S-02 / 08On-Line Characterization of Ethylene
Furnace Liquid Feedstock
  • Rod Spitler Andy Weber
  • The Dow Chemical Company

2
Summary of Discussion
  • Analysis requirements
  • Available analytical methods
  • Description of on-line system
  • Validation of on-line results
  • Conclusions

3
Analysis Requirements
Total analysis of naphtha
4
Available Analytical Technology
5
On-Line System
GC System Mounted in Climate Controlled Purged
Enclosure Class 1, Division 2, Group B,C,D
6
Valve / Column Configuration
  • Analysis Time 195 min
  • Approximately 45 min. dedicated to bake-out
  • n-C16 elutes at 130 min.
  • New data acquired approximately every 3.5 hours
    (including oven cool down time)
  • May be able to shorten bake-out time to decrease
    time between updates

7
Liquid Injection Valve
8
Sample Conditioning System
Sample Acquisition Steps 1. V1 energized, V3
open, V4 open samples flows to inj. valve
(2 mins.) 2. V1 de-energized, V3 open, V4
closed sample pressurized with He 3.
Sample Injected to GC 4. V1 de-energized, V3
closed, V4 open inj. valve purged with He
9
Chromatogram 1
10
Chromatogram 2
11
Chromatogram 3
12
Measurement Precision
13
n-pentane and iso-pentaneLab vs. On-line
14
PIONALab vs. On-Line
15
Conclusions
  • A modified laboratory GC has successfully been
    configured for on-line use
  • Non continuous slipstream flow with He purge
  • Minimizes opportunities for flammable liquid
    leaks
  • Minimizes potential for injection valves
    exposure to particulates, corrosive agents or
    pressure/temperature extremes.
  • Measurement precision of lt 1 RSD (1 std. dev.)
    was demonstrated
  • Good agreement between on-line and lab analyses
    was achieved