The influence that ophthalmic viscosurgical device gives intra ocular pressure during small incision and ultra small incision cataract surgeries - PowerPoint PPT Presentation

About This Presentation
Title:

The influence that ophthalmic viscosurgical device gives intra ocular pressure during small incision and ultra small incision cataract surgeries

Description:

Viscous Cohesive. Opegan Hi (Santen, Japan) Viscoadaptive. Healon 5 (AMO, USA) ... Viscous Cohesive (Opegan Hi, etc) 3.0 2.4 1.5. Midium Viscosity Dispersive ... – PowerPoint PPT presentation

Number of Views:297
Avg rating:3.0/5.0
Slides: 10
Provided by: takuya5
Category:

less

Transcript and Presenter's Notes

Title: The influence that ophthalmic viscosurgical device gives intra ocular pressure during small incision and ultra small incision cataract surgeries


1
The influence that ophthalmic viscosurgical
device gives intra ocular pressure during small
incision and ultra small incision cataract
surgeries
  • Takuya Shiba, Takako Torisu, Akiko Hayama Katsuya
    Mitooka, Hiroshi Tsuneoka
  • Department of Ophthalmology
  • Jikei University School of Medicine
  • Tokyo, Japan

The authors have no financial interest in the
products mentioned in this poster
2
Objective
  • Using the experimental model we originally
    prepared, we estimated the difference in the
    inner pressure of the anterior chamber filling a
    different type of ophthalmic viscosurgical
    devices (OVD) in hydrodissection during small
    incision and ultra small incision cataract
    surgeries.

3
Methods 1/3
  • We made the diameter of the top of the pipettes
    1.5mm (micro bimanual phaco group), 2.4mm (micro
    coaxial phaco group), and 3.0mm (coaxial phaco
    group).

4
Methods 2/3
  • Filled the pipettes with OVD, and applied 5mmHg
    of constant pressure, using a pressure pump to
    imitate the process of hydrodissection, and then
    measured the inner pressure of the pipettes.

5
Methods 3/3
  • Used 4 types of OVD
  • Low Viscosity Dispersive
  • Opegan (Santen, Japan)
  • Midium Viscosity Dispersive
  • Viscoat (Alcon, USA).
  • Viscous Cohesive
  • Opegan Hi (Santen, Japan)
  • Viscoadaptive
  • Healon 5 (AMO, USA)

6
Results 1/2
3.0mm Coaxial Phaco
Viscoat
Healon 5
Opegan
Opegan Hi
2.4mm Micro Coaxial Phaco
Opegan Hi
Opegan
Viscoat
Healon 5
1.5mm Micro Bimanual Phaco
Opegan Hi
Opegan
Viscoat
Healon 5
7
Results 2/2
  • The micro coaxial and micro bimanual groups have
    shown a difference in the inner pressure with
    Opegan and other OVDs that is greater than that
    of the coaxial group.

8
Discussion
  • Intraocular Pressure / Incision Size (mm)
  • Low Viscosity Dispersive (Opegan, etc)
  • 3.0 2.4 1.5
  • Viscous Cohesive (Opegan Hi, etc)
  • 3.0 2.4 lt 1.5
  • Midium Viscosity Dispersive (Viscoat, etc)
  • 3.0 2.4 ltlt1.5
  • Viscoadaptive (Healon 5, etc)
  • 3.0 lt 2.4 ltltlt1.5

9
Conclusion
  • Our results suggested the necessity of selecting
    a low-molecular dispersive type of ophthalmic
    viscosurgical device for anterior capsulotomy
    when we implement small-incision and
    micro-incision cataract surgeries.
Write a Comment