SPECTROPHOTOMETRY IN BIOTECHNOLOGY - PowerPoint PPT Presentation

1 / 74
About This Presentation
Title:

SPECTROPHOTOMETRY IN BIOTECHNOLOGY

Description:

SPECTROPHOTOMETRY IN BIOTECHNOLOGY TOPICS Spectrophotometers in Biotechnology Light and its Interactions with Matter Spectrophotometer Design Spectrophotometer ... – PowerPoint PPT presentation

Number of Views:492
Avg rating:3.0/5.0
Slides: 75
Provided by: austinccE80
Category:

less

Transcript and Presenter's Notes

Title: SPECTROPHOTOMETRY IN BIOTECHNOLOGY


1
SPECTROPHOTOMETRY IN BIOTECHNOLOGY
2
TOPICS
  • Spectrophotometers in Biotechnology
  • Light and its Interactions with Matter
  • Spectrophotometer Design
  • Spectrophotometer Operation
  • Qualitative Spectrophotometry
  • Quantitative Spectrophotometry
  • UV Spectrophotometry of DNA, RNA and Proteins
  • Calibration of Spectrophotometers

3
(No Transcript)
4
(No Transcript)
5
  • Spectrophotometers in Biotechnology
  • Light and its Interactions with Matter
  • Spectrophotometer Design
  • Spectrophotometer Operation
  • Qualitative Spectrophotometry
  • Quantitative Spectrophotometry
  • UV Spectrophotometry of DNA, RNA and Proteins
  • Calibration of Spectrophotometers

6
LIGHT IS A TYPE OF ELECTROMAGNETIC RADIATION
  • Imagine electromagnetic radiation like waves on a
    pond
  • But instead of water, electromagnetic radiation
    is energy moving through space
  • Distance from one crest to the next is the
    wavelength

7
WAVELENGTH AND COLOR
  • Different wavelengths of light correspond to
    different colors
  • All colors blended together is called white light
  • The absence of all light is black
  • Light of slightly shorter wavelengths is
    ultraviolet
  • Eyes do not perceive UV light

8
(No Transcript)
9
WAVELENGTH OF VISIBLE LIGHT AND COLOR
10
INTERACTION OF LIGHT WITH MATERIALS IN SOLUTION
  • When light shines on a solution, it may pass
    through be transmitted or
  • Some or all of the light energy may be absorbed

11
(No Transcript)
12
(No Transcript)
13
THE ABSORPTION OF LIGHT AND COLOR OF SOLUTIONS
14
BIOLOGICAL SOLUTIONS
  • Usually appear clear to our eyes have no color
  • DNA, RNA, most proteins do not absorb any visible
    light
  • But they do absorb UV light, so UV
    spectrophotometers are useful to biologists
  • Example, can use a detector that measures
    absorbance at 280 nm, or 254 nm to detect proteins

15
  • Spectrophotometers in Biotechnology
  • Light and its Interactions with Matter
  • Spectrophotometer Design
  • Spectrophotometer Operation
  • Qualitative Spectrophotometry
  • Quantitative Spectrophotometry
  • UV Spectrophotometry of DNA, RNA and Proteins
  • Calibration of Spectrophotometers

16
SPECTROPHOTOMETERS
  • Are instruments that measure the interaction of
    light with materials in solution

17
(No Transcript)
18
Monochromator Separates Light into Its Component
Wavelengths. Modern Specs Use Diffraction
Gratings
19
  • Spectrophotometers in Biotechnology
  • Light and its Interactions with Matter
  • Spectrophotometer Design
  • Spectrophotometer Operation
  • Qualitative Spectrophotometry
  • Quantitative Spectrophotometry
  • UV Spectrophotometry of DNA, RNA and Proteins
  • Calibration of Spectrophotometers

20
THE BLANK
  • Spectrophotometers compare the light transmitted
    through a sample to the light transmitted through
    a blank.
  • The blank is treated just like the sample
  • The blank contains everything except the analyte
    (the material of interest)
  • Contains solvent
  • Contains whatever reagents are added to the
    sample

21
(No Transcript)
22
WHEN OPERATING SPEC
  • Blank is inserted into the spectrophotometer
  • Instrument is set to 100 transmittance or zero
    absorbance

23
PROPER SELECTION, USE, AND CARE OF CUVETTES
  • Cuvettes are made from plastic, glass, or quartz.
  • Use quartz cuvettes for UV work.
  • Glass, plastic or quartz are acceptable visible
    work.
  • There are inexpensive plastic cuvettes that may
    be suitable for some UV work.

24
(No Transcript)
25
  • 2. Cuvettes are expensive and fragile (except
    for disposable plastic ones). Use them properly
    and carefully.
  • a. Do not scratch cuvettes do not store them
    in wire racks or clean with brushes or abrasives.
  • b. Do not allow samples to sit in a cuvette for
    a long period of time.
  • c. Wash cuvettes immediately after use.

26
  • 3. Disposable cuvettes are often recommended for
    colorimetric protein assays, since dyes used for
    proteins tend to stain cuvettes and are difficult
    to remove.
  • 4. Matched cuvettes are manufactured to absorb
    light identically so that one of the pair can be
    used for the sample and the other for the blank.

27
  • 5. Do not touch the base of a cuvette or the
    sides through which light is directed.
  • 6. Make sure the cuvette is properly aligned in
    the spectrophotometer.
  • 7. Be certain to only use clean cuvettes.

28
(No Transcript)
29
(No Transcript)
30
  • Spectrophotometers in Biotechnology
  • Light and its Interactions with Matter
  • Spectrophotometer Design
  • Spectrophotometer Operation
  • Qualitative Spectrophotometry
  • Quantitative Spectrophotometry
  • UV Spectrophotometry of DNA, RNA and Proteins
  • Calibration of Spectrophotometers

31
(No Transcript)
32
EXAMPLES
  • Some examples of qualitative spectrophotometry
  • The absorbance spectra of various common
    solvents. Note that some solvents absorb light
    at the same wavelengths as DNA, RNA, and proteins
  • Hemoglobin bound to oxygen versus carbon monoxide
  • Native versus denatured bovine serum albumin (a
    protein commonly used in the lab)

33
(No Transcript)
34
(No Transcript)
35
  • Spectrophotometers in Biotechnology
  • Light and its Interactions with Matter
  • Spectrophotometer Design
  • Spectrophotometer Operation
  • Qualitative Spectrophotometry
  • Quantitative Spectrophotometry
  • UV Spectrophotometry of DNA, RNA and Proteins
  • Calibration of Spectrophotometers

36
(No Transcript)
37
OVERVIEW OF QUANTITIVE SPECTROPHOTOMETRY
  • A. Measure the absorbance of standards
    containing known concentrations of the analyte
  • B. Plot a standard curve with absorbance on the
    X axis and analyte concentration on the Y axis
  • C. Measure the absorbance of the unknown(s)
  • D. Determine the concentration of material of
    interest in the unknowns based on the standard
    curve

38
(No Transcript)
39
LINEAR RANGE
  • If there is too much or too little analyte,
    spectrophotometer cannot read the absorbance
    accurately

40
(No Transcript)
41
COLORIMETRIC ASSAYS
  • Quantitative assays of materials that do not
    intrinsically absorb visible light
  • Combine the sample with reagents that make the
    analyte colored
  • The amount of color is proportional to the amount
    of analyte present

42
BRADFORD PROTEIN ASSAY
  • A quantitative colorimetric assay
  • Used to determine the concentration, or amount,
    of protein in a sample

43
  • Prepare standards with known protein
    concentrations
  • Add Bradford Reagent to the samples and to
    standards
  • Read absorbances
  • Create a standard curve
  • Determine the concentration of protein in the
    samples based on the standard curve

44
MORE ABOUT THE CALIBRATION LINE ON A STANDARD
CURVE
  • Three things determine the absorbance of a
    sample
  • The concentration of analyte in the sample
  • The path length through the cuvette
  • The intrinsic ability of the analyte to absorb
    light at the wavelength of interest

45
(No Transcript)
46
BEER-LAMBERT LAW
  • A ? B C
  • Where
  • A absorbance at a particular wavelength
  • ? E absorptivity constant intrinsic
    ability of analyte to absorb light at a
    particular wavelength
  • B path length through cuvette
  • C concentration of analyte

47
APPLYING THE EQUATION
  • Suppose you have a sample
  • And you know the path length
  • And you know the absorptivity constant for the
    analyte of interest at a particular wavelength
  • Then, measure the samples absorbance at the
    specified wavelength

48
  • Can calculate the concentration of the analyte
    from the Beer-Lambert equation
  • A ? B C
  • But this is a shortcut that may give inaccurate
    results!

49
EQUATION FOR A LINE
  • A ? B C
  • y m x 0

50
  • Y intercept should be zero because of the blank
  • Blank has no analyte (zero concentration) and is
    used to set transmittance to 100 absorbance to
    zero

51
SLOPE
  • Slope relates to the absorptivity constant
  • A ? B C
  • y m x 0

52
DETERMINATION OF THE ABSORPTIVITY CONSTANT
  • 1. Prepare a calibration line based on a series
    of standards
  • Plot concentration on the X axis and absorbance
    on the Y axis
  • 2. Calculate the slope of the calibration line
  • Y2 Y1
  • X2 - X1

53
  • Determine the path length for the system (assume
    1 cm for a standard sample holder and cuvette)

54
  • A ? B C
  • y m x 0
  • Slope absorptivity constant X path length
  • Absorptivity constant slope
  • path length
  • (Observe that the constant has units that depend
    on how concentration was expressed in the
    standards)

55
  • Spectrophotometers in Biotechnology
  • Light and its Interactions with Matter
  • Spectrophotometer Design
  • Spectrophotometer Operation
  • Qualitative Spectrophotometry
  • Quantitative Spectrophotometry
  • UV Spectrophotometry of DNA, RNA and Proteins
  • Calibration of Spectrophotometers

56
UV SPECTROPHOTOMETRY
  • DNA, RNA and proteins are commonly analyzed with
    UV spectrophotometry because these molecules
    absorb UV light

57
(No Transcript)
58
UV METHODS
  • Used to evaluate the quality of DNA or RNA in a
    sample
  • Used to estimate the quantity of DNA or RNA in a
    sample
  • Procedure Take single wavelength readings of
    samples at 260 and 280 nm

59
CONCENTRATION
  • The absorbance at 260 nm is related to the
    concentration of DNA or RNA in the sample
  • Pure, double-stranded DNA has an absorbance of
    about 1 at 260 nm when it is at a concentration
    of about 50 micrograms/mL
  • Pure, single-stranded DNA has an absorbance of
    about 1 at 260 nm when it is at a concentration
    of about 33 micrograms/mL
  • Values for proteins vary considerably from
    protein to protein

60
PURITY
  • The ratio of the absorbance at 260 and 280 nm is
    related to the purity of the sample
  • An A260/A280 ratio of 2.0 is characteristic of
    pure RNA
  • An A260/A280 ratio of 1.8 is characteristic of
    pure DNA
  • An A260/A280 ratio of 0.6 is characteristic of
    pure protein

61
UV METHODS
  • These UV methods for estimating concentration and
    purity of DNA, RNA, and proteins are very
    commonly used, are very quick, and easy to
    perform
  • However, they values obtained are not very
    accurate they are rough estimates

62
  • Spectrophotometers in Biotechnology
  • Light and its Interactions with Matter
  • Spectrophotometer Design
  • Spectrophotometer Operation
  • Qualitative Spectrophotometry
  • Quantitative Spectrophotometry
  • UV Spectrophotometry of DNA, RNA and Proteins
  • Calibration of Spectrophotometers

63
CALIBRATION OF A SPECTROPHOTOMETER
  • Brings the readings of the spectrophotometer into
    accordance with nationally accepted values
  • Part of routine quality control/maintenance

64
CALIBRATION
  • Two parts
  • 1. Wavelength accuracy, the agreement between
    the wavelength selected by the operator and the
    actual wavelength of light that shines on sample
  • 2. Photometric accuracy, or absorbance scale
    accuracy, the extent to which a measured
    absorbance or transmittance value agrees with an
    accepted reference value

65
  • Wavelength accuracy is determined using certified
    standard reference materials (SRMs) available
    from NIST or traceable to NIST
  • An absorbance spectrum for the reference material
    is prepared
  • The absorbance peaks for reference standards are
    known, so the wavelengths of the peaks generated
    by the instrument can be checked

66
  • Manufacturers specify the wavelength accuracy of
    a given instrument
  • For example, a high performance instrument may be
    specified to have a wavelength accuracy with a
    tolerance of 0.5 nm
  • A less expensive instrument may be specified to
    have a wavelength accuracy of 3 nm

67
(No Transcript)
68
PHOTOMETRIC ACCURACY
  • Assures that
  • If the absorbance of a given sample is measured
    in two spectrophotometers at the same wavelength
    and under identical conditions
  • then the readings will be the same
  • and the readings will correspond to nationally
    accepted values

69
  • Photometric accuracy is difficult to achieve due
    to different instrument designs and optics
  • Usually photometric accuracy is not critical if
    the same instrument is used consistently and if
    its readings are linear and reproducible

70
  • Photometric accuracy is required where values
    from different labs and instruments are compared
  • Required if rely on published absorptivity
    constants
  • Likely required in a GMP-compliant facility

71
PROBLEM
  • Assume that a spectrophotometer is able to read
    accurately in the range from 0.1 to 1.8 AU. The
    molar absorptivity constant for NADH is 15,000
    L/mole-cm at 260 nm. Using Beer's Law, calculate
    the concentration range of NADH that can be
    accurately quantitated at 260 nm based on the
    limits of the spectrophotometer.

72
ANSWER
  • This involves calculation of the molar
    concentrations which will produce absorbances of
    0.1 and 1.8. From Beer's Law
  • C A
  • e b
  • Substituting 0.1 and 1.8 into the equation

73
  • C 0.1
    6.7 X 10-6 mole/L
  • (15,000 L)(1 cm)
  • Mole-cm
  • C 1.8
    120.0 X 10-6 mole/L
  • (15,000 L)(1 cm)
  • mole-cm

74
  • Thus, the range of NADH concentrations that can
    be detected at this wavelength with this
    spectrophotometer is from 6.7 X 10-6 mole/L to
    120.0 X 10-6 mole/L. These are very dilute
    solutions of NADH.
Write a Comment
User Comments (0)
About PowerShow.com