  • The most commonly used electrode type
  • Silver is interfaced with its salt
    silver-chloride
  • Choice of materials helps to reduce junction
    potentials
  • Junction potentials are the result of the
    dissimilar electrolytic interfaces
  • Electrolytic gel enhances conductivity and also
    reduces junction potentials
  • Typically based on sodium or potassium chloride,
    concentration in the order of 0.1 M weak enough
    to not irritate the skin
  • The gel is typically soaked into a foam pad or
    applied directly in a pocket produced by
    electrode housing
  • Relatively low-cost and general purpose electrode
  • Particularly suited for ambulatory or long term
    use

28
Gold Electrodes
  • Very high conductivity ? suitable for low-noise
    meas.
  • Inertness ? suitable for reusable electrodes
  • Body forms cavity which is filled with
    electrolytic gel
  • Compared to Ag-AgCL greater expense,
    higherjunction potentials and motion artifacts
  • Often used in EEG, sometimes in EMG

Conductive polymer electrodes
  • Made out of material that is simultaneously
    conductive and adhesive
  • Polymer is made conductive by adding monovalent
    metallic ions
  • Aluminum foil allows contact to external
    instrumentation
  • No need for gel or other adhesive substance
  • High resistivity makes unsuitable for low-noise
    meas.
  • Not as good connection as with traditional
    electrodes

29
Metal or carbon electrodes
  • Other metals are seldom used as high-quality
    noblemetal electrodes or low-cost carbon or
    polymericelectrodes are so readily available
  • Historical value. Bulky and awkward to use
  • Carbon electrodes have high resistivity and are
    noisier but they are also flexibleand reusable
  • Applications in electrical stimulation and
    impedance plethysmography

Needle electrodes
  • Obviously invasive electrodes
  • Used when measurements have to be taken from the
    organ itself
  • Small signals such as motor unit potentials can
    be measured
  • Needle is often a steel wire with hooked tip

30
  • Thats it,
  • Now for QA

SQUID Superconducting Quantum Interference
Device
Write a Comment
User Comments (0)
About PowerShow.com