Title: Recent news about SiPM based applications R
1Recent news about SiPM based applications RD in
DESY
- Nicola DAscenzo
- University of Hamburg - DESY
2Calorimetry at International Linear Collider
- The ILC physics requires a calorimeter with
- High granularity (Good jet energy resolution)
- Ermeticity and good missing energy resolution
- Compactness (The detector has to stay inside the
magnetic coil)
3The scintillator tile hadron calorimeter
- Sampling calorimeter absorber stainless steel,
active medium plastic organic scintillator. - Along the magnetic field direction it extends in
-220 cm lt z lt220 cm - It covers the detector along the radius in the
region 180 cmlt r lt 290 cm - The test beam prototype 1 m3 .
- Test beam prototype 38 layers, 216 channel per
layer.
4SiPM a good photodetector for hadron calorimetry
- The good SiPM for this application
- Low noise above the suggested cut on the mip
spectrum (Controlled dark rate and cross talk) - Light Yield not exceeding the saturation
behaviour of the device. - Stable and controlled production
5The scintillator direct readout
- The direct readout is easier to extend to
- a large scale production.
- The solution which respects the required bounds
- on the parameters
- Blue Sensitive?
- Green sensitive and large?
6The MPPC by Hamamatsu
Blue sensitive Silicon PhotoMultiplier, Suitable
for the direct coupling
7Direct readout with MPPC
- Readout of a plastic scintillator tile directly
and comparison with the mediated green fiber - Tile wrapped with a 3M reflector. In case of the
direct readout, only a window of 1mm x 1mm is
open in front of the MPPC - Source Ru106
- Integration gate 80 ns
- External trigger (Photomultiplier tube)
- Reproducibility of the measurement - 3
- systematic (coupling, positioning etc ...)
8Direct readout with MPPC Most Probable Value of
the mip response spectrum
400 px
1600 px
The green dots correspond to the traditional
wavelength shifter fiber readout. The
performances of the MEPHI SiPMs used are totally
reproduced. THe blue dots correspond to the
direct readout a good light yield is obtained
9The direct readout with MPPC mip collection
efficiency
1600 px
Cut at 3 KHz
400 px
10Direct Readout with MPPC application to
electromagnetic calorimetry
GLD-CAL group, Kobe University
This is the first attempt of using MPPC in the
direct readout configuration in the calorimeter
design. The study of the impact of this
technology in the physics reconstruction is
still in progress.
11The direct readout discussion
- The direct readout looks promising for the
simplification of the design of the calorimeter
and the extension of the concept to a large
scale. - Still some conceptual problems
- Uniformity of the light response according to the
impact of the particle in the scintillator - Montecarlo studies and experiments will test this
soon - What is the effect on the performance of the
calorimeter? - What is the best position for the coupling?
- The blue sensitive MPPC opened good perspectives
to the design of calorimeters ready to test new
photodetectors.
12Application of MPPC in PET motivations and goals
- The calorimeter si just a big PET machine
- The hadron calorimeter has milions of channels
- The hadron calorimeter has a very high
granularity imaging detector! - It is only needed to
- Change scintillator (from organic to inorganic)
- Change granularity (from 3x3 cm2 to 1x1 mm2)
- The technology and the needs of the photodetector
are the same ! - Energy resolution and timing measurements
This study is just at the beginning!
13Application of MPPC in PET setup
2 LSO Crystals of 1mm x 1mm (3mm x3mm) Read out
by 1x1 mm (3x3 mm) Hamamatu MPPC No
amplification needed!!!
QDC
14Application of MPPC in PET energy resolution
PRELIMINARY
PRELIMINARY
3mmx3mm LSO crystal Coupled with a 3mmx3mm MPPC
1mmx1mm LSO crystal Coupled with a 1mmx1mm MPPC
Resolution 14
Resoultion 20
15Application of MPPC in PET timing resolution
Cut at 7 photoelectrons
PRELIMINARY
4 GHz Digital oscilloscope used to calculate the
time difference (50 ps precision) Still the
system needs to be improved with a lower
discrimination.
16Application of MPPC in PETdiscussion
- The application of MPPC in Positron Emission
Tomography looks promising the blue sensitivity
of the photodetectors is excellent. - In DESY we will test different combinations of
crystals and photodetectors YAP, LYSO... With
SenSL, new Hamamatsu .... - The enrgy resolution measurement will be
continuosly improved in reproducibility and
stability of the system - Timing measurement (important also for time of
flight) is currently beeing improved.
17After the serious conclusions...
The spin off of the technlogical developement
around the new high energy Phyics experiments is
very important... We are dedicating lots of
efforts in order to keep open and fruictful the
interplay between high energy physics and medical
physics. If we will not find the Higgs Boson, at
least, we can tell we were useful for
something...