IPPB
2
IPPB Introduction
  • Intermittent Positive Pressure Breathing
  • Def. as the therapeutic application of
    inspiratory positive pressure to the airway

3
Physiologic Effects of IPPB
  • Increase in MAP - Def. as the average
    pressure applied to the lungs over a defined
    period - potential for decreased venous
    return
  • Decreased work of breathing
  • Manipulation of IE ratio
  • Increased Vt

4
Physiologic Effects (contd)
  • Mechanical bronchodilation
  • Collateral ventilation

5
Clinical Goals of IPPB
  • To improve and promote the cough mechanism
  • Improve distribution of ventilation
  • To deliver medication

6
Indications for IPPB
  • Acute or chronic hypoventilation
  • Atelectasis
  • Decreased cough
  • Increased airway resistance
  • Increased WOB
  • Accumulated secretions
  • Pulmonary congestion
  • Cardinal indication is the inability to inspire
    adequately

7
Contraindications for IPPB
  • Tension pneumothorax without chest tube
  • Pulmonary hemorrhage
  • Active Tb
  • Hyperventilation
  • Increased ICP
  • Decreased C.O.
  • Uncooperative patient
  • More economical methods

8
Hazards of IPPB
  • Excessive ventilation
  • - decreased PaCO2 may lead to dizziness,
    loss of consciousness
  • Excessive oxygenation
  • Decreased cariac output
  • Increased I.C.P.
  • Pneumothorax
  • Hemoptysis - usually due to improved cough

9
Considerations for an effective IPPB Tx.
  • Usually volume-oriented
  • - use 10 ml/kg of ideal body weight
  • Machine must be sensitive to respond to pts
    insp. effort (-2 cmH2O)
  • Tight seal must be maintained between the pt. and
    the apparatus
  • Suitable control pressure and flow should be
    delivered
  • Patient must relax, let machine do work

10
Considerations (contd)
  • Nebulizer must function properly
  • Vital signs before, during, after
  • Monitor pt. and make necessary adjustments (pulse
    increases 20 from baseline, stop tx.)

11
Basic Components of IPPB Machines
  • Gas pressure source - 45 - 55 psi
  • Main control valve
  • Breathing circuit
  • Automatic cycling control

12
Bird Mark 7
  • Is a pneumatically powered, assist/controller,
    and is pressure cycled
  • Incorporates magnetism vs. gas pressure

13
Bennett PR 2
  • Is pneumatically powered, pressure limited at low
    terminal flow, time limited, patient cycled, and
    time cycled
  • Has controls for adjusting sensitivity, peak
    flow, independent insp. and exp. nebulizer
    controls, terminal flow for minor leak
    compensation, and an automatic rate with adj. IE
    ratios
  • Can deliver 100 or air-diluted source gas