Title: SPECTROPHOTOMETRY IN BIOTECHNOLOGY
1SPECTROPHOTOMETRY IN BIOTECHNOLOGY
2TOPICS
- Spectrophotometers in Biotechnology
- Light and its Interactions with Matter
- Spectrophotometer Design
- Spectrophotometer Operation
- Qualitative Spectrophotometry
- Quantitative Spectrophotometry
- UV Spectrophotometry of DNA, RNA and Proteins
- Calibration of Spectrophotometers
3(No Transcript)
4(No Transcript)
5- Spectrophotometers in Biotechnology
- Light and its Interactions with Matter
- Spectrophotometer Design
- Spectrophotometer Operation
- Qualitative Spectrophotometry
- Quantitative Spectrophotometry
- UV Spectrophotometry of DNA, RNA and Proteins
- Calibration of Spectrophotometers
6LIGHT IS A TYPE OF ELECTROMAGNETIC RADIATION
- Imagine electromagnetic radiation like waves on a
pond - But instead of water, electromagnetic radiation
is energy moving through space - Distance from one crest to the next is the
wavelength
7WAVELENGTH AND COLOR
- Different wavelengths of light correspond to
different colors - All colors blended together is called white light
- The absence of all light is black
- Light of slightly shorter wavelengths is
ultraviolet - Eyes do not perceive UV light
8(No Transcript)
9WAVELENGTH OF VISIBLE LIGHT AND COLOR
10INTERACTION OF LIGHT WITH MATERIALS IN SOLUTION
- When light shines on a solution, it may pass
through be transmitted or - Some or all of the light energy may be absorbed
11(No Transcript)
12(No Transcript)
13THE ABSORPTION OF LIGHT AND COLOR OF SOLUTIONS
14BIOLOGICAL SOLUTIONS
- Usually appear clear to our eyes have no color
- DNA, RNA, most proteins do not absorb any visible
light - But they do absorb UV light, so UV
spectrophotometers are useful to biologists - Example, can use a detector that measures
absorbance at 280 nm, or 254 nm to detect proteins
15- Spectrophotometers in Biotechnology
- Light and its Interactions with Matter
- Spectrophotometer Design
- Spectrophotometer Operation
- Qualitative Spectrophotometry
- Quantitative Spectrophotometry
- UV Spectrophotometry of DNA, RNA and Proteins
- Calibration of Spectrophotometers
16SPECTROPHOTOMETERS
- Are instruments that measure the interaction of
light with materials in solution
17(No Transcript)
18Monochromator Separates Light into Its Component
Wavelengths. Modern Specs Use Diffraction
Gratings
19- Spectrophotometers in Biotechnology
- Light and its Interactions with Matter
- Spectrophotometer Design
- Spectrophotometer Operation
- Qualitative Spectrophotometry
- Quantitative Spectrophotometry
- UV Spectrophotometry of DNA, RNA and Proteins
- Calibration of Spectrophotometers
20THE BLANK
- Spectrophotometers compare the light transmitted
through a sample to the light transmitted through
a blank. - The blank is treated just like the sample
- The blank contains everything except the analyte
(the material of interest) - Contains solvent
- Contains whatever reagents are added to the
sample
21(No Transcript)
22WHEN OPERATING SPEC
- Blank is inserted into the spectrophotometer
- Instrument is set to 100 transmittance or zero
absorbance
23PROPER SELECTION, USE, AND CARE OF CUVETTES
- Cuvettes are made from plastic, glass, or quartz.
- Use quartz cuvettes for UV work.
- Glass, plastic or quartz are acceptable visible
work. - There are inexpensive plastic cuvettes that may
be suitable for some UV work.
24(No Transcript)
25- 2. Cuvettes are expensive and fragile (except
for disposable plastic ones). Use them properly
and carefully. - a. Do not scratch cuvettes do not store them
in wire racks or clean with brushes or abrasives. - b. Do not allow samples to sit in a cuvette for
a long period of time. - c. Wash cuvettes immediately after use.
-
26- 3. Disposable cuvettes are often recommended for
colorimetric protein assays, since dyes used for
proteins tend to stain cuvettes and are difficult
to remove. - 4. Matched cuvettes are manufactured to absorb
light identically so that one of the pair can be
used for the sample and the other for the blank.
27- 5. Do not touch the base of a cuvette or the
sides through which light is directed. - 6. Make sure the cuvette is properly aligned in
the spectrophotometer. - 7. Be certain to only use clean cuvettes.
28(No Transcript)
29(No Transcript)
30- Spectrophotometers in Biotechnology
- Light and its Interactions with Matter
- Spectrophotometer Design
- Spectrophotometer Operation
- Qualitative Spectrophotometry
- Quantitative Spectrophotometry
- UV Spectrophotometry of DNA, RNA and Proteins
- Calibration of Spectrophotometers
31(No Transcript)
32EXAMPLES
- Some examples of qualitative spectrophotometry
- The absorbance spectra of various common
solvents. Note that some solvents absorb light
at the same wavelengths as DNA, RNA, and proteins - Hemoglobin bound to oxygen versus carbon monoxide
- Native versus denatured bovine serum albumin (a
protein commonly used in the lab)
33(No Transcript)
34(No Transcript)
35- Spectrophotometers in Biotechnology
- Light and its Interactions with Matter
- Spectrophotometer Design
- Spectrophotometer Operation
- Qualitative Spectrophotometry
- Quantitative Spectrophotometry
- UV Spectrophotometry of DNA, RNA and Proteins
- Calibration of Spectrophotometers
36(No Transcript)
37OVERVIEW OF QUANTITIVE SPECTROPHOTOMETRY
- A. Measure the absorbance of standards
containing known concentrations of the analyte - B. Plot a standard curve with absorbance on the
X axis and analyte concentration on the Y axis - C. Measure the absorbance of the unknown(s)
- D. Determine the concentration of material of
interest in the unknowns based on the standard
curve
38(No Transcript)
39LINEAR RANGE
- If there is too much or too little analyte,
spectrophotometer cannot read the absorbance
accurately
40(No Transcript)
41COLORIMETRIC ASSAYS
- Quantitative assays of materials that do not
intrinsically absorb visible light - Combine the sample with reagents that make the
analyte colored - The amount of color is proportional to the amount
of analyte present
42BRADFORD PROTEIN ASSAY
- A quantitative colorimetric assay
- Used to determine the concentration, or amount,
of protein in a sample
43- Prepare standards with known protein
concentrations - Add Bradford Reagent to the samples and to
standards - Read absorbances
- Create a standard curve
- Determine the concentration of protein in the
samples based on the standard curve
44MORE ABOUT THE CALIBRATION LINE ON A STANDARD
CURVE
- Three things determine the absorbance of a
sample - The concentration of analyte in the sample
- The path length through the cuvette
- The intrinsic ability of the analyte to absorb
light at the wavelength of interest
45(No Transcript)
46BEER-LAMBERT LAW
- A ? B C
-
- Where
- A absorbance at a particular wavelength
- ? E absorptivity constant intrinsic
ability of analyte to absorb light at a
particular wavelength - B path length through cuvette
- C concentration of analyte
47APPLYING THE EQUATION
- Suppose you have a sample
- And you know the path length
- And you know the absorptivity constant for the
analyte of interest at a particular wavelength - Then, measure the samples absorbance at the
specified wavelength
48- Can calculate the concentration of the analyte
from the Beer-Lambert equation - A ? B C
- But this is a shortcut that may give inaccurate
results!
49EQUATION FOR A LINE
50- Y intercept should be zero because of the blank
- Blank has no analyte (zero concentration) and is
used to set transmittance to 100 absorbance to
zero
51SLOPE
- Slope relates to the absorptivity constant
- A ? B C
- y m x 0
52DETERMINATION OF THE ABSORPTIVITY CONSTANT
- 1. Prepare a calibration line based on a series
of standards - Plot concentration on the X axis and absorbance
on the Y axis - 2. Calculate the slope of the calibration line
- Y2 Y1
- X2 - X1
53- Determine the path length for the system (assume
1 cm for a standard sample holder and cuvette)
54- A ? B C
- y m x 0
- Slope absorptivity constant X path length
- Absorptivity constant slope
- path length
- (Observe that the constant has units that depend
on how concentration was expressed in the
standards)
55- Spectrophotometers in Biotechnology
- Light and its Interactions with Matter
- Spectrophotometer Design
- Spectrophotometer Operation
- Qualitative Spectrophotometry
- Quantitative Spectrophotometry
- UV Spectrophotometry of DNA, RNA and Proteins
- Calibration of Spectrophotometers
56UV SPECTROPHOTOMETRY
- DNA, RNA and proteins are commonly analyzed with
UV spectrophotometry because these molecules
absorb UV light
57(No Transcript)
58UV METHODS
- Used to evaluate the quality of DNA or RNA in a
sample - Used to estimate the quantity of DNA or RNA in a
sample - Procedure Take single wavelength readings of
samples at 260 and 280 nm
59CONCENTRATION
- The absorbance at 260 nm is related to the
concentration of DNA or RNA in the sample - Pure, double-stranded DNA has an absorbance of
about 1 at 260 nm when it is at a concentration
of about 50 micrograms/mL - Pure, single-stranded DNA has an absorbance of
about 1 at 260 nm when it is at a concentration
of about 33 micrograms/mL - Values for proteins vary considerably from
protein to protein
60PURITY
- The ratio of the absorbance at 260 and 280 nm is
related to the purity of the sample - An A260/A280 ratio of 2.0 is characteristic of
pure RNA - An A260/A280 ratio of 1.8 is characteristic of
pure DNA - An A260/A280 ratio of 0.6 is characteristic of
pure protein
61UV METHODS
- These UV methods for estimating concentration and
purity of DNA, RNA, and proteins are very
commonly used, are very quick, and easy to
perform - However, they values obtained are not very
accurate they are rough estimates
62- Spectrophotometers in Biotechnology
- Light and its Interactions with Matter
- Spectrophotometer Design
- Spectrophotometer Operation
- Qualitative Spectrophotometry
- Quantitative Spectrophotometry
- UV Spectrophotometry of DNA, RNA and Proteins
- Calibration of Spectrophotometers
63CALIBRATION OF A SPECTROPHOTOMETER
- Brings the readings of the spectrophotometer into
accordance with nationally accepted values - Part of routine quality control/maintenance
64CALIBRATION
- Two parts
- 1. Wavelength accuracy, the agreement between
the wavelength selected by the operator and the
actual wavelength of light that shines on sample - 2. Photometric accuracy, or absorbance scale
accuracy, the extent to which a measured
absorbance or transmittance value agrees with an
accepted reference value
65- Wavelength accuracy is determined using certified
standard reference materials (SRMs) available
from NIST or traceable to NIST - An absorbance spectrum for the reference material
is prepared - The absorbance peaks for reference standards are
known, so the wavelengths of the peaks generated
by the instrument can be checked
66- Manufacturers specify the wavelength accuracy of
a given instrument - For example, a high performance instrument may be
specified to have a wavelength accuracy with a
tolerance of 0.5 nm - A less expensive instrument may be specified to
have a wavelength accuracy of 3 nm
67(No Transcript)
68PHOTOMETRIC ACCURACY
- Assures that
- If the absorbance of a given sample is measured
in two spectrophotometers at the same wavelength
and under identical conditions - then the readings will be the same
- and the readings will correspond to nationally
accepted values
69- Photometric accuracy is difficult to achieve due
to different instrument designs and optics - Usually photometric accuracy is not critical if
the same instrument is used consistently and if
its readings are linear and reproducible
70- Photometric accuracy is required where values
from different labs and instruments are compared - Required if rely on published absorptivity
constants - Likely required in a GMP-compliant facility
71PROBLEM
- Assume that a spectrophotometer is able to read
accurately in the range from 0.1 to 1.8 AU. The
molar absorptivity constant for NADH is 15,000
L/mole-cm at 260 nm. Using Beer's Law, calculate
the concentration range of NADH that can be
accurately quantitated at 260 nm based on the
limits of the spectrophotometer.
72ANSWER
- This involves calculation of the molar
concentrations which will produce absorbances of
0.1 and 1.8. From Beer's Law - C A
- e b
- Substituting 0.1 and 1.8 into the equation
73- C 0.1
6.7 X 10-6 mole/L - (15,000 L)(1 cm)
- Mole-cm
- C 1.8
120.0 X 10-6 mole/L - (15,000 L)(1 cm)
- mole-cm
74- Thus, the range of NADH concentrations that can
be detected at this wavelength with this
spectrophotometer is from 6.7 X 10-6 mole/L to
120.0 X 10-6 mole/L. These are very dilute
solutions of NADH.