Title: Fundamentals of Mass spectrometry
1Fundamentals of Mass spectrometry
February 6, 2008 Case Center for Proteomics
2What is mass spectrometry?
- Mass spectrometry is an instrumental approach
that allows for the mass measurement of ions
derived from molecules. - Mass spectrometers are capable of forming,
separating and detecting molecular ions on their
mass-to-charge ratio.
3Sample Introduction
Sample
4Vacuum in MS
5Sample Introduction - Ionization tasks
- Need to transfer sample from atmospheric pressure
to vacuum - Samples solid, liquid, gas
- Need to create ions in the gas phase
- Gas ionize
- Liquid evaporate, than ionize OR
- both at the same time (spray)
- Solid sublime, than ionize OR
- both at the same time (desorption)
6- Gaseous samples
- Easy to transfer from A.P. to vacuum
- Liquid/solid (high volatility) vapor can be used
as well must be free of air (freeze-pump-thaw
cycles) - Controlled flow leak valves, capillaries
- Mixed with carrier gas
- GC interface (GC/MS)
7- Liquid and solid samples
- Liquid samples analyte solutions spray
ionization techniques - Solid samples Solids probe
- Low volatility can be put in the vacuum
- Vacuum lock to keep high vacuum intact
- Evaporation by heating, then ionization OR
- Desorption by laser / particle beam
8Once analyte molecules are in the gas phase they
can easily be ionized
- Electron Impact (EI) and Chemical Ionization (CI)
- FAB-fast atom bombardment ionization
- MALDI-TOF
- ESI
- Electron-Capture ionization
9(No Transcript)
10EI(top) vs CI of ephedrine
11EI and CI Summary
- EI gives extensive fragmentation (70eV) but has
extensive database for fingerprinting - CI is more suitable for molecular weight
determination but structural info is limited - Alternate EI/CI can combine advantages
- Excellent sensitivity and natural GC/MS
interface - Major limitation MW lt 1000 Da, volatility must
be high, not very suitable for most biological
compounds - See Interpretation of Mass Spectra by
McLafferty and Turecek (FRL) for EI/CI of organic
molecules
12Fast Atom Bombardment (FAB)
To mass analyzer
13FAB
- Higher mass range (7 kDa)
- Easy to get molecular ion MH or (MNa), little
fragmentation - Can be done on Hi-Res mass spectrometers
- Both positive and negative ion modes possible
- Bad sensitivity, especially at high masses
- High background of matrix peaks
- Sample must be soluble in matrix
- Not very useful for non-polar species
- Was a big step up from EI/CI for the analysis
of - biomolecules, but now is almost obsolete
14 Matrix-Assisted Laser Desorption/Ionization
(MALDI)
MH
15MALDI
- Tanaki, Chemistry Nobel Prize 2002 (shared)
- Karas Hillenkamp, 1985-1988
- Laser UV _at_ 337, 355 or 266 nm
- or IR _at_ 3 or 10 mm
- Matrix must absorb well in UV (IR) region
- Subst. aromatic acids, alcohols
- Sample is mixed with matrix solution and dried,
almost anything can be analyzed, including very
high MW compounds
16Typical MALDI matricies
- UV MALDI
- Sinapinic acid (SA) 3,5-Dimethoxy-4-hydroxycinnam
ic acid - DHB 2,5-Dihydroxybenzoic acid
- CHCA a-Cyano-4-hydroxycinnamic acid
- Ferulic acid (FA) 4-Hydroxy-
- 3-methoxycinnamic acid
- p-Nitrophenol
- Solvent
- Water/ ACN in diffferent proportions (0-100)
- Sometimes Ethanol or other organic solvents
17Merits of MALDI
- Low resolution
- Matrix background for MW lt 1000 Da
- (DIOS no matrix
- and 700 amol sensitivity !)
- Other surfaces are being tested
- High accessible mass range (up to 1,000,000)
- Femtomole and better sensitivity
- Soft ionization almost no fragmentation
- Very good for mixture analysis, can work in the
presence of salts
18Electrospray Ionization (ESI)
lt- 5 kV dif. -gt
Analyte solution
Analyte solution, 5 mL/min
-gtIons to MS
Fenn and others, 1984 Fenn, Nobel Prize 2002
(shared)
19Ion Desolvation in ESI
20Comparison of Ionization Methods
21Mass Analysis
Sorting of Ions
Ion Detection
Ion Source
Mass Analyzer
Based on their m/z
Mass Spectrum
First mass analyzers early 1900s
22Performance of mass analyzer
- Mass range (low to high m/z)
- Resolution (separation of two adjacent peaks in
the mass spectrum) - Detection sensitivity (lowest amount of analyte
that can be detected) - Scan speed (time required to scan all ions in the
m/z range of interest)
23Resolution
(Resolving Power) R M/DM Where
DMM2-M1 (2 peaks) or FWHM (1 peak)
M2
M1
24Types of mass analyzers
- Magnetic / Sector
- Quadrupole (Q)
- Quadrupole Ion Trap (IT), Linear Trap (LT)
- Time-of-Flight (TOF)
- Fourier-Transform Ion Cyclotron Resonance (FT-ICR
or FT-MS) - Ion Mobility
25Magnetic mass analyzer
Kinetic energy zV ½ mv2 In magnetic field
mv2/r zvB Combine m/z B2r2/2V
Alternatively, with a fixed slit position, Scan
magnetic field B gt mass spectrum
26Merits of sector instruments
- Benefits
- High resolution (gt100, 000)
- Good mass range (gt15,000 m/z)
- Reasonable scan speed and dynamic range, good
mass accuracy - High sensitivity
- CID MS/MS spectra are very reproducible
- Limitations
- Huge cost, hard to operate
- Low ion transmission efficiency
- Applications
- All organic MS analysis methods
- Accurate mass measurements
- Quantitation
- Isotope ratio measurements
27Quadrupole mass filter
Combination of DC and RF voltages on 4
rods Allows to pass ions of only certain m/z
Scan RF field to obtain MS
28RF-only Quadrupole
- In the absence of DC field, quadrupole operates
as a wide-band mass filter. It can be used as an
ion guide or reaction chamber - Higher-order multipoles (hexapoles, octopoles,
etc.) are used for ion-guiding and
ion-accumulating purposes
DC-only Quadrupole
- In the absence of RF field, quadrupole operates
as a lens element in some ion optics designs.
29Quadrupole features
- Benefits
- Upper mass limit up to 4000 m/z
- m 0.136 V / (f2r2)
- Resolution typically low (unit mass) R ? mf2L2/
(zV) - Can tolerate pressure up to 5x10-5 Torr
- Good for GC, LC, CE interfaces
- Natural analyzer for ESI sources
- Low cost gt most popular mass analyzer
- Limitations
- Limited resolution
- Peak heights variable as a function of mass
- CID depends on energy, collision gas, pressure
30Quadrupole Ion Traps
Paul,Steinwedel 1953
Principle similar to Quadrupole MS
Ions are trapped in mTorr of He
-V pulses open V closed to inject ions
31IT Modes of operation
Ion storage Ring electrode RF only Endcaps
grounded Ions of all m/z are trapped
m/z 4V / (qmaxw2r02)
32Trapping Ions in the Axial (Z) Direction
3 130 VDC
Creates a Potential Energy Well to Confine the
Ions
33Heliums Role as a Damping Gas
Potential Well
34The Effects of No Helium
Helium flowing into trap
Helium shut off and not flowing into trap
35Linear Ion Trap Stability Diagram
- Ions stability depends upon
- m/z
- Frequency of fundamental RF
- Amplitude of the V on the ring electrode
- Ion trap size
1.0
36Mass Selective Instability
q - Axis
At a particular RF voltage, every mass moves
with a specific (secular) frequency
q .908
- Ramp RF, Ions Leave Low m/z to High m/z
- CID RF resonance excitation voltage to end caps
electrodes to enable energetic collisions with He
37Ion Traps merits
- Easily coupled to EI/CI (GC) ESI (LC, CE)
through ion guides other mass analyzers (TOF,
Ion Mobility) - Low cost, good sensitivity
- R5,000 (gt30,000 at low scan speeds)
- Mass range 2-4,000 m/z (gt20,000)
- Ion chemistry, MSn capability
38Fourier-transform Ion Cyclotron Resonance Mass
Spectrometer
Another type of ion-trapping MS
Ions are trapped by combination of electric(DC)
and magnetic fields
Ions can be excited by applying RF voltage to
transmitter plates Ion selection, MSn, Ion
detection
39Ion excitation
w B z/m
40Ion Detection in FT-MS
Unique type of ion detection, one measures not
the ion but image current. Non-destructive !
Receiver plate
Receiver plate
41Fourier transformation
Pulse covers whole frequency range
- B z/m or
- f 15357 B z/m
- kHz T
42Benefits High Resolution
R lt B T z/m
43FT-MS figures of merit
- Ultrahigh resolution (gt1,000,000 for
biomolecules, 108 achieved for m/z18) - Unsurpassed mass accuracy, good mass range
(gt20,000 m/z with 4.7 T magnet) - Extremely versatile in terms of ion chemistry
(MSn, dissociation, ion-molecule reactions), ions
can be trapped for days! - Cost more than one would like
- Require low pressures (lt10-7 Torr)
- Supercon. magnets require space/LHe/LN2
44Time-of-Flight MS
mv2/2 zV v (2zV/m)1/2 vL/t t L/v
L(m/2zV)1/2
Naturally couples to MALDI, can be adopted to ESI
45Resolution in TOF
Rm/Dmt/Dt
Problems Temporal and spatial ion distribution
Use short pulse and high extraction V (gt20 kV) R
used to be lt500, now lt 5000
46Reflectron TOF
(reflectron)
47Delayed Extraction
In a regular TOF experiment, ions are extracted
and detected promptly before they can fragment
gt Lack of structural information Ion extraction
can be time-delayed (DE) to give the ions time
to fragment. (Post-source decay PSD).
Fragments. DE not only allows detection of
source-formed fragment ions but also improves
mass accuracy and resolution.
48Merits of TOF
- Virtually unlimited mass range
- Very high scan speed (as low as 25 ms)
- High ion transmission, low detection limits
- Simple design, low cost
- Decent resolution
- Pulsed mode, hard to
- interface with ESI
49TOF summary
- Unlimited mass range in linear mode
- Resolution 3000 (L) 10,000 (R), more
- Extremely quick scan
- Excellent sensitivity
- Simple design, portability
- DE for structure elucidation
- OA for ESI interface
50Mass Analyzers Comparison