Title: Comparison and Implications of Three Optical Microscopy Data Acquisition Modalities


1
Comparison and Implications of Three Optical
Microscopy Data Acquisition Modalities
Widefield Fluorescence
Laser Scanning Cytometry (LSC)
Confocal
LSC Images Courtesy CompuCyte Corporation
James Butler Ph.D. Nikon Instruments, Inc.
2
Widefield Fluorescence
Widefield Point Spread Function (PSF)
Tube Lens
Dichromatic Mirror
CCD
0.23µm
  • Light from the Focal Plane --------
  • Light from outside the Focal Plane --------

1.0µm using
60x Plan Apochromat 1.4NA Oil Objective
Objective Lens
Focal Plane
Depth 2-20µm
Specimen
3
Confocal
Confocal PSF
Pinhole
PMT
Scanning Mirrors
  • Confocal or in Focus Fluorescence

0.4µm using
  • Fluorescence from Below Focus
  • Fluorescence from Above Focus

60x Plan Apochromat 1.4NA Oil Objective
  • Laser Point Scanning, e.g. Nikon C1 Confocal
  • Fast Scanning, e.g Nikon Livescan SFC Confocal
  • Other techniques deconvolution, structured
    illumination, TIRF

Depth lt5µm
4
Laser Scanning Cytometry (LSC)
  • Including CompuCyte iCyte, iCys

PMT
  • Scan line width 2.5 to 10µm possibly smaller
    depending on objective used

Scanning Mirror (1 direction line scan)
  • Light from extended depth of field above and
    below objective focal plane gt 20µm

Depth 20-100µm
Stage Translation (gt0.25µm steps) Perpendicular
to Scan line
Laser Scan Image
Confocal Image
5
Observations
  • Data collected is very much dependent on modality
    particularly in z
  • There are benefits/tradeoffs for each of the
    modalities i.e. resolution, speed, quantity of
    data collected, etc.
  • Application specific implications - required data
    may be present regardless of resolution