Preliminary Proposal For Insulin Pump Standards - PowerPoint PPT Presentation

About This Presentation
Title:

Preliminary Proposal For Insulin Pump Standards

Description:

Preliminary Proposal For Insulin Pump Standards Standards to improve insulin pump use and medical outcomes These proposals are not yet final but are made available ... – PowerPoint PPT presentation

Number of Views:34
Avg rating:3.0/5.0
Slides: 76
Provided by: diabetesn1
Category:

less

Transcript and Presenter's Notes

Title: Preliminary Proposal For Insulin Pump Standards


1
Preliminary Proposal For Insulin Pump Standards
Standards to improve insulin pump use and medical
outcomes These proposals are not yet final but
are made available for review and editorial
comment. Any suggestions you have for
improvements or changes, or for additional
approaches to improve diabetes care are welcomed.
Any major contributions will be attributed.
2
These Standards Are Supported By
  • John Walsh, PA, CDE, and Ruth Roberts, MA
  • You (4,6)

Any reservations you have about a particular
standard will be noted
3
ReviewDefinitions
  • TDD total daily dose of insulin (all basals and
    boluses)
  • Basal background insulin pumped slowly through
    the day to keep BG flat
  • Bolus a quick surge of insulin as
  • Carb boluses to cover carbs
  • Correction boluses to lower high readings that
    arise from too little basal insulin delivery or
    insufficient carb boluses
  • Bolus On Board (BOB) the units of bolus insulin
    or glucose-lowering activity still working from
    recent boluses
  • Duration of Insulin Action (DIA) time that a
    bolus will lower the BG. This is used to
    calculate BOB.

4
The Role Of Insulin Pumps
  • Insulin pumps should
  • Lower A1c and eAG levels
  • Decrease the frequency and severity of
    hypoglycemia
  • Provide safe and reliable insulin dosing to users
  • Reduce complications
  • Improve quality of life

5
Proposal For Insulin Pump Standards
  • These standards to improve pump use and outcomes
    are designed to
  • Reduce inconsistencies in pump settings between
    pump manufacturers
  • Improve the accuracy and safety of carb and
    correction factors in use
  • Improve safety of DIA time increments and
    defaults that are in use
  • Consistently account for and apply BOB
  • Improve monitoring and identification of infusion
    set failure
  • Improve monitoring of hypoglycemia
    hyperglycemia
  • Identify excessive use of correction boluses
  • Reduce blind bolusing and non-entry of glucose
    values into pumps

6
ReviewInconsistent Dosing From Insulin Pumps
  • There are several significant sources for error
    in bolus doses from todays insulin pumps.
  • Widespread use of inaccurate carb factors
  • Excessively large carb factor increments
  • Widespread use of inaccurate correction factors
  • Wide variations in how BOB is handled and in DIA
    default times between pump manufacturers
  • Wide variations in DIA defaults between pump
    manufacturers

7
Intellectual Property
1
  • Issue Optimal glucose values for those who have
    diabetes is critical to prevention of disability
    and early death. Devices owned and used by those
    with diabetes contain unique information that can
    be used to improve control and reduce
    complication risks.

8
Standard ForIntellectual Property
1
  • We recommend that all existing and future patents
    that may contribute to improvements in glucose
    control be made available at a reasonable cost to
    any device manufacturer who wants to use them to
    improve glucose control.

9
Carb Factor (CarbF) Increments
2
  • Issue Current carb factor increments are too
    large for smaller carb factor numbers. This lack
    of precision for carb boluses may create excess
    hyperglycemia or hypoglycemia for many pump
    users.

10
ExampleCarb Factor Increments
2
  • Most pumps offer 1 gram per unit as their
    smallest CarbF increment. This increment becomes
    relatively large for CarbFs below 15 or 20 g/u.
    For instance, when the carb factor is reduced
    from 10 to 9 g/u, all subsequent carb boluses are
    increased by 11.1 in most pumps.
  • A shift in the carb factor from 1u/5g to 1u/4g
    causes each subsequent carb bolus to increase by
    25.
  • When carb boluses make up 50 of the TDD, a
    change in the a carb factor larger than 2.5
    would be expected to create more than a 25 mg/dl
    shift in the glucose following each meal.
  • Finer CarbF increments would allow safer and more
    precise adjustment of subsequent carb boluses.

Example TDD 40u, carb factor 1u/11g, corr
factor 1u/60 mg/dl carb boluses 20u or
6u/meal x 6 .36u x 60 a 22 mg/dl change in
BG
11
What Current Changes In CarbFs Do
2
  • Table shows how a one-step reduction in the CarbF
    using various CarbF increments affect the size of
    subsequent carb boluses. Yellow area shows impact
    from most pumps. Green (preferred) areas show
    increments that impact subsequent boluses less
    than 5.

12
ExampleImpact On BG From CarbF Adjustments
2
  • This table shows the average additional fall in
    glucose that is likely after each meal of the day
    when a carb factor is reduced from 10 grams per
    unit to 9 grams per unit (for appropriate weight
    TDD), and again a reduction from 5 grams per
    unit to 4 grams per unit.

Calculated as carb grams per day X increase in
avg. carb bolus size 3
13
2
Standard For Carb Factor Increments
  • We recommend that carb factor increments be small
    enough that a single step adjustment in a factor
    causes subsequent carb boluses to change by no
    more than 5 from previous bolus doses for the
    same number of grams of carb.

14
Inaccurate Carb Factors
3
  • Issue A carb factor that does not match the
    individual using it will significantly magnify
    other sources of error in the calculation of carb
    bolus doses. Many carb factors used in insulin
    pumps today are poorly tuned to users.

15
Review Actual Carb Factors In Use 1
3
  • Avg. carb factors for 468 consecutive Cozmo
    insulin pump downloads (gt126,000 boluses) are
    shown in blue
  • Note that they are NOT bell-shaped or physiologic
  • People prefer magic numbers 7, 10, 15, and 20
    g/unit for their carb factors
  • Determined directly from grams of carb divided
    by carb bolus units for each carb bolus

10
7
115
20
1
16
ReviewActual Carb Factors In Use 1
3
  • MANY magic carb factors, shown in blue, are
    inaccurate. A more normal or physiologic
    distribution is shown in green
  • Use of magic numbers creates major, though
    consistent bolus errors that magnify other
    sources of error

10
7
115
20
1
17
3
Standards For Carb Factor Settings
  • To encourage use of consistent and accurate carb
    factors in pumps, we request that insulin pump
    companies jointly determine what range of carb
    factor rule numbers (CarbF x TDD) provide optimal
    glucose results that lead to a lower eAG and less
    hypoglycemia for various TDD ranges and carb
    intakes as a percentage of calories.
  • We request that insulin pump companies measure
    and publish each year the carb factors in use for
    200 random downloads from pumps that use carb
    factors. This information is needed as an
    overview to guide interventions directed at
    reducing errors in carb factor settings.
  • We recommend that carb factors be monitored
    within each pump for accuracy and effectiveness
    with a report available to users or clinicians.

18
ExampleCarb Factor Settings
3
  • To assist users in setting accurate CarbFs,
    insulin pumps should allow the user to compare
    their current CarbF against an optimal settings
    range of CarbF Rule Numbers. Proposed CarbF Rule
    Numbers for various TDDs

Optimal ranges would be determined from
research studies of best practices
19
Inaccurate Correction Factors
4
  • Issue A correction factor that does not match
    the individual using it will significantly
    magnify other sources of error in the calculation
    of correction bolus doses. Many correction
    factors used in insulin pumps today are poorly
    tuned to users.

20
Review Actual Correction Factors In Use 1
4
  • Avg. correction factors in use for 452
    consecutive Cozmo insulin pump downloads
  • Like carb factors, correction factors in use are
    NOT bell-shaped or physiologic
  • Users or clinicians often select magic numbers
    for their correction factors.

10
7
115
20
1
21
4
Standards For Correction Factor Settings
  • To encourage use of consistent and accurate
    correction factors in pumps, we request that
    insulin pump companies jointly determine what
    range of correction factor rule numbers (CorrF x
    TDD) provide optimal glucose results that lead to
    a lower eAG and less hypoglycemia for various TDD
    ranges, and publish them for users and clinicians
    to use.
  • We request that insulin pump companies
    voluntarily measure and publish each year the
    correction factors in use for 200 consecutive
    downloads from pumps that use correction factors.
  • We recommend that correction factors be monitored
    within each pump for accuracy and effectiveness
    with a report available to users or clinicians.

22
ExampleCorrection Factor Settings
4
  • To assist users in setting accurate CorrFs,
    insulin pumps should allow the user to compare
    their current CorrF against an optimal settings
    range of CorrF Rule Numbers. Proposed CorrF Rule
    Numbers for various TDDs

Optimal ranges would be determined from
research studies of best practices
23
Basal/Carb Bolus Balance
5
  • Issue Correction boluses are used to correct for
    deficits in insulin that arise from inadequate
    basal delivery or inadequate carb boluses.
    Because the reason for their use cannot be
    clearly identified as basal or bolus, they should
    not be included in basal/bolus balance.

24
Standard ForBasal/Carb Bolus Balance
5
  • We propose that basal/carb bolus balance is a
    more definitive term and should replace
    basal/bolus balance. Basal/carb bolus balance
    should NOT include correction bolus doses which
    will be listed separately to more clearly define
    and understand control issues.

25
ExampleBasal/Carb Bolus Balance
5
  • How a pump might display insulin information
  • TDD 40.0 u in last 7 days
  • of TDD units
  • Basal 30 12 u
  • Carb boluses 50 20 u
  • Corr. boluses 20(12) 8 u ( 4.8 u)
  • Basal/Carb bolus balance 0.6 (12u/20u) or 60.
    This particular imbalance would typically favor
    adding more of the correction bolus excess to
    basal delivery.

26
Duration Of Insulin Action Default Times
6
  • Issues
  • The default DIA times in current pumps vary
    widely between 3 and 6 hours
  • Many users shorten their default DIA to increase
    the size of their boluses without realizing that
    this introduces significant errors into bolus
    (and ultimately basal) doses.

DIA times that are too short hide bolus insulin
activity and create insulin stacking. DIA times
that are too long overestimate bolus activity.
27
Standard ForDuration Of Insulin Action (DIA)
Time
6
  • We recommend that a panel of experts in insulin
    action review existing pharmacodynamic studies,
    consider differences between pharmacodynamics and
    DIA time, and provide guidance on an acceptable
    range of DIA times to recommend to clinicians and
    users to improve the accuracy of bolus
    calculations.

28
DIA Time Increments
7
  • Issue Current DIA time increments vary from 15
    minutes to 1 hour in different pumps
  • When a DIA time is changed in a pump, a larger
    time increment, such as 1 hr, can create an
    excessive change in subsequent estimates of BOB.
  • For example, when the DIA is reduced from 5 hours
    to 4 hours, subsequent BOB estimates are
    decreased, while recommendations for carb boluses
    would increase.

29
ReviewGlucose Infusion Rate (GIR) Studies
7
  • Most GIR studies suggest that pharmacodynamic
    action of insulin varies about 25 between
    individuals. For a DIA time of 5 hr, a 25 range
    is equivalent to 1 hr and 15 min, such as from 4
    hrs and 15 min to 5hr and 30 min.
  • A pump that has only 1 hr DIA increments would
    enable the user to select only one setting within
    a physiologic range, while a 30 min increment
    would allow only 2 or 3 choices that are close to
    a physiologic range.

30
7
Standard For DIA Time Increments
  • For safety and accuracy, we recommend that DIA
    time increments be no greater than 15 minutes to
    allow more accurate estimation of residual BOB.

31
ReviewBolus On Board (BOB)
8
  • An accurate measurement of the glucose-lowering
    activity that remains from recent boluses helps
  • Prevent insulin stacking
  • Improve bolus accuracy
  • Allows the current carb or insulin deficit to be
    determined

aka insulin on board, active insulin, unused
insulin Introduced as Unused Insulin in 1st ed
of Pumping Insulin (1989)
32
ExampleAn Accurate BOB Can Avoid Insulin Stacking
8
  • Bedtime BG 173
  • Is there an insulin or a carb deficit?

Bedtime BG 173 mg/dl
Correction
Dessert
Dinner
6 pm
8 pm
10 pm
12 am
33
ReviewInsulin Stacking Is Common
8
  • CDA1 Study Results
  • Of 201,538 boluses, 64.8 were given within 4.5
    hrs of a previous bolus
  • An accurate DIA shows that some BOB is present
    for MOST boluses
  • Note that 4.5 hours may underestimate true DIA

4.5 hrs
34
ReviewHow Current Pumps Handle BOB
8
Except when BG is below target BG
35
ExampleBolus Recommendations Differ Significantly
8
  • Situation BOB 3.0 u and 30 gr. of
    carb will be eaten at these glucose levels
  • Carb factor 1u / 10 gr
  • Corr. Factor 1 u / 40 mg/dl over
    100
  • Target BG 100
  • TDD 50 u

The graphic shows how widely bolus
recommendations vary from one pump to another for
the same situation.
units
mg/dl
Omnipod bolus cannot be determined - it counts
only correction bolus insulin as BOB
36
ExampleUnsafe BOB Handling
8
  • If a pump user gets frustrated with high BGs and
    they overdose to speed the drop in their BG, or
    they exercise longer or more intensely than they
    anticipated, they can acquire a significant
    excess in BOB.
  • In this situation, most pumps bolus for all carb
    intake regardless of how much BOB is present. A
    subsequent bolus will deliver an excess of
    insulin if the glucose is not high enough to
    offset the excess BOB.1
  • This introduces a significant risk for
    hypoglycemia from the pumpsexcessive bolus
    recommendations.

1 Pumping Insulin, 1st ed, 1989, Chap 12, pgs
70-73 The Unused Insulin Rule
37
Current BOB Handling
8
  • Issue Most bolus calculators in current insulin
    pumps assume that excess BOB does not need to be
    taken into account when determining the next carb
    bolus.
  • Because of the way they are determined, bolus
    dose recommendations from most pumps can cause
    unexplained and unnecessary insulin stacking and
    hypoglycemia.

38
Standard ForBOB Handling
8
  • For safe and accurate BOB measurement, we
    recommend that
  • BOB include all carb and correction boluses
  • Residual BOB be subtracted from both carb and
    correction bolus recommendations delivered within
    the DIA

Assumes that the DIA time chosen by the
clincian or user is accurate.
39
BOB Handling
8
  • Exception to usual BOB handling
  • When a second bolus is taken for unplanned carb
    intake or desert that is consumed within 60
    minutes or so of a meal bolus, BOB should not be
    taken into account for the second bolus because
    the impact of the first bolus cannot be
    accurately determined.
  • Given that, it is wise to account for BOB as soon
    after a meal as possible, such as within 60 to 90
    minutes (adjustable), to provide early warning if
    the bolus given was excessive or inadequate.

It is always safer, though not always more
accurate, to account for and apply all BOB in
subsequent boluses.
Adjustable
40
Blind Bolusing
9
  • Issue Pump users often bolus for carbs without
    checking their glucose first. With no glucose
    reading, the pump does not account for BOB that
    may be present at the time, and the bolus is not
    appropriately adjusted for potentially high or
    low glucose levels.
  • Blind bolusing often leads to insulin stacking.

41
Standard ForBlind Bolusing
9
  • We recommend that insulin pumps alert the wearer
    when there is sufficient insulin stacking to
    introduce a significant error in a current bolus.

Adjustable for an expected mg/dl drop in
glucose with visible, audio, or vibratory output.
42
ExampleInsulin Stacking or BOB Alert
9
  • When a carb bolus is planned without a recent BG
    check, but BOB is more than 1.25 of the average
    TDD (enough to cause about a 25 mg/dl drop in the
    glucose), the pump will recommend that the wearer
    do a BG check due to an excess in BOB.
  • For instance, for someone with
  • Avg TDD 1.25 of TDD
  • 43 units 0.54 units
  • This individual would be alerted whenever they
    give a bolus but have 0.54 u or more of BOB
    present.

1.25 of TDD provides a reasonable degree of
safety but may need modification
43
Inadequate Manual Entry Of BGs
10
  • Issue Pump users often do not enter BG values
    into their pump if they must do it manually.

44
Inadequate Manual Entry Of BGs
10
  • Issue In the CDA study where BG values can be
    entered either manually or automatically, users
    entered only 2.6 BG values per day manually
    versus 4.1 values per day for pumps that had an
    attached glucose meter. This means that BOB may
    be taken into account for 1.5 additional boluses
    per day when BG readings are not automatically
    entered.

45
Standard ForInadequate Manual Entry Of BGs
10
  • Due to a significant decrease in glucose entry
    when BGs must be entered manually, and the
    benefit to control that this provides, we
    recommend that all pumps be enabled to have
    direct BG entry of BG test results from two or
    more meters.

46
Correction Bolus Excess
11
  • Issue Hyperglycemia is more common than
    hypoglycemia for most people on insulin pumps.
  • When glucose levels consistently run high, many
    pump users address the problem by giving frequent
    correction boluses rather than increasing their
    basal rates or carb boluses.
  • In these cases, the correction bolus of the TDD
    can become excessive, but this information is
    either not shown in some pumps or no alert is
    given regarding the excess.

47
Standards ForCorrection Bolus Excess
11
  • We suggest that the pump wearer and clinician be
    alerted when the wearer uses more than 8
    (adjustable) of their TDD for correction bolus
    doses for at least 4 days in a row (adjustable).
  • We recommend that, once an excess in correction
    bolus is identified, that the user be given
    instruction in how to safely distribute any
    excess into carb boluses or basal rates.

48
Excess Hypoglycemia
12
  • Issue Current insulin pumps and glucose monitors
    do not warn users that they are experiencing
    hypoglycemia that is too severe or too frequent.
  • Although most insulin pumps contain adequate data
    to do so, they do not provide sufficient guidance
    for correcting this serious problem.

49
Standard ForExcess Hypoglycemia
12
  • We recommend that insulin pumps which store
    glucose and insulin dosing data alert users when
    they experience severe or excessive hypoglycemia
    and provide specific guidance regarding likely
    causes.

50
Excess Hyperglycemia
13
  • Issue Current insulin pumps and glucose monitors
    do not warn users that they are experiencing
    hyperglycemia that is too severe or too frequent.
  • Although most insulin pumps contain adequate data
    to do so, they do not provide sufficient guidance
    for correcting this serious problem.

51
Standard ForExcess Hyperglycemia
13
  • We recommend that insulin pumps which store
    glucose and insulin dosing data alert users when
    they experience severe or excessive hyperglycemia
    and provide specific guidance regarding likely
    causes.

52
Duration Of Insulin Action
14
  • Issue Although DIA is designed to measure the
    glucose-lowering activity of a carb or correction
    bolus at any time, various clinicians recommend
    DIA times that vary from 2 to 6 hours or more,
    despite the fact that interindividual variation
    in pharmaco-dynamics time is generally less than
    25.
  • There are also questions about whether
    pharmaco-dynamic time from GIR studies is
    equivalent to DIA time?

53
ReviewHow Long Do Boluses Lower The BG?
14
  • Novolog claims 3 to 5 hours 10, but numerous
    studies show rapid insulin lowers the glucose for
    5 hours or more.
  • With Novolog (aspart) at 0.2 u/kg (0.091 u/lb),
    23 of glucose lowering activity remained after 4
    hours.12
  • Another study found Novolog (0.2 u/kg) lowered
    the glucose for 5 hours and 43 min. /- 1 hour.13
  • After 0.3 u/kg or 0.136 u/lb of Humalog (lispro),
    peak glucose-lowering activity was seen at 2.4
    hours and 30 of activity remained after 4 hours.
    11
  • These times would be longer if the unmeasured
    basal suppression in pharmacodynamic studies were
    accounted for.

10 Novolog product labeling information, October
21, 2005. 11 From Table 1 in Humalog Mix50/50
product information, PA 6872AMP, Eli Lilly and
Company, issued January 15, 2007. 12 Mudaliar S,
et al Insulin aspart (B28 Asp-insulin) a
fast-acting analog of human insulin. Diabetes
Care 1999 221501-1506. 13 L Heinemann, et al
Time-action profile of the insulin analogue
B28Asp. Diabetic Med 199613683-684.
54
An Accurate DIA Can Prevent Lows
14
  • Accurate DIA Time
  • Accurate BOB
  • Accurate Boluses Accurate HypoManager
  • Better Readings, Fewer Lows

Prevention
Prediction
55
ReviewShort DIAs Hide Bolus Insulin Activity
14
  • A short DIA time creates significant problems
    because it hides true BOB level and its
    glucose-lowering activity. This
  • Leads to unexplained lows
  • Leads to incorrect adjustments in basal rates,
    carb factors, and correction factors
  • Causes user to start ignoring their smart
    pumps advice
  • In contrast, an inappropriately long DIA
    overestimates bolus insulin activity. DIA should
    be selected, based on its real insulin action
    time.
  • Do NOT modify the DIA time to fix a control
    problem

56
ReviewDuration Of Insulin Action (DIA)
14
Accurate bolus estimates require an accurate DIA.
DIA times shorter than 4 to 7 hrs will hide BOB
and its glucose lowering activity
Glucose-lowering Activity
6 hrs
2 hrs
0
4 hrs
57
ReviewDIA
14
  • Large doses (0.3 u/kg 15 u for 110 lb. person)
    of rapid insulin in 18 non-diabetic, obese
    people
  • Med. doses (0.2 u/kg 10 u for 110 lb. person)

Regular
Apidra product handout, Rev. April 2004a
58
ReviewDoes Dose Size Affect Duration Of Action?
14
  • This graphic suggests that smaller boluses do not
    lower the BG as long as larger boluses.
  • However, this may not be true see next 2
    slides.
  • For a 154 lb or 70 kg person
  • 0.05 u/kg 3.5 u
  • 0.1 u/kg 7 u
  • 0.2 u/kg 14 u
  • 0.3 u/kg 21 u

Woodworth et al. Diabetes. 199342(Suppl. 1)54A
59
ReviewPharmacodynamics Is Not DIA
14
  • The DIA time entered into an insulin pump is
    based on studies of insulin pharmacodynamics.
  • However, the traditional method used to determine
    the pharmacodynamics of insulin action routinely
    underestimates insulins true duration of action.
    See next slide.

60
ReviewPharmacodynamics Underestimates DIA And
Overestimates Impact Of Bolus Size
14
  • To measure pharmacodynamics, glucose clamp
    studies are done in healthy individuals (0.05 to
    0.3 u/kg)
  • Injected insulin ALSO SUPPRESSES normal basal
    release from the pancreas (grey area in figure)
  • The basal suppression makes smaller boluses
    appear to have a shorter DIA

61
ReviewPharmacodynamic Time Does Not Equal DIA
14
  • After accounting for the lack of basal
    suppression,
  • True DIA times become longer than the
    pharmacodynamic times derived from typical
    research
  • At least some of the apparent variation in DIA
    due to relative bolus size disappears
  • Some of the apparent inter-individual variation
    in pharmacodynamics may also disappear

62
Standard ForDuration Of Insulin Action
14
  • We recommend that a panel of researchers and
    clinicians who are familiar with insulin
    pharmacodynamics recommend consistent and safe
    guidelines for DIA times in pumps for children
    and adults.
  • These guidelines will be used to advise
    clinicians, train pump users, and as a reference
    on the DIA setting screen in insulin pumps.

63
Linear Versus Curvilinear DIA
15
  • Issue The current straight-linear method of
    measuring DIA is less accurate than curvilinear
    method for estimation of residual BOB.
  • DIA times selected in pumps which use a linear
    method must be shorter to approximate the DIA in
    a pump that uses a curvilinear method.
  • Pump manufacturers currently use at least 3
    different methods to measure glucose-lowering
    activity

64
Linear And Curvilinear DIA Compared
15
  • Set

5 hr Linear
5 hr Curvilinear
From Pumping Insulin, 4th ed., adapted fom
Mudaliar et al Diabetes Care, 22 1501, 1999
65
The Modified Triple-Linear DIA
15
  • An alternative and more accurate approach would
    be to modify the linear method into a
    triple-linear method to provide more precise BOB
    estimates.

66
ExampleTriple-Linear DIA Times
15
  • A triple linear line can more closely imitate a
    curvilinear DIA.
  • For a 5 hr DIA
  • 30 min no change
  • 3 hrs fall 75
  • 1.5 hrs fall last 25 (approximate values)

5 hr Triple Linear
From Pumping Insulin, 4th ed., adapted fom
Mudaliar et al Diabetes Care, 22 1501, 1999
67
Standard ForLinear Versus Curvilinear DIA
15
  • We recommend that insulin pumps use either a 100
    curvilinear or a triple-linear method to improve
    the accuracy of BOB estimates.

68
Infusion Sets
16
  • Issue Infusion set design issues and inadequate
    site preparation training introduce erratic
    losses of control for a significant number of
    pump wearers.

69
ReviewInfusion Set Failure
16
  • Loss of glucose control caused by infusion set
    failure can occur due unrecognized pullout of the
    set.
  • A more common error occurs when Teflon infusion
    sets come loose and some insulin leaks back to
    the skin surface. This causes unexplained high
    readings rather than the complete loss of control
    typically seen with a complete pullout.
  • Metal needle sets can also cause occasional
    bleeding under the skin that interferes with
    insulin delivery and leads to elevated glucose
    readings.
  • Selecting the right infusion set plus good site
    technique can significantly reduce this
    unnecessary loss of control.

70
ReviewCauses For Infusion Sets Failure
16
  • Certain infusion sets are more prone to failure
    due to their design.
  • Other sets fail when tugging and pulling on an
    unanchored infusion line during routine wear
    loosens the Teflon beneath the skin.
  • In a review of dozens of pictures of infusion
    sets online and insulin pump manuals, anchoring
    of the infusion line is usually not recommended
    and is not generally done. Anchoring of the
    infusion line can
  • Stop movement of Teflon catheter under the skin
  • Stop unexplained highs caused when insulin
    leaks back to surface
  • Reduce skin irritation
  • Prevent many pull outs

Example
71
ReviewDetection Of Bad Infusion Set Or Site
16
  • If a pump user has unexplained highs, ask
  • How often do unexplained highs happen?
  • Do they usually correct when you replace your
    infusion set?
  • For yes answers
  • Always use tape to anchor the infusion line
  • Consider changing to a different infusion set

The right infusion set and good site technique
prevents headaches and improves the A1c
72
ToolInfusion Set Monitor
16
  • Insulin pumps with direct BG entry can identify
    those who may be having consistent but
    intermittent loss of glucose control secondary to
    infusion set failure. The pump can
  • Show avgerage time and interval variation between
    use of reservoir loads or the prime function in
    the pump.
  • Show average BGs for each 12 hour segment
    following set changes (indicated by the prime
    function) over at least the last 7 set changes
    (or as soon as statistical significance is
    reached).

73
Need For An Infusion Set Monitor
16
  • Many pump wearers have random erratic readings
    that are greatly reduced in number when they
    change to a different infusion set or start to
    anchor their infusion lines with tape to stop
    line tugging.
  • However, there is currently no tool for clincians
    or pump users to tell who is having problems with
    their infusion sets.

74
Standards ForInfusion Sets
16
  • We recommend that infusion sets be expected to
    perform at least 72 hours without a loss of
    glucose control.
  • We recommend that monitoring be provided in all
    insulin pumps to detect consistent patterns of
    infusion set problems or failure for individual
    pump users.

75
Standards ForInfusion Sets
16
  • 3. We recommend that insulin pump manuals and
    training cover methods to identify and prevent
    infusion set failure.
  • 4. We recommend that future infusion set designs
    incorporate easy to use methods to anchor
    infusion lines and minimize tugging of the
    infusion line near the infusion site.
Write a Comment
User Comments (0)
About PowerShow.com