Title: Observer Performance Using Monitors With Different Phosphors


1
Observer Performance Using Monitors With
Different Phosphors
  • Elizabeth A. Krupinski, PhD
  • Hans Roehrig, PhD
  • University of Arizona
  • Tucson, AZ

2
P104 vs P45
  • P104
  • bigger grain size
  • fewer grains per pixel
  • two components
  • more non-uniformity
  • Lower SNR for P104

3
Examples
P45 Phosphor P104 Phosphor
4
Max SNR / CRT Pixel
120
P45 Phosphor
80
.
P 104 Phosphor
40
0
0
20
40
60
80
Luminance (cd/m2)
5
Display Function
200.00
160.00
120.00
P104
P45
80.00
40.00
0.00
0.00
100.00
200.00
300.00
Command Level (ADU)
6
Veiling Glare
0.020
0.016

0.012
Veiling glare ratio

P45
0.008
0.004
P104
0
0
200
400
600
800
1000
Disk Diameter in of Pixels
7
Pixel Emitted intensity
1.2
1.0
P45
0.8
P104
0.6
0.4
0.2
0
2
4
6
8
10
Pixels in Pixel String
8
MTF
P45 P104
1.2
1.2
113 ADU
113 ADU
0.8
0.8
-
Vertical MTF
Vertical
MTF
200 ADU

200 ADU
0.4
0.4
113 ADU
113 ADU
Horizontal
MTF
Horizontal MTF
200 ADU
200 ADU
0.0
0.0
Frequency(lp/mm)
9
Luminance Uniformity
10
JND Performance
F 28.55 df 23 p 11
Methods Materials
  • 20 chest images
  • 5 lesion free
  • 15 solitary pulmonary nodule
  • Image Systems P45 P104 monitors
  • 1600 x 1200 resolution
  • Pixel size 0.245 mm
  • 6 radiologists as observers

12
Images
- Conventional radiographs 150 mA 125 kVp -
CT confirmation - Digitized with VIDAR VXR -
2100 x 2552 x 16-bits
13
Viewing Conditions
  • Eye-position recorded
  • ASL 4000 SU with head-tracker
  • Ambient lights off
  • No time restrictions
  • 2 weeks between sessions

14
ROC Az Results
  • Mean Az P45 0.9235
  • P104 0.8851
  • ANOVA
  • Treatment F 0.4516, p 0.5097
  • Cases F 0.4038, p 0.9864
  • Readers F 0.1548, p 0.9781
  • No significant interaction effects

15
Eye-Position Results
16
Eye-Position Results
  • Mean Viewing Times Overall
  • - P45 22.36 seconds
  • - P104 27.41 seconds
  • 75 of the data. Future will split out
    lesion vs lesion-free images residents vs
    radiologists

17
Eye-Position Results
  • Median Dwell Times x Decision (msec)
  • P45 P104
  • TP 2043
  • FN 1528 1411
  • FP 1986
  • TN 470

18
Eye-Position Results
  • Mean Time to First Hit nodules (msec)
  • P45 P104
  • TP 312
  • FN 563

19
Conclusions
  • Phosphor may not affect diagnostic accuracy
  • May affect search parameters
  • Might influence purchase clinical use choices

20
Caveats
  • Detection not classification
  • Pulmonary nodules only
  • No image processing (window / level) allowed
  • Task stimuli may influence results
  • Further studies should be done