Title: FDG-PET/CT Technical Committee


1
FDG-PET/CT Technical Committee
  • Quality Control Subcommittee

2
Goals
  • Bridge the gap between PET/CT scanner
    acceptance/qualification/certification and
    routine (daily) quality control
  • Summarize methods and parameters for evaluating
    daily PET and CT QC on each vendors PET/CT
    scanner
  • Long term goal Establish common standards for
    recording and distributing QC parameters for
    individual patient scans (e.g. extended DICOM
    tags)

3
Process
  1. Identify the daily quality control tests that are
    performed on each vendors PET/CT scanners
  2. Specify the range of acceptable parameters for
    each test as recommended by the vendor
  3. Describe the methods for obtaining QC parameters
    and the output format for each vendor
  4. Ideally, develop standard for daily QC output
    that could be exported with individual patient
    scan, e.g. extended DICOM tags or standard file
    format (CSV, XML)

4
CT Quality Control
  • Siemens automated test (pass/fail)
  • Check HU value of water.
  • Check pixel noise
  • Check tube voltages
  • GE manual test with acceptable range
  • Check mean CT value of water (range 0 3 HU) in
    3 ROIs of uniform section.
  • Check variance of water (range 3 0.4) in 3 ROIs
    of uniform section

5
Siemens PET Quality Control
  • Ge-68 cylinder compared to normalization standard
  • Visually inspect sinograms and fansums.
  • If a sinogram shows a dark or bright streak or a
    fansum block is not uniform, notify physicist
  • If Chi-square gt 2.5, notify physicist. (Original
    limit, adopted as DFCI threshold.)
  • If Chi-square gt 10, call Service. Notify physicist

6
GE PET Quality Control
  • Ge-68 pin source compare daily tests to baseline
    with color coded results (green/yellow/red)
  • Coincidence mean (sensitivity test Average of
    coincidences for all crystals)
  • Coincidence variance
  • Singles mean (Average of singles for all
    crystals).
  • Singles variance
  • Mean deadtime (Average deadtime of all blocks).
  • Timing mean (Average timing adjustment of all
    crystals)
  • Energy shift (Average difference in energy peak
    location)

7
QC Output formats
  • Mostly based on visual display
  • Siemens has some files available in html format
    on local folder
  • GE results can be printed as postscript

8
Resources Needed
  • Information about Philips Daily QC from end users
    and vendor
  • Additional clarifications from GE and Siemens
  • Feedback from RSNA QIBA members
  • Potential interactions with DICOM working group
    to discuss inclusion of QC results