1DYNAMIC IMRT WITH PEACOCK
THE UNIVERSITY OF CALIFORNIA, IRVINE EXPERIENCE
Robert S. Kwon, M.D., Muthana S. A. L. Al-Ghazi,
Ph.D., Jeffrey V. Kuo, M.D., Nilam S.
Ramsinghani, M.D., Richard H. Yakoob, M.S.
Department of Radiation Oncology
University of California, Irvine
2(No Transcript)
3(No Transcript)
4Multileaf Intensity Modulating Collimator (MIMiC)
5CraneTM attachment for precision treatment table
indexing
6Treatment Planning Computer
7CT Imaging
8Dosimetric Measurements Required Prior to
Clinical Use
- 1. Two-dimensional scans of the beam profile for
the beam produced by opening a single central
leaf at 10cm depth in water and 100cm SSD
(OARi(r) for leaf i) - 2. Beam profiles at a number of depths for a
field produced by fully opening the MIMiC ( 2x20
cm2 ) Enable PDD and TMR determination. - 3. ROFs at central axis for different field
sizes. - (1x1 - 2x20 cm2 )
- 4. Dose / MU to determine a calibration factor.
Instruments used Diode (p type Si 0.09mm3),
Markus Chamber (0.045cm3), Film
9Dose Calculation Algorithm
10Patient Characteristics
11INDICATIONS FOR PEACOCK
12Site Distribution
13CNS Histology
14Head and Neck Treatment Sites
15Patient Immobilization
16Immobilization Devises Cranial and HeadNeck
17Immobilization Devices Body Mask
18TREATMENT PLANNING PARAMETERS
19Craniopharyngioma 15 Gy/1f post 32P treatment, 2
targets.
20Recurrent Glioma 60Gy/30f Post 55.8Gy/31f
EBRTResection
21Lung tumor 2 targets, 1 abutting the heart, 20
Gy/10 f boost after 40 Gy EBRT.
22Chordoma of the Buttock 30Gy/15f (Post-op.
50.4Gy EBRT 62Gy Ir-192)
23Ca. Prostate 32Gy/16f Boost (Post 40Gy/20f EBRT)
24Field Placement Verification Cranial Case
25Field Placement Verification Cranial Case
26Dosimetric Verification in Phantom Prostate
(a) Calculated
(b) Film
27Dose Distribution Verification Prostate
(a) Calculated
(b) Measured (Film)
28CONCLUSIONS
- 1. Relevant dosimetric data were acquired to
enable clinical use of the system. - 2. Patient immobilization is achieved using the
Talon system for selected cranial cases or an
acquaplast mask for the remainder of sites
treated. - 3. Accuracy of positioning and delivery is
within 1mm for Talon cases and 2mm for mask
cases. This is verified by weekly port films. - 4. The dose/fractionation schemes employed are in
the range 15Gy/1f - 80Gy/40f. - 5. Multiple target (up to 3) can be treated in
one set-up.
29CONCLUSIONS (contd.)
- 6. The range of target volumes treated is 2-500
cm3 - 7. To date 100 patients have been treated.
- 8. Dosimetric accuracy of treatment delivery was
verified by film, TLD and ion chamber
measurements in phantom to be within /- 5. - 9. This approach provides an opportunity for dose
escalation studies. - 10.The system is fitted to a standard linac
without major modifications
30Acknowledgements
- Robert Sanford, PhD, Heeteak Chung and Rina Lim
B.S. for assistance with dosimetric verification
and preparation of the manuscript.