Title: Luminescence is rare in nonaromatic hydrocarbons'
1Nonaromatic Unsaturated Hydrocarbons
Luminescence is rare in nonaromatic hydrocarbons.
Possible if highly conjugated due to p p
transitions.
Seyhan Ege, Organic Chemistry, D.C. Heath and
Company, Lexington, MA, 1989.
2Aromatic Hydrocarbons
Fluorescent
Low lying p p singlet state
Phosphorescence is weak because there are no n
electrons
Ingle and Crouch, Spectrochemical Analysis
3Heterocyclic Aromatics
Aromatics containing carbonyl or heteroatoms are
more likely to phosphoresce
n p promotes intersystem crossing.
Fluorescence is often weaker.
Skoog, Hollar, Nieman, Principles of Instrumental
Analysis, Saunders College Publishing,
Philadelphia, 1998.
4Aromatic Substituents
- Electron donating groups usually increase fF.
- Electron withdrawing groups usually decrease fF.
Ingle and Crouch, Spectrochemical Analysis
5Halogen Substituents
Internal Heavy Atom Effect
Promotes intersystem crossing. fF decreases as MW
increases. fP increases as MW increases. tP
decreases as MW increases.
Ingle and Crouch, Spectrochemical Analysis
6Increased Conjugation
fF increases as conjugation increases. fP
decreases as conjugation increases. Hypsochromic
effect and bathochromic shift.
Ingle and Crouch, Spectrochemical Analysis
7Rigid Planar Structure
fF 1.0
fF 0.2
fF 0.8
not fluorescent
Ingle and Crouch, Spectrochemical Analysis
Skoog, Hollar, Nieman, Principles of Instrumental
Analysis, Saunders College Publishing,
Philadelphia, 1998.
8Metals
Metals other than certain lanthanides and
actinides (with f-f transitions) are usually not
themselves fluorescent.
A number of organometallic complexes are
fluorescent.
Skoog, Hollar, Nieman, Principles of Instrumental
Analysis, Saunders College Publishing,
Philadelphia, 1998.
9Fluorescence and Phosphorescence
Which effect is used more regularly?
SciFinder Scholar Citations
Fluorescence Phosphorescence
Labels/Tags 60642 231 Dyes 86673 412
www.wikipedia.org
10Fluorescence or Phosphorescence Labels?Answer
from a Commercial View
http//www.invitrogen.com/
11Fluorescence or Phosphorescence?Commercially
Available Phosphorescence Labels
Erythrosin derivative
Eosin derivative
http//www.invitrogen.com/
12Fluorescence or Phosphorescence?Publications in
Analytical Chemistry
- Fluorescence Phosphorescence
- 10847 7927
- Phosphorescence is rarer than fluorescence gt
Higher selectivity. - Phosphorescence Analysis of aromatic compounds
in environmental samples.
13Solvent Effects
Increased viscosity can increase luminescence
intensity.
H-bonding and dipole interactions with the
solvent contribute to the Stokes shift.
Ashutosh Sharma and Stephen Schulman,
Fluorescence Spectroscopy, John Wiley Sons, New
York, 1999.
14Solvent Polarity
Increasing solvent polarity usually causes a
red-shift in fluorescence.
http//micro.magnet.fsu.edu/primer/techniques/fluo
rescence/fluorescenceintro.html
15Solvent Polarity
Joseph Lakowicz, Principles of Fluorescence
Spectroscopy, Kluwer Academic / Plenum
Publishers, New York, 1999.
16Temperature
Increasing temperature increases ks
Joseph Lakowicz, Principles of Fluorescence
Spectroscopy, Kluwer Academic / Plenum
Publishers, New York, 1999.
17Decreasing temperature can induce a blue-shift in
fluorescence.
Joseph Lakowicz, Principles of Fluorescence
Spectroscopy, Kluwer Academic / Plenum
Publishers, New York, 1999.
18Shpolskii Spectroscopy
- Analytical potential of fluorescence spectroscopy
often limited by unresolved band structure (5-50
nm) - homogeneous band broadening depends directly on
radiative deactivation properties of the excited
state (usually 10-3 nm) - inhomogeneous band broadening various analyte
microenvironments yields continuum of bands
(usually few nm) - Solution Incorporate molecules in rigid matrix
at low temperature to minimize broadening - Result Very narrow luminescence spectra with
each band representing different substitution
sites in the host crystalline matrix
19Shpolskii Spectroscopy
- Requirements
- T lt 77K with rapid freezing rate
- Matrix with dimension match
- Low analyte concentration
- Instrumentation
- Xe lamp excitation
- Cryogenerator with sample cell
- High resolution monochromator with PMT
Analytes polycyclic aromatic compounds in
environmental, toxicological, or geochemical
systems
Garrigues and Budzinski, Trends in Analytical
Chemistry, 14 (5), 1995, pages 231-239.
20Epi-Fluorescence Microscopy
- Light Source - Mercury or xenon lamp (external
to reduce thermal effects) - Dichroic mirror reflects one range of
wavelengths and allows another range to pass. - Barrier filter eliminates all but fluorescent
light.
http//micro.magnet.fsu.edu/primer/techniques/fluo
rescence/fluorosources.html
21Fluorescence Microscopy
Need 3 filters Exciter Filters Barrier
Filters Dichromatic Beamsplitters
http//microscope.fsu.edu/primer/techniques/fluore
scence/filters.html
22Are you getting the concept?
You plan to excite catecholamine with the 406 nm
line from a Hg lamp and measure fluorescence
emitted at 470 15 nm. Choose the filter cube
you would buy to do this. Sketch the transmission
profiles for the three optics.
http//microscope.fsu.edu/primer/techniques/fluore
scence/fluorotable3.html
23Fluorescence Microscopy Objectives
Image intensity is a function of the objective
numerical aperture and magnification
Fabricated with low fluorescence glass/quartz
with anti- reflection coatings
http//micro.magnet.fsu.edu/primer/techniques/fluo
rescence/anatomy/fluoromicroanatomy.html
24Fluorescence Microscopy Detectors
No spatial resolution required PMT or
photodiode Spatial resolution required CCD
http//micro.magnet.fsu.edu/primer/digitalimaging/
digitalimagingdetectors.html
25Special Fluorescence Techniques
LIF
TIRF
http//microscopy.fsu.edu/primer/techniques/fluore
scence/tirf/tirfintro.html