Practical HPLC - PowerPoint PPT Presentation

About This Presentation
Title:

Practical HPLC

Description:

Practical HPLC In This Section, We Will Discuss: How to set up an HPLC System for a sample injection including: Solvent Handling Mobile Phase preparation Priming the ... – PowerPoint PPT presentation

Number of Views:1255
Avg rating:3.0/5.0
Slides: 32
Provided by: hk40
Category:
Tags: hplc | practical

less

Transcript and Presenter's Notes

Title: Practical HPLC


1
Practical HPLC
2
In This Section, We Will Discuss
  • How to set up an HPLC System for a sample
    injection including
  • Solvent Handling
  • Mobile Phase preparation
  • Priming the HPLC
  • Column Handling - Equilibration
  • System Performance Checks

3
Solvent Handling
  • Solvent Characteristics (Specifications)
  • Purity
  • Viscosity
  • Refractive index
  • Boiling Point
  • Toxicity
  • UV Transparency/UV-Cutoff
  • Solubility

4
Solvent Miscibility
5
Solvent UV-Cutoff/Transparency
Solvent
UV Cutoff (nm)
Acetonitrile
190
Water
190
UV cutoff is the wavelength at which absorbance
equals 1, measured in a 1 cm cell with air as a
reference.
Cyclohexane
195
Hexane
200
Methanol
210
Ethanol
210
Diethyl Ether
220
Dichloromethane
220
Chloroform
240
Carbon Tet
265
Tetrahydrofuran
280 (220)
Toluene
285
6
Mobile Phase Preparation
  • Major Steps
  • Measure appropriate volume of each solvent
  • Mix solvents
  • Add buffers and additives
  • Filter mobile phase
  • Degas mobile phase

7
Priming HPLC System
8
Priming HPLC System
9
Priming HPLC System
10
Priming HPLC System
11
Priming HPLC System
12
Priming HPLC System
13
Priming HPLC System
14
Priming HPLC System
15
Priming HPLC System
16
Priming HPLC System
17
Priming the HPLC
18
Sample Preparation
  • At a minimum - filter samples
  • Nylon - hydrophilic nature works with aqueous and
    solvent based samples, autoclavable to 121ÂșC, pH
    range 3-12, no concentrated acids.
  • PTFE- a hydrophobic membrane which is highly
    resistant to solvents, acids, and alkalis. This
    filter is generally used for non-aqueous samples.
    pH range 1-14.
  • Cellulose Acetate- good filter for aqueous
    biological samples with very low protein
    retention. pH range 4-8.
  • PVDF- highly resistant to most solvents, exhibits
    low protein binding. pH 2-12.
  • Ultrafilter Membranes- molecular weight cut-off
    filters for biological samples.
  • Nitrocellulose- exhibits high protein retention.
  • Solid Phase Extraction.

19
Sample Preparation
  • Dissolve the sample in the mobile phase or in a
    solvent weaker than the mobile phase.
  • The sample volume should be kept as small as
    possible.

20
Column Storage
  • Avoid any physical stress to the column.
  • Close on both ends to avoid dryness.
  • Store the column well flushed with the
    appropriate solvent.
  • Record the history of the column .

21
Column Installation
  • Each column has a defined flow direction!
  • The flow direction is shown by the arrow or
    direction of writing.
  • Dont change the flow direction, this will
    decrease column performance.

22
Column Installation
23
Column Installation cont.
Practical hints
  • Finger tighten
  • 1/4 turn with wrench

24
Column Equilibration
  • Equilibrate with mobile phase
  • Do not pressure shock the column.
  • 5-10 column volumes for reversed-phase
    equilibration.
  • Assures reproducible results.

25
Column Check
  • New columns should be delivered with a
    performance certificate.
  • Each additional use should be documented
    including
  • Back pressure
  • Mobile Phase
  • Temperature
  • Sample type
  • Storage condition (Solvent)

Based on that history the column can be checked
with defined compound mixture.
26
Column Care and Handling
  • Wash the column after use with selected solvents
    flush highly retained sample components from the
    column, eliminate buffers.
  • Do not store a column in 100 water. Microbes
    may grow and clog the column.
  • Do not store the column in 100 Acetonitrile.
  • Dont open the column and repack the material if
    you want to maintain performance.
  • Use the column at its optimal flow rate - avoid
    high flow rates.
  • Do not operate silica or bonded phases for
    extended periods at high temperature.
  • Keep the pH of the mobile phase in an appropriate
    range for the column.

27
System Check - Routinely
  • Principle
  • The HPLC system (including the column) can be
    checked out using a defined test sample and
    method. Use at least three replicates.
  • Preparations for a system check
  • HPLC system is primed with mobile phase.
  • Column is equilibrated.
  • Detector shows a stable response.
  • There are no leaks.
  • System is ready for injection

28
System Check - routinely cont.
  • Test sample requirements
  • Sample is well characterized.
  • Detector response is known.
  • Sample contains multiple components.
  • Test design
  • The test sample is analyzed using a defined test
    method. The results are compared with the
    expected results. If the results are in the
    defined range, than the system is ready for use.
  • This is not comparable to an OQ test or PV test!!

29
Summary
  • Prepare mobile phase
  • Prime the HPLC system
  • Install the column
  • Turn on the detector (warm-up at least 20 minutes
    for UV)
  • Equilibrate the column
  • Prepare the samples
  • Record the detector response - stable response
  • Perform a system check using a test sample and
    test method
  • Compare the results with the expectations
    (limits)
  • Document the results (Control Chart)
  • Record any failures/errors if appropriate
  • If system check is OK, then

You are Ready for Sample Analyses
30
Review
1. You are running a routine analysis when you
notice a periodic perturbation in the
baseline. The pressure reading is fluctuating
up and down. What is the problem? How would
you correct it?
31
Review
2. You decide to run a reversed-phase analysis
on an instrument in lab. The previous
operator does not indicate the solvents last used
on the instrument. You place water in
channel A and turn on the pump. You
cannot get a stable baseline. Suggest a possible
reason for this dilemma.
Write a Comment
User Comments (0)
About PowerShow.com