Electron Paramagnetic Resonance Biodosimetry in Teeth and Fingernails - PowerPoint PPT Presentation

1 / 50
About This Presentation
Title:

Electron Paramagnetic Resonance Biodosimetry in Teeth and Fingernails

Description:

Electron Paramagnetic Resonance Biodosimetry in Teeth and Fingernails – PowerPoint PPT presentation

Number of Views:451
Avg rating:3.0/5.0
Slides: 51
Provided by: nnm4
Category:

less

Transcript and Presenter's Notes

Title: Electron Paramagnetic Resonance Biodosimetry in Teeth and Fingernails


1
Electron Paramagnetic Resonance Biodosimetry in
Teeth and Fingernails
  • A. Romanyukha1,2, R.A. Reyes2, F. Trompier3, L.A.
    Benevides1, H.M. Swartz4

1Naval Dosimetry Center, 8901 Wisconsin Ave.,
Bethesda, MD, 20889, USA, 2Uniformed Services
University, 4301 Jones Bridge Rd., Bethesda, MD,
20814, USA, 3Institut de Radioprotection et de
Sûreté Nucléaire, Fontenay-aux-roses,
France, 4Dartmouth Medical School, Hanover, NH,
03755, USA
2
Outline
  • EPR dosimetry basics
  • In vitro X and Q dosimetry in tooth enamel
  • In vivo tooth L-band dosimetry
  • EPR dosimetry in fingernails
  • Conclusions

3
What is Electron Paramagnetic Resonance (EPR) ?
  • Non-destructive magnetic resonance technique used
    to detect and quantify unpaired electrons.
  • Absorption of ionizing radiation generates
    unpaired electrons (i.e., paramagnetic centers).
  • The concentration of radiation-induced
    paramagnetic centers is proportional to the
    absorbed dose.

4
EPR Fundamentals and Principles
  • There is a net absorption of energy from the
    microwave field at resonance because of a greater
    population of electrons are in the lower energy
    state.
  • The process is non-destructive because the
    population difference reestablishes itself after
    the microwave field is turned off.
  • Thus, the history of radiation exposure is not
    destroyed by EPR measurements.

5
Optical Imaging
6
Typical frequencies and wavelengths required for
resonance of a free electron in EPR measurements
7
EPR dosimeters for partial body exposure
Radiation-induced radicals are stable only in
hard tissues teeth, bone, fingernails and hairs.
Depending on mw band EPR can be measured in vivo
or in vitro using specially prepared samples from
human hard tissues
Finger- and toenails
8
Characteristics of EPR dosimetry
  • Non-invasive
  • Based on a physical process
  • Not affected by biological processes such as
    stress
  • Not affected by simultaneous damage that is
    likely to occur with irradiation such as wounds
    burns
  • Applicable to individuals
  • Measurements can be made at any interval after
    irradiation up to at least 2 weeks (fingernails)
    or indefinately (teeth)
  • Can provide output immediately after the
    measurement
  • Unaffected by dose rate
  • Can operate in a variety of environments
  • Systems can be developed so that they can be
    operated by minimally trained individuals

9
In vitro measurements in tooth enamel samples (X
and Q-bands)
10
Extracted teeth can be available for in vitro
EPR measurements
11
EPR dosimetry with teeth is the only method which
can reconstruct external gamma radiation doses
(lt100 mGy) individually.
Validation and Standardization
Four successful International Dose
Intercomparisons with totally more than 20
participating labs
ICRU, 2002. Retrospective Assessment of Exposures
to Ionizing Radiation. Report 68 (Bethesda, MD
ICRU).
IAEA, 2002. Use of electron paramagnetic
resonance dosimetry with tooth enamel for
retrospective dose assessment. International
Atomic Energy Agency, Vienna, IAEA-TECDOC-1331.
12
Steps of the method
  • Tooth collections
  • Tooth enamel sample preparation
  • EPR measurements of radiation response
  • Calibration of EPR radiation response

13
EPR Biodosimetry(Teeth)
14
EPR Biodosimetry(Teeth)
  • Hydroxyapatite constitutes
  • 95 by weight of tooth enamel
  • 70-75 of dentin
  • 60-70 of compact bones

Romanyukha, et. al, Appl. Radiat. Isot. (2000)
and IAEA-TECDOC-1331
15
EPR Biodosimetry(Dose Calibration)
16
EPR Biodosimetry Applications (Epidemiological
Investigations Using Tooth EPR)
17
Conclusion
  • EPR X-band (9 GHz) dosimetry in tooth enamel
    works excellent (LLDlt100 mGy, time after exposure
    when dose measurements are possible from 0.01 hr
    to 106 yr.
  • But it requires to have extracted or exfoliated
    teeth available for preparation of tooth enamel

18
Alternatives to exfoliated/extracted teeth
L-band (1.2 GHz) non-Invasive in vivo
measurements
Q-band (35 GHz) measurements in enamel biopsy
samples (2 mg) with followed up tooth restoration
19
Q-band (35 GHz) measurements in enamel biopsy
samples (2 mg) with followed up tooth
restoration
20
Tooth enamel powder samples for test 0 0.1 Gy
0.5 Gy 1 Gy 3 Gy 5 Gy
Description of Q-band feasibility test
Each sample was recorded 3 times in X (100 mg)
and Q bands (2, 4 mg)
Recent publication Romanyukha A. et al. Q-band
EPR biodosimetry in tooth enamel microprobes
Feasibility test and comparison with X band.
Health Physics. 93, 631-635, (2007).
21
X-band spectrum vs Q-band spectrum
X-band (100 mg), 0.1 Gy
Q-band, (4 mg) 0.1 Gy
  • Q-band has significantly lesser amount of the
    sample required for dose measurements
  • Q-band has significantly better spectral
    resolution of dose response

22
Dose dependence X vs Q
23
Dental Biopsy Technique
  • With the enamel biopsy technique a small enamel
    chip is removed from a tooth crown with minimal
    damage to the structural integrity of the tooth.
  • A high-speed compressed-air driven dental hand
    piece is used with appropriate dental burs for
    this purpose.
  • Standard techniques for tooth restoration using
    light-cured composite resins rapidly restore the
    small enamel defect in the biopsied enamel
    surface of the crown.
  • Preliminary study on discarded teeth have
    demonstrated the feasibility of removing 2 mg
    enamel chips, the desired size for sufficient
    sensitivity with Q-band EPR dosimetry.

In collaboration with B. Pass, P. Misra, T. De
(Howard University)
24
Q-band biopsy experiment
  • Tooth enamel biopsy sample 2.2 mg was irradiated
    4 times to the same dose - 4.3 Gy
  • After each irradiation angle dependence (12
    positions) of biopsy sample was studied
  • Using average, maximum, minimum and median values
    of EPR radiation response at each dose (e.g. 4.3,
    8.6, 12.9 and 17.1 Gy) and linear back
    extrapolation attempt to reconstruct dose of 4.3
    Gy was made

25
Angle dependence of radiation response
Possible approaches 1. Use average value of
radiation response at each dose 2. Use maximum
value of radiation response at each dose 3. Use
minimum value of radiation response at each
dose 4. Use median value of radiation response
at each dose.
26
Spectra in biopsy sample at different doses and
dose dependences
Appearance of tooth enamel spectrum (maximum) of
the same biopsy sample 2.2 mg at different doses
Dose dependences for average, maximum, minimum
and median values of radiation response at each
dose
27
Results of attempt to reconstruct 4.3 Gy in
biopsy sample (2.2 mg) using different approaches
28
Preliminary conclusions
  • Tooth enamel biopsy spectra have slightly
    different shape from powder spectra, they are
    more narrow and have higher signal-to-noise ratio
    for the same dose than powder spectra. However
    existence of angle dependence for biopsy spectra
    makes difficult dose reconstruction. Possible
    solution is to use average, maximum, minimum or
    median values for each dose for dose
    reconstruction
  • Use of average and minimum EPR radiation response
    values gives the best results to reconstruct 4.3
    Gy, e.g. 5.5 0.8 Gy and 5.4 0.7 Gy,
    respectively
  • A possible reason for some dose offset (1 Gy) is
    a slope of a base line of the spectra for this
    sample
  • A possible solution is to apply base line
    correction to spectra before measurements of
    peak-to-peak amplitude of radiation response

29
L-band in vivo
30
Recent publications
  • Swartz H.M. et al. Measurements of clinically
    significant doses of ionizing radiation using
    non-invasive in vivo EPR spectroscopy of teeth in
    situ. Appl. Radiat. Isot. 62, 293-299 (2005)
  • Swartz H.M. et al. In Vivo EPR Dosimetry to
    Quantify Exposures to Clinically Significant
    Doses of Ionizing Radiation. Radiat. Prot. Dosim.
    120, 163-170 (2006).
  • Swartz H.M. et al. In Vivo EPR for Dosimetry.
    Radiat. Meas. 42, 1075-1084, (2007).

31
  • L-band (1 GHz) of microwaves is better for
    realization of in vivo EPR than standard X-band
    (9 GHz) because it has
  • Greater tolerance for the presence of water
  • Relatively large sample volume sufficient for
    whole tooth.

32
Components of in vivo EPR spectrometer
  • Resonators that will probe teeth in vivo
  • Magnet system that can comfortably and
    effectively encompass the human head
  • Software for EPR dose response determination
  • Dose calibration for in vivo L-band measurements

33
Clinical EPR Spectrometers
34
Retrospective Radiation Dosimetry
35
In Vivo EPR Radiation Dosimetry
Under practical conditions with an irradiated
tooth in the mouth of a volunteer, the dose
dependent signal amplitude is clearly observed.
(Acq. time 4.5 minutes/spectrum)
36
EPR Dose Response
37
(No Transcript)
38
Dose-dependence for 6 in vivo teeth, with each
tooth irradiated to a different dose and measured
on 3 separate days. Linear regression analysis
shows that the standard error of dose prediction
is  46 cGy.
39
EPR biodosimetry in tooth enamel for partial body
dose assessment
  • X-band EPR is ready to use for forensic dose
    assessment. Could be carried out on compact and
    transportable (lt 150 kg) EPR spectrometer. Dose
    level lt100 mGy.
  • Q-band biopsy potentially is able to measure
    doses lt 500 mGy in biopsy tooth enamel samples
    2-4 mg.
  • L-band in vivo EPR potentially is able to measure
    doses as low as 3 Gy. Needs some additional
    development.

40
Finger-and toenails facts
  • Typical available amounts of nail parings are up
    to 120 mg for fingernails and up to 160 mg for
    toe nails
  • Nails grow all the time, but their rate of
    growth slows down with age and poor circulation
  • Fingernails grow at an average of one-tenth of
    an inch (3 mm) a month. It takes 6 months for a
    nail to grow from the root to the free edge
  • Toenails grow about 1 mm per month and take
    12-18 months to be completely replaced
  • The nails grow faster on your dominant hand, and
    they grow more in summer than in winter

The major component of fingernails is a
a-keratin. This protein is built up from three,
long a-helical peptide chains that are twisted
together in a left-handed coil, strengthened by S
S bridges formed from adjacent cisteine groups.
41
Recent development
Romanyukha A. et al. EPR dosimetry in chemically
treated fingernails. Radiat. Meas. 42, 1110-1113,
(2007). Trompier F. et al. Protocol for emergency
EPR dosimetry in fingernails. Radiat. Meas. 42,
1085-1088, (2007). Reyes R.A. et al. Electron
paramagnetic resonance in human fingernails the
sponge model implication. To be published in
Radiat. Env. Biophys. (2008)
42
New insights in EPR fingernail dosimetry
  • Fingernails can be considered as a sponge-like
    tissue which behaves differently from in vivo
    fingernails when mechanically-stressed after
    clipping
  • Most of previously published results on EPR
    fingernail dosimetry were obtained on stressed
    samples and not applicable to life-scenario
    situation
  • Unstressed fingernails have more significantly
    stable and sensitive radiation response which can
    be measured with EPR

43
Radiation-induced signal in unstressed
fingernails
RIS spectra obtained by subtraction of BKS
spectrum recorded prior irradiation
RIS parameters g2.0088 DH9 G
44
Result of dose reconstruction in the sample
irradiated to 4 Gy 5 days before reconstruction
Reconstructed dose 3.66 Gy, reduction
45
Variability of dose dependence in fingernails
46
Dosimetric properties of fingernails
  • Optimal sample mass is 15-20 mg (nail-parings
    from 2-3 fingers)
  • Measurements time 5 minutes (10 scans)
  • Achievable lower dose threshold 1 Gy
  • RIS fading half-time 300 hr (2 weeks)

47
Conclusions
48
Acknowledgements
  • G. Burke, E. Demidenko, C. Calas, I. Clairand, T.
    De, O. Grinberg, A. Iwasaki, M. Kmiec, L. Kornak,
    B. LeBlanc, P. Lesniewski, P. Misra, C. Mitchell,
    R.J. Nicolalde, B. Pass, A. Ruuge, D.A. Schauer,
    J. Smirniotopoulos, A. Sucheta, T. Walczak

49
Disclaimer
  • The views expressed in this presentation are
    those of the author and do not reflect the
    official policy or position of the Navy and
    Marine Corps Public Health Center, Navy Bureau of
    Medicine and Surgery, Department of the Navy,
    Department of Defense, or the U.S. Government.

50
www.Biodose-2008.org
Write a Comment
User Comments (0)
About PowerShow.com