Title: How to programme and follow up a pacemaker BCS Glasgow 2006 Technical Aspects of a Pacemaker
1How to programme and follow up a pacemakerBCS
Glasgow 2006 Technical Aspects of a Pacemaker
- Mr Stuart Allen.
- Southampton General Hospital.
2(No Transcript)
3Goals of a Pacemaker ClinicTechnical
Physiological
- Satisfy BPEG/HRuk Guidelines
- Electrical characteristics/Battery Check
- Histograms/Diagnostics/ Arrhythmia
analysis/evaluation - Physiological needs of patient
- Prediction of potential problems
Clinical/electrical - Troubleshooting
- Majority of patients fall in to the category of
routine follow up
4Common Referrals for Pacemaker Checks
- My SHO/Registrar told me to ask for it!
- No Pacing spikes on ECG.
- Patient has a pacemaker implanted. Can we check
it? (no other reason given) - Syncope/Dizzy Spells/Palpitations. ?Pacemaker
working. - If the ECG shows normal pacing and the patient
is free of any symptoms there is unlikely to be a
pacemaker/lead problem - Pre/ Post surgery
- Augment cardiac output
- .
5Ask the right questions!
- Pacemakers can help answer key clinical questions
regarding device status, clinical status and
arrhythmia status of the patient. - Pacemakers can give useful insight into disease
progression and long-term patient management
questions.
6Most Pacemakers Perform Four Functions
- Stimulate cardiac depolarization
- Sense intrinsic cardiac function
- Respond to increased metabolic demand by
providing rate responsive pacing - Provide diagnostic information stored by the
pacemaker
7Potential Problems Identifiable on an ECG Can
Generally Be Assigned to Five Categories
- Undersensing
- Oversensing
- Noncapture
- No output
- (Pseudomalfunctions)
8Paced Rhythm Recognition
9Electrical Characteristics
- Lead threshold(s)
- Lead Impedance(s)
- Sensing threshold(s)
- Battery voltage/Current drain
- Battery impedance
- Longevity
10Typical Follow-Up Activities
Percent of Follow-Up Time
Marshall M, Butts, L, Flaim, G, et al. Predictors
of time requirements for pacemaker clinic
evaluation. PACE 1995 18 Pt II952.
11(No Transcript)
12Stimulation Threshold
2.0 v
1.5 v
1 v
- The minimum output pulse neededto consistently
capture the heart
13Threshold Trending
14Sensitivity
- The degree that the pacing system sees or
senses signals, controlled by the sensitivity
setting which is graduated in millivolts (mV)
Sensitivity (mV)
5 (mV)
2.5 (mV)
1.25 (mV)
15Undersensing
- An intrinsic depolarization that is present, yet
not seen or sensed by the pacemaker
Atrial Undersensing
16Undersensing May Be Caused By
- Inappropriately programmed sensitivity
- Lead dislodgment
- Lead failure
- Insulation break conductor fracture
- Lead maturation
- Change in the native signal
17Oversensing
- The sensing of an inappropriate signal
- Can be physiologic or non-physiologic
18Noncapture May Be Caused By
- Lead dislodgment
- Lead perforation
- Low output
- Lead maturation
- Poor connection in the header
- Lead failure
19Less Common Causes of Noncapture May Include
- Twiddlers syndrome
- Electrolyte abnormalities
e.g. hyperkalemia - Myocardial infarction
- Drug therapy
- Battery depletion
- Exit block
20Non Capturedue to Atrial Lead Displacement
21Lead Impedance Measurement
- High Lead Impedance
- Open Circuit e.g. lead not connected to device
(Set Screw) - Lead fracture
- Low Lead Impedance
- Partial lead/insulation break fluid ingress
- Fluid/blood in header
22An Insulation Break Around the Lead Wire Can
Cause Impedance Values to Fall
- Insulation breaks expose the wire to body fluids
which have a low resistance and cause impedance
values to fall - Current drains through the insulation break into
the body which depletes the battery - An insulation break can cause impedance values to
fall below 300 ohms
23A Wire Fracture Within the Insulating Sheath May
Cause Impedance Values to Rise
- Impedance values across a break in the wire will
increase - Current flow may be too low to be effective
- Impedance values may exceed 3,000 ohms
24Normal Impedance Trend
25High Lead Impedance indicating possible lead
fracture
26Pseudomalfunctions
- Pseudomalfunctions are defined as
- Unusual
- Unexpected
- ECG findings that appear to result from pacemaker
malfunction but that represent normal pacemaker
function eg Pacemaker Mediated Tachycardia (PMT).
27Pseudomalfunctions Pacemaker Mediated Tachycardia
PVC
Retrograde P waves
PVARP
PVARP
PVARP
PVARP
28Pacemaker DiagnosticsLead Monitoring
- Automatically included on this initial
interrogation report
Measured Data includes Lead impedance lead Status
29Diagnostics, Histograms Arrhythmia analysis
- Majority of pacemaker
- patients fall into the category of routine
follow-up. - Routine Follow-up
- Interrogate Review
- Remaining Longevity
- A and V Thresholds
- A and V Impedances
- P- and R-Wave Ampl
- Histograms and Pace/Sense History
30Atrial High Rate Episode
- AF Verified
- Effect on Ventricular Rate suspected
31Ventricular Rate Histogram DuringAtrial High
Rate Episodes
- Evaluate
- Rate Distribution
- Time in High Rate
- V-pace vs. V-sense
- Ventricular Histogram
32Patient Presenting with TIA12 lead ECG shows
NSR/ 24hr Holter monitor NAD
33Patient Presenting With Syncope
34Physiological Programming
- Rate Rate Response
- Promotion of intrinsic conduction
- Anti Arrhythmic functions
35Rate Responsive Pacing
- When the patients exercise increases, the
pacemaker ensures that the heart rate increases
to provide additional cardiac output
Adjusting Heart Rate to Activity
Normal Heart Rate
Rate Responsive Pacing
Fixed-Rate Pacing
Daily Activities
36A Variety of Rate Response Sensors Exist
- Those most accepted in the market place are
- Activity sensors that detect physical movement
and increase the rate according to the level of
activity - Minute ventilation sensors that measure the
change in respiration rate and tidal volume via
transthoracic impedance readings - Blended sensor of activityMV
- QT interval
- Sensitivity of the Rate Response sensor can be
set in each patient
37Why Promote Intrinsic Conduction
- Right ventricular (RV) pacing alters the normal
myocardial electrical activation sequence leading
to regional wall motion abnormalities of the left
ventricle (LV). - Histopathologic abnormalities of the myocardium
have been identified with prolonged ventricular
pacing of the mature and immature heart.
38Promoting Intrinsic Conduction
- Programming AV interval (Search AV)
39Arrhythmia Management
- Atrial Pacing Preference for suppression of APCs
that may cause AF - Anti Tachycardia pacing for Atrial flutter
- Ventricular Rate Regulation for AF
40The future.Remote Patient Management
- Internet-based system providing full device
interrogation data - All programmed parameters
- Stored episodes with electrograms
- 10 second electrogram captured at interrogation
- All reports available on programmer available for
display and printing
Secure Network
CareLink? Monitor
Patient Website
41Thank You