2nd TUG Meeting - PowerPoint PPT Presentation

1 / 31
About This Presentation
Title:

2nd TUG Meeting

Description:

results of the 1st TUG Meeting. presentation on XEDS (F. Ernst) arrangements for plasma cleaner demonstration. demand ... Ansatz (Cliff Lorimer): 'k factor' kAB ... – PowerPoint PPT presentation

Number of Views:46
Avg rating:3.0/5.0
Slides: 32
Provided by: frank176
Category:
Tags: 2nd | tug | lorimer | meeting

less

Transcript and Presenter's Notes

Title: 2nd TUG Meeting


1
2nd TUG Meeting
  • August 19, 2002

2
Presentations
3
Presentations
4
Agenda
  • results of the 1st TUG Meeting
  • presentation on XEDS (F. Ernst)
  • arrangements for plasma cleaner demonstration
  • demand for further action

5
Results of 1st TUG Meeting
  • what we achieved
  • newsgroup (thanks Leah Lucas)
  • simulator (thanks Yeonsop Yu)
  • XEDS detector geometry (thanks Yeonsop Yu)
  • Gatan Duo Mill (thanks Lichun Zhang)
  • demo of plasma cleaner by Fischione tomorrow!
  • schedule of presentations
  • what we have not (yet) achieved
  • decision about purchase of a wire saw
  • acquisition of tripods

6
XEDS
  • X-ray Energy-Dispersive Spectroscopy

7
Inelastic Scattering
8
Which Elements?
  • ratio between X-ray emission and emission of
    Auger electrons fluorescence yield ?
  • XEDS is not particularly suited to detect light
    elements
  • ? EELS

9
Detector
10
Detector Efficiency
11
Pulse Processing
  • ideally high counting rate, without shifting or
    distorting the spectrum
  • adjustable parameters
  • time constant ?
  • dead-time ratio
  • time constant
  • time available for processing the last pulse
  • typically 10 - 50µs
  • larger time constant ? better accuracy, but
    smaller counting rate
  • dead-time ratio
  • percentage of time the detector needs to switch
    off for pulse processing
  • closely related to ?
  • the more X-ray photons arrive at the detector per
    second, the larger the dead-time ratio
  • the dead-time ratio can be expressed as
  • Ri input pulse rate Re output pulse rate (rate
    of counts).

12
Dead-Time Ratio
  • a high dead-time ratio (larger than 25) means
  • the detector receives too many pulses per second
  • the detection of X-ray photons tends to be
    inefficient
  • reduce the beam current!
  • reduce spot size!
  • remove objective aperture before inserting the
    detector!

13
Energy Resolution
  • definition of the energy resolution, R
  • R energy resolution P parameter
    characterizing the quality of he associated
    electronics X effect of leakage current and
    incomplete charge collection I intrinsic line
    width of the detector.
  • intrinsic line width of the detector
  • F Fano factor (Poisson statistics) ? energy
    to create an electron hole pair in the detector
    E energy of the X-ray photons.
  • ? energy resolution can only be specified for
    specific analysis conditions
  • standard (IEEE) Mn-K? line, generated by a
    well-defined source (Fe55 with 103 cps and 8 µs
    pulse processor time constant)
  • typical energy resolution
  • Si detectors  140 eV Ge detectors 114 eV
    (record)
  • Tecnai F30 CWRU 129.9 eV (at Mn-K?)

14
Artifacts in XEDS Spectra
  • Escape-Peaks
  • phantom peaks, caused by fluorescence
  • a small fraction of the X-ray photons absorbed by
    the detector is not converted to electron-hole
    pairs
  • these photons cause the detector material to
    fluorescence (emit photons that are
    characteristic of the detector material)
  • case of Si detectors
  • Si-K? photon escapes from the detector
  • energy of Si-K? photons 1.74eV
  • ? detector does not register energy E, but
    E-1.74 keV
  • ? additional peak (escape peak), 1.74 keV below
    the actual peak at E

15
Escape Peaks
16
Sum Peaks
  • Sum Peaks
  • phenomenon instead of evaluating a single X-ray
    photon, the detector simultaneously evaluates two
    photons
  • reason for this problem electronics cannot
    switch-off the detector fast enough
  • ? strong peaks may be accompanied by a phantom
    peak at twice the energy

17
Sum Peaks Example
18
XEDSTEM Interface
19
Detector Geometry Tecnai F30
  • window
  • Super Ultra Thin type
  • 0.3 µm thick
  • polymer type over Al support grid
  • detector
  • dead layer thickness 0.085 µm.
  • Al thickness 0.04 µm
  • area 30mm2
  • geometry
  • detector distance to the sample 10.72 mm
  • corresponds to scale setting of 10.3 mm
  • elevation angle 0
  • azimuth 0

20
Acceptance Angle
  • ? angle between plane of specimen (untilted) and
    line to center of detector
  • minimize absorption of X-rays in the specimen ?
    maximize acceptance angle ?
  • problem increasing ? requires to decrease ?
  • ideal acceptance angle of ?  90 position the
    detector above upper pole piece
  • ? ? small, detector needs to view specimen
    through pole piece hole
  • if detector below the upper pole piece, ? is
    limited to typically 20
  • for ideally thin TEM specimens absorption does
    not cause problems
  • thicker specimens
  • increase acceptance angle by tilting specimen
    towards detector
  • Tecnai F30 13

21
Optimum Geometry
  • reduce absorption
  • use adequately thin specimens
  • appropriately position the specimen with respect
    to the detector
  • region of interest should not be on the same side
    of the specimen hole as the detector, but on the
    opposite side
  • rotation holders facilitate this setup
  • when using standard specimen holders, load
    specimen with the region of interest rotated to
    the appropriate position

22
Optimum Geometry?
23
Optimum Geometry for Interfaces
24
Spurious X-Rays
25
Qualitative XEDS
  • optimize the conditions for recording spectra
  • identify all element-characteristic peaks
  • identify artifacts
  • identify all relevant elements in the specimen

26
Peak Significance
  • significance of peaks in XEDS spectra
  • intensity IA of the peak above the background
  • integrate signal and background over the same
    interval of N channels
  • the peak is significant at the 99  level of
    confidence if
  • if necessary, improve sensitivity
  • not by inadequate increase of the intensity (beam
    current, specimen thickness  this reduces the
    energy resolution and favors the formation of sum
    peaks)
  • instead extend recording time!

27
Example Increased Recording Time
28
Quantitative XEDS
  • quantitative analysis of an alloy AB
  • what are the relative parts of A and B?
  • distinguish
  • mole fractions ?Ax, ?Bx dimension-less
  • mass fractions XAx, XBx in wt  (weight
    percent should be mass percent!)
  • ? determine intensities IA and IB of the
    corresponding peaks, above the background of the
    spectrum
  • assumption absorption and fluorescence may be
    neglected
  • Ansatz (CliffLorimer)
  • k factor kAB
  • conversion from relative peak intensities to
    relative amounts of elements A and B at x

29
Background Bremsstrahlung
30
Background Subtraction
  • two windows of equal width, before and one after
    peak(s)
  • window width should roughly correspond to peak
    width (FWHM)
  • integrate counts over both windows N
  • ? average the results to obtain an estimate for
    the background below the peak(s)
  • ? this method works well only for slowly varying
    background

31
Example of an XEDS Linescan
Write a Comment
User Comments (0)
About PowerShow.com