DYNAMIC IMRT WITH PEACOCK
THE UNIVERSITY OF CALIFORNIA, IRVINE EXPERIENCE
Robert S. Kwon, M.D., Muthana S. A. L. Al-Ghazi,
Ph.D., Jeffrey V. Kuo, M.D., Nilam S.
Ramsinghani, M.D., Richard H. Yakoob, M.S.
Department of Radiation Oncology
University of California, Irvine
2
(No Transcript)
3
(No Transcript)
4
Multileaf Intensity Modulating Collimator (MIMiC)
5
CraneTM attachment for precision treatment table
indexing
6
Treatment Planning Computer
7
CT Imaging
8
Dosimetric Measurements Required Prior to
Clinical Use
  • 1. Two-dimensional scans of the beam profile for
    the beam produced by opening a single central
    leaf at 10cm depth in water and 100cm SSD
    (OARi(r) for leaf i)
  • 2. Beam profiles at a number of depths for a
    field produced by fully opening the MIMiC ( 2x20
    cm2 ) Enable PDD and TMR determination.
  • 3. ROFs at central axis for different field
    sizes.
  • (1x1 - 2x20 cm2 )
  • 4. Dose / MU to determine a calibration factor.

Instruments used Diode (p type Si 0.09mm3),
Markus Chamber (0.045cm3), Film
9
Dose Calculation Algorithm
10
Patient Characteristics
11
INDICATIONS FOR PEACOCK
12
Site Distribution
13
CNS Histology
14
Head and Neck Treatment Sites
15
Patient Immobilization
16
Immobilization Devises Cranial and HeadNeck
17
Immobilization Devices Body Mask
18
TREATMENT PLANNING PARAMETERS
19
Craniopharyngioma 15 Gy/1f post 32P treatment, 2
targets.
20
Recurrent Glioma 60Gy/30f Post 55.8Gy/31f
EBRTResection
21
Lung tumor 2 targets, 1 abutting the heart, 20
Gy/10 f boost after 40 Gy EBRT.
22
Chordoma of the Buttock 30Gy/15f (Post-op.
50.4Gy EBRT 62Gy Ir-192)
23
Ca. Prostate 32Gy/16f Boost (Post 40Gy/20f EBRT)
24
Field Placement Verification Cranial Case
25
Field Placement Verification Cranial Case
26
Dosimetric Verification in Phantom Prostate
(a) Calculated
(b) Film
27
Dose Distribution Verification Prostate
(a) Calculated
(b) Measured (Film)

28
CONCLUSIONS
  • 1. Relevant dosimetric data were acquired to
    enable clinical use of the system.
  • 2. Patient immobilization is achieved using the
    Talon system for selected cranial cases or an
    acquaplast mask for the remainder of sites
    treated.
  • 3. Accuracy of positioning and delivery is
    within 1mm for Talon cases and 2mm for mask
    cases. This is verified by weekly port films.
  • 4. The dose/fractionation schemes employed are in
    the range 15Gy/1f - 80Gy/40f.
  • 5. Multiple target (up to 3) can be treated in
    one set-up.

29
CONCLUSIONS (contd.)
  • 6. The range of target volumes treated is 2-500
    cm3
  • 7. To date 100 patients have been treated.
  • 8. Dosimetric accuracy of treatment delivery was
    verified by film, TLD and ion chamber
    measurements in phantom to be within /- 5.
  • 9. This approach provides an opportunity for dose
    escalation studies.
  • 10.The system is fitted to a standard linac
    without major modifications

30
Acknowledgements
  • Robert Sanford, PhD, Heeteak Chung and Rina Lim
    B.S. for assistance with dosimetric verification
    and preparation of the manuscript.