Title: Visible Spectrophotometer
1Visible Spectrophotometer
2Visible Spectrophotometer
3Visible Spectrophotometer
100 Adjust
I0
I
white light
detector/recorder
100 adjust blank distilled water
prism
I0 I transmittance (I / I0) x 100
100 absorbance 0
4Visible Spectrophotometer
Zero Adjust
I0
I
white light
detector/recorder
zero adjust (barrier inserted)
prism
I 0 transmittance 0 absorbance maximum
5Visible Spectrophotometer
Sample
I0
I
white light
detector/recorder
sample solution
prism
I lt I0 transmittance (I / I0) x 100
50 absorbance 1.7
6Visible Spectrophotometer
I0
I
white light
detector/recorder
sample solution
prism
I0 I transmittance (I0 / I) x 100
100 absorbance 0
In order for blue solution to absorb light,
wavelength must be changed to blues complement
7Visible Spectrophotometer
white light
I0
I
detector/recorder
sample solution
prism
Wavelength is changed by either rotating prism
or position of slit
8Spec 20 Visible Spectrophotometer
digital
analog
91. Adjust wavelength
102. Zero adjust with sample compartment empty
0 T
113. 100 adjust with distilled water (blank)
in sample compartment 100 transmittance
124. Record sample absorbance (Be certain to
switch MODE to absorbance)
13Beers Law
- As the intensity of the color of the sample
solution increases - the absorbance increases.
- A ? b c
absorbance
constant
path length
concentration
14Determination of the Concentration of an Unknown
Solution
- Prepare a set of standard solutions of known
concentrations - Select the appropriate wavelength
- Determine the absorbance of each
- Determine the absorbance of the unknown solution
15(No Transcript)
1613 ppm
17Recall Beers Law
A ? b c
0
y mx b
.0051 0.00037 x C 0.00030
13.0