The Design and Test of a Pulsed 250 MHz EPR Spectrometer PowerPoint PPT Presentation

presentation player overlay
1 / 42
About This Presentation
Transcript and Presenter's Notes

Title: The Design and Test of a Pulsed 250 MHz EPR Spectrometer


1
The Design and Test of a Pulsed 250 MHz EPR
Spectrometer
10th International Workshop on Bio-Medical ESR
Spectroscopy and Imaging Fukuoka, Japan April
1-3, 2003
Gareth R. Eaton, Richard W. Quine, George A.
Rinard, and Sandra S. Eaton Department of
Chemistry and Biochemistry and Department of
Engineering, University of Denver Howard Halpern
and Colin Mailer, University of Chicago
NIH-funded Center for Electron Paramagnetic
Resonance Imaging for in Vivo Physiology. H.
Halpern, University of Chicago, G. Eaton,
University of Denver, and G. Rosen, University of
Maryland
2
(No Transcript)
3
Low Frequency EPR
  • Low EPR frequencies being used include
  • 200, 220, 250, 280, 300, 600, 700, 1000, 1200 MHz
  •  
  • RF penetration as a function of frequency depends
    on
  • dielectric properties
  • conductivity
  • sample shape
  • homogeneity
  •  
  • RF penetration increases
  • the lower the RF frequency
  • the more heterogeneous the sample
  • ?deeper penetration in animals than in spheres or
    cylinders of water
  •  

4
Pulsed 250 MHz EPR  
  • Is being developed to
  • Exploit the high sensitivity of electron spin
    relaxation times to O2
  • Gate data acquisition to physiologic motion
  • Detect transient phenomena
  • Electron spin relaxation times are at least
    1000-fold faster than the nuclear spin relaxation
    times important in MRI. Consequently, we need to
    develop several crucial enabling technologies
  • Resonators with sub-microsecond dead time
    following the high-power pulse
  • High-power pulsed RF amplifiers
  • Methodology for fast data acquisition

5
Comparison of CW and Pulsed EPR - 1
  • For a sample for which
  • all of the spins in the sample form the echo
  • the CW spectrum is not saturated
  • the CW magnetic field modulation is approximately
    equal to the line width
  •  
  • The relative CW and spin echo signal intensities
    are given by the ratio

If the CW spectrum saturates, then the B1 has to
be reduced below the maximum available in the
spectrometer, reducing the CW intensity. If the
magnetic field modulation is reduced to avoid
distorting the CW line shape, then the CW
intensity if reduced roughly proportionately.
6
Comparison of CW and pulsed EPR - 2
If noise is thermal noise, then the noise in the
signal is proportional to the bandwidth of the
spectrometer. In a CW spectrometer the bandwidth
is usually limited by a filter at a late stage of
the detection system. Commonly a time constant
of a few ms to 1 s is used for CW EPR,
corresponding to bandwidths of 160 Hz to 0.16
Hz.   The -3dB bandwidth of a single pole filter
with time constant TC is
In a pulsed spectrometer, the filter is wider to
pass all relevant frequencies, and commonly is
several MHz.   If the bandwidths were, e.g., 10
Hz (time constant 0.016 s) for the CW spectrum
and 10 MHz for the pulse spectrum, the noise in
the pulse spectrum would be 103 greater.
However, the pulse spectrum might be acquired at
a repetition of 104 per second, so in 100 seconds
of signal averaging, the noise would be about the
same in the pulse spectrum as in a CW spectrum
acquired as quickly as the line width
permitted.   from Rinard, Quine, Song, Eaton and
Eaton, JMR 140, 69-83 (1999)
7
Relative Benefits of Pulsed and CW EPR
  • CW
  • can be performed on many radicals
  • including radicals that have wide spectra
  • can be performed for large gradients, to achieve
    high spatial resolution
  • Pulse
  • can be more sensitive per unit time than CW for
    very-narrow-line spin probes such as the trityl
    radicals
  • is limited by dead time following the pulse
  • FID detection is limited by the effective T2
    caused by the gradient field
  •  
  • see experimental comparison at 300 MHz in
  • K.-I. Yamada, R. Murugesan, N. Devasahayam, J. A.
    Cook, J. B. Mitchell, S. Subramanian, and M. C.
    Krishna. Evaluation and Comparison of Pulsed and
    Continuous Wave Radiofrequency Electron
    Paramagnetic Resonance Techniques for in Vivo
    Detection and Imaging of Free Radicals. J. Magn.
    Reson. 154, 287-297 (2002).

8
Selection of Experiment - 1
  • The goal of the measurement dictates the
    conditions for spectroscopy
  •  
  • Optimum detection occurs when the line is
    saturated and over-modulated
  • The maximum CW signal for a single Lorentzian
    line occurs when ?2B12T1T2 1
  • The line broadens when saturated. It is about
    1.2 times broader at the power that gives the
    maximum in the signal amplitude.
  • The maximum intensity occurs when the modulation
    amplitude is approximately equal to the line
    width, at which point the width is increased by
    about a factor of 2 (depends on details of line
    shape).

9
Selection of Experiment - 2
  • One might use less saturation and lower
    modulation if the goal is to
  • identify the radical
  • discriminate between radicals - also use
    multi-frequency, differential relaxation
  • measure viscosity
  • measure pH
  • However, if the goal is to measure O2
    concentration, the spectroscopy becomes more
    demanding.
  • Small changes in line width measure O2
    concentration
  • Over-modulation by a factor of 1 to 2x to improve
    S/N is permitted if Colin Mailers program is
    used to deconvolute the excess modulation.
    (Robinson, Mailer, Reese, J. Magn. Reson. 138,
    199, 210 (1999)).
  • Relaxation times are sensitive to changes in O2
    collision rate, so pulsed EPR may be valuable.

10
MRI Comparison
Concentrations 1H in vivo is about 100 M, but
electron spin concentrations for EPR typically
are less than 1 mM Relaxation Times about 1000
times longer for NMR than EPR Echo vs. FID MRI
uses spin echos but most pulsed EPR uses
FID's Maximum Frequency MRI at 8T, 340 MHz
(Robataille et al. J. Computer Assisted
Tomography 23821-831 (1999).
1.5 T 4.7 T 8 T
11
MRI Field Distributions
Robataille et al., J. Computer Assisted
Tomography 23821-831 (1999).
12
C. M. Collins, S. Li, M. B. Smith, Magn. Reson.
Med. 40 847-856 (1998).
13
Absolute Signal-to-Noise
The EPR signal voltage, Vs, is given by
  • C"(w) is the imaginary component of the effective
    RF susceptibility
  • is the filling factor
  • Q is the loaded quality factor of the resonator
  • Zo is the characteristic impedance of the
    transmission line
  • P is the RF microwave power to the resonator

14
Resonator B1
  • d is the diameter of the LGR
  • z is the length of the LGR
  • ?0 is the permeability in a vacuum
  • is the conductivity of the surface of the
    resonator
  • ? is the resonance frequency
  • P is the power to the resonator

15
Frequency Dependence of EPR Parameters
P is RF power, and B1 is the RF magnetic field
intensity.
G. A. Rinard, R. W. Quine, S. S. Eaton, and G. R.
Eaton J. Magn. Reson. 156, 113 (2002).
16
Frequency Dependence of Signal Intensity
Experimental Results
  • VHF vs. L-band FID of trityl normalized signals
    agreed with prediction within 5
  • VHF, L-band, X-band resonator dimensions scaled
    by factors of 6
  • For each pair of frequencies the CW signal at
    lower frequency was predicted to be larger by
    factor of 1.57.
  • 250 MHz / 1.5 GHz was 1.52
  • 1.5 GHz / 9 GHz was 1.14

17
Frequency Dependence of EPR S/N when sample loss
dominates the noise
r is sample resistivity
18
(No Transcript)
19
CW operation with Bruker console
20
Pulse operation with DU hardware
21
Pulse operation with Bruker console
22
Cross-Loop Resonator and Pre-Amplifier Modules
23
Magnet and Gradient Coil Specifications
24
Location of gradient coils
25
Hall-probe/current control adapter
a Two six-turn spiral coils of no. 12 copper
magnet wire with 2.2 cm mean diameter spaced
1.25 cm apart b Aluminum coil forms c 5 cm
aluminum channel d Aluminum spacers e
Fiberglass insulating spacers f insulated
Plexiglas mount for Hall probe
26
Magnet and cross-loop resonator
The 250 MHz crossed loop resonator is shown in
the very homogeneous (lt 40 ppm variation over 15
cm diameter) 4-coil resistive magnet designed for
EPR imaging. The outer diameter of the large
coil is ca. 1 meter. The picture also shows the
z-gradient coils.
The x, y gradient coils have been removed to show
the other coils more clearly. The crossed loop
resonator can isolate the time-domain EPR signal
from the high power RF pulse by 60 dB (typical).
27
Cross-Loop Resonator
a Sample tube resonator b Cross resonator
c,d Brackets to adjust orthogonality e
Isolation adjustment screw f Sample tube g
Sample tube holder h Coupling screw for
resonator a i Coupling screw for resonator
b j Frequency tuning for resonator b
Resonators machined from tellurium-copper alloy
28
Cross-Loop Resonator
  • Sample tube
  • Sample tube holder
  • Loop (inductor) of sample tube resonator
  • gap (capacitor) or sample tube resonator
  • Re-entrant loop for sample tube resonator
  • Loop for cross resonator
  • Re-entrant loop for cross resonator
  • Sample access hole

29
Phase Noise Comparison
Blue CLR Red - LGR
30
Isolation of cross-loop resonator
Loop gap resonator
Cross-loop resonator
31
Q-switching circuit
32
Q-switching sequence
The Q of the excitation resonator is held high
and the detection resonator Q is spoiled during
the RF pulses. The Q states are reversed during
the time when the EPR signal is recorded.
33
Active Q-spoiling
Without Q-spoiling Q1 (excitation) 74 Q2
(detection) 469
With Q-spoiling
34
High-power amplifier
35
High-power amplifier
There was no RF amplifier commercially available
that would meet the needs of pulsed 250 MHz EPR
imaging.   Two companies, Communications Power
Corporation (CPC) and Tomco, have contracted to
develop and produce amplifiers with       
Output power gt 400 W        lt 30 ns rise and
fall times        blanking of output noise to
lt10dB above thermal within 80 ns   The
Tomco amplifier (top) meets the speed
specifications and was tested at 550 W.   A
400 W prototype of the CPC amplifier is shown in
the bottom figure.
36
Design Philosophy of Pulsed VHF Bridge for
University of Chicago
  • Use of a crossed-loop resonator (CLR)
  • Stand-alone testing of resonators
  • Stand-alone FID
  • Electron spin echo (ESE) when controlled by a
    Bruker PatternJet
  • 1 kW maximum RF pulse power
  • 2 kW maximum RF power
  • To be used with Bruker E540 console
  • Data collection via Bruker SpecJet

37
Time-domain data acquisition methodology at 250
MHz
       The isolation of the crossed loop
resonator reduces the dead time of the system
following the high-power pulse, because the pulse
power reaching the detection system is reduced by
the amount of the isolation.        We have
decreased dead time even more by active
Q-switching of the excitation and detection
resonators in synchrony with the RF pulses and
detection times.        The signal on the
following slide is the free induction decay (FID)
of a sample of Nycomed triarylmethyl (trityl)
radical, whose electron spin relaxation time is a
measure of oxygen concentration in vivo.  Full
FIDs were digitized with a repetition time of 100
?s
38
Pulse Sequences

39
FID, no gradient
FID for two 0.3 mm i.d. tubes of 0.2 mM
trityl-CD3 separated by 1.8 cm 4096 data points,
199,680 averages Decay time constant is 2.88 ms.
Zero-filled to 16k, FFT Full-width half-height
35 mG DBpp 21 mG
40
FID-detected projection
Two 0.3 mm i.d. tubes of 0.2 mM trityl-CD3
separated by 1.8 cm. Gradient of 0.34
G/cm 99,328 averages 4096 points, zero-filled to
16,384 Phase determined by starting point in the
FID. Signals are sharper in the real component of
the absorption spectrum than in the absolute
value spectrum.
41
Comparison of spin echo and FID
Same two 3 mm id tubes of 0.2 mM sym-trityl
separated by 1.8 cm. Gradient of 0.55 G/cm and
99,328 acquisitions At this higher gradient the
FID decays so quickly that the resolution in the
Fourier transform is poorer than for the
transform of the spin echo data.
42
Block Diagram of DU Bridge
Write a Comment
User Comments (0)
About PowerShow.com