Ultrasound Physics - PowerPoint PPT Presentation

1 / 41
About This Presentation
Title:

Ultrasound Physics

Description:

Ultrasound Physics Artifacts Our Hospital Physics Group George David, M.S. Associate Professor of Radiology Nothing to do with Anything Artifacts Assumptions can ... – PowerPoint PPT presentation

Number of Views:378
Avg rating:3.0/5.0
Slides: 42
Provided by: George57
Category:

less

Transcript and Presenter's Notes

Title: Ultrasound Physics


1
Ultrasound Physics
  • Artifacts

Our Hospital Physics Group
George David, M.S. Associate Professor of
Radiology
2
Nothing to do with Anything
3
Artifacts
  • Assumptions can cause artifacts when assumed
    conditions are not true
  • sound travels at 1540 m/s
  • sound travels in a straight line
  • All sound attenuation exactly0.5 dB/cm/MHz

4
Distance from Transducer
  • Calculation of Distance
  • scanner accurately measures time delay between
    sound generation echo reception

Distance Assumed Speed X Measured Delay / 2
Actual Distance to interface
58 usec
1380 m/s X 58usec / 2 4 cm
4 cm
Calculated Distance to interface
V 1380 m/s
1540 m/s X 58usec / 2 4.47 cm
(Average speed of sound in soft tissue)
5
Distance from Transducer
  • Echo positioning on image
  • distance from transducer calculated from assumed
    speed of sound
  • can place reflector too close to or too far from
    transducer
  • can alter size or shape of reflector

V 1380 m/s
X
Actual Object Position
Position of Object on Image
X
V 1540 m/s
6
Attenuation
  • For all scanning your scanner assumes
  • soft tissue attenuation
  • .5 dB/cm per MHz
  • Your scanners action
  • compensate for assumedattenuation
  • allow operator fine tuning
  • TGC

7
Shadowing
  • Clinical Manifestation
  • reduction in imaged reflector amplitude
  • Cause
  • object between this reflector transducer
    attenuates ultrasound more than assumed
  • assumed compensation not enough to provide proper
    signal amplitude
  • intensity under-compensated
  • Opposite of Enhancement

Attenuates more than .5 dB/cm/MHz
Shadowed Reflector
8
Shadowing
Attenuates more than .5 dB/cm/MHz
Shadowed Reflector
http//raddi.uah.ualberta.ca/hennig/teach/cases/a
rtifact/noframe/imag2-f2.htm
9
Enhancement
  • Clinical Manifestation
  • increase in imaged reflector amplitude
  • Cause
  • object between reflector transducer attenuates
    ultrasound less than assumed
  • assumed compensation more than needed to provide
    proper signal amplitude
  • intensity over-compensated
  • Opposite of Shadowing

Attenuates less .5 dB/cm/MHz
Enhanced reflector
10
Enhancement
Attenuates less .5 dB/cm/MHz
Enhanced reflector
http//raddi.uah.ualberta.ca/hennig/teach/cases/a
rtifact/noframe/imag6-f1.htm
11
Shadowing / Enhancing
  • these artifacts not necessarily bad
  • can help in determining nature of masses
    upstream of artifact which caused shadowing /
    enhancing

12
Scanner Assumptions
  • Echo positioning on image
  • direction of all sound travel assumed to be
    direction that sound was transmitted

X
Refraction
13
Refraction Artifact
  • refraction alters beam direction
  • direction of sound travel assumed to be direction
    sound transmitted

X
Refraction
14
Refraction Artifact
  • refraction alters beam direction
  • scanner places dot in wrong location along line
    of assumed beam direction
  • can alter reflector shape

15
Lobe Artifacts
  • Side Lobes
  • beams propagating from a single element
    transducer in directions different from primary
    beam
  • reflections from objects here will be placed on
    main sound transmission line
  • Grating Lobes
  • same as above except for transducer arrays

X
16
Range Ambiguity
  • Reflection from 1st pulse reaches transducer
    after 2nd pulse emitted
  • scanner assumes this is reflection from 2nd pulse
  • places echo too close in wrong direction

1
2
17
Range Ambiguity
  • To improve any 1 of 3, at least 1 of other 2 must
    be reduced.
  • many scanners automatically reduce frame rate as
    depth increases

Depth
RangeAmbiguityTrade-off
Lines / Frame
Frames / sec(dynamics)
18
Scanner Assumptions
Multipath Artifact
Actual Object Position
Position of Object on Image
X
X
19
Multiple Reflection Scenario
  • reflection from reflector B splits at A
  • some intensity re-reflected toward B
  • Result
  • later false echoes heard
  • scanner places dots behind reflector B

2
3
1
A
B
real
1
2
false
3
20
Artifacts
  • Reverberation (multiple echo) artifact
  • comet tail effect is 1 example
  • can have dozens of multiple reflections between
  • transducer reflector
  • 2 reflectors
  • Mirror Image
  • common around diaphragm pleura

21
Artifacts
http//raddi.uah.ualberta.ca/hennig/teach/cases/a
rtifact/noframe/imag1-f1.htm
Caused by Shotgun Pellets
22
Multiple Reflection Scenario
http//raddi.uah.ualberta.ca/hennig/teach/cases/a
rtifact/noframe/imag5-f2.htm
23
Resolution Artifacts
  • Axial and Lateral Resolution Limitations
  • results in failure to resolve 2 adjacent
    structures as separate
  • minimum image size equal to resolution in each
    direction

24
Section Thickness Artifact
  • anatomy may not be uniform over its thickness
  • universal problem of imaging 3D anatomy
  • in CT MRI this is known as partial volume effect

Thickness
25
Constructive Interference
  • 2 echoes received at same time
  • in phase
  • Result
  • higher intensity



26
Destructive Interference
  • 2 echoes received at same time
  • Exactly 180o out of phase
  • Result
  • flat (zero) wave

-

27
Acoustic Speckle
  • texture seen on image may not correspond to
    tissue texture
  • Results from interference effects between
    multiple reflectors received simultaneously which
    can
  • add together
  • constructive interference
  • subtract from one another
  • destructive interference

28
Mirror Image Doppler
  • Analogous to mirror image artifact discussed
    previously
  • mirrored structures can include mirrored vessel
  • duplicate image visible on opposite side of
    strong reflector
  • example bone
  • Doppler data also duplicated
  • flow spectrum copied from original vessel

29
Spectral Duplication
  • mirror image of Doppler spectrum appears on
    opposite side of baseline
  • causes
  • electronic duplication caused by receiver gain
    set too high
  • overloads receiver
  • True sensing caused by too large Doppler angle
  • beam covers flow in both directions

Blood flows toward transducer
Blood flows away from transducer
30
Doppler Artifacts
  • Doppler spectrum speckle
  • Cause
  • same as acoustic speckle
  • random constructive destructive interference
    from sound scattered in blood

31
Aliasing
  • Results in detection of improper flow direction
  • occurs because sampling rate too slow
  • Similar to wagon wheels rotating backwards in
    movies

32
Aliasing
Sufficient Sampling
Insufficient Sampling
33
Aliasing
  • Which way is this shape turning?

OR
1
2
3
34
Aliasing
  • Did the shape turn 1/4 turn right or 3/4 turn
    left?
  • 1 1/4 turn right?

1
2
3
35
Aliasing
  • Does it help to sample more often?

2
1
1A
2A
3A
3
36
Aliasing
  • Maximum detectable Doppler shift equals half the
    pulse repetition frequency
  • Sampling rate
  • Same as pulse repetition frequency
  • Must be at least twice highest frequency to be
    sensed
  • Aliasing occurs when Doppler shift exceeds 0.5
    PRF

37
Aliasing
  • Maximum detectable Doppler Shift not limited for
    continuous wave Doppler
  • Maximum detectable Doppler Shift is limited for
    pulsed instruments

Maximum Detectable Doppler Shift half pulse
repetition frequency
38
Coping with Aliasing
  • decrease transducer frequency
  • reduces Doppler shift
  • shift proportional to operating frequency
  • increase pulse repetition frequency
  • decreases maximum imaging depth
  • increases likelihood of range ambiguity for
    pulsed instruments

77 X fD (kHz) v (cm/s)
-------------------------- fo
(MHz) X cosq
39
Coping with Aliasing
  • increase Doppler angle
  • Reduces relative flow rate between blood
    transducer
  • Reduces Doppler shift sensed by scanner

77 X fD (kHz) v (cm/s)
-------------------------- fo
(MHz) X cosq
40
Coping with AliasingBaseline Shifting
  • operator instructs scanner to assume that
    aliasing is occurring
  • scanner does calculations based on operators
    assumption
  • scanner has no way of determining where in image
    aliasing occurs

Yes
No
41
Any Questions?
Write a Comment
User Comments (0)
About PowerShow.com