Title: Ingen bildrubrik
17 YEARS EXPERIENCES WITH HIGHLY CROSS-LINKED PE
Two Randomized studies using radiostereometry
Johan Kärrholm, Georgios Digas, Peter
Herberts, Per-Erik Johansson, Jonas Thanner
Department of Orthopaedics, Sahlgrenska
Hospital, Gotheburg University, SE-431 80
Mölndal, Sweden. Division of Orthopaedics,
General Hospital of Xanthi Greece
2highly cross-linked polyethylene
- - First trial dates back to the 70-ties
- In general practice since for about 10 years
- Low wear rates up to 5-7 years
Oonishi and Kadoya 2000
Bragdon et al 2007, Digas et al 2003, 2004,
2007 Dorr et al 2007 Engh et al 2006 Röhrl et al
2007.
3Present studies - both randomized
control or study
dec. 1998 cemented cup, unilateral operation
oct. 2000 polyliners in uncemented cups,
bilateral operation
control study
- highly cross-linked vs. standard irradiated
polyethylene
4control or study
Cemented study
- 30 male, 30 female 61 hips
- 55 years (median, range35-70)
- 49 primary arthrosis
- 12 sec. Arthrosis (non inflammatory)
5Cups made of compression molded GUR 1050
DURASULTM highly crosslinked PE - study group
- - Heated (1250) in air
- - Irradiated at 1250C - absorbed dose 9,5 Mrad
- - Melt - annealed at 1500
- - EtO sterilized
Muratoglu and Harris ORS 1999
SULENETM - control group
- Gamma irradiated 2,5-4,0 Mrad
6Study design
Randomized trial Cemented arthroplasty
- Weber cup (Sulzer)
- Spectron stem (Smith Nephew)
- cobalt-chromium head (Ø 28 mm)
Ra 2.8mm
Ra 0.7mm
7Methods
Radiostereometry (RSA) Bone mineral density
(DEXA) Conventional radiography Harris Hip
Score Activity rating
8Radiostereometry
Supine Postop, 3, 6 months, 1, 2, 3, 5, 7
years Standing 3, 6 months, 1, 2, 3, 5, 7 years
9(No Transcript)
10Cemented study at 7 years
57 attended 1 revised at 5.5 (and one at 8 yrs)
(loose stems) 1 declined (not revised) 2
deceased with implants in situ
11Cemented study
RSA results up to 7 years head penetration
proximal
3D
12Cemented study
Proximal penetration up to 7 yrs (mean, SE 7 yrs
plt0.0001)
133-5 5-7
p lt 0.001
p 0.008
5-7 years wear in both groups but sign.
difference 3 - 5 yrs vs. 5 - 7 yrs only in
control group (0.04 vs 0.5) (Wilcoxon
signed rank test)
14Cemented study
Proximal penetration 3 m. - 7 yrs (mean SE 7
yrs plt0,001)
15p lt 0.001
p 0.3
increase of wear only in control group 5 - 7
years (Wilcoxon signed rank test)
- but 10 missing observations.
16Cemented study
Total (3D) penetration up to 7 yrs (mean SE 7
yrs plt0.0001)
p lt 0.0001
p 0.011
172 revisions
Revised at 5.5 yrs Revised at 8 yrs (stem and
cup) (stem)
RSA data
Prox wear 0.09 mm (Durasul) 0,02
mm (Durasul) Prox cup mig. 0.34 mm
-0.05 mm Stem subsidence 6.6 mm
4.4 mm
18(No Transcript)
19Cemented study
Proximal-Distal migration mean SE, 7 yrs p0,8
20Cemented study
Harris hip score1
Mann-Whitney test pgt0.1
difference preop - 7 yrs
total 7 yrs
95
38
Durasul
63-100
22-77
97
47
Control
61-100
12-81
1median values
21 Hybrid study
control
study side side
- 32 patients, 64 hips
- 20 female, 12 male
- median age 48 years (29 - 70)
- median weight 75 kg (51 - 97)
- 17 bilat prim. OA, 15 bilat sec. OA
-
patients
22Liners made of GUR 1050
LONGEVITYTM highly crosslinked PE
- electron beam irradiation - 10 MRad at room
- temperature
- melt-annealed at 150º
- gas plasma sterilized
CONTROLS
- Gamma irradiated at 2,5 Mrad in inert
atmosphere
differs from Durasul
23STUDY DESIGN
Hybrid arthroplasty
- Trilogy cup (Zimmer)
- Spectron stem (Smith Nephew)
- 28 mm cobalt-chromium head
24STUDY DESIGN
- bilateral one - stage procedure
- most painful side - randomized
- least painful side - different polyethylene
25METHODS
radiostereometry (RSA) radiography
bone mineral density (DEXA)
Harris hip score, MACTAR, SF 36
activity rating
26Radiostereometry
Supine Postop, 3, 6 months, 1, 2, 3, 5, 7
years Standing 3, 6 months, 1, 2, 3, 5, 7 years
27(No Transcript)
28Hybrid study at 7 years
- 26 attended (52 hips)
- 3 delayed (not revised)
- 2 declined (not revised)
- deceased with implants in situ
25 /20 (supine/standing) cases with bilat.
successful RSA
29Hybrid study
RSA results head penetration
proximal
3D
30Hybrid study
Proximal penetration mean SE 7 yrs plt0.00001
313-5 5-7
p lt 0.00004
p 0.002
5-7 years wear in both groups but sign.
difference 3 - 5 yrs vs. 5 - 7 yrs only in
control group (0.005 vs 0.5) (Wilcoxon
signed rank test)
32Hybrid study
Total (3D) penetration Mean SE 7 yrs plt0.00001
p lt 0.0001
p 0.06
33Hybrid study
Proximal penetration mean SE 7 yrs plt0.00003
p 0.03
p 0.8
34RESULTS - RSA Cup migration
supine examinations
35Hybrid study
Proximal cup migration mean SE 7 yrs p0.3
36Hip Score and pain scores 7 yrs and 0 to 7 yrs.,
median, range
Hip Score 100 88-100 diff 0 7 yrs 57 31 - 83
37Conclusions after 7 years
- highly x-linked reduced head penetration
- with 55-80
- there was no consistent influence on cup
- migration
38Conclusion
Lossening of a cemented rough stem had no
measureable effect on wear (2 cases) of highly
x-linked cemented cups
39Conclusion
- this quality of highly cross-linked poly
- not resistant to wear/creep
- past 3 years - no certain evidence for
- accelerated wear/creep
40CONCLUSION
- The crosslinked materials studied
- has hitherto worked well.
- They should be
- continuously studied for another decade
- used as judiciously as we learnt from
- previous experiences with polyethylene.
41