Title: Introductory Lecture Series: The Anesthesia Machine
1Introductory Lecture Series The Anesthesia
Machine
- Kevin Ewing PGY-4
- July 7, 2004
2Objectives
- Anesthesia Machine
- Vaporizers
- Circuits
- Ventilators
- Scavenging Systems
- System Checkout
3The Anesthesia Machine
High
Intermediate
Low Pressure Circuit
4High Pressure System
- Receives gasses from the high pressure E
cylinders attached to the back of the anesthesia
machine (2200 psig for O2, 745 psig for N2O) - Consists of
- Hanger Yolk (reserve gas cylinder holder)
- Check valve (prevent reverse flow of gas)
- Cylinder Pressure Indicator (Gauge)
- Pressure Reducing Device (Regulator)
- Usually not used, unless pipeline gas supply is
off
5E Size Compressed Gas Cylinders
6Hanger Yolk
- Hanger Yolk orients and supports the cylinder,
providing a gas-tight seal and ensuring a
unidirectional gas flow into the machine - Index pins Pin Index Safety System (PISS) is gas
specific?prevents accidental rearrangement of
cylinders (e.g.. switching O2 and N2O)
7Pressure Reducing Device
- Reduces the high and variable pressures found in
a cylinder to a lower and more constant pressure
found in the anesthesia machine (45 psig) - Reducing devices are preset so that the machine
uses only gas from the pipeline (wall gas), when
the pipeline inlet pressure is 50 psig. This
prevents gas use from the cylinder even if the
cylinder is left open (i.e. saves the cylinder
for backup if the wall gas pipeline fails) - Cylinders should be kept closed routinely.
Otherwise, if the wall gas fails, the machine
will automatically switch to the cylinder supply
without the anesthetist being aware that the wall
supply has failed (until the cylinder is empty
too).
8Intermediate Pressure System
- Receives gasses from the regulator or the
hospital pipeline at pressures of 40-55 psig - Consists of
- Pipeline inlet connections
- Pipeline pressure indicators
- Piping
- Gas power outlet
- Master switch
- Oxygen pressure failure devices
- Oxygen flush
- Additional reducing devices
- Flow control valves
9Pipeline Inlet Connections
- Mandatory N2O and O2, usually have air and
suction too - Inlets are non-interchangeable due to specific
threading as per the Diameter Index Safety System
(DISS) - Each inlet must contain a check valve to prevent
reverse flow (similar to the cylinder yolk)
10Oxygen Pressure Failure Devices
- Machine standard requires that an anesthesia
machine be designed so that whenever the oxygen
supply pressure is reduced below normal, the
oxygen concentration at the common gas outlet
does not fall below 19 - A Fail-Safe valve is present in the gas line
supplying each of the flowmeters except O2. This
valve is controlled by the O2 supply pressure and
shuts off or proportionately decreases the supply
pressure of all other gasses as the O2 supply
pressure decreases - Historically there are 2 kinds of fail-safe
valves - Pressure sensor shut-off valve (Ohmeda)
- Oxygen failure protection device (Drager)
11Pressure Sensor Shut-Off Valve
- The Ohmeda pressure sensor shut-off valve is a
threshold valve which is either open or closed. - Oxygen supply pressure opens the valve as long as
it is above a pre-set minimum value (e.g.. 20
psig). - If the oxygen supply pressure falls below the
threshold value the valve closes and the gas in
that limb (e.g.. N2O), does not advance to its
flow-control valve.
12Oxygen Failure Protection Device (OFPD)
- Based on a proportioning principle rather than a
shut-off principle - The pressure of all gases controlled by the OFPD
will decrease proportionately with the oxygen
pressure
13Oxygen Supply Failure Alarm
- The machine standard specifies that whenever the
oxygen supply pressure falls below a
manufacturer-specified threshold (usually 30
psig) a medium priority alarm shall blow within 5
seconds. - Electronic alarms A pressure operated electric
switch operates this alarm\ - Ohmeda 28 psig
- Drager 30-37 psig
- Pneumatic alarms (aka Bowmans Whistle) Uses a
pressurized canister that is filled with oxygen
when the anesthesia machine is turned on. When
the oxygen pressure falls below a certain value,
the alarm directs a stream of oxygen through a
whistle
14Limitations of Fail-Safe Devices/Alarms
- Fail-safe valves do not prevent administration of
a hypoxic mixture because they depend on pressure
and not flow. - These devices prevent hypoxia from some problems
occurring upstream in the machine circuitry
(disconnected oxygen hose, low oxygen pressure in
the pipeline and depletion of the oxygen
cylinder) - These devices do not prevent hypoxia from
accidents such as pipeline crossovers or a
cylinder containing the wrong gas - Equipment problems that occur downstream (for
example leaks or partial closure of the oxygen
flow control valve) are not prevented by these
devices.
15Oxygen Flush Valve (O2)
- Receives O2 from pipeline inlet or cylinder
reducing device and directs high, unmetered flow
directly to the common gas outlet (downstream of
the vaporizer) - Machine standard requires that the flow be
between 35 and 75 L/min (AS/3 is 35 L/min) - The ability to provide jet ventilation via the O2
flush valve is presence of a check valve between
the vaporizer and the O2 flush valve (otherwise
some flow would be wasted retrograde) - Hazards
- May cause barotrauma
- Dilution of inhaled anesthetic
16Second-Stage Reducing Device
- Located just upstream of the flow control valves
- Receives gas from the pipeline inlet or the
cylinder reducing device and reduces it further
to 26 psig for N2O and 14 psig for O2 - Purpose is to eliminate fluctuations in pressure
supplied to the flow indicators caused by
fluctuations in pipeline pressure
17Low Pressure System
- Extends from the flow control valves to the
common gas outlet - Consists of
- Flow meters
- Vaporizer mounting device
- Check valve
- Common gas outlet
18Flowmeter assembly
- When the flow control valve is opened the gas
enters at the bottom and flows up the tube
elevating the indicator - The indicator floats freely at a point where the
downward force on it (gravity) equals the upward
force caused by gas molecules hitting the bottom
of the float - Because the tube is tapered the annular opening
around the indicator increases with height and
more gas flows around the float
19Flowmeter Physics
- The rate of flow through the flowmeter tube
depends on 3 things - Pressure drop across the constriction The loss
of energy as gas passes the float is reflected in
a pressure drop across the float. This pressure
drop is given by - weight of float/cross sectional area
- Size of annular opening The larger the annular
opening the greater the flow of gas - Physical characteristics of the gas
- Low Flow Small annular space, therefore flow is
laminar, therefore flow is a function of gas
viscosity (Poiseuilles Law) - High Flow Large annular space, therefore flow is
turbulent, therefore the flow is a function of
gas density (Grahams Law)
20AS/3 Flowmeter
- Electric flow measurement devices determine the
gas flow according to the pressure difference
caused by a flow restrictor - Each flow measurement unit has a
- Pressure sensor sense pressure before and after
flow restrictor - Flow restrictor causes a pressure drop which is
measured by the pressure sensor the pressure
difference is then used to calculate the flow
through the flow measurement unit - There are 4 flow measurement devices in the AS/3
ADU. - Flow control section
- O2 flow measurement
- N2O/air flow measurement
- Vaporizer section
- Bypass flow measurement
- Aladin cassette flow
21Arrangement of the Flow-Indicator Tubes
- In the presence of a flowmeter leak (either at
the O ring or the glass of the flow tube) a
hypoxic mixture is less likely to occur if the O2
flowmeter is downstream of all other flowmeters - In A and B a hypoxic mixture can result because a
substantial portion of oxygen flow passes through
the leak, and all nitrous oxide is directed to
the common gas outlet - Note that a leak in the oxygen flowmeter tube
can cause a hypoxic mixture, even when oxygen is
located in the downstream position
22Proportioning Systems
- Ohmeda
- Mechanical integration of the N2O and O2
flow-control valves - Automatically intercedes to maintain a minimum
25 concentration of oxygen with a maximum N2OO2
ratio of 31 - The Link-25 Proportion Limiting Control System
does this by automatically increasing the O2 flow
to prevent a hypoxic mixture
23Proportioning Systems
- Drager
- The Oxygen Ratio Monitor Controller maintains a
minimum oxygen concentration of at least 25 /-
3 - Back pressure from resistors located in the N20
and O2 flowmeters provides pneumatic input to the
N2O slave control valve - The value of the O2 resistor is 3-4 times that of
the N2O resistor, therefore if O2 flow falls
below 25, the N2O slave control valve reduces
the flow of N2O
24AS/3 Proportional Valve
- Datex-Ohmeda AS/3 Anesthetic Delivery Unit (ADU)
now uses a linear, electrically driven solenoid
valve controlled by the CPU. - The CPU receives input from the O2 and N2O flow
measurement devices and sends and electric
current to the solenoid valve - The valve limits the N2O flow to a maximum of 75
of the total flow (assuring a minimum of 25
oxygen)
N2O
25Limitations of Proportioning Systems
- Machines equipped with proportioning systems can
still deliver a hypoxic mixture under the
following conditions - Wrong supply gas
- Defective pneumatics or mechanics (e.g.. The
Link-25 depends on a properly functioning second
stage regulator) - Leak downstream (e.g.. Broken oxygen flow tube)
- Inert gas administration Proportioning systems
generally link only N2O and O2 - In general, an oxygen analyzer is the only
machine safety device that can detect these
problems (gas sampling done at the Y-piece of the
patient circuit)
26 Low Pressure Limb Check Valve
- Built into Ohmeda (including the AS/3) machines,
but not Drager machines, between the vaporizer
and the common gas outlet to prevent positive
pressure from the patient circuit from being
transmitted back into the vaporizer and affecting
the amount of volatile issued from the vaporizer - Necessary to permit jet ventilation from the
common gas outlet using the O2 flush valve - Significant when checking for leaks (see later)
27Vaporizers
- A vaporizer is an instrument designed to change a
liquid anesthetic agent into its vapor and add a
controlled amount of this vapor to the fresh gas
flow
28Vaporizer Physics
- Partial Pressure The vapor of a liquid exerts
pressure inside a closed container. That part of
the total pressure exerted by gasses in a
container which is due to a specific gas is that
gass partial pressure - Vapor Pressure The highest partial pressure that
can be exerted by a gas at a given temperature is
its vapor pressure - Vapor pressure is dependent only on 1) the
identity of the gas and 2) temperature. It is not
dependent on pressure - Boiling Point That temperature at which the
vapor pressure of the substance equals the
atmospheric pressure - Latent Heat of Vaporization Number of calories
required to change 1 g of liquid to a vapor
without a temperature change - The energy for this must either come from the
liquid itself or an outside source. If it comes
from the liquid itself then the temperature of
the liquid decreases, therefore its vapor
pressure decreases, therefore the amount of
vaporization decreases. - Specific Heat Number of calories required to
increase the temperature of 1 g of a substance by
1 degree Celsius. - This tells us how much heat we need to add to a
liquid to offset the heat lost by vaporization - Allows us to choose vaporizer materials with high
specific heats which therefore have minimal
temperature changes with vaporization
29Classification of Vaporizers
30Generic Bypass Vaporizer
- Flow from the flowmeters enters the inlet of the
vaporizer - The function of the concentration control valve
is to regulate the amount of flow through the
bypass and vaporizing chambers - Splitting Ratio flow though vaporizing
chamber/flow through bypass chamber - Examples include the Tec 3, Tec 4, Tec 5 and the
Drager 19.1
31Concentration of Inhaled Anesthetic
- Depending on the temperature and vapor pressure
of the inhaled anesthetic, the flow through the
vaporizing chamber entrains a specific flow of
inhaled anesthetic vapor - The final concentration of inhaled anesthetic at
the vaporizer outlet is given by - entrained anesthetic flow
- entrained anesthetic flow bypass flow
vaporizing chamber flow - This is a volume concentration (for example,
the MAC of isoflurane is 1.5 volume ) - Volume is related to partial pressure in the
following way - (volume gas A/total volume) (partial pressure
gas A/total pressure) - Patient uptake and depth of anesthesia depend on
partial pressure, not volume concentration
32Factors That Influence Vaporizer Output
- Flow Rate The output of the vaporizer is
generally less than the dial setting at very low
(lt 200 ml/min) or very high (gt 15 L/min) flows - Temperature Automatic temperature compensating
mechanisms in bypass chambers maintain a constant
vaporizer output with varying temperatures - Back Pressure Intermittent back pressure (e.g.
positive pressure ventilation causes a higher
vaporizer output than the dial setting) - Atmospheric Pressure Changes in atmospheric
pressure affect variable bypass vaporizer output
as measured by volume concentration, but not
(or very little) as measured by partial pressure
(lowering atmospheric pressure increases volume
concentration and vice versa) - Carrier Gas Vaporizers are calibrated for 100
oxygen. Carrier gases other than this result in
decreased vaporizer output.
33Desflurane
- Volatility
- The vapor pressure of desflurane is 3-4 times
that of the other inhaled anesthetics that we use
and its boiling point is 22.8 degrees Celsius - Normal flow through a traditional vaporizer would
vaporize 10-20 times as much desflurane as the
other volatiles (and thats assuming it doesnt
boil) - Potency and Heat of Vaporization
- The MAC of desflurane is 4-9 times that of the
other volatiles, therefore the absolute amount of
desflurane vaporized over a given time period is
considerably higher than the other volatiles?this
would lead to excessive cooling of the desflurane
vaporizer - Desflurane would be very intolerant of this
cooling because it has a very steep
temperature/vapor pressure curve
34Tec-6 Vaporizer
- Electronically heated and pressurized to achieve
controlled vaporization of desflurane - 2 independent circuits (fresh gas and vaporizer)
- Vaporizer output is controlled by adjusting the
concentration control valve (R2) - Pressure in the two limbs is equalized by the
pressure regulating valve
35Tec-6 and Altitude
- The Tec-6 works at absolute pressures, therefore
ambient pressure makes no difference to its
performance per se. It will accurately deliver
the desired vol of desflurane, but - When this gas is brought to ambient pressure at
high altitudes, this volume percent will
represent an absolute decrease, in the partial
pressure of the anesthetic - To compensate for this the following compensation
must be made - Required dial setting Normal dial setting (vol
) x (760 mm Hg)/(ambient pressure mmHg)
36Datex-Ohmeda Aladin Cassette Vaporizer
The ADU identifies the agent type by sensing the
position of identification magnets on the Aladin
cassette?the removable aladin cassette is the
only agent-specific part of the Aladin cassette
vaporizer
37The Aladin Cassette Vaporizer
- Agent Control Fresh gas enters the Aladin
cassette from the upstream flow controls. It is
divided into two streams by a fixed restrictor in
the bypass unit - Bypass Flow measured by the bypass flow
measurement unit - Cassette Flow measured by the cassette flow
measurement unit - Agent concentration is altered by adjusting the
cassette outflow. This done by the
electronically-controlled flow-control valve
controlled by the CPU - The CPU receives input from 6 sources
- The concentration control dial (that the
anesthetist sets) - The flow measurement units of the bypass chamber
- The flow measurement unit of the vaporizing
chamber - A temperature sensor located in the vaporizing
chamber - A pressure sensor located in the vaporizing
chamber - The composition of the carrier gases
-
38The Aladin Cassette Vaporizer and Desflurane
- In the AS/3 the same vaporizer is used for
desflurane as is used for all the other
volatilesthis poses a problem given desfluranes
volatility - At higher temperatures the cassette becomes
pressurized (desflurane boils at 22.8
degrees)?this closes the inlet check valve
effectively shunting all of the carrier gas
through the bypass chamber, skipping the
vaporizer chamber entirely. Under these
conditions the electronically controlled
flow-control valve simply meters in the
appropriate flow of pure desflurane vapor - To avoid the problem of cooling due to
desfluranes high MAC, the vaporizer is equipped
with a fan which warms the vaporizer back up
towards room temperature
39The Circuit Circle System
- So-called because the components are arranged in
a circular manner - Arrangement is variable, but to prevent
re-breathing of CO2, the following rules must be
followed - Unidirectional valves between the patient and the
reservoir bag - Fresh-gas-flow cannot enter the circuit between
the expiratory valve and the patient - Adjustable pressure-limiting valve (APL) cannot
be located between the patient and the
inspiratory valve
40The AS/3 Circle System
41Circle System
- Advantages
- Relative stability of inspired concentration
- Conservation of respiratory moisture and heat
- Prevention of operating room pollution
- PaCO2 depends only on ventilation, not fresh gas
flow - Low fresh gas flows can be used
- Disadvantages
- Complex design potential for malfunction
- High resistance (multiple one-way valves)
higher work of breathing
42Efficiency of a circuit
- Efficiency of a circuit refers to conserving dead
space gas and venting alveolar gas - i.e. we want to get rid of alveolar gas which is
high in CO2 and low in useful gases and keep dead
space gas which hasn't been used up or polluted
with CO2 - In the circle system the dead space extends from
the patient port of the Y-piece to the partition - The most efficient set-up of the circle system is
to put the APL valve and one-way inspiratory/
expiratory valves very close to the Y-piece - This is because the first gas out of the body
during exhalation is dead space gas, and the
latter gasses are alveolar gas. Overflow (i.e.
venting of excess gas through the APL valve)
occurs during the latter part of expiration
during spontaneous ventilation, therefore putting
the APL close to the patient would vent the
alveolar gases and conserve the dead space gases
before they had a chance to mix in the expiratory
limb of the circuit - This isnt done in the AS/3 system because it
would be too bulky
43The Adjustable Pressure Limiting (APL) Valve
- User adjustable valve that releases gases to the
scavenging system and is intended to provide
control of the pressure in the breathing system - Increased pressure in the breathing system (from
patient) pushes the diaphragm off its seat
venting the excess gas into the scavenging system - The control knob controls the position of the
diaphragm - Bag-mask Ventilation Valve is usually left
partially open. During inspiration the bag is
squeezed pushing gas into the inspiratory limb
until the pressure relief is reached, opening the
APL valve. At this point the additional volume
the patient receives is determined by the
relative resistances to flow exerted by the
patient and the APL valve - Mechanical Ventilation The APL valve is excluded
from the circuit when the selector switch is
changed from manual to automatic ventilation
44Anesthetic Ventilators
- Power Source Either compressed gas (pneumatic),
electricity, or both (AS/3). In the event of a
power failure older pneumatic ventilators will
continue to function, but the AS/3 relies on its
limited battery supply - Drive Mechanism Most anesthesia machines are
pneumatically driven. A compressed gas (O2 or
air) compresses a bag or bellows, which in turn
delivers a volume of gas to the patient - Note that the AS/3 will automatically switch to
the gas not in use if the primary driving gas
fails - Bellows classification Direction of bellows
movement during the expiratory phase determines
the bellows classification - Ascending Bellows Ascend during the expiratory
phase. Will not fill in the event of a
disconnection (safer) - Descending Bellows Descend during the expiratory
phase. Will continue up and down movement during
a disconnection by entraining room air
45Generic Ascending Bellows Ventilator
- Bellows physically separates the driving gas
circuit from the patient gas circuit - During the inspiratory phase the driving gas
enters the bellow chamber resulting in - Compression of bellows delivering the anesthetic
gases within the bellows to the patient - Closure of the overflow valve, preventing
anesthetic gas from escaping into the scavenging
system - During the expiratory phase the driving gas exits
the bellows chamber. - Exhaled gas fills the bellows
- Excess gas opens the overflow valve (PEEP of 2-3
cmH2O) allowing scavenging of excess gases to
occur
46AS/3 Ascending Bellows Ventilator
47Scavenging Systems
- Scavenging Interface Protects the breathing
circuit or ventilator from excessive positive or
negative pressure. There are 2 kinds of
scavenging interfaces - Open Contains no valves and is open to the
atmosphere allowing both positive and negative
pressure relief - Closed Communicates with the atmosphere through
valves - Gas Disposal Assembly This assembly ultimately
eliminates the waste gas. There are 2 kinds of
gas disposal assemblies - Passive Uses the pressure of the waste gas
itself to produce flow through the system - Active Uses a central vacuum to induce flow in
the system, moving the waste gas along. A
negative pressure relief valve is mandatory (in
addition to positive pressure relief)
48AS/3 Scavenging System
- In the AS/3, the scavenging assembly shown in the
previous slide is internal to the machine - Located on the back of the ADU are 5 connectors,
3 for N2O(1), air(2) and O2(3) pipeline supply,
one for suction(5) and one for EVAC(4). - The EVAC connector is the outlet for the
scavenging system, and is connected to the LHSC
central scavenging system via pink hosing
49Checking Anesthesia Machines
- Anesthesia Apparatus Checkout Recommendations,
FDA. 1993. - 8 Categories of check
- Emergency ventilation equipment
- High-Pressure system
- Low-Pressure system
- Scavenging system
- Breathing system
- Manual and automatic ventilation system
- Monitors
- Final Position
50Low Pressure Circuit Leak Test
- Checks the integrity of the anesthesia machine
from the flow control valves to the common outlet
(e.g. leaks at flow tubes, O-rings, vaporizer) - Two types of leak test (depending on presence or
absence of check valve) - Oxygen Flush Positive-Pressure Leak Test Only
used in machines without check valves basically
just pressurize the low pressure circuit using
the O2 flush valve and look for leak - Negative Pressure Leak Test Used in machines
with or without check valves (i.e. Ohmeda).
Attach suction bulb to common gas outlet, squeeze
repeatedly until fully collapsed and ensure that
it remains collapsed for 10 seconds. Will detect
leaks as small as 30 ml/min.
51Circle System Test
- Evaluates the integrity of the circle breathing
system, which spans from the common gas outlet to
the Y-piece. - Tests for leaks, valve integrity and obstruction
- Two parts
- Leak Test Close the APL valve, occlude the
Y-piece and pressurize the circuit to 30 cmH2O
using the O2 flush valve. In the absence of a
leak the pressure should remain at 30 cmH2O for
10 seconds - Flow Test (tests for valve integrity and absence
of obstruction) Breathe through the expiratory
limb (ensuring that you can exhale but not inhale
against the unidirectional valve) and the
inspiratory limb (ensuring you can inhale but not
exhale) looking for proper unidirectional valve
motion.
52AS/3 Automated System Check
- N2O delivery and hypoxic mixture control check
checks for adequate flow of N2O and for hypoxic
mixture prevention - Agent delivery check checks the function
(including leakage) of the anesthetic agent
delivery system - AUTO ventilation check
- Checks the amount of internal and external leak
(low pressure, bellows, patient circuit). Pass
leakage less than 150 ml/min - Calculates the compressible volume of the patient
circuit. Pass compressible volume less than 10
ml/cmH2O - MAN ventilation check checks the amount of
leakage in the bag hose and in the manual bag.
Pass leakage less than 100 cmH2O - Checklist Suction, CO2 absorbent, gas cylinders,
insp/exp valve, O2 flush, gas monitor
53The Virtual Anesthesia Machine