An introduction to Ultraviolet/Visible Absorption Spectroscopy - PowerPoint PPT Presentation

1 / 100
About This Presentation
Title:

An introduction to Ultraviolet/Visible Absorption Spectroscopy

Description:

An introduction to Ultraviolet/Visible Absorption Spectroscopy Chapter 13 In this chapter, absorption by molecules, rather than atoms, is considered. – PowerPoint PPT presentation

Number of Views:651
Avg rating:3.0/5.0
Slides: 101
Provided by: monzirpal
Category:

less

Transcript and Presenter's Notes

Title: An introduction to Ultraviolet/Visible Absorption Spectroscopy


1
An introduction to Ultraviolet/Visible Absorption
Spectroscopy
  • Chapter 13

2
  • In this chapter, absorption by molecules, rather
    than atoms, is considered. Absorption in the
    ultraviolet and visible regions occurs due to
    electronic transitions from the ground state to
    excited state. Broad band spectra are obtained
    since molecules have vibrational and rotational
    energy levels associated with electronic energy
    levels. The signal is either absorbance or
    percent transmittance of the analyte solution
    where

3
An Introduction to Ultraviolet/VisibleMolecular
Absorption Spectrometry
  • Absorption measurements based upon ultraviolet
    and visible radiation find widespread application
    for the quantitative determination of a large
    variety species.
  • Beers Law
  • A -logT logP0/P ?bc
  • A absorbance
  • ? molar absorptivity M-1 cm-1
  • c concentration M
  • P0 incident power
  • P transmitted power (after passing through
    sample)

4
(No Transcript)
5
(No Transcript)
6
(No Transcript)
7
  • Measurement of Transmittance and Absorbance
  • The power of the beam transmitted by the analyte
    solution is usually compared with the power of
    the beam transmitted by an identical cell
    containing only solvent. An experimental
    transmittance and absorbance are then obtained
    with the equations.
  • P0 and P refers to the power of radiation after
    it has passed through the solvent and the
    analyte.

8
Beers law and mixtures
  • Each analyte present in the solution absorbs
    light!
  • The magnitude of the absorption depends on its e
  • A total A1A2An
  • A total e1bc1e2bc2enbcn
  • If e1 e2 en then simultaneous determination
    is impossible
  • Need nls where es are different to solve the
    mixture

9
Limitations to Beers Law
  • Real limitations
  • Chemical deviations
  • Instrumental deviations

10
  • 1.      Real Limitations
  •  
  • a.       Beers law is good for dilute analyte
    solutions only. High concentrations (gt0.01M) will
    cause a negative error since as the distance
    between molecules become smaller the charge
    distribution will be affected which alter the
    molecules ability to absorb a specific
    wavelength. The same phenomenon is also observed
    for solutions with high electrolyte
    concentration, even at low analyte concentration.
    The molar absorptivity is altered due to
    electrostatic interactions.

11
  • b.      In the derivation of Beers law we have
    introduced a constant (e). However, e is
    dependent on the refractive index and the
    refractive index is a function of concentration.
    Therefore, e will be concentration dependent.
    However, the refractive index changes very
    slightly for dilute solutions and thus we can
    practically assume that e is constant.
  • c.       In rare cases, the molar absorptivity
    changes widely with concentration, even at dilute
    solutions. Therefore, Beers law is never a
    linear relation for such compounds, like
    methylene blue.

12
  • 2.      Chemical Deviations
  •  
  • This factor is an important one which largely
    affects linearity in Beers law. It originates
    when an analyte dissociates, associates, or
    reacts in the solvent. For example, an acid base
    indicator when dissolved in water will partially
    dissociate according to its acid dissociation
    constant

13
(No Transcript)
14
(No Transcript)
15
(No Transcript)
16
(No Transcript)
17
  • Chemical deviations from Beers law for
    unbuffered solutions of the indicator Hln. Note
    that there are positive deviations at 430 nm
    and negative deviations at 570 nm. At 430 nm, the
    absorbance is primarily due to the ionized In-
    form of the indicator and is proportional to the
    fraction ionized, which varies nonlinearly with
    the total indicator concentration. At 570 nm, the
    absorbance is due principally to the
    undissociated acid Hln, which increases
    nonlinearly with the total concentration.

18
Calculated Absorbance Data for Various Indicator
Concentrations
19
(No Transcript)
20
  • 3.      Instrumental Deviations
  •  
  • a.       Beers law is good for monochromatic
    light only since e is wavelength dependent. It is
    enough to assume a dichromatic beam passing
    through a sample to appreciate the need for a
    monochromatic light. Assume that the radiant
    power of incident radiation is Po and Po while
    transmitted power is P and P. The absorbance of
    solution can be written as

21
(No Transcript)
22
(No Transcript)
23
  • The effect of polychromatic radiation on Beers
    law. In the spectrum at the top, the absorptivity
    of the analyte is nearly constant over Band A
    from the source. Note in the Beers law plot at
    the bottom that using Band A gives a linear
    relationship. In the spectrum, Band B corresponds
    to a region where the absorptivity shows
    substantial changes. In the lower plot, note the
    dramatic deviation from Beers law that results.

24
(No Transcript)
25
  • Therefore, the linearity between absorbance and
    concentration breaks down if incident radiation
    was polychromatic. In most cases with UV-Vis
    spectroscopy, the effect small changes in
    wavelengths is insignificant since e differs only
    slightly especially at the wavelength maximum.

26
  • b.      Stray Radiation
  •  
  • Stray radiation resulting from scattering or
    various reflections in the instrument will reach
    the detector without passing through the sample.
    The problem can be severe in cases of high
    absorbance or when the wavelengths of stray
    radiation is in such a range where the detector
    is highly sensitive as well as at wavelengths
    extremes of an instrument. The absorbance
    recorded can be represented by the relation
  • A log (Po Ps)/(P Ps)
  • Where Ps is the radiant power of stray radiation.

27
(No Transcript)
28
Instrumental Noise as a Function in Transmittance
29
(No Transcript)
30
(No Transcript)
31
(No Transcript)
32
  • Therefore, an absorbance between 0.2-0.7 may be
    advantageous in terms of a lower uncertainty in
    concentration measurements. At higher or lower
    absorbances, an increase in uncertainty is
    encountered. It is therefore advised that the
    test solution be in the concentration range which
    gives an absorbance value in the range from
    0.2-0.7 for best precision.
  • However, it should also be remembered that we
    ended up with this conclusion provided that sT is
    constant. Unfortunately, sT is not always
    constant which complicates the conclusions above.

33
EFFECT OF bandwidth WIDTH
  • Effect of bandwidth on spectral detail for a
    sample of benzene vapor. Note that as the
    spectral bandwidth increases, the fine structure
    in the spectrum is lost. At a bandwidth of 10 nm,
    only a broad absorption band is observed.

34
  • Effect of slit width (spectral bandwidth) on peak
    heights. Here, the sample was s solution of
    praseodymium chloride. Note that as the spectral
    bandwidth decreases by decreasing the slit width
    from 1.0 mm to 0.1 mm, the peak heights increase.

35
Effect of Scattered Radiation at Wavelength
Extremes of an Instrument
  • Wavelength extremes of an instrument are
    dependent on type of source, detector and optical
    components used in the manufacture of the
    instrument. Outside the working range of the
    instrument, it is not possible to use it for
    accurate determinations. However, the extremes of
    the instrument are very close to the region of
    invalid instrumental performance and would thus
    be not very accurate. An example may be a visible
    photometer which, in principle, can be used in
    the range from 340-780 nm. It may be obvious that
    glass windows, cells and prism will start to
    absorb significantly below 380 nm and thus a
    decrease in the incident radiant power is
    significant.

36
B UV-VIS spectrophotometer A VIS
spectrophotometer
EFFECT OF SCATTERED RADIATION
  • Spectrum of cerium (IV) obtained with a
    spectrophotometer having glass optics (A) and
    quartz optics (B). The false peak in A arises
    from transmission of stray radiation of longer
    wavelengths.

37
  • The output from the source at the low wavelength
    range is minimal. Also, the detector has best
    sensitivities around 550 nm which means that away
    up and down this value, the sensitivity
    significantly decrease. However, scattered
    radiation, and stray radiation in general, will
    reach the detector without passing through these
    surfaces as well as these radiation are
    constituted from wavelengths for which the
    detector is highly sensitive. In some cases,
    stray and scattered radiation reaching the
    detector can be far more intense than the
    monochromatic beam from the source. False peaks
    may appear in such cases and one should be aware
    of this cause of such peaks.

38
  • Instrumentation
  • Light source
  • ? - selection
  • Sample container
  • Detector
  • Signal processing
  • Light Sources (commercial instruments)
  • D2 lamp (UV 160 375 nm)
  • W lamp (vis 350 2500 nm)

39
SourcesDeuterium and hydrogen lamps (160 375
nm)
  • D2 Ee ? D2 ? D D h?

Excited deuterium molecule with fixed quantized
energy
Dissociated into two deuterium atoms with
different kinetic energies
Ee ED2 ED ED hv
Ee is the electrical energy absorbed by the
molecule. ED2 is the fixed quantized energy of
D2, ED and ED are kinetic energy of the two
deuterium atoms.
40
(No Transcript)
41
Sources
Deuterium lamp UV region
  • (a) A deuterium lamp of the type used in
    spectrophotometers and (b)
  • its spectrum. The plot is of irradiance E?
    (proportional to radiant power) versus
  • wavelength. Note that the maximum intensity
    occurs at 225 m.Typically,
  • instruments switch from deuterium to tungsten at
    350 nm.

42
Visible and near-IR region
  • (a) A tungsten lamp of the
  • type used in spectroscopy
  • and its spectrum (b).
  • Intensity of the tungsten
  • source is usually quite low
  • at wavelengths shorter than about 350 nm. Note
    that the intensity reaches a maximum in the
    near-IR
  • region of the spectrum
  • (1200 nm in this case).

43
  • The tungsten lamp is by far the most common
    source in the visible and near IR region with a
    continuum output wavelength in the range from
    350-2500 nm. The lamp is formed from a tungsten
    filament heated to about 3000 oC housed in a
    glass envelope. The output of the lamp approaches
    a black body radiation where it is observed that
    the energy of a tungsten lamp varies as the
    fourth power of the operating voltage.

44
  • Tungsten halogen lamps are currently more popular
    than just tungsten lamps since they have longer
    lifetime. Tungsten halogen lamps contain small
    quantities of iodine in a quartz envelope. The
    quartz envelope is necessary due to the higher
    temperature of the tungsten halogen lamps (3500
    oC). The longer lifetime of tungsten halogen
    lamps stems from the fact that sublimed tungsten
    forms volatile WI2 which redeposits on the
    filament thus increasing its lifetime. The output
    of tungsten halogen lamps are more efficient and
    extend well into the UV.

45
SourcesTungsten lamps (350-2500 nm)
  • Why add I2 in the lamps?
  • W I2 ? WI2
  • Low limit 350 nm
  • Low intensity
  • Glass envelope

46
  • 3. Xenon Arc Lamps
  •  
  • Passage of current through an atmosphere of high
    pressured xenon excites xenon and produces a
    continuum in the range from 200-1000 nm with
    maximum output at about 500 nm. Although the
    output of the xenon arc lamp covers the whole UV
    and visible regions, it is seldom used as a
    conventional source in the UV-Vis. The radiant
    power of the lamp is very high as to preclude the
    use of the lamp in UV-Vis instruments. However,
    an important application of this source will be
    discussed in luminescence spectroscopy which will
    be discussed later

47
(No Transcript)
48
Sample Containers
  • Sample containers are called cells or cuvettes
    and are made of either glass or quartz depending
    on the region of the electromagnetic spectrum.
    The path length of the cell varies between 0.1
    and 10 cm but the most common path length is 1.0
    cm. Rectangular cells or cylindrical cells are
    routinely used. In addition, disposable
    polypropylene cells are used in the visible
    region. The quality of the absorbance signal is
    dependent on the quality of the cells used in
    terms of matching, cleaning as well as freedom
    from scratches.

49
  • Instrumental Components
  • Source
  • ? - selection (monochromators)
  • Sample holders
  • Cuvettes (b 1 cm typically)
  • Glass (Vis)
  • Fused silica (UVVis)
  • Detectors
  • Photodiodes
  • PMTs

50
  • 1. Single beam
  • Place cuvette with blank (i.e., solvent) in
    instrument and take a reading ? 100 T
  • Replace cuvette with sample and take reading ?
    T for analyte (from which absorbance is calcd)

51
Instrumentation
  • Most common spectrophotometer Spectronic 20.
  • On/Off switch and zero transmission adjustment
    knob
  • Wavelength selector/Readout
  • Sample chamber
  • Blank adjustment knob
  • Absorbance/Transmittance scale

52
(No Transcript)
53
  • End view of the exit slit of the Spectronic 20
  • spectrophotometer pictured earlier

54
  • Single-Beam Instruments for the
    Ultraviolet/Visible Region

55
  • Single-Beam Computerized Spectrophotometers

Inside of a single-beam spectrophotometer
connected to a computer.
56
Types of Instruments
  • Instrumental designs for UV-visible photometers
  • or spectrophotometers. In (a), a single-beam
    instrument is shown. Radiation from the filter
    or monochromator passes through either the
    reference cell or the sample cell before
    striking the photodetector.

57
(No Transcript)
58
  • 2. Double beam (most commercial instruments)
  • Light is split and directed towards both
    reference cell (blank) and sample cell
  • Two detectors electronics measure ratio (i.e.,
    measure/calculate absorbance)
  • Advantages
  • Compensates for fluctuations in source intensity
    and drift in detector
  • Better design for continuous recording of spectra

59
General Instrument Designs Double Beam In - Space
Needs two detectors
60
General Instrument Designs Double Beam In - Time
61
(No Transcript)
62
  • Merits of Double Beam Instruments
  • Compensate for all but the most short term
    fluctuation in radiant output of the source
  • Compensate drift in transducer and amplifier
  • Compensate for wide variations in source
    intensity with wavelength

63
Location of Sample cell
  • In all photometers and scanning
    spectrophotpmeters described above, the cell has
    been positioned after the monochromators. This is
    important to decrease the possibility of sample
    photodecomposition due to prolonged exposure to
    all frequencies coming from the source. However,
    the sample is positioned before the monochromator
    in multichannel instruments like a photodiode
    array spectrophotometer. This can be done without
    fear of photodecomposition since the sample
    exposure time is usually less than 1 s.
    Therefore, it is now clear that in UV-Vis where
    photodecomposition of samples can take place, the
    sample is placed after the monochromators in
    scanning instruments while positioning of the
    sample before the monochromators is advised in
    multichannel instruments.

64
  • 3. Multichannel Instruments
  • Photodiode array detectors used (multichannel
    detector, can measure all wavelengths dispersed
    by grating simultaneously).
  • Advantage scan spectrum very quickly snapshot
    lt 1 sec.
  • Powerful tool for studies of transient
    intermediates in moderately fast reactions.
  • Useful for kinetic studies.
  • Useful for qualitative and quantitative
    determination of the components exiting from a
    liquid chromatographic column.

65
Multi-channel Design
66
(No Transcript)
67
  • A multichannel diode-array spectrophotometer, the
    Agilent Technologies 8453.

68
4. Probe Type Instruments
  • These are the same as conventional single beam
    instruments but the beam from the monochromators
    is guided through a bifurcated optical fiber to
    the sample container where absorption takes
    place. The attenuation in reflected beam at the
    specified wavelength is thus measured and related
    to concentration of analyte in the sample.
  • A fiber optic cable can be referred to as a light
    pipe where light can be transmitted by the fiber
    without loss in intensity (when light hits the
    internal surface of the fiber at an angle larger
    than a critical angle). Therefore, fiber optics
    can be used to transmit light for very long
    distances without losses. A group of fibers can
    be combined together to form a fiber optic cable
    or bundle. A bifurcated fiber optic cable has
    three terminals where fibers from two separate
    cables are combined at one end to form the new
    configuration.

69
(No Transcript)
70
Fiber optic probe
71
(No Transcript)
72
Double Dispersing Instruments
  • The instrument in this case has two gratings
    where the light beam leaving the first
    monochromators at a specified wavelength is
    directed to the second grating. This procedure
    results in better spectral resolution as well as
    decreased scattered radiation. However, double
    dispersing instruments are expensive and seem to
    offer limited advantages as compared to cost
    especially in the UV-Vis region where exact
    wavelength may not be crucial.

73
  • Optical diagram of the Varian Cary 300
    double-dispersing spectrophotometer. A second
    monochromator is added immediately after the
    source.

74
(No Transcript)
75
Molar absorptivities
  • e 8.7 x 10 19 P A
  • A cross section of molecule in cm2 (10-15)
  • P Probability of the electronic transition (0-1)
  • Pgt0.1-1 ? allowable transitions
  • Plt0.01 ? forbidden transitions

76
Molecular Absorption
  • M hn ? M (absorption 10-8 sec)
  • M ? M heat (relaxation process)
  • M ? ABC (photochemical decomposition)
  • M ? M hn (emission)

77
Visible Absorption Spectra
78
  • The absorption of UV-visible radiation generally
    results from excitation of bonding electrons.
  • can be used for quantitative and qualitative
    analysis

79
  • Molecular orbital is the nonlocalized fields
    between atoms that are occupied by bonding
    electrons. (when two atom orbitals combine,
    either a low-energy bonding molecular orbital or
    a high energy antibonding molecular orbital
    results.)
  • Sigma (?) orbital
  • The molecular orbital associated with single
    bonds in organic compounds
  • Pi (?) orbital
  • The molecular orbital associated with parallel
    overlap of atomic P orbital.
  • n electrons
  • No bonding electrons

80
Molecular Transitions for UV-Visible Absorptions
  • What electrons can we use for these transitions?

81
MO Diagram for Formaldehyde (CH2O)
H
C
O
H
s
p
n
82
Singlet vs. triplet
  • In these diagrams, one electron has been excited
    (promoted) from the n to ? energy levels
    (non-bonding to anti-bonding).
  • One is a Singlet excited state, the other is a
    Triplet.

83
Type of Transitions
  • s ? s
  • High energy required, vacuum UV range
  • CH4 ? 125 nm
  • n ? s
  • Saturated compounds, CH3OH etc (? 150 - 250 nm)
  • n ? ? and ? ? ?
  • Mostly used! ? 200 - 700 nm

84
Examples of UV-Visible Absorptions
LOW!
85
UV-Visible Absorption Chromophores
86
Effects of solvents
  • Blue shift (n- p) (Hypsocromic shift)
  • Increasing polarity of solvent ? better solvation
    of electron pairs (n level has lower E)
  • ? peak shifts to the blue (more energetic)
  • 30 nm (hydrogen bond energy)
  • Red shift (n- p and p p) (Bathochromic shift)
  • Increasing polarity of solvent, then increase the
    attractive polarization forces between solvent
    and absorber, thus decreases the energy of the
    unexcited and excited states with the later
    greater
  • ? peaks shift to the red
  • 5 nm

87
UV-Visible Absorption Chromophores
88
Typical UV Absorption Spectra
Chromophores?
89
The effects of substitution
Auxochrome function group
Auxochrome is a functional group that does not
absorb in UV region but has the effect of
shifting chromophore peaks to longer wavelength
as well As increasing their intensity.
90
Now solvents are your container
  • They need to be transparent and do not erase the
    fine structure arising from the vibrational
    effects

Polar solvents generally tend to cause this
problem
Same solvent must be Used when comparing absorptio
n spectra for identification purpose.
91
(No Transcript)
92
Summary of transitions for organic molecules
  • s ? s transition in vacuum UV (single bonds)
  • n ? s saturated compounds with non-bonding
    electrons
  • l 150-250 nm
  • e 100-3000 ( not strong)
  • n ? p, p ? p requires unsaturated functional
    groups (eq. double bonds) most commonly used,
    energy good range for UV/Vis
  • l 200 - 700 nm
  • n ? p e 10-100
  • p ? p e 1000 10,000

93
List of common chromophores and their transitions
94
Organic Compounds
  • Most organic spectra are complex
  • Electronic and vibration transitions superimposed
  • Absorption bands usually broad
  • Detailed theoretical analysis not possible, but
    semi-quantitative or qualitative analysis of
    types of bonds is possible.
  • Effects of solvent molecular details complicate
    comparison

95
Rule of thumb for conjugation
If greater then one single bond apart - e are
relatively additive (hyperchromic shift) - l
constant CH3CH2CH2CHCH2 lmax 184 emax
10,000 CH2CHCH2CH2CHCH2 lmax185 emax
20,000 If conjugated - shifts to higher ls
(red shift) H2CCHCHCH2 lmax217 emax
21,000
96
Spectral nomenclature of shifts
97
What about inorganics?
  • Common anions n?p nitrate (313 nm), carbonate
    (217 nm)
  • Most transition-metal ions absorb in the UV/Vis
    region.
  • In the lanthanide and actinide series the
    absorption process results from electronic
    transitions of 4f and 5f electrons.
  • For the first and second transition metal series
    the absorption process results from transitions
    of 3d and 4d electrons.
  • The bands are often broad.
  • The position of the maxima are strongly
    influenced by the chemical environment.
  • The metal forms a complex with other stuff,
    called ligands. The presence of the ligands
    splits the d-orbital energies.

98
Transition metal ions
99
Charge-Transfer-Absorption
  • A charge-transfer complex consists of an
    electron-donor group bonded to an electron
    acceptor. When this product absorbs radiation, an
    electron from the donor is transferred to an
    orbital that is largely associated with the
    acceptor.
  • Large molar absorptivity (emax gt10,000)
  • Many organic and inorganic complexes

100
(No Transcript)
Write a Comment
User Comments (0)
About PowerShow.com