Technical Updates on UTEVA Resin and TEVA Resin - PowerPoint PPT Presentation

1 / 18
About This Presentation
Title:

Technical Updates on UTEVA Resin and TEVA Resin

Description:

Extrude resin from the column in a LSC vial. Add cocktail and count on LSC. ... Extrude resin in LSC vial. Added 15 ml of cocktail and count ... – PowerPoint PPT presentation

Number of Views:226
Avg rating:3.0/5.0
Slides: 19
Provided by: atha8
Category:

less

Transcript and Presenter's Notes

Title: Technical Updates on UTEVA Resin and TEVA Resin


1
Technical Updates on UTEVA Resin and TEVA Resin
Denver May 2, 2002
2
Topics
  • UTEVA Resin
  • Decontamination of Po-210
  • U-232 yields
  • TEVA Resin
  • Decontamination of U-nat/Th in Tc-99 region

3
U-232 Yields
  • Unrealistic high U-232 yields reported with UTEVA
    resin in samples with high in Pb-210

4
Current Procedure
  • 1) Load sample on UTEVA resin with 10 ml of 3M
    HNO3/1M Al(NO3)3. with 2 mL of 0.6M Ferrous
    sulfamate and 200 mg of ascorbic acid.
  • 2) Rinse sample beaker with 5 mL of 3M HNO3
  • 3) Rinse column with 5 mL of 3M HNO3
  • 4) Rinse column with 5 mL of 9M HCl
  • 5) Rinse column with 20 mL of 5M HCl/0.05M Oxalic
  • 6) Elute U with 15 mL of 0.01M HCl

5
Effect of Ferrous210Pb/210Po spike (1100 pCi)
6
Effect of Additional 3M HNO3 Rinse 210Pb/210Po
spike
7
(No Transcript)
8
Effect of Uranium Strip Solution no ferrous
added 210Pb/210Po spike (1100 pCi)
9
Additional Rinse/1M HCl Strip Combo
10
Revised Procedure
  • 1) Load sample on UTEVA with 10 ml of 3M HNO3/1M
    Al(N03)3 with 2 mL of 0.6M Ferrous sulfamate and
    200 mg of ascorbic acid
  • 2) Rinse sample beaker with 5 mL of 3M HNO3
  • 3) Rinse column with 5 mL of 3M HNO3
  • 4) Rinse column with 10 mL of 3M HNO3 (3x)
  • 5) Rinse column with 5 mL of 9M HCl
  • 6) Rinse column with 20 mL of 5M HCl/0.05M Oxalic
  • 7) Elute U with 15 mL of 1M HCl

11
Summary
  • A good decontamination of Po-210 in U-232 region
    can be achieved by additional 3M HNO3 rinse on
    UTEVA and eluting uranium selectively with 1M HCl

12
TEVA Column Tc-99 Issues
  • Problem reported by Darrin Mann (BWXT, Oak Ridge,
    Tn) with bias high Tc-99 recoveries in samples
    with natural uranium.
  • Th-234 interference was detected in Tc-99 region.

13
BWXTs Tc-99 Procedure
  • Solid samples are digested and dissolved finally
    in 20 ml of 0.1M HNO3
  • Load sample on TEVA column
  • Rinse column with 25 ml of 0.1M HNO3
  • Extrude resin from the column in a LSC vial.
  • Add cocktail and count on LSC.

14
BWXTs Procedure tested at Eichrom
  • Spiked 20 ml of 0.01M HNO3 with 2000 dpm of
    U-234/238
  • Load on TEVA Resin column
  • Rinse with 25 ml of 0.1M HNO3
  • Extrude resin in LSC vial
  • Added 15 ml of cocktail and count

15
Possible Solutions for Decontamination
  • Use of fluoride to complex the residual Th-234 on
    the column
  • Two fluoride solutions tested
  • Hydrofluoric acid
  • Sodium fluoride

16
Decontamination of Th-234
  • Spiked 20 ml of 0.01HNO3 with 2000 dpm of
    U-234/238
  • Load on TEVA Resin column
  • Rinse with 5 ml of 0.1M HNO3
  • Rinse with 25 ml of 0.5M HF/0.02M HNO3 OR with
    25 ml of 1M NaF/0.02 HNO3
  • Rinse with 5 ml of 0.1 HNO3
  • Extrude resin in LSC vial
  • Added 15 ml of cocktail and count

17
BWXTs ProcedureTest ResultsRinse volume
used25 mL
18
Conclusion
  • Decontamination of Th-234 in Tc-99 region can be
    accomplished with additional fluoride rinse on
    the TEVA Resin columns
  • Fluoride in the form of 1M NaF or 0.5M HF
    solutions could be used to remove the residual
    amounts of Th-234
Write a Comment
User Comments (0)
About PowerShow.com