How to programme and follow up a pacemaker BCS Glasgow 2006 Technical Aspects of a Pacemaker - PowerPoint PPT Presentation

1 / 37
About This Presentation
Title:

How to programme and follow up a pacemaker BCS Glasgow 2006 Technical Aspects of a Pacemaker

Description:

Syncope/Dizzy Spells/Palpitations. ? Pacemaker working. ... Patient Presenting With Syncope. S ALLEN 06. Physiological Programming. Rate & Rate Response ... – PowerPoint PPT presentation

Number of Views:1006
Avg rating:3.0/5.0
Slides: 38
Provided by: stuart80
Category:

less

Transcript and Presenter's Notes

Title: How to programme and follow up a pacemaker BCS Glasgow 2006 Technical Aspects of a Pacemaker


1
How to programme and follow up a pacemakerBCS
Glasgow 2006 Technical Aspects of a Pacemaker
  • Mr Stuart Allen.
  • Southampton General Hospital.

2
(No Transcript)
3
Goals of a Pacemaker ClinicTechnical
Physiological
  • Satisfy BPEG/HRuk Guidelines
  • Electrical characteristics/Battery Check
  • Histograms/Diagnostics/ Arrhythmia
    analysis/evaluation
  • Physiological needs of patient
  • Prediction of potential problems
    Clinical/electrical
  • Troubleshooting
  • Majority of patients fall in to the category of
    routine follow up

4
Common Referrals for Pacemaker Checks
  • My SHO/Registrar told me to ask for it!
  • No Pacing spikes on ECG.
  • Patient has a pacemaker implanted. Can we check
    it? (no other reason given)
  • Syncope/Dizzy Spells/Palpitations. ?Pacemaker
    working.
  • If the ECG shows normal pacing and the patient
    is free of any symptoms there is unlikely to be a
    pacemaker/lead problem
  • Pre/ Post surgery
  • Augment cardiac output
  • .

5
Ask the right questions!
  • Pacemakers can help answer key clinical questions
    regarding device status, clinical status and
    arrhythmia status of the patient.
  • Pacemakers can give useful insight into disease
    progression and long-term patient management
    questions.

6
Most Pacemakers Perform Four Functions

  • Stimulate cardiac depolarization
  • Sense intrinsic cardiac function
  • Respond to increased metabolic demand by
    providing rate responsive pacing
  • Provide diagnostic information stored by the
    pacemaker

7
Potential Problems Identifiable on an ECG Can
Generally Be Assigned to Five Categories
  • Undersensing
  • Oversensing
  • Noncapture
  • No output
  • (Pseudomalfunctions)

8
Paced Rhythm Recognition
9
Electrical Characteristics
  • Lead threshold(s)
  • Lead Impedance(s)
  • Sensing threshold(s)
  • Battery voltage/Current drain
  • Battery impedance
  • Longevity

10
Typical Follow-Up Activities
Percent of Follow-Up Time
Marshall M, Butts, L, Flaim, G, et al. Predictors
of time requirements for pacemaker clinic
evaluation. PACE 1995 18 Pt II952.
11
(No Transcript)
12
Stimulation Threshold
2.0 v
1.5 v
1 v
  • The minimum output pulse neededto consistently
    capture the heart

13
Threshold Trending
14
Sensitivity
  • The degree that the pacing system sees or
    senses signals, controlled by the sensitivity
    setting which is graduated in millivolts (mV)

Sensitivity (mV)
5 (mV)
2.5 (mV)
1.25 (mV)
15
Undersensing
  • An intrinsic depolarization that is present, yet
    not seen or sensed by the pacemaker

Atrial Undersensing
16
Undersensing May Be Caused By
  • Inappropriately programmed sensitivity
  • Lead dislodgment
  • Lead failure
  • Insulation break conductor fracture
  • Lead maturation
  • Change in the native signal

17
Oversensing
  • The sensing of an inappropriate signal
  • Can be physiologic or non-physiologic

18
Noncapture May Be Caused By
  • Lead dislodgment
  • Lead perforation
  • Low output
  • Lead maturation
  • Poor connection in the header
  • Lead failure

19
Less Common Causes of Noncapture May Include
  • Twiddlers syndrome
  • Electrolyte abnormalities
    e.g. hyperkalemia
  • Myocardial infarction
  • Drug therapy
  • Battery depletion
  • Exit block

20
Non Capturedue to Atrial Lead Displacement
21
Lead Impedance Measurement
  • High Lead Impedance
  • Open Circuit e.g. lead not connected to device
    (Set Screw)
  • Lead fracture
  • Low Lead Impedance
  • Partial lead/insulation break fluid ingress
  • Fluid/blood in header

22
An Insulation Break Around the Lead Wire Can
Cause Impedance Values to Fall
  • Insulation breaks expose the wire to body fluids
    which have a low resistance and cause impedance
    values to fall
  • Current drains through the insulation break into
    the body which depletes the battery
  • An insulation break can cause impedance values to
    fall below 300 ohms

23
A Wire Fracture Within the Insulating Sheath May
Cause Impedance Values to Rise
  • Impedance values across a break in the wire will
    increase
  • Current flow may be too low to be effective
  • Impedance values may exceed 3,000 ohms

24
Normal Impedance Trend
25
High Lead Impedance indicating possible lead
fracture
26
Pseudomalfunctions
  • Pseudomalfunctions are defined as
  • Unusual
  • Unexpected
  • ECG findings that appear to result from pacemaker
    malfunction but that represent normal pacemaker
    function eg Pacemaker Mediated Tachycardia (PMT).

27
Pseudomalfunctions Pacemaker Mediated Tachycardia
PVC
Retrograde P waves
PVARP
PVARP
PVARP
PVARP
28
Pacemaker DiagnosticsLead Monitoring
  • Automatically included on this initial
    interrogation report

Measured Data includes Lead impedance lead Status
29
Diagnostics, Histograms Arrhythmia analysis
  • Majority of pacemaker
  • patients fall into the category of routine
    follow-up.
  • Routine Follow-up
  • Interrogate Review
  • Remaining Longevity
  • A and V Thresholds
  • A and V Impedances
  • P- and R-Wave Ampl
  • Histograms and Pace/Sense History

30
Atrial High Rate Episode
  • AF Verified
  • Effect on Ventricular Rate suspected

31
Ventricular Rate Histogram DuringAtrial High
Rate Episodes
  • Evaluate
  • Rate Distribution
  • Time in High Rate
  • V-pace vs. V-sense
  • Ventricular Histogram

32
Patient Presenting with TIA12 lead ECG shows
NSR/ 24hr Holter monitor NAD
33
Patient Presenting With Syncope
34
Physiological Programming
  • Rate Rate Response
  • Promotion of intrinsic conduction
  • Anti Arrhythmic functions

35
Rate Responsive Pacing
  • When the patients exercise increases, the
    pacemaker ensures that the heart rate increases
    to provide additional cardiac output

Adjusting Heart Rate to Activity
Normal Heart Rate
Rate Responsive Pacing
Fixed-Rate Pacing
Daily Activities
36
A Variety of Rate Response Sensors Exist
  • Those most accepted in the market place are
  • Activity sensors that detect physical movement
    and increase the rate according to the level of
    activity
  • Minute ventilation sensors that measure the
    change in respiration rate and tidal volume via
    transthoracic impedance readings
  • Blended sensor of activityMV
  • QT interval
  • Sensitivity of the Rate Response sensor can be
    set in each patient

37
Why Promote Intrinsic Conduction
  • Right ventricular (RV) pacing alters the normal
    myocardial electrical activation sequence leading
    to regional wall motion abnormalities of the left
    ventricle (LV).
  • Histopathologic abnormalities of the myocardium
    have been identified with prolonged ventricular
    pacing of the mature and immature heart.

38
Promoting Intrinsic Conduction
  • Programming AV interval (Search AV)

39
Arrhythmia Management
  • Atrial Pacing Preference for suppression of APCs
    that may cause AF
  • Anti Tachycardia pacing for Atrial flutter
  • Ventricular Rate Regulation for AF

40
The future.Remote Patient Management
  • Internet-based system providing full device
    interrogation data
  • All programmed parameters
  • Stored episodes with electrograms
  • 10 second electrogram captured at interrogation
  • All reports available on programmer available for
    display and printing

Secure Network
CareLink? Monitor
Patient Website
41
Thank You
Write a Comment
User Comments (0)
About PowerShow.com