Title: Chapter 40 Radiation Protection Procedures
1Chapter 40 Radiation Protection Procedures
- ALARA and Occupational Exposure
2ALARA
- ALARA stands for As Low As Reasonably Achievable.
It is the basic principle of radiation protection
procedures. - There is much that we can do to keep exposure to
the patient and the operator as low as possible. - The chiropractor is unique in the fact that you
can perform radiography and refer your patients
for other types of examinations.
3Occupational Exposures
- In radiologic technology, 95 of the occupational
exposure comes from fluoroscopy and mobile
radiography. - Neither would be used in your office so the worst
case scenario is that you would receive 5 of the
exposure that a technologist would receive. -
4Occupational Exposures
- During radiography, the operator should be behind
a protective barrier. - These barriers are usually considered as
secondary barriers so protection would be from
tube leakage and scatter from the patient. The
tube should never be pointed toward this type
barrier. - If the barrier can have the tube angled toward
the barrier. It must be a primary barrier. - Staying behind the barrier effectively eliminates
the source of occupational exposure if the
shielding is adequate.
5Occupational Exposures
- Medical Imaging Exposures
- Fluoroscopy All personnel will wear protective
apron. If extremities get into the beam lead
gloves can be worn. - The radiologist will usually be close to the
machine during fluoroscopy so their exposure will
be higher than that the technologist. Aprons
between the Image intensifier and Bucky Slot
covers reduce radiologist exposure. - The technologist should stand as far away from
the table as possible during the exam and move
closed only when necessary. - The radiologist will use short burst of exposure
and keep the exposure time as short as possible.
The 5 minute clock timer will alarm when 5
minutes of fluoroscopy has been used.
6Occupational Exposures
- Medical Imaging Exposures
- Mobile radiography
- The technologist must wear a lead apron during
mobile plain film or fluoroscopy examinations. - An apron must be assigned to each portable
machine. - The exposure cord for portable radiographic
machines must be 2 meters long to maximize
distance from the tube during exposures.
7Occupational Exposures
- Radiology Ancillary Staff
- Assuming the rooms are adequately shielded, the
receptionist, file room and darkroom staff
should not receive any occupational exposure. - Radiology ancillary staff should not be used to
hold patients during radiography.
8Occupational Radiation Monitoring
- Occupational Radiation Monitoring is required if
there is any likelihood that an individual will
receive more than 1/10 of the recommended dose. - With just plain film radiography, monitoring may
not be required as long as the operator stays in
the control booth during all exposures. - There are some exams such as stress views of the
ankle where the operator would be in the room
with the patient. If this is done, monitoring
would be necessary. - If the operator ever holds a patient monitoring
would be necessary.
9Occupational Radiation Monitoring
- Occupational radiation monitoring offers no
protection against exposure. It merely records
the exposure received. - If needed, find a certified laboratory to process
the dosimeters. - Types of monitors
- Film badges
- TLD
- OSL
10Film Badges
- Film badges have been used since the 1940s and
are still used today. - Exposures below 10 mR are not measured on the
film. - Filters along with the window in the badge allow
estimation of the energy of the exposure. - The must be worn with the proper side to the
front. - They are typically worn on the collar so they
would remain outside the lead apron.
11Film Badges
- Advantages
- Inexpensive
- Easy to handle and process
- Reasonably accurate
- Disadvantages
- Can not be reused
- Sensitive to heat and humidity
- Must be changes monthly
12TLD
- TLD has several advantages over film badges.
- Not sensitive to heat or humidity
- Measure exposures to 5mR More sensitive and
accurate. - Can be changed quarterly instead of monthly
- Disadvantages
- Cost but changing badges less frequently than
monthly eliminates cost problem.
13Optically Stimulated Luminescence
- All of the advantages of the TLD over film badges
plus - Can be re-read to confirm exposure
- More accurate than TLD
14Where to wear the monitor
- The whole body badge is typically worn at collar
level so it can be outside the lead apron. - Fetal monitoring badges used during pregnancy are
worn at waist level under the apron. - Hand or finger TLDs are worn on the extremity.
15Occupational Radiation Monitoring Reports
- State and federal regulations require that the
results of the occupational radiation monitoring
program be recorded in a precise fashion and
maintained for review. - Specific information is required to be on the
report including current and cumulative exposure. - Each site of monitoring must be identified
separately. - There will also be a control monitor to measure
the background exposure during transport,
handling and storage.
16Occupational Radiation Monitoring Reports
- The supplier of the badges must know the type of
radiation for proper calibration of the
equipment. - The badges are control are shipped back to the
supplier in a timely manner. - For lost or damaged badges, a health physicist
will estimate the exposure. - The annual exposure is discussed with each worker
and receipt of the information is documented.
Monthly reports may be posted but care must be
taken with sensitive information.
17Protective Apparel
- Lead apron used for operator or patient
protection must be the equivalent of 0.5mm of
lead. - They must be worn when in a room during the
exposure or during fluoroscopy. - Half aprons are effective means to provide gonad
protection during radiography. - 0.25 mm of lead aprons should be avoided as they
only attenuate 66 of the beam at 76 kVp.
18Protective Apparel
- Aprons used in interventional radiology should be
a wrap around type. Thyroid shields may also be
worn. - Lead gloves can be worn when the hands are in the
beam. - Aprons must be stored on specially designed racks
or laid flat on the floor. They are never folded. - Aprons are tested annually for cracks or holes in
the lead, usually by fluoroscopy.
19Position
- During fluoroscopy radiologic technologist should
stand as far as possible from the machine. - Standing behind the radiologist offers added
protection. - If you must be in the room, position your body as
far away from the primary beam as possible.
20Patient holding
- Many patients will find the x-ray examination to
be physically demanding. Some may not be capable
of staying in position. - This is a particular challenge for weight-bearing
radiography. Mechanical supportive devices are
limited for erect studies. - If you have a radiographic table, the patient may
be examined recumbent. Sponges may be used as
supportive devices.
21Patient holding
- Radiology or office staff should never hold a
patient. Family or friends may be called upon to
assist the patient. - The person assisting the patient must wear a lead
apron and if their hands will be in the beam lead
gloves. - Position the person as far away from the primary
beam as possible. - Since the person holding the patient may be a
parent, make sure they are not pregnant.
22Reducing Unnecessary Patient Dose
- As a doctor, you have the responsibility to
determine if the radiography is necessary and
justified. - There are more practice guidelines available
every year to assist in determining if the
examination will yield necessary diagnostic
information. - There are many examinations that are performed
knowing that they will yield little helpful
information so they in no way justify the patient
radiation dose.
23Reducing Unnecessary Patient Dose
- Check to see if the patient has previous
examinations that may make the new examination
not necessary. - You may be sued if you dont take films and the
treatment plan fails because you missed something
the films would show. - The yield of information must be greater than the
risk of radiation exposure.
24Reducing Unnecessary Patient Dose
- Routine x-ray examinations should not be
performed. - Used the most accurate tests to confirm or rule
out your working diagnosis. - Consider using MR instead of CT
25Repeat Examinations
- One area of unnecessary patient exposure is
repeated x-ray examinations. Past estimates of
frequency has been as high as 10 but they should
normally not exceed 5. - Most of the retakes are of the lumbar spine,
abdomen and thoracic spine. - Most retake are due to the exposure factors being
incorrect resulting in an over exposed or under
exposed film. Proper measurement are important.
26Repeat Examinations
- Positioning errors account for about 25 of
retakes. Proper training and practice is
important to fine tune positioning skills. - Motion causes about 11 of retakes so proper
patient communication during the exam is
important. - But do not be afraid to retake a poor quality
film. If you can not see a problem makes it
likely you will miss it. Poor quality exams are
never justified.
27Radiographic Technique
- Use as high kVp as possible to get adequate
contrast and reduce patient exposure. - Collimate the beam to slightly smaller than film
size or the area of interest, whichever is
smaller. - Use the fastest-speed screen-film combination
consistent with the nature of the examination.
28Positioning
- When taking films with the patient seated, do not
allow the gonad to be in the primary beam.
Position the patient lateral to the beam. - For female patients turn the patient PA to reduce
breast and gonad exposures when possible.
29Patient shielding
- Some form of patient shielding should be used on
all patients able to reproduce. - All children should have shielding.
- Pre-menopausal women should be shielded except
when the shield would interfere with the
examination. - Men should be shielded beyond 50 years.
30Patient shielding
- Patient shielding includes contact shields and
shadow shields. - Contact shields are placed on the patient and
include aprons, the heart shaped filter and the
bell. - Shadow shields are placed between the patient and
the tube. Here we attach it to the tube.
31Patient shielding
- Shielding must be used when the gonads lie in or
near the useful beam and when it does not
interfere with obtaining the required diagnostic
information. - Accurate placement is extremely important.
Repeated examinations can result form improper
placement of the shield. - Proper patient positioning and collimation should
not be relaxed when gonad shields are in use.
32Ten Commandments of ALARA
- Understand and apply the cardinal principles of
radiation control time, distance and shielding. - Do not allow familiarity to result in a false
security. - Never stand in the primary beam.
- Always wear protective apparel when not behind a
protective barrier. - Always wear a radiation monitor and position it
outside the protective apron at collar level.
33Ten Commandments of ALARA
- Never hold a patient during radiographic
examinations. Use mechanical restraining devices
when possible. Otherwise, use patients or friends
to hold the patient. - The person holding the patient must wear
protective apron and if possible, gloves. - Use gonadal protective on all people of
childbearing age when it will not interfere with
the examination.
34Ten Commandments of ALARA
- Examinations of the pelvis or lower abdomen of a
pregnant patient should be avoided whenever
possible, especially during the first trimester. - Always collimate to the smallest field size
appropriate to the examination.
35Chapter 31 Quality Control
- Two areas of activity are designed to ensure the
best possible image quality with the lowest
possible exposure and minimum costs. - Quality Assurance deals with people
- Quality Control deals with instrumentation and
equipment.
36Chapter 31 Quality Control
- Two areas of activity are designed to ensure the
best possible image quality with the lowest
possible exposure and minimum costs. - Quality Assurance deals with people
- Quality Control deals with instrumentation and
equipment.
37Ten Step Approach to Quality Assurance
- Assign responsibility
- Delineate scope of care
- Identify aspects of care
- Identify outcomes that effect the aspects of
care. - Establish limits of the scope of assessment.
38Ten Step Approach to Quality Assurance
- Collect and organize data.
- Evaluate care when outcomes are reached.
- Take action to improve care
- Assess and document actions
- Communicate information to organization-wide QA
Program
39QA Projects
- Things that QA can evaluate includes
- Scheduling of patients
- Instructions given to patients
- Wait times in the office
- Interpretation of films
- Retake analysis
- Record accuracy
40QA Program
- Quality Assurance deals with people and processes
used to complete tasks. - QA involves training and record keeping.
- As the owner of the equipment, you will be
responsible for your radiology services. - The State of California Department of Radiologic
Health established the Standards of Good Practice
that is the foundation of QA and QC in
radiography.
41QA and QC Requirements
- Degree of requirements vary by state. California
and New York have very tight standards for
quality control of the radiographic and
processing equipment. - We are required by statue to teach QA and QC in
the radiology program. It is covered in detail in
9th Quarter. My textboook covers QC in detail.
42Quality Control
- An acceptable QC program has three steps
- Acceptance Testing
- Routine performance monitoring
- Maintenance
43Acceptance Testing
- The x-ray machine, cassettes and film processor
or digital system are the largest capital expense
you may experience. - It makes economic sense to make sure that the
equipment meets the performance standards. - It is recommended that a third party such as a
health physicist do the testing.
44Acceptance Testing
- Areas that should be tested include on the x-ray
machine - Shielding of Room
- Focal spot size
- Calibration of mA, timer or mAs
- Calibration of kVp
- Linearity of exposure
- Beam alignment
- Grid centering
- Collimation
- Filtration (HVL)
45Acceptance Testing
- Areas that should be tested on the x-ray
cassettes - Screen contact
- Screen speed
- Light leaks
- Light spectrum matching
46Acceptance Testing
- Areas that should be tested on the x-ray film
processor - Developer temperature
- Replenishment rates
- Travel time
- Water flow
- Hypo retention
47Quality Control
- The acceptance testing ensures that the machine
was installed and calibrated properly. - The performance may drift or deteriorate over
time. Consequently, periodic testing is required
to monitor the performance. - With the exception of film processing most
testing is annual or semiannual.
48Quality Control
- After a major repair, the machine should be
retested to ensure that it was repaired properly. - When the testing shows that the machine is not
performing properly, service or preventive
maintenance is required. - Manufactures establish recommended preventive
service schedules. When these are followed many
repairs become unnecessary.
49Radiographic Quality Control
- Areas of concern in x-ray machine
- Focal Spot Size will impact spatial resolution
- Filtration will impact patient exposure
- Collimation will impact patient exposure
- kVp calibration will impact image quality and
exposure. - Exposure timer accuracy will impact image quality
and exposure
50Radiographic Quality Control
- Areas of concern in x-ray machine
- Exposure linearity will impact exposure and image
quality - Exposure reproducibility will impact exposure and
image quality. - Alignment of tube and image receptor will impact
exposure and image quality.
51Focal Spot Testing
- When the machine is installed or the tube is
replaced, the focal spot size should be measured.
Then annually thereafter. - A pin hole camera, star test pattern or line pair
test tool. - As the tube ages, the focal spot tends to grow
and spatial resolution is lost.
52Filtration
- The filtration is measured but determining the
half value layer of the beam at specific exposure
levels. Minimum filtration is 2.5mm aluminum. - As a tube ages, tungsten will plate the x-ray
port and increase filtration. This can cause
technique problems. Inadequate filtration will
significantly increase patient exposure.
53Collimation
- If the collimation is misaligned, intended
anatomy can be missed. - It can be tested in many ways from using pennies
to using test patterns. - Misalignment can not exceed 2 of the SID.
- It is tested semiannually and after the
replacement of the collimator lamp.
54kVp Calibration
- In diagnostic radiology, any change will impact
patient exposure. A variation of about 3 will
impact contrast and image density. - Can be tested with filtered ion chambers,
filtered photodiodes or even a cassette with
calibrated filters. - Tested annually.
55Exposure Timer Accuracy
- The exposure time is the responsibility of the
operator. It will impact the density of the image
and spatial resolution. - Tested with an ion chamber, multi-meter
internally or even a spinning top. - Exposure time must be within 5 for exposure
times greater than 10 ms and 20 less than 10 ms.
56Exposure Linearity
- Many combinations of mA and time will produce the
same mAs value. The ability of the machine to
produce a constant level of exposure with various
combinations of mA and time is called exposure
linearity. - Can be tested with a step wedge and densitometer
or rate meter. - Should be within 10 for adjacent stations.
57Exposure Reproducibility
- Any exposure using the same factors should
produce the same level of density and contrast on
the image. - Sequential exposure should be reproducible to
within 5 - Can be tested with a rate meter or step wedge and
densitometer.
58Performance standards for x-ray equipment
Measurement Frequency Tolerance US Tolerance Ca
Filtration Annually 2.5 mm Al 2.7mm Al
Collimation Semiannually 2 of SID 2 of SID
Focal Spot Annual 50 50
kVp Annual 10 2 kVp between 60 100 kVp
Timer Annual 5 gt 10 ms 20 20 ms 3 phase 5 1 phase 10
Linearity Annually 10 10
Reproducibility Annual 5 5
59Darkroom and Processing
- The development of the image is dependent upon
the temperature of the developer, its
concentration and how long the film is in the
developer. - The film is sensitive to variations in the
environment and processing from the time it is
manufactured until it is processed. - Darkroom and Processor QC is the key process of
Quality Control.
60Processing
- Processor densitometry is performed daily before
the first patient is exposed. - A sensitometer is used to produce a step wedge
image on the film that is evaluated with a
densitometer. - The densitometer reads the optical density of the
processed image. - A digital thermometer is used to test the
chemical temperatures in the processor.
61Processing
- Key densities on the processed film are measured
and then graphed. - Base plus Fog is measured on an area of unexposed
film to check the darkroom environment. - Speed is tested at the level of exposure that
produces a density of 1.25OD - Contrast is tested at the level that produced a
density of 0.40 OD and one that produced a
density of 2.20.
62Processing
- By monitoring these densities, problems with film
processing can be detected before image quality
deteriorates. - In 9th Quarter we will cover how to perform
processor QC and problem solve.
63Waste Records
- Since used fixer is classified as a hazardous
waste material, it is important to maintain
accurate records of usage and disposal. - The extent of records vary by city, county and
state. You are responsible for the proper
disposal of the waste. Some regions include
developer as hazardous waste.
64Silver Recovery
- If the silver ions are removed from the fixer, it
may be disposed of in the normal waste when
diluted with water. - There are two primary types of silver recovery
systems. - Metallic replacement uses steel wool and can
recover 95 of the silver in the effluent - Electrolytic recovery passes direct current
through the solution and nearly pure metallic
silver is deposited on the cathode.
65Silver Recovery
- Old radiographic films and repeated films are
retained for silver recovery. X-ray images can
not be disposed of in normal trash. - They also can not be used to clean the processor
rollers. - Waste recovery companies will either burn or
chemically remove the silver from the film.
66Performance standards for film processor and
darkroom equipment
Measurement Frequency Tolerance CA Tolerance US
Sensitometry Daily BF 0.05 MD 0.10 OD Contrast 0.10OD BF0.08 OD MD 0.15 OD Contrast 0.15 OD
Safelight Semiannual lt 0.05 OD in 2 minutes n/a
Darkroom temp Monthly 70ºF 5º n/a
Darkroom humidity Monthly 50 10 n/a
Developer temp Daily 0.5ºF 2-3º
Replenishment Daily 5
Transport Annual 3 3
67Accessory QC
- The cassettes and screens are the area of chief
concern. Problems with either will result in
artifacts on the images and increased retakes. - The screens need to be properly cleaned
frequently. Mammography screens are cleaned
daily. California recommends monthly cleaning.
68Accessory QC
- Dirty screens produce white artifacts on the
image. - Multiple white artifacts indicate the need to
replace the screens. - Screen contact is tested semiannually. A problem
with screens contact will cause a loss of
resolution. - As screens age, they loose speed so this is also
tested.
69Accessory QC
- The other important accessory is the gonad
protection devices and lead aprons. - Improper care of the apron can result in cracks
and holes in the lead that reduces their
effectiveness. - Aprons and shields are tested semiannually. The
easiest was to test them is with
video-fluoroscopy but film can be used.
70Performance Standards for Accessories
Measurement Frequency Tolerance CA Tolerance US
Screen contact Semi annual No problems detected No problems detected
Aprons/shields Annually No holes No holes
Screen matching Annually 0.05 OD for all cassettes used
Screen cleaning Monthly Semimonthly
71Record keeping
- All electromechanical devices need periodic
service. - There are three types of maintenance.
- Scheduled maintenance such as processor monthly
or weekly service. It includes observing moving
parts and lubrication. - Preventive maintenance is planned service and
replacement of parts at regular intervals before
they fail at inopportune times.
72Record keeping
- Non-scheduled maintenance is the worst type of
service because it impacts patient service. It
may also be very expensive. - With proper scheduled and preventive maintenance,
non-scheduled service can be minimized. - All service schedules should meet manufacture
recommendations. - All service should be documented as part of the
quality assurance program.
73Retake Analysis
- Required part of a QA program in California.
- Evaluation includes
- View repeated
- Cause of the repeat
- Rate of retakes should be less than 5.
- Information can be gathered from the log that the
state mandates for patients being exposed to
radiation.
74Retake Analysis
- Done every three months using a relatively large
sample of data to see trends in - Type of examination being repeated.
- Reasons for the repeated films.
- Determine if additional training or review is
needed. - Determine if equipment service might be required.
75Repeat Analysis
- Your patient log can be designed to capture both
films usage and repeated films. - Data is gathered from log for analysis
- Repeated films can be put into two main
categories - X-ray Personnel Errors
- Equipment malfunctions
76X-ray Personnel Errors
- Failure to measure patient.
- Use of improper technical factors (mAs, kVp or
distance) - Incorrect positioning
- Improper Collimation
- Improper use of accessories such as cassettes,
grids or filters
77X-ray Personnel Errors
- Improper handling of exposed or unexposed films.
- Failure to clearly communicate to the patient
breathing instructions and to remain still. - Failure to observe patient during exposure.
78X-ray Equipment and Accessory Failure or
Malfunctions
- Inaccurate calibration of kVp and mA.
- Inaccurate timer calibration
- Dirty or damaged cassettes
- Improperly labeled or damaged grids
- Malfunctioning collimator
- Improper film storage or processor function.
79Reasons for Retake Films
- Over or under exposure accounts for over 50 of
retakes nationally. - Errors in positioning (25)
- Patient motion 11
- Processing errors 6
- Wrong view, beam alignment,cassette screen or
grid errors and artifacts.
80Retake Films by Region
- Cervical Spine Exams 7
- Thoracic Spine Exams 17
- Lumbar and Abdominal X-rays 40
- Skull, Chest,Lower Extremities 15
- Majority or retake results in unnecessary
exposure to gonads or blood forming organs of the
body.
81Daily Log Design
- Most regulatory agencies will require a log of
patients having radiation exposure. - Columns can be added to capture film usage,
repeats, views repeated and reason. - This data can be gathered and analyzed.
82Retake Analysis
- In this example, most of the retakes were of the
T-spine. Potential reasons include - Improper use of filters
- Incorrect measurements
- Faulty technique chart
83Retake Analysis
- The reason of each retake is recorded and
percentages are computed to determine the
overall rate and rate by reason. - Less than 5 is ideal.
84End of Lecture