Title: A Patient Safety Initiative For Insulin Pumps
1A Patient Safety Initiative For Insulin Pumps
Manufacturing Standards to improve insulin pump
safety and medical outcomes Suggestions for
improvements and editorial changes are welcome.
Send comments to John Walsh, PA, CDE at
jwalsh_at_diabetesnet.com and Ruth Roberts, MA
rroberts_at_diabetesnet.com (619) 497-0900
Version 2 09/28/09Version 1 12/07/08
2Definitions
- TDD total daily dose of insulin (all basals and
boluses) - Basal background insulin pumped slowly through
the day to keep BG flat - Bolus a quick surge of insulin as
- Carb boluses to cover carbs
- Correction boluses to lower high readings that
arise from too little basal insulin delivery or
insufficient carb boluses - Bolus On Board (BOB) the units of bolus insulin
with glucose-lowering activity still working from
recent boluses - Duration of Insulin Action (DIA) time that a
bolus will lower the BG. This is used to
calculate BOB.
3Background
- Although insulin pumps were first designed to
improve insulin delivery, new technology has
transformed them into data collection centers.
Additional data from continuous glucose
monitoring devices increases the datas value. - Data necessary for monitoring and clinical
decisions is now directly accessible in the pump
itself. With regular glucose testing and data
analysis, a pump can provide important clinical
information to users and clinicians regarding
glucose and insulin dosing patterns with
recommendations for improvement.
Pump collectively refers to the pump body,
PDA, cell phone, and other devices that control
insulin delivery and store data.
4Introduction
- Over 500,000 insulin pumps are in use around the
world, yet there are no formal guidelines
regarding manufacturing standards and medical
practice in the diabetes clinical community and
pump manufacturing industry. - These suggested standards are intended as
imprements for future insulin pumps and in
current pumps where software changes allow. - They are designed to
- Provide safer dosing increments and problem
solving to pump users - Allow clinicians to make safer and more informed
dosing decisions when managing a variety of pumps - Facilitate the training of adjunctive medical and
accessory personnel (ER, surgical, school nurses,
etc.) to improve their familiarity and
interactions with insulin pumps
5Benefits Of Insulin Pump Standards
- These mechanical standards are designed for
- Consistent use of pump settings between pump
manufacturers - Accuracy and safety of carb and correction factor
increments - Safety of DIA defaults and consistent use of DIA
increments - Consistent handling of BOB and insulin stacking
- Improved monitoring for hypoglycemia
hyperglycemia - Direct entry of glucose values into bolus
calculations - Notification when use of correction boluses is
excessive - Faster identification of control issues caused by
infusion sets
6Overview
Slides are numbered by topic for easy reference.
- Topic
- Carb Factor Increments
- Correction Factor Increments
- Carb Factor Accuracy
- Correction Factor Accuracy
- DIA Default Times
- DIA Time Increments
- Handling Of BOB
Topic 8. Multi-Linear And Curvilinear DIA
9. Hypoglycemia Alert 10. Hyperglycemia
Alert 11. Correction Bolus Alert 12. Insulin
Stacking Alert 13. Automatic Entry Of BG
Values 14. Infusion Set Monitoring
7Safety
- Increments and Accuracy
- Carb factor
- Correction factor
- DIA
- Defaults
- DIA
- Handling Of BOB
- Hypo Manager
- Topic
- Alerts
- Hypo frequency
- Hyper frequency
- Unusual highs
- Correction Bolus Excess
- Insulin Stacking
- Overview table of alert frequency
- 13. Infusion Set Monitor
8Easy Data Access
- BOB
- TDD
- Correction bolus
- Carb bolus
- Basal
- Avg glucose
- SD
- Carbs/day with approximate calorie equivalent
- Accurate carb and correction factors
- Automatic basal and factor testing
- How often bolus recommendation is overridden and
direction - Frequency of occlusions and stops (and length of
stops)
9ReviewChange in TDD Changes The BG
- For some later slides, it helps to know how a
change in the TDD affects the glucose levels. - Using a modified Davidson Rule for the carb
factor and a 2000 Rule for the correction factor - 1.25 change in the TDD changes glucose levels
about 25 mg/dl when given as a single dose - A 5 change in the TDD is equivalent to about a
25 mg/dl rise or fall in the glucose through the
day - A 5 to 6 change in the carb factor (approx. 2.5
to 3 of the TDD) changes the glucose about 20
mg/dl per meal.
10Carb Factor Increments
1
Carb Factor (CarbF) the number of grams of carb
covered by 1 u of insulin for an individual The
smaller the carb factor, the larger a carb bolus
111
Standard For Carb Factor Increments
- Carb factor increments shall be less than or
equal to 5 of the next larger whole number so
that each single step adjustment causes
subsequent carb boluses to change by no more than
5 from previous doses. - Recommended minimum carb factor increments 5
- 1.0 g/u above 20 g/u
- 0.5 g/u for 10 to 20 g/u
- 0.2 g/u for 5 to 9.8 g/u
- 0.1 g/u for 3 to 4.9 g/u
- 0.05 g/u for 0.1 to 2.95 g/u
5 Improved carb factor increments recommended by
Gary Scheiner, MS, CDE
12Carb Factor (CarbF) Increments
1
- Issue Existing carb factor increments are too
large. - Current carb factors are a safety concern because
they lack the precision required to avoid
excessive hyperglycemia and hypoglycemia,
especially for smaller carb factor numbers.
13ExampleCarb Factor Increments
1
- 1 gram per unit is the smallest CarbF increment
in most pumps. This increment is relatively large
for CarbFs lower than 15 or 20 g/u. - For instance, when the carb factor is reduced
from 10 to 9 g/u, all subsequent carb boluses are
increased by 11.1. A shift in the carb factor
from 1u/5g to 1u/4g causes each subsequent carb
bolus to increase by 25. - For most pump users, a change in carb factor of 5
to 6 causes the average glucose to rise or fall
by at least 20 mg/dl after all meals.
14ExampleImpact On BG From CarbF Adjustment
1
- Table shows average additional fall in glucose
when a carb factor is reduced from 10 gram/u to 9
gram/u.
More carbs greater glucose fall
Calculated as carbs in meal carbs in
meal X 1900 new carb
factor old carb factor TDD
15ReviewMedian Carb Factor
1
- In unpublished data from the Cozmo Data Analysis
Study for 135 pump users in good control (avg BG
144.0 mg/dl with 4.92 tests/day) - The average carb factor was 10.4 g/u
- Nearly all use a carb factor below 20 g/u
- Almost 50 of carb factors are 10.0 g/u or less.
- The smaller the carb factor, the more vulnerable
the pumper is to error when an increment is
changed
Authors unpublished data from subset of over
1,000 complaint-free insulin pumps turned in for
software upgrade in 2007.
16What Current Changes In CarbFs Do
1
- Table shows how subsequent carb boluses are
affected by a one-step reduction in the CarbF
using different CarbF increments. Yellow area
shows values for most current pumps. Green areas
show safer increments that impact subsequent
boluses less than 5.
17Correction Factor Increments
2
Corr Factor (CorrF) the number of mg/dl (mmol)
drop in glucose that is produced by 1 u of
insulin for an individual The smaller the corr.
factor, the larger a corr. bolus
182
Standard For Correction Factor Increments
- For similar reasons, correction factor increments
shall be 5 or less of the next larger whole
number so that each single step adjustment causes
subsequent correction boluses to change by no
more than 5 from previous doses. - Recommended minimum correction factor increments
- 5.0 mg/dl per u above 80 mg/dl per u
- 2.0 mg/dl per u for 40 to 78 mg/dl per u
- 1.0 mg/dl per u for 20 to 39 mg/dl per u
- 0.5 mg/dl per u for 10 to 19.5 mg/dl per u
- 0.2 mg/dl per u for 5 to 9.8 mg/dl per u
- 0.1 mg/dl per u for 3 to 4.9 mg/dl per u
- 0.05 mg/dl per u for 0.1 to 2.95 mg/dl per u
19Carb Factor Accuracy
3
Carb Factors are often not physiologically
appropriate
203
Standard For Verification Of Carb Factor Accuracy
- Insulin pump companies shall record and publish
each year the carb factors used in insulin pumps
returned for upgrade or repair. This report will
include sufficient numbers of pumps to ensure
statistical significance for commonly used carb
factors between 5 and 20 grams per unit. - This data is designed to ensure that pump
training and clinical followup are assisting in
the selection of accurate carb factors. - To improve accuracy of carb factors, efforts
shall be taken to better train clinicians and
users in appropriate selection of physiologic
carb factors.
21Personal Carb Factors
3
- Issue Many carb factor settings in todays
insulin pumps today are physiologically
inappropriate. - Inappropriate carb factors introduce a
significant source for error in carb bolus
calculations.
22Review Carb Factors In Use 1
3
10
7
115
20
- Carb factor settings from 405 Cozmo pumps do NOT
have a bell-shaped (physiologic) distribution.
Easy numbers 5, 10, 15, and 20 g/unit are
preferred.
1
23Correction Factor Accuracy
4
244
Standard For Verification Of Corr Factor Accuracy
- Insulin pump companies shall record and publish
each year the correction factors used in insulin
pumps returned for upgrade or repair. This report
will include sufficient numbers of pumps to
ensure statistical significance for commonly used
correction factors between 20 and 80 mg/dl per
unit. - This data is designed to ensure that pump
training and clinical followup are assisting in
the selection of accurate correction factors. - For more accurate selection of corr. factors,
efforts shall be taken to improve corr. factor
selection and to automate corr. factor testing.
25Personal Correction Factors
4
- Issue Many correction factors used in insulin
pumps today are poorly tuned to the users need.
This inaccuracy significantly magnifies other
sources of error in correction bolus
calculations.
26Review Correction Factors In Use 1
4
- Avg. correction factors in use for 452
consecutive Cozmo insulin pump downloads - Like carb factors, correction factors in use are
NOT bell-shaped or physiologic. A more accurate
choice of correction factors would create a
bell-shaped curve. - Users or clinicians appear to frequently select
magic numbers for correction factors.
10
7
115
20
1
27DIA Default Times
5
28Standard ForDIA Default Times
5
- Default duration of insulin action (DIA) times in
current pumps vary between 3 and 6 hours. In
bolus calculations, a short DIA time shall be set
no shorter than 4.5 hours in pumps that determine
DIA in a linear fashion and no shorter than 5
hours in pumps that determine DIA in a
curvilinear or multi-linear fashion for fast
insulins (lispro, aspart, and glulisine) in use
at this time.
29DIA Default Time Settings
5
- Issue DIA measures the glucose-lowering activity
of a carb or correction bolus over time. Current
default times for DIA range from 3 to 6 hours in
different pumps. - The DIA is often considered another tool to
improve control rather than being set at an
appropriate value and focusing on more
appropriate changes in basal rates or carb and
correction factors to improve control. - A DIA that is too short allows excess
unrecognized bolus insulin to accumulate, usually
in the afternoon and evening hours. - Example for a DIA of 2.5 hrs, a bolus given at 9
am appears to have no activity after 1130 am. If
a high BG occurs at 1145 am, more bolus than
needed will be given. At lunch, the bolus will be
excessive, regardless of the BG at that time,
creating stacking and a high likelihood of
hypoglycemia.
30ReviewHow Long Do Boluses Lower The BG?
5
- Numerous GIR studies show rapid insulins lower
the glucose for 5 hours or more. - With Novolog (aspart) at 0.2 u/kg (0.091 u/lb),
23 of glucose lowering activity remained after 4
hours.12 - Another study found Novolog (0.2 u/kg) lowered
the glucose for 5 hours and 43 min. /- 1 hour.13 - After 0.3 u/kg or 0.136 u/lb of Humalog (lispro),
peak glucose-lowering activity was seen at 2.4
hours and 30 of activity remained after 4 hours.
11 - These times would be longer if the unmeasured
basal suppression in pharmacodynamic studies were
accounted for.
11 From Table 1 in Humalog Mix50/50 product
information, PA 6872AMP, Eli Lilly and Company,
issued January 15, 2007. 12 Mudaliar S, et al
Insulin aspart (B28 Asp-insulin) a fast-acting
analog of human insulin. Diabetes Care 1999
221501-1506. 13 L Heinemann, et al Time-action
profile of the insulin analogue B28Asp. Diabetic
Med 199613683-684.
31ReviewShort DIAs Hide Bolus Insulin Activity
5
- A short DIA time hides true BOB level and its
glucose-lowering activity. This can be a safety
issue in that it - Leads to unexplained lows
- Leads to incorrect adjustments in basal rates,
carb factors, and correction factors - Causes users to start ignoring their smart
pumps advice - An inappropriately long DIA time overestimates
bolus insulin activity this leads to
underdosing rather than overdosing on subsequent
boluses. - DIA should be based on an insulins real action
time. - Do NOT modify the DIA time to fix a control
problem
32ReviewDuration Of Insulin Action (DIA)
5
Accurate bolus estimates require an accurate DIA.
DIA times shorter than 4.5 to 7 hrs may hide BOB
and its glucose lowering activity
Glucose-lowering Activity
6 hrs
2 hrs
0
4 hrs
33ReviewDIA
5
- Large doses (0.3 u/kg 30 u for 220 lb. person)
of rapid insulin in 18 non-diabetic, obese
people - Med. doses (0.2 u/kg 20 u for 220 lb. person)
- This shows a glucose lowering activity for 7-8
hours
Regular
Apidra product handout, Rev. April 2004a
34ReviewDoes Dose Size Affect DIA?
5
- This graphic suggests that smaller boluses do not
lower the BG as long as larger boluses. - However, this may not be true see next 2
slides. - Size of the injected Humalog dose for a 154 lb or
70 kg person - 0.05 u/kg 3.5 u
- 0.1 u/kg 7 u
- 0.2 u/kg 14 u
- 0.3 u/kg 21 u
Woodworth et al. Diabetes. 199342(Suppl. 1)54A
35ReviewPharmacodynamics Is Not DIA
5
- The DIA time entered into an insulin pump is
based on studies of insulin pharmacodynamics. - However, the traditional method used to determine
insulin pharmacodynamics may underestimate
insulins true duration of action, as shown in
the next two slides.
36ReviewPharmacodynamics Underestimates DIA And
Overestimates Impact Of Bolus Size
5
- To measure pharmaco-dynamics, glucose clamp
studies are done in healthy individuals who
receive an injection of fast insulin (0.05 to 0.3
u/kg) - Because there is no basal insulin replacement,
the injected insulin dose ALSO SUPPRESSES normal
basal release from the pancreas (grey area in
figure)
37ReviewPharmacodynamic Time Does Not Equal DIA
5
- After accounting for the lack of basal insulin
replacement, - True DIA times become longer than the PD times
derived in traditional research - If basal suppression activity is accounted for,
small boluses may be found to have a longer DIA
than it currently appears, erasingsome of the
apparent variation in DIA related to bolus size - Some of the apparent inter-individual variation
in pharmacodynamics may also disappear
38DIA Time Increments
6
396
Standard For DIA Time Increments
- For safe and accurate estimates of residual BOB,
DIA time increments shall be no greater than 15
minutes.
40DIA Time Increments
6
- Issue Current DIA time increments vary from 15
minutes to 1 hour in different pumps - When a pumps DIA time is adjusted, large time
increments, such as 1 hr, can introduce large
changes in subsequent estimates of BOB. - For example, when the DIA is reduced from 5 hours
to 4 hours, subsequent BOB estimates are
decreased and recommendations for carb boluses
are increased by about 25.
41ReviewGlucose Infusion Rate (GIR) Studies
6
- Most GIR studies suggest that pharmacodynamic
action of insulin varies only about 25 to 40
between individuals. - For a DIA time of 5 hr and 15 min, a 25 range is
equivalent to 1 hr and 20 min, such as from 4 hrs
and 30 min to 5hr and 50 min. - A pump that has 1 hr DIA increments would enable
the user to select only 1 or 2 settings within
this physiologic range, while a 30 min increment
would allow only 2 or 3 choices that are close to
a physiologic range.
42Handling Of BOB
7
aka insulin on board, active insulin, unused
insulin Introduced as Unused Insulin in 1st ed
of Pumping Insulin (1989)
43Standard ForHandling Of BOB
7
- For safe and accurate BOB measurement
- BOB measurements shall include all carb and
correction boluses given within the selected DIA - When residual BOB is present at the time of a
bolus, the BOB shall be subtracted from both carb
and correction bolus recommendations. - When BOB exceeds the current correction bolus
need or the current carb plus correction need,
the user will be alerted to how many grams of
carb they need to eat to avert a low.(BOB
correction carb bolus need) X carb factor
44Handling Of BOB
7
- Issue Current pumps differ significantly in what
is counted as BOB and in whether or not BOB is
subtracted from subsequent carb boluses. - Most insulin pumps assume that excess BOB does
not need to be taken into account when
determining the next carb bolus. - Though commonly determined in this way, the
resulting bolus dose recommendations can cause
unexplained and unnecessary insulin stacking and
hypoglycemia.
45ExampleInsulin Stacking
7
- With a bedtime BG of 173 mg/dl,
- is there an insulin deficit or a carb deficit?
Bedtime BG 173 mg/dl
Correction
Dessert
Dinner
6 pm
8 pm
10 pm
12 am
46ReviewFrequency Of Insulin Stacking
7
- CDA1 Study Results
- Of 201,538 boluses, 64.8 were given within 4.5
hrs of a previous bolus - Although 4.5 hours may underestimate true DIA,
use of this minimal DIA time shows that some BOB
is present for MOST boluses
4.5 hrs
47ReviewBolus On Board (BOB)
7
- An accurate measurement of the glucose-lowering
activity that remains from recent boluses - Prevents insulin stacking
- Improves bolus accuracy
- Allows the current carb or insulin deficit to be
determined -
48ReviewHow Current Pumps Handle BOB
7
Except when BG is below target BG
A Yes response is generally considered safer
49ExampleUnsafe BOB1 Handling
7
- If a pump user gets frustrated with a high BG and
they overdose to speed its fall, or they exercise
longer or more intensely than anticipated, they
can acquire a significant excess in BOB. - In these situations, most current pumps recommend
that a bolus be given for all carb intake
regardless of how much BOB is actually present. - When the BOB is larger than the correction bolus
required at the time, the pumps bolus
recommendation introduces an unnecessary risk of
hypoglycemia.
1 Pumping Insulin, 1st ed, 1989, Chap 12, pgs
70-73 The Unused Insulin Rule
50ExampleDifferences In Bolus Recommendations
7
- Situation BOB 3.0 u and 30 gr. of
carb will be eaten at these glucose levels - Carb factor 1u / 10 gr
- Corr. Factor 1 u / 40 mg/dl over
100 - Target BG 100
- TDD 50 u
The graphic shows how widely bolus
recommendations vary from one pump to another for
the same situation.
units
mg/dl
Omnipod bolus cannot be determined - it counts
only correction bolus insulin as BOB
51ReviewTrack BOB Or Carb Digestion?
7
- For safety after a bolus, it is more important to
account for the glucose-lowering action of BOB
than the glucose-raising action of digesting meal
carbs - When a BG is taken after a meal, the BOB times
the correction factor ideally represents the
maximum anticipated fall in glucose. - Accounting for the impact of the BOB on the
current glucose provides the safest approach in
the determination of bolus recommendations. - Low glycemic index meals, gastroparesis, Symlin,
and other issues may counteract a predicted fall
in glucose based on BOB, but the user can more
easily judge and remedy this situation than
dealing with an unknown excess of insulin.
52Exceptions To Usual Handling Of BOB
7
- When a second bolus is taken for an unplanned
carb intake or a desert that is consumed within
60 minutes or so of a meal bolus, BOB should not
be taken into account for the second bolus
because the impact of the first bolus cannot be
accurately determined. - Given that, it is wise to account for BOB as soon
after a meal as possible, such as within 60 to 90
minutes, to provide early warning if the bolus
given was excessive or inadequate.
Accounting for all BOB and applying it to
subsequent boluses is generally safer, although
not always more accurate.
Adjustable setting in pump/controller
53Multi-Linear And Curvilinear DIA
8
54Standard ForMulti-Linear And Curvilinear DIA
8
- Insulin pumps shall use either a 100 curvilinear
or a multi-linear method to improve accuracy and
consistency of BOB estimates.
55ReviewLinear And Curvilinear DIA
8
- Issue Pump manufacturers use at least 3
different methods (100 curvilinear, 95 of
curvilinear, and straight linear) to measure DIA
and BOB. - When a realistic DIA time is selected, a linear
determination of residual BOB will not be as
accurate as a curvilinear method that
incorporates the slow onset of insulin action and
its longer tailing off in activity. In most
situations, an accurate determination of
insulins tailing activity will be most important
to the pump user.
56Linear And Curvilinear DIA Compared
8
- Note how values for the 5 hr linear line in red
and the thinner 5 hr curvilinear line diverge in
value at several points along the graph.
5 hr Linear
5 hr 95 Curvilinear
From Pumping Insulin, 4th ed., adapted fom
Mudaliar et al Diabetes Care, 22 1501, 1999
57ExampleA Multi-Linear DIA
8
- Use of a multi-linear method to measure DIA
improves accuracy. The next page shows a
triple-linear example for measurement of BOB.
58ExampleA Triple-Linear Approximation Of DIA
8
- A triple-linear line in red can more closely
approximate a curvilinear DIA. - For a 5 hr DIA
- 1st 10 no change
- Mid 65 fall 75
- Last 25 fall 25 ( adjustable as needed
in device)
5 hr Triple Linear
modification suggested by Gary Scheiner, MS,
CDE
59Hypoglycemia Analysis And Alert
9
60Standards ForHypoglycemia Analysis And Alert
9
- Insulin pumps that store BG and insulin dosing
data can present this BG control data in a
readily accessible form on the pump or
controller. - The pump shall alert the user when the BG data
from their glucose meter or continuous monitor
suggests they are experiencing frequent or
severe patterns of hypoglycemia, based on
preselected, adjustable criteria.
Adjustable settings in pump/controller
61Frequent/Severe Hypoglycemia Alert
9
- Issue Although most current insulin pumps
contain sufficient data to determine when the
user is experiencing patterns of frequent or
severe hypoglycemia, pumps give no warning of
this to the user.
62ExamplePump Screen Hypoglycemia Display 1
9
To improve statistics for a better understanding
of the severity and frequency of low glucose
events over time (weeks), the pump can provide a
screen such as the following
Modified to display per Gary Scheiner, MS, CDE
Adjustable settings in pump/controller
63ExamplePump Screen Hypoglycemia Display 2
9
The daily timing for severity and frequency of
hypoglycemia can be shown in a screen such as the
following
Modified to display per Gary Scheiner, MS, CDE
Adjustable settings in pump/controller
64Frequent/Severe Hypoglycemia Alert
9
- Solution Depending on the settings selected by
the user and clinician, an alert would sound when
a pattern of more than one hypoglycemia event
exceeds their threshold for frequency or
severity of hypoglycemia. - In addition, typical ways to resolve the pattern
of hypoglycemia will be presented.
Adjustable settings in pump/controller
65Hyperglycemia Analysis And Alert
10
66Standards ForHyperglycemia Analysis And Alert
10
- Insulin pumps that store BG and insulin dosing
data can present this BG control data in a
readily accessible form on the pump or
controller. - The pump shall alert the user when the BG data
from their glucose meter or continuous monitor
suggests they are experiencing frequent or
severe patterns of hyperglycemia, based on
preselected, adjustable criteria.
Adjustable settings in pump/controller
67Frequent/Severe Hyperglycemia
10
- Issue Although most current insulin pumps
contain sufficient data to determine when the
user is experiencing patterns of frequent or
severe hyperglycemia, pumps give no warning of
this to the user.
68ExamplePump Screen Hyperglycemia Display 1
10
To improve statistics for a better understanding
of the severity and frequency of high glucose
events over time (weeks), the pump can provide a
screen such as the following
Modified to display per Gary Scheiner, MS, CDE
Adjustable settings in pump/controller
69ExamplePump Screen Hyperglycemia Display 2
10
The daily timing for severity and frequency of
hyperglycemia can be shown in a screen such as
the following
Modified to display per Gary Scheiner, MS, CDE
Adjustable settings in pump/controller
70Frequent/Severe Hyperglycemia Alert
10
- Solution Depending on the settings selected by
the user and clinician, an alert would sound when
a pattern of more than one hyperglycemia events
exceeds their threshold for frequency or
severity. - In addition, typical ways to resolve the pattern
of hyperglycemia will be presented.
Adjustable settings in pump/controller
71Correction Bolus Alert
11
72Standard ForCorrection Bolus Alert
11
- Insulin pumps shall show in a readily accessible
history screen the percentage of the TDD that is
used for correction boluses over time, and alarm
based on preselected, adjustable criteria. - For instance, the insulin pump shall alert the
wearer when they use more than 8 of their TDD
for correction bolus doses over the most recent 4
day period.
Adjustable settings in pump/controller
73Correction Bolus Alert
11
- Issue Hyperglycemia is more common than
hypoglycemia for most people on insulin pumps. - When glucose levels consistently run high, many
pump users address the problem by giving frequent
correction boluses rather than correcting the
core problem through an increase in their basal
rates or carb boluses. - If the correction bolus becomes excessive
relative to the TDD, the pump user is made aware
of their excess use of correction boluses.
74Insulin Stacking Alert
12
- Accurate accounting of the BOB is important to
those who bolus frequently, those who experience
frequent or severe hypoglycemia, and those whose
average glucose levels are closer to normal
values.
75Standard ForInsulin Stacking Alert
12
- Insulin pumps shall alert the user when he/she
gives a bolus and no glucose value has been
entered in the pump, but there is sufficient
insulin stacking to significantly alter the
bolus they would otherwise give. - The alert is on by default once a DIA time is
selected to measure BOB, but may be turned off if
the user desires.
Such as when the BOB is greater than 1.25 of
the avg. TDD, sufficient to change the glucose
about 25 mg/dl. ( Adjustable setting in
pump/controller for a certain fall in glucose
selected by the user or clinician)
76Insulin Stacking Alert
12
- Issue Pump users often bolus for carbs without
checking their glucose first. With no glucose
reading, the pump cannot account for BOB, nor
appropriately adjust a recommended bolus for the
BOB or the current BG. - Even though no BG test is done, data available in
the pump when the bolus is given can inform the
user when there is sufficient BOB to
substantially change their bolus dose. - The pump can alert the user to this unseen,
substantial insulin stacking.
77ExampleInsulin Stacking or BOB Alert
12
- When a carb bolus is planned without a recent BG
check, but BOB is more than 1.25 of the average
TDD (enough to cause about a 25 mg/dl drop in the
glucose), the pump will recommend that the wearer
do a BG check due to the substantial presence of
BOB. -
- For instance, for someone with
- Avg TDD 1.25 of TDD
- 40 units 0.5 units
- 50 mg/dl per u (corr factor) X 0.5 u 25
mg/dl - This individual would be alerted when they do not
check their glucose and want to give a bolus but
have 0.5 u or more of BOB present.
Adjustable in pump/controller for a reasonable
degree of safety
78Automatic Entry Of BG Values
13
79Standard ForAutomatic Entry Of BG Values
13
- For completion of the BG history, improved
handling of BOB, and more accurate bolus
recommendations, insulin pumps shall be enabled
to have wireless or direct entry of BG test
results from a glucose meter. - Automatic glucose entry from two or more major
brands of meters is recommended to increase the
likelihood of insurance coverage for test strips.
80Automatic Entry Of BG Values
13
- Issue Pump users do not enter as many BG values
into their pump when readings must be entered
manually rather than automatically entered from a
meter. - When a BG is checked but not entered, the lost
data cannot be used to account for BOB, warn of
insulin stacking (see 12), and are not as
accurate when attampting to analyze glucose
patterns or the frequency of hypoglycemia and
hyperglycemia. - Relatively normal and hypoglycemia values are
less likely to be entered manually, but are more
likely to be influenced by BOB.
81ReviewAutomatic Entry Of BG Values
13
- In a study of over 500 insulin pumps where BG
values could be entered either manually or
automatically, users entered 2.6 BG values per
day manually, compared to 4.1 values per day for
pumps in which glucose values wre automatically
entered. - BOB could typically be taken into account for 1.5
additional boluses per day with automatic entry
of BG values. - When glucose values are not automatically
entered, BOB cannot be determined and bolus
recommendations will not be as accurate.
J. Walsh, D. Wroblewski, T.S. Bailey
unpublished data
82ReviewAutomatic Entry Of BG Values
13
- Automatic entry of glucose values into pumps
offers a significant clinical advantage to users
and clinicians because more boluses will be
adjusted for high and low BGs, and residual BOB
is more likely to taken into accout in bolus
calculations. - Automatic entry of glucose values (along with
accurate accounting of BOB, see 7) ensures a
greater degree of safety for those who experience
frequent or sever hypoglycemia, and those whose
glucose values are closer to normal.
83Infusion Set Monitoring
14
84Standards ForInfusion Sets
14
- Insulin pumps shall monitor and record in easily
accessible history the duration of infusion set
usage recorded as mean, median, and SD of time of
use. - Insulin pumps shall monitor and report average
glucose values over 6 or 12 hour time intervals
between set changes with the ability to change
the observation interval, such as the last 1 to
30 set changes. - This monitor allows HCPs and users to identify
infusion set problems through loss of glucose
control related to length of set use and
variations in length of infusion set use.
Adjustable setting in pump/controller
85Infusion Sets
14
- Issue A significant number of pump wearers
encounter partial or complete infusion set
failure. Failure may arise from poor infusion set
design or poor site preparation. Which users are
encountering infusion set problems, how often
infusion sets may be responsible for erratic
glucose readings, and how significantly infusion
set problems impact glucose control are currently
difficult for users and clinicians to identify.
86ReviewInfusion Set Failure
14
- Infusion set failure can arise due to occlusions
or from a Teflon infusion set coming loose
beneath the skin due to movement or tugging. When
loosened, some of the infused insulin can leak
back to the skin surface and result in
unexplained high readings. - A complete loss of glucose control can also occur
when an infusion set is pulled out entirely. - Selecting the right infusion set plus good site
technique, especially taping the infusion line to
the skin, can significantly minimize unexplained
high readings for many pump wearers.
87Why Tubing Needs To Be Taped
14
- Many problems with infusion sets come from
loosening of the Teflon under the skin, not from
a complete pullout. A 1 tape placed on the
infusion line - Stops tugging on the Teflon catheter under the
skin - Prevents loosening of the Teflon catheter under
the skin - Avoids many unexplained highs caused when
insulin leaks back to the skin surface - Reduces skin irritation
- And prevents many pull outs
88Tape The Tubing
14
- This helps prevent
- Tugging
- Irritation
- Bleeding
89ExamplesLack Of Anchoring Of Sets
14
A review of dozens of pictures of infusion sets
online and pump manuals finds that anchoring of
the infusion line with tape is rarely recommended
or practiced.
No tape!
90Review Infusion Set Monitor
14
- Many pump wearers experience random erratic
readings until they change to a different
infusion set or start to anchor their infusion
lines with tape to stop line tugging. - However, insulin pumps offer no mechanism for
clinicians or pump users to detect who may be
having problems with their infusion sets.
91ToolInfusion Set Monitor
14
- Insulin pumps with direct BG entry can identify
those who may be having intermittent loss of
glucose control secondary to infusion set
failure. The pump - Shows the average time and variation in time of
use between reservoir loads or use of the priming
function. - Shows average BGs for each full or partial 12 or
24 hour time interval following set changes
(indicated by the prime function) over a various
number of set changes or as soon as statistical
significance is reached.
Adjustable setting in pump/controller