Title: Liquid argon as an active medium


1
Liquid argon as an active medium
Better title What has been learned by ICARUS
collaboration from 20 years of drifting
electrons in LAr
Carl Bromberg Michigan State University
Fermilab (2004)
2
Sources
See list of ICARUS talks and papers on FLARE web
site /www-off-axis/flare/documents.htm
Particularly
  • Talk by F. Sergiampietri, at NuFact 01 (LANNDD)
  • Talk by A. Para at Fermilab (Wine Cheese)
  • Publications
  • Design, construction and tests of the ICARUS
    T600 detector, NIM A527, 329 (2004)
  • Study of electron recombination in liquid Argon
    with the ICARUS TPC NIM A523, 275 (2004)
  • Measurement of the muon decay spectrum with the
    ICARUS T600 liquid Argon TPC, Euro Phys J C33,2
    (2004) 233

3
Liquid Argon basics
4
(No Transcript)
5
TPC operation
  • Time Projection
  • electrons drift toward transparent wire planes
    (minimum 1), and collection wire plane
  • planes with wires at various angles for stereo
  • Ambiguities
  • track parallel to wire planes, all hits have same
    time, i.e., x.
  • track dip resolved by space points vertices,
    delta rays or ionization bumps
  • distance between space points changes in the two
    views

delta ray
vertex
6
Signal characteristics
D tracks 16 mm
FWHM 4.5 mm
D
  • Diffusion not a big factor in signal shape for d
    2 m
  • Fit for x-position, resolution lt 1 mm (2.5 MHz
    clock), helps with ambiguity resolution.
  • Even with significant overlap in time,
    distinguish 1 track from 2, by consistent
    ionization/pulse width over many wires.
  • But tracks in drift direction have very distorted
    pulse shapes

7
Energy Resolution
8
FLARE is not ICARUS
  • Typical ICARUS capacitance is 20 pF/m 200 pF
    cable 400 pF (max) induction wire
  • Induction signals are smaller than collection
    signals by 30
  • FLARE will have wires 4 times as long, noise will
    be 1.6 times larger, yielding S/N 8
  • Does increasing wire spacing (to say 5 mm) help
    by lowering capacitance and increasing the
    signal?
  • Cables for horizontal or stereo wires? Long
    cables mean more capacitance. Restrict to
    collection wires.
  • Pixels anyone? K. McDonald.

9
For what is needed?
  • Energy resolution of 2 great for EM shower. For
    total energy, make corrections based on particle
    ID via ionization changes, range, p-gtm-gte decays,
    nuclear fragments n-stars
  • Full reconstruction is probably not needed.
    Automated scanning perhaps possible in 2D
  • In scanning simulations, stereo views were not
    that useful. Helped in large angle particle ID,
    but not often in electron ID.
  • Automatic space point finding is a challenge
  • Cosmic Rays (during 1.3 ms readout). Should do
    simulation. Get ICARUS raw data? Interest by
    others trying to do surface calorimetry and
    tracking.