Title: Pulmonary Artery Pressure Measurement
1Pulmonary Artery Pressure Measurement
2Lecture Content
- Patient Positioning
- Identify Phelbostatic Axis
- Leveling
- Square Wave Test
- Obtain PA Reading
-
3Patient Positioning
Angels 45º 30º 0º
4Leveling
- Eliminates effects of hydrostatic forces on the
observed hemodynamic pressures - Ensure air-fluid interface of the transducer is
leveled before zeroing and/or obtaining pressure
readings - Phlebostatic axis
- Level of left atrium
- 4th ICS ½ AP diameter
- Mark the chest with washable felt pen
5Identify Phlebostatic Axis
Intersection of the 4th ICS and ½ the
anterior-posterior diameter of the chest
McHale DL, Carlson KK. AACN Procedure Manual for
Critical Care 4th ed WB Saunders Philadelphia,
Pa 2001 (479) With permission from Elsevier
6Mark location chest wall with washable felt pen
Used with permission of PACEP Collaborative
7Level Transducer System
Ensure air-fluid interface of the transducer is
level to phelbostatic axis.
Used with permission of PACEP Collaborative
8Level Transducer System
- Relevel the transducer with any
- change in the patients position
- Referencing the system 1 cm
- above the left atrium decreases
- the pressure by 0.73 mm Hg
- Referencing the system 1 cm
- below the left atrium increases
- the pressure by 0.73 mm Hg
Angels 45º 30º 0º
9Square Wave Test
- Determines the ability of the
- transducer to correctly reflect
- pressures
- Perform at the beginning of
- each shift
10Square Wave Test
Figure A Expected square wave test
Figure B Over damped
Reprinted from Darovic GO. Hemodynamic
Monitoring Invasive and Noninvasive Clinical
Application 2nd ed. Philadelphia,Pa WB Saunders
Co1995161-162 Used with permission from Elsevier
Figure C Under damped
11Under Damped System
- Over responsive,
- exaggerated, artificially
- spiked waveform
- SBP erroneously high
- DBP erroneously low
- Causes small air
- bubbles, too long of
- tubing, defective
- transducer
Reprinted from Darovic GO. Hemodynamic
Monitoring Invasive and Noninvasive Clinical
Application 2nd ed. Philadelphia,Pa WB Saunders
Co1995161-162 Used with permission from Elsevier
12Over Damped System
- Sluggish, artificially
- rounded blunted
- appearance
- SBP erroneously low
- DBP erroneously high
- Causes large air
- bubbles in system,
- compliant tubing,
- loose/open connections,
- low fluid level in flush bag
Reprinted from Darovic GO. Hemodynamic
Monitoring Invasive and Noninvasive Clinical
Application 2nd ed. Philadelphia,Pa WB Saunders
Co1995161-162 Used with permission from Elsevier
13PAP Documentation
- Measure at end expiration
- Measure pressures from a graphic tracing
- Measure pulmonary capillary wedge pressure at
end-expiration using the mean of the a wave - a wave indicates atrial contraction and falls
- within the P QRS interval of the
corresponding - ECG complex
14Respiratory Component
- Changes in intrathoracic pressure
- during respiration change PAP
- readings
- Record and trend pressure readings at end
expiration
15Used with permission of PACEP Collaborative
16Used with permission of PACEP Collaborative
17Used with permission of PACEP Collaborative
18Respiratory Variation
Spontaneous ventilation
Mechanical ventilation
Aherns TS, Taylor LK. Hemodynamic Waveform
Analysis. Philadelphia, Pa WB Saunders 1992
27 Used with permission from Elsevier
19Used with permission of PACEP Collaborative
20Used with permission of PACEP Collaborative
21Used with permission of PACEP Collaborative
22Used with permission of PACEP Collaborative
23Need Further Assistance?
For more information or further assistance,
please contact a clinical practice specialist
with the AACN Practice Resource Network.
Email practice_at_aacn.org Phone (800) 394-5995,
x217