Quantitative Arterial Spin Labelling ASL at UltraHigh Field 7.0 T PowerPoint PPT Presentation

presentation player overlay
1 / 1
About This Presentation
Transcript and Presenter's Notes

Title: Quantitative Arterial Spin Labelling ASL at UltraHigh Field 7.0 T


1
Quantitative Arterial Spin Labelling (ASL) at
Ultra-High Field (7.0 T) A. G. Gardener, P. A.
Gowland and S. T. Francis Sir Peter Mansfield
Magnetic Resonance Centre, School of Physics and
Astronomy, University of Nottingham, University
Park, Nottingham, NG7 2RD, UK.
  • Introduction and Theory
  • At ultra-high field arterial spin labelling
    (ASL) using magnetically-labelled blood as an
    endogenous tracer - benefits from an overall
    increase in SNR and T1 relaxation times of blood
    and tissue. This leads to increased perfusion
    weighted (PW) signals occurring at longer delay
    times (TIs).
  • Here two Pulsed ASL schemes, STAR (Signal
    Targeting with Alternating Radiofrequency) 1
    and FAIR (Flow sensitive Alternating Inversion
    Recovery) 2 are assessed in humans at 7.0 T.
  • STAR and FAIR PASL techniques are based on the
    subtraction of two consecutively acquired label
    and non-label images. STAR (Fig. 1A) uses a label
    inversion slab (tag) applied proximal and a
    non-label slab (control) applied distal to the
    imaging slab. FAIR (Fig. 1B) uses a selective
    inversion slab for the label, and a non-selective
    inversion pulse for the non-labelled image,
    centred over the imaging slab. In both cases the
    difference between the label and non-label images
    provides the PW signal.
  • However B1 and B0 inhomogeneity at 7.0 T can
    lead to imperfect inversion pulse profiles,
    resulting in offsets between label and non-label
    images in the absence of perfusion.
  • One solution is to increase the gap between the
    inversion slab and image slice, by altering the
    width of the label/non-label, and/or by applying
    pre-/post-saturation pulses to the imaging slices
    immediately before/after the inversion pulses.
    However, these can all have an adverse effect on
    transit time.
  • These factors are assessed to determine the
    optimal PASL sequence for use at 7.0 T.

Figure 1 (A) STAR and (B) FAIR ASL
Schemes.
  • Method and Analysis
  • Data were acquired on a 7.0 T Philips Achieva
    scanner. For inversion a hyperbolic secant (hsc)
    adiabatic pulse was used, whilst for
    pre/post-saturation an optimized sinc-gauss pulse
    was applied. Initial experiments performed on
    phantoms assessed inversion pulse efficiency, a,
    as a function of position for various slab
    widths/offsets. Profiles of labelling slabs were
    measured transaxially using sagittal SE-EPI
    images.
  • For the ASL sequences, an image slab of five
    GE-EPI slices with in-plane resolution of 3 x 3
    mm2, 3 mm slice thickness with 1 mm slice gap, 64
    x 64 matrix and echo time of 20 ms were acquired
    in ascending order after each labelling scheme.
  • STAR used 70 mm wide label and non-label slabs
    label gaps of 10-25 mm between image/inversion
    slabs. FAIR used selective label slabs of 30 -
    50 mm and 150 mm non-selective non-label slabs.
    Widths of labels/non-labels were chosen to allow
    for full refreshment of blood in the TR period.
    The effect of the addition of pre-/post-saturation
    pulses immediately before and after the
    inversion pulse (up to 2 of each) and different
    crusher gradient areas after each saturation
    pulse was assessed.
  • The optimal STAR and FAIR labelling schemes were
    performed on 4 healthy human volunteers, who gave
    informed written consent. Images were acquired at
    five TIs, with sixty ASL sets acquired at each
    TI, with a TR of 3 s. Images were acquired both
    with and without velocity crushing (diffusion
    weighting 2.5 smm-1, critical velocity 20 mms-1).
    A T1 map and M0 map were also generated for
    perfusion quantification. PW images were formed
    from (label - non-label) images and
    quantification of perfusion performed using a
    multi-compartment system model, assuming the T1
    of arterial blood to be 2.0 s.

Figure 3 (A) STAR vs. (B) FAIR at TI
800 ms. For these scans no vascular crushing was
applied.
  • Results and Discussion
  • Figure 2 shows inversion efficiency as a
    function of label width for a label positioned at
    coil isocentre and also off-centre by label
    width. The inversion efficiency is gt0.98 over the
    central 10 cm of the label, but profile sharpness
    is poor and varies depending on position in the
    coil. The hsc pulse had a simulated inversion
    profile Mz lt -0.98 M0 for 86 of slice width,
    which gave significant static tissue artefact for
    STAR in the difference images, but not for FAIR.
  • The effect of using pre- and post-saturation
    pulses to suppress the offset signal for STAR
    label gap of 25 mm was assessed. It was found
    that the application of a single pre- and post-
    saturation pulse with variable crushers gave the
    optimal reduction of offset signal (0.15 ). The
    same scheme reduced the offset signal to 0.08
    in FAIR and this saturation scheme was used in
    subsequent human subject studies. STAR labels
    were set to be 70 mm wide, whereas FAIR could use
    a 50 mm selective 15 mm effective label gap and
    a 150 mm non-selective slab, resulting in shorter
    transit time delays for FAIR.
  • Figure 3 shows averaged difference images at a
    TI of 800 ms for the two schemes, showing the
    unwanted static tissue signal artefact in the
    STAR PW image set. Figure 4 shows a perfusion map
    formed from fitting five increasing TI FAIR
    data-sets. Figure 5 shows corresponding signal
    change with TI in two grey matter ROIs with
    vascular crushing applied for FAIR.
  • Conclusions
  • At 7.0 T STAR gives large artefactual signal
    from static tissue, due to poor label slab
    profiles caused by B1 inhomogeneity. This problem
    was worse away from the isocentre of the coil.
    The STAR static tissue artefact could not be
    fully suppressed by saturation pulses even at
    wider label gaps, which led to longer long
    transit times and so reduced label efficiency.
  • The FAIR labelling scheme is less sensitive to
    offset effects as the image slab is positioned at
    the centre of the labels, and hence the label
    centres are generally in the homogeneous region
    of the coil. With pre- and post-saturation pulses
    good static tissue suppression could be obtained
    for FAIR at small label gaps.
  • At 7.0 T quantitative measures of perfusion are
    best obtained with FAIR using in-plane
    saturation, with PW signal intensity in grey
    matter of 1.5 and signal change peaking at TIs
    of 1800 ms.

References and Acknowledgments 1) Edelman et al,
MRM, 40, 513 (1994). 2) Kim et al, MRM, 34,
293-301 (1995). 3) Pfeuffer et al, MRM, 47, 903
(2002). Funded by the MRC and EPSRC.
Write a Comment
User Comments (0)
About PowerShow.com