Introductory Lecture Series: The Anesthesia Machine - PowerPoint PPT Presentation

1 / 53
About This Presentation
Title:

Introductory Lecture Series: The Anesthesia Machine

Description:

Introductory Lecture Series: The Anesthesia Machine Kevin Ewing PGY-4 July 7, 2004 * Mention how the machine is divided into high, intermediate, and low pressure ... – PowerPoint PPT presentation

Number of Views:938
Avg rating:3.0/5.0
Slides: 54
Provided by: biobrain1
Category:

less

Transcript and Presenter's Notes

Title: Introductory Lecture Series: The Anesthesia Machine


1
Introductory Lecture Series The Anesthesia
Machine
  • Kevin Ewing PGY-4
  • July 7, 2004

2
Objectives
  • Anesthesia Machine
  • Vaporizers
  • Circuits
  • Ventilators
  • Scavenging Systems
  • System Checkout

3
The Anesthesia Machine
High
Intermediate
Low Pressure Circuit
4
High Pressure System
  • Receives gasses from the high pressure E
    cylinders attached to the back of the anesthesia
    machine (2200 psig for O2, 745 psig for N2O)
  • Consists of
  • Hanger Yolk (reserve gas cylinder holder)
  • Check valve (prevent reverse flow of gas)
  • Cylinder Pressure Indicator (Gauge)
  • Pressure Reducing Device (Regulator)
  • Usually not used, unless pipeline gas supply is
    off

5
E Size Compressed Gas Cylinders
6
Hanger Yolk
  • Hanger Yolk orients and supports the cylinder,
    providing a gas-tight seal and ensuring a
    unidirectional gas flow into the machine
  • Index pins Pin Index Safety System (PISS) is gas
    specific?prevents accidental rearrangement of
    cylinders (e.g.. switching O2 and N2O)

7
Pressure Reducing Device
  • Reduces the high and variable pressures found in
    a cylinder to a lower and more constant pressure
    found in the anesthesia machine (45 psig)
  • Reducing devices are preset so that the machine
    uses only gas from the pipeline (wall gas), when
    the pipeline inlet pressure is 50 psig. This
    prevents gas use from the cylinder even if the
    cylinder is left open (i.e. saves the cylinder
    for backup if the wall gas pipeline fails)
  • Cylinders should be kept closed routinely.
    Otherwise, if the wall gas fails, the machine
    will automatically switch to the cylinder supply
    without the anesthetist being aware that the wall
    supply has failed (until the cylinder is empty
    too).

8
Intermediate Pressure System
  • Receives gasses from the regulator or the
    hospital pipeline at pressures of 40-55 psig
  • Consists of
  • Pipeline inlet connections
  • Pipeline pressure indicators
  • Piping
  • Gas power outlet
  • Master switch
  • Oxygen pressure failure devices
  • Oxygen flush
  • Additional reducing devices
  • Flow control valves

9
Pipeline Inlet Connections
  • Mandatory N2O and O2, usually have air and
    suction too
  • Inlets are non-interchangeable due to specific
    threading as per the Diameter Index Safety System
    (DISS)
  • Each inlet must contain a check valve to prevent
    reverse flow (similar to the cylinder yolk)

10
Oxygen Pressure Failure Devices
  • Machine standard requires that an anesthesia
    machine be designed so that whenever the oxygen
    supply pressure is reduced below normal, the
    oxygen concentration at the common gas outlet
    does not fall below 19
  • A Fail-Safe valve is present in the gas line
    supplying each of the flowmeters except O2. This
    valve is controlled by the O2 supply pressure and
    shuts off or proportionately decreases the supply
    pressure of all other gasses as the O2 supply
    pressure decreases
  • Historically there are 2 kinds of fail-safe
    valves
  • Pressure sensor shut-off valve (Ohmeda)
  • Oxygen failure protection device (Drager)

11
Pressure Sensor Shut-Off Valve
  • The Ohmeda pressure sensor shut-off valve is a
    threshold valve which is either open or closed.
  • Oxygen supply pressure opens the valve as long as
    it is above a pre-set minimum value (e.g.. 20
    psig).
  • If the oxygen supply pressure falls below the
    threshold value the valve closes and the gas in
    that limb (e.g.. N2O), does not advance to its
    flow-control valve.

12
Oxygen Failure Protection Device (OFPD)
  • Based on a proportioning principle rather than a
    shut-off principle
  • The pressure of all gases controlled by the OFPD
    will decrease proportionately with the oxygen
    pressure

13
Oxygen Supply Failure Alarm
  • The machine standard specifies that whenever the
    oxygen supply pressure falls below a
    manufacturer-specified threshold (usually 30
    psig) a medium priority alarm shall blow within 5
    seconds.
  • Electronic alarms A pressure operated electric
    switch operates this alarm\
  • Ohmeda 28 psig
  • Drager 30-37 psig
  • Pneumatic alarms (aka Bowmans Whistle) Uses a
    pressurized canister that is filled with oxygen
    when the anesthesia machine is turned on. When
    the oxygen pressure falls below a certain value,
    the alarm directs a stream of oxygen through a
    whistle

14
Limitations of Fail-Safe Devices/Alarms
  • Fail-safe valves do not prevent administration of
    a hypoxic mixture because they depend on pressure
    and not flow.
  • These devices prevent hypoxia from some problems
    occurring upstream in the machine circuitry
    (disconnected oxygen hose, low oxygen pressure in
    the pipeline and depletion of the oxygen
    cylinder)
  • These devices do not prevent hypoxia from
    accidents such as pipeline crossovers or a
    cylinder containing the wrong gas
  • Equipment problems that occur downstream (for
    example leaks or partial closure of the oxygen
    flow control valve) are not prevented by these
    devices.

15
Oxygen Flush Valve (O2)
  • Receives O2 from pipeline inlet or cylinder
    reducing device and directs high, unmetered flow
    directly to the common gas outlet (downstream of
    the vaporizer)
  • Machine standard requires that the flow be
    between 35 and 75 L/min (AS/3 is 35 L/min)
  • The ability to provide jet ventilation via the O2
    flush valve is presence of a check valve between
    the vaporizer and the O2 flush valve (otherwise
    some flow would be wasted retrograde)
  • Hazards
  • May cause barotrauma
  • Dilution of inhaled anesthetic

16
Second-Stage Reducing Device
  • Located just upstream of the flow control valves
  • Receives gas from the pipeline inlet or the
    cylinder reducing device and reduces it further
    to 26 psig for N2O and 14 psig for O2
  • Purpose is to eliminate fluctuations in pressure
    supplied to the flow indicators caused by
    fluctuations in pipeline pressure

17
Low Pressure System
  • Extends from the flow control valves to the
    common gas outlet
  • Consists of
  • Flow meters
  • Vaporizer mounting device
  • Check valve
  • Common gas outlet

18
Flowmeter assembly
  • When the flow control valve is opened the gas
    enters at the bottom and flows up the tube
    elevating the indicator
  • The indicator floats freely at a point where the
    downward force on it (gravity) equals the upward
    force caused by gas molecules hitting the bottom
    of the float
  • Because the tube is tapered the annular opening
    around the indicator increases with height and
    more gas flows around the float

19
Flowmeter Physics
  • The rate of flow through the flowmeter tube
    depends on 3 things
  • Pressure drop across the constriction The loss
    of energy as gas passes the float is reflected in
    a pressure drop across the float. This pressure
    drop is given by
  • weight of float/cross sectional area
  • Size of annular opening The larger the annular
    opening the greater the flow of gas
  • Physical characteristics of the gas
  • Low Flow Small annular space, therefore flow is
    laminar, therefore flow is a function of gas
    viscosity (Poiseuilles Law)
  • High Flow Large annular space, therefore flow is
    turbulent, therefore the flow is a function of
    gas density (Grahams Law)

20
AS/3 Flowmeter
  • Electric flow measurement devices determine the
    gas flow according to the pressure difference
    caused by a flow restrictor
  • Each flow measurement unit has a
  • Pressure sensor sense pressure before and after
    flow restrictor
  • Flow restrictor causes a pressure drop which is
    measured by the pressure sensor the pressure
    difference is then used to calculate the flow
    through the flow measurement unit
  • There are 4 flow measurement devices in the AS/3
    ADU.
  • Flow control section
  • O2 flow measurement
  • N2O/air flow measurement
  • Vaporizer section
  • Bypass flow measurement
  • Aladin cassette flow

21
Arrangement of the Flow-Indicator Tubes
  • In the presence of a flowmeter leak (either at
    the O ring or the glass of the flow tube) a
    hypoxic mixture is less likely to occur if the O2
    flowmeter is downstream of all other flowmeters
  • In A and B a hypoxic mixture can result because a
    substantial portion of oxygen flow passes through
    the leak, and all nitrous oxide is directed to
    the common gas outlet
  • Note that a leak in the oxygen flowmeter tube
    can cause a hypoxic mixture, even when oxygen is
    located in the downstream position

22
Proportioning Systems
  • Ohmeda
  • Mechanical integration of the N2O and O2
    flow-control valves
  • Automatically intercedes to maintain a minimum
    25 concentration of oxygen with a maximum N2OO2
    ratio of 31
  • The Link-25 Proportion Limiting Control System
    does this by automatically increasing the O2 flow
    to prevent a hypoxic mixture

23
Proportioning Systems
  • Drager
  • The Oxygen Ratio Monitor Controller maintains a
    minimum oxygen concentration of at least 25 /-
    3
  • Back pressure from resistors located in the N20
    and O2 flowmeters provides pneumatic input to the
    N2O slave control valve
  • The value of the O2 resistor is 3-4 times that of
    the N2O resistor, therefore if O2 flow falls
    below 25, the N2O slave control valve reduces
    the flow of N2O

24
AS/3 Proportional Valve
  • Datex-Ohmeda AS/3 Anesthetic Delivery Unit (ADU)
    now uses a linear, electrically driven solenoid
    valve controlled by the CPU.
  • The CPU receives input from the O2 and N2O flow
    measurement devices and sends and electric
    current to the solenoid valve
  • The valve limits the N2O flow to a maximum of 75
    of the total flow (assuring a minimum of 25
    oxygen)

N2O
25
Limitations of Proportioning Systems
  • Machines equipped with proportioning systems can
    still deliver a hypoxic mixture under the
    following conditions
  • Wrong supply gas
  • Defective pneumatics or mechanics (e.g.. The
    Link-25 depends on a properly functioning second
    stage regulator)
  • Leak downstream (e.g.. Broken oxygen flow tube)
  • Inert gas administration Proportioning systems
    generally link only N2O and O2
  • In general, an oxygen analyzer is the only
    machine safety device that can detect these
    problems (gas sampling done at the Y-piece of the
    patient circuit)

26
Low Pressure Limb Check Valve
  • Built into Ohmeda (including the AS/3) machines,
    but not Drager machines, between the vaporizer
    and the common gas outlet to prevent positive
    pressure from the patient circuit from being
    transmitted back into the vaporizer and affecting
    the amount of volatile issued from the vaporizer
  • Necessary to permit jet ventilation from the
    common gas outlet using the O2 flush valve
  • Significant when checking for leaks (see later)

27
Vaporizers
  • A vaporizer is an instrument designed to change a
    liquid anesthetic agent into its vapor and add a
    controlled amount of this vapor to the fresh gas
    flow

28
Vaporizer Physics
  • Partial Pressure The vapor of a liquid exerts
    pressure inside a closed container. That part of
    the total pressure exerted by gasses in a
    container which is due to a specific gas is that
    gass partial pressure
  • Vapor Pressure The highest partial pressure that
    can be exerted by a gas at a given temperature is
    its vapor pressure
  • Vapor pressure is dependent only on 1) the
    identity of the gas and 2) temperature. It is not
    dependent on pressure
  • Boiling Point That temperature at which the
    vapor pressure of the substance equals the
    atmospheric pressure
  • Latent Heat of Vaporization Number of calories
    required to change 1 g of liquid to a vapor
    without a temperature change
  • The energy for this must either come from the
    liquid itself or an outside source. If it comes
    from the liquid itself then the temperature of
    the liquid decreases, therefore its vapor
    pressure decreases, therefore the amount of
    vaporization decreases.
  • Specific Heat Number of calories required to
    increase the temperature of 1 g of a substance by
    1 degree Celsius.
  • This tells us how much heat we need to add to a
    liquid to offset the heat lost by vaporization
  • Allows us to choose vaporizer materials with high
    specific heats which therefore have minimal
    temperature changes with vaporization

29
Classification of Vaporizers
30
Generic Bypass Vaporizer
  • Flow from the flowmeters enters the inlet of the
    vaporizer
  • The function of the concentration control valve
    is to regulate the amount of flow through the
    bypass and vaporizing chambers
  • Splitting Ratio flow though vaporizing
    chamber/flow through bypass chamber
  • Examples include the Tec 3, Tec 4, Tec 5 and the
    Drager 19.1

31
Concentration of Inhaled Anesthetic
  • Depending on the temperature and vapor pressure
    of the inhaled anesthetic, the flow through the
    vaporizing chamber entrains a specific flow of
    inhaled anesthetic vapor
  • The final concentration of inhaled anesthetic at
    the vaporizer outlet is given by
  • entrained anesthetic flow
  • entrained anesthetic flow bypass flow
    vaporizing chamber flow
  • This is a volume concentration (for example,
    the MAC of isoflurane is 1.5 volume )
  • Volume is related to partial pressure in the
    following way
  • (volume gas A/total volume) (partial pressure
    gas A/total pressure)
  • Patient uptake and depth of anesthesia depend on
    partial pressure, not volume concentration

32
Factors That Influence Vaporizer Output
  • Flow Rate The output of the vaporizer is
    generally less than the dial setting at very low
    (lt 200 ml/min) or very high (gt 15 L/min) flows
  • Temperature Automatic temperature compensating
    mechanisms in bypass chambers maintain a constant
    vaporizer output with varying temperatures
  • Back Pressure Intermittent back pressure (e.g.
    positive pressure ventilation causes a higher
    vaporizer output than the dial setting)
  • Atmospheric Pressure Changes in atmospheric
    pressure affect variable bypass vaporizer output
    as measured by volume concentration, but not
    (or very little) as measured by partial pressure
    (lowering atmospheric pressure increases volume
    concentration and vice versa)
  • Carrier Gas Vaporizers are calibrated for 100
    oxygen. Carrier gases other than this result in
    decreased vaporizer output.

33
Desflurane
  • Volatility
  • The vapor pressure of desflurane is 3-4 times
    that of the other inhaled anesthetics that we use
    and its boiling point is 22.8 degrees Celsius
  • Normal flow through a traditional vaporizer would
    vaporize 10-20 times as much desflurane as the
    other volatiles (and thats assuming it doesnt
    boil)
  • Potency and Heat of Vaporization
  • The MAC of desflurane is 4-9 times that of the
    other volatiles, therefore the absolute amount of
    desflurane vaporized over a given time period is
    considerably higher than the other volatiles?this
    would lead to excessive cooling of the desflurane
    vaporizer
  • Desflurane would be very intolerant of this
    cooling because it has a very steep
    temperature/vapor pressure curve

34
Tec-6 Vaporizer
  • Electronically heated and pressurized to achieve
    controlled vaporization of desflurane
  • 2 independent circuits (fresh gas and vaporizer)
  • Vaporizer output is controlled by adjusting the
    concentration control valve (R2)
  • Pressure in the two limbs is equalized by the
    pressure regulating valve

35
Tec-6 and Altitude
  • The Tec-6 works at absolute pressures, therefore
    ambient pressure makes no difference to its
    performance per se. It will accurately deliver
    the desired vol of desflurane, but
  • When this gas is brought to ambient pressure at
    high altitudes, this volume percent will
    represent an absolute decrease, in the partial
    pressure of the anesthetic
  • To compensate for this the following compensation
    must be made
  • Required dial setting Normal dial setting (vol
    ) x (760 mm Hg)/(ambient pressure mmHg)

36
Datex-Ohmeda Aladin Cassette Vaporizer
The ADU identifies the agent type by sensing the
position of identification magnets on the Aladin
cassette?the removable aladin cassette is the
only agent-specific part of the Aladin cassette
vaporizer
37
The Aladin Cassette Vaporizer
  • Agent Control Fresh gas enters the Aladin
    cassette from the upstream flow controls. It is
    divided into two streams by a fixed restrictor in
    the bypass unit
  • Bypass Flow measured by the bypass flow
    measurement unit
  • Cassette Flow measured by the cassette flow
    measurement unit
  • Agent concentration is altered by adjusting the
    cassette outflow. This done by the
    electronically-controlled flow-control valve
    controlled by the CPU
  • The CPU receives input from 6 sources
  • The concentration control dial (that the
    anesthetist sets)
  • The flow measurement units of the bypass chamber
  • The flow measurement unit of the vaporizing
    chamber
  • A temperature sensor located in the vaporizing
    chamber
  • A pressure sensor located in the vaporizing
    chamber
  • The composition of the carrier gases

38
The Aladin Cassette Vaporizer and Desflurane
  • In the AS/3 the same vaporizer is used for
    desflurane as is used for all the other
    volatilesthis poses a problem given desfluranes
    volatility
  • At higher temperatures the cassette becomes
    pressurized (desflurane boils at 22.8
    degrees)?this closes the inlet check valve
    effectively shunting all of the carrier gas
    through the bypass chamber, skipping the
    vaporizer chamber entirely. Under these
    conditions the electronically controlled
    flow-control valve simply meters in the
    appropriate flow of pure desflurane vapor
  • To avoid the problem of cooling due to
    desfluranes high MAC, the vaporizer is equipped
    with a fan which warms the vaporizer back up
    towards room temperature

39
The Circuit Circle System
  • So-called because the components are arranged in
    a circular manner
  • Arrangement is variable, but to prevent
    re-breathing of CO2, the following rules must be
    followed
  • Unidirectional valves between the patient and the
    reservoir bag
  • Fresh-gas-flow cannot enter the circuit between
    the expiratory valve and the patient
  • Adjustable pressure-limiting valve (APL) cannot
    be located between the patient and the
    inspiratory valve

40
The AS/3 Circle System
41
Circle System
  • Advantages
  • Relative stability of inspired concentration
  • Conservation of respiratory moisture and heat
  • Prevention of operating room pollution
  • PaCO2 depends only on ventilation, not fresh gas
    flow
  • Low fresh gas flows can be used
  • Disadvantages
  • Complex design potential for malfunction
  • High resistance (multiple one-way valves)
    higher work of breathing

42
Efficiency of a circuit
  • Efficiency of a circuit refers to conserving dead
    space gas and venting alveolar gas
  • i.e. we want to get rid of alveolar gas which is
    high in CO2 and low in useful gases and keep dead
    space gas which hasn't been used up or polluted
    with CO2
  • In the circle system the dead space extends from
    the patient port of the Y-piece to the partition
  • The most efficient set-up of the circle system is
    to put the APL valve and one-way inspiratory/
    expiratory valves very close to the Y-piece
  • This is because the first gas out of the body
    during exhalation is dead space gas, and the
    latter gasses are alveolar gas. Overflow (i.e.
    venting of excess gas through the APL valve)
    occurs during the latter part of expiration
    during spontaneous ventilation, therefore putting
    the APL close to the patient would vent the
    alveolar gases and conserve the dead space gases
    before they had a chance to mix in the expiratory
    limb of the circuit
  • This isnt done in the AS/3 system because it
    would be too bulky

43
The Adjustable Pressure Limiting (APL) Valve
  • User adjustable valve that releases gases to the
    scavenging system and is intended to provide
    control of the pressure in the breathing system
  • Increased pressure in the breathing system (from
    patient) pushes the diaphragm off its seat
    venting the excess gas into the scavenging system
  • The control knob controls the position of the
    diaphragm
  • Bag-mask Ventilation Valve is usually left
    partially open. During inspiration the bag is
    squeezed pushing gas into the inspiratory limb
    until the pressure relief is reached, opening the
    APL valve. At this point the additional volume
    the patient receives is determined by the
    relative resistances to flow exerted by the
    patient and the APL valve
  • Mechanical Ventilation The APL valve is excluded
    from the circuit when the selector switch is
    changed from manual to automatic ventilation

44
Anesthetic Ventilators
  • Power Source Either compressed gas (pneumatic),
    electricity, or both (AS/3). In the event of a
    power failure older pneumatic ventilators will
    continue to function, but the AS/3 relies on its
    limited battery supply
  • Drive Mechanism Most anesthesia machines are
    pneumatically driven. A compressed gas (O2 or
    air) compresses a bag or bellows, which in turn
    delivers a volume of gas to the patient
  • Note that the AS/3 will automatically switch to
    the gas not in use if the primary driving gas
    fails
  • Bellows classification Direction of bellows
    movement during the expiratory phase determines
    the bellows classification
  • Ascending Bellows Ascend during the expiratory
    phase. Will not fill in the event of a
    disconnection (safer)
  • Descending Bellows Descend during the expiratory
    phase. Will continue up and down movement during
    a disconnection by entraining room air

45
Generic Ascending Bellows Ventilator
  • Bellows physically separates the driving gas
    circuit from the patient gas circuit
  • During the inspiratory phase the driving gas
    enters the bellow chamber resulting in
  • Compression of bellows delivering the anesthetic
    gases within the bellows to the patient
  • Closure of the overflow valve, preventing
    anesthetic gas from escaping into the scavenging
    system
  • During the expiratory phase the driving gas exits
    the bellows chamber.
  • Exhaled gas fills the bellows
  • Excess gas opens the overflow valve (PEEP of 2-3
    cmH2O) allowing scavenging of excess gases to
    occur

46
AS/3 Ascending Bellows Ventilator
47
Scavenging Systems
  • Scavenging Interface Protects the breathing
    circuit or ventilator from excessive positive or
    negative pressure. There are 2 kinds of
    scavenging interfaces
  • Open Contains no valves and is open to the
    atmosphere allowing both positive and negative
    pressure relief
  • Closed Communicates with the atmosphere through
    valves
  • Gas Disposal Assembly This assembly ultimately
    eliminates the waste gas. There are 2 kinds of
    gas disposal assemblies
  • Passive Uses the pressure of the waste gas
    itself to produce flow through the system
  • Active Uses a central vacuum to induce flow in
    the system, moving the waste gas along. A
    negative pressure relief valve is mandatory (in
    addition to positive pressure relief)

48
AS/3 Scavenging System
  • In the AS/3, the scavenging assembly shown in the
    previous slide is internal to the machine
  • Located on the back of the ADU are 5 connectors,
    3 for N2O(1), air(2) and O2(3) pipeline supply,
    one for suction(5) and one for EVAC(4).
  • The EVAC connector is the outlet for the
    scavenging system, and is connected to the LHSC
    central scavenging system via pink hosing

49
Checking Anesthesia Machines
  • Anesthesia Apparatus Checkout Recommendations,
    FDA. 1993.
  • 8 Categories of check
  • Emergency ventilation equipment
  • High-Pressure system
  • Low-Pressure system
  • Scavenging system
  • Breathing system
  • Manual and automatic ventilation system
  • Monitors
  • Final Position

50
Low Pressure Circuit Leak Test
  • Checks the integrity of the anesthesia machine
    from the flow control valves to the common outlet
    (e.g. leaks at flow tubes, O-rings, vaporizer)
  • Two types of leak test (depending on presence or
    absence of check valve)
  • Oxygen Flush Positive-Pressure Leak Test Only
    used in machines without check valves basically
    just pressurize the low pressure circuit using
    the O2 flush valve and look for leak
  • Negative Pressure Leak Test Used in machines
    with or without check valves (i.e. Ohmeda).
    Attach suction bulb to common gas outlet, squeeze
    repeatedly until fully collapsed and ensure that
    it remains collapsed for 10 seconds. Will detect
    leaks as small as 30 ml/min.

51
Circle System Test
  • Evaluates the integrity of the circle breathing
    system, which spans from the common gas outlet to
    the Y-piece.
  • Tests for leaks, valve integrity and obstruction
  • Two parts
  • Leak Test Close the APL valve, occlude the
    Y-piece and pressurize the circuit to 30 cmH2O
    using the O2 flush valve. In the absence of a
    leak the pressure should remain at 30 cmH2O for
    10 seconds
  • Flow Test (tests for valve integrity and absence
    of obstruction) Breathe through the expiratory
    limb (ensuring that you can exhale but not inhale
    against the unidirectional valve) and the
    inspiratory limb (ensuring you can inhale but not
    exhale) looking for proper unidirectional valve
    motion.

52
AS/3 Automated System Check
  • N2O delivery and hypoxic mixture control check
    checks for adequate flow of N2O and for hypoxic
    mixture prevention
  • Agent delivery check checks the function
    (including leakage) of the anesthetic agent
    delivery system
  • AUTO ventilation check
  • Checks the amount of internal and external leak
    (low pressure, bellows, patient circuit). Pass
    leakage less than 150 ml/min
  • Calculates the compressible volume of the patient
    circuit. Pass compressible volume less than 10
    ml/cmH2O
  • MAN ventilation check checks the amount of
    leakage in the bag hose and in the manual bag.
    Pass leakage less than 100 cmH2O
  • Checklist Suction, CO2 absorbent, gas cylinders,
    insp/exp valve, O2 flush, gas monitor

53
The Virtual Anesthesia Machine
  • http//vam.anest.ufl.edu/
Write a Comment
User Comments (0)
About PowerShow.com