Title: Instrument Qualification DQ, IQ, OQ, PQ vs. Validation and Routine MaintenanceCalibration
1Instrument Qualification (DQ, IQ, OQ, PQ) vs.
Validation and Routine Maintenance/Calibration
- Tom Layloff
- Management Sciences for Health
- www.msh.org
- The views expressed here are those of the author
and not necessarily those of MSH
2Drivers
- Society
- Litigious
- Risk-free
- ICH
3Regulators Want
- Assurance Of Product Safety, Efficacy And Quality
- Assurance Of Reliable Data And Information
- Repeatable
- Reproducible
4Analytical Instrument Users Want
- Versatile And Reliable Instruments Which Meet
Their Requirements. - Make Them The Envy Of All Other Analysts.
- Best Toy On The Block.
- Their Budgeted Instrument Request Delivered
Yesterday. - More
- Sensitivity
- Selectivity
- Resolution
- Accuracy
- Precision
5Instrument Manufacturers
- Successful Analytical Instrument Manufacturers
Want To Sell Lots Of Equipment To Satisfied
Users. - Successful Analytical Instrument Manufacturers
Constantly Monitor The Law, Regulation, And
Guidances Seeking To Identify As Soon As Possible
New Analytical Instrument Requirements.
6Instrument Company Success
- A Successful Instrument Manufacturer Must
Anticipate The Availability Of Funds And Drivers
For Additional Data. - The Societal Values Must Be Served.
- Instrument Companies Which Do Not Adequately
Address The Needs Of The Testing Community Do Not
Survive.
7USER REQUIREMENT SPECIFICATION (URS)
- Sales And Marketing Personnel Visit With Users To
Identify Needs And Bells And Whistles. - Minimum Performance Requirements Must Be
Identified. - Focus Groups May Aid In Refining The URSs.
8The Road Forward
- The URSs Drive The Development Of Functional
Requirement Specifications (FRS) Which Establish
The Design Qualification (DQ). - The FRS Are Developed Into The Detailed Design
Specification (DDS). The DDS Establishes The
Basis For The Installation Qualification (IQ). - The DDS Provide The Basis For Construction.
9(No Transcript)
10FDA CMC Guidance
- Methods validation is the process of
demonstrating that analytical procedures are
suitable for their intended use. The methods
validation process for analytical procedures
begins with the planned and systematic collection
by the applicant of the validation data to
support the analytical procedures.
11Instrument ValidationToms Revision
- Instrument validation is the process of
demonstrating that analytical instruments are
suitable for their intended use. The validation
process for analytical instruments begins with
the planned and systematic collection by the
applicant of the validation data to support the
analytical applications.
1221 CFR 211.160 (b)(4)
- The calibration of instruments, apparatus,
gauges, and recording devices at suitable
intervals in accordance with an established
written program containing specific directions,
schedules, limits for accuracy and precision, and
provisions for remedial action in the event
accuracy and/or precision limits are not met.
Instruments, apparatus, gauges, and recording
devices not meeting established specifications
shall not be used.
13GUIDELINE ON GENERAL PRINCIPLES OF PROCESS
VALIDATION, MAY, 1987
- 1. Equipment and Process
- a. Equipment Installation Qualification
- b. Process Performance Qualification
- Installation qualification - Establishing
confidence that process equipment and ancillary
systems are capable of consistently operating
within established limits and tolerances.
14Coup De Grâce
- It is important that the manufacturer prepare a
written validation protocol which specifies the
procedures (and tests) to be conducted and the
data to be collected. The purpose for which data
are collected must be clear, the data must
reflect facts and be collected carefully and
accurately. - In assessing the suitability of a given piece of
equipment, it is usually insufficient to rely
solely upon the representations of the equipment
supplier, - Information obtained from these studies should be
used to establish written procedures covering
equipment calibration, maintenance, monitoring,
and control.
15Continuing 1
- Installation qualification studies establish
confidence that the process equipment and
ancillary systems are capable of consistently
operating within established limits and
tolerances. After process equipment is designed
or selected, it should be evaluated and tested to
verify that it is capable of operating
satisfactorily within the operating limits
required by the process.
16Two Prongs of Equipment Purchasing/Receipt
- Good Business Practices ISO Standards
- Purchased Goods Must Conform to The Purchasers
Requirements and Specifications - cGMP
- Equipment Must Be Fit For Intended Use
- Equipment Must Conform to Requesters
Specifications - Performance Assessment
17ISO 9001-2000
- 7.4 Purchasing
- 7.4.1 Purchasing process
- The organization shall ensure that purchased
product conforms to specified purchase
requirements. The type and extent of control
applied to the supplier and the purchased product
shall be dependent upon the effect of the
purchased product on subsequent product
realization or the final product. The
organization shall evaluate and select suppliers
based on their ability to supply product in
accordance with the organization's requirements.
Criteria for selection, evaluation and
re-evaluation shall be established. Records of
the results of evaluations and any necessary
actions arising from the evaluation shall be
maintained (see 4.2.4).
18- 7.4.2 Purchasing information
- Purchasing information shall describe the product
to be purchased, including where appropriate - a) requirements for approval of product,
procedures, processes and equipment, - b) requirements for qualification of personnel,
and - c) quality management system requirements.
- The organization shall ensure the adequacy of
specified purchase requirements prior to their
communication to the supplier.
19- 7.4.3 Verification of purchased product
- The organization shall establish and implement
the inspection or other activities necessary for
ensuring that purchased product meets specified
purchase requirements. Where the organization or
its customer intends to perform verification at
the supplier's premises, the organization shall
state the intended verification arrangements and
method of product release in the purchasing
information.
20ISO 17025 Section 4.6.2
- (a) All incoming supplies and reagents that can
affect the quality of test results shall be shown
to be suitable for their intended use through
acceptable control sample testing or other system
suitability testing prior to their release for
routine use or supplier certifications may be
accepted after the reliability of the supplies
has been established. A program to periodically
monitor the supplies must be implemented to
assure the continued adherence to those standards.
21(No Transcript)
22(No Transcript)
23(No Transcript)
24Fused Silica Capillary 1
- The small improvement are overshadowed by
the fact that most of the adverse quality issues
of two decades ago remain. variations in
capillary bore and outer dimensions, ovality,
random brittleness, and surface activity.
25Fused Silica Capillary 2
- Although fused-silica capillary enables
separation methods unimaginable prior to its
introduction, most manufactured capillaries are
highly variable in quality and less than optimal
for any application.
26Fused Silica Capillary 3
- Not only have fused-silica capillary columns
revolutionized GC, but the newer separation
science technologies capillary electrophoresis
(CE), capillary LC, and capillary
electrochromatography (CEC)perhaps would not
have been explored without the availability - Steve Griffin, LCGC North America, 2002, 20,
928-938.
27Waters 717 plus Autosampler
- Precision lt0.5 RSD (5 µL to 100 µL)
- Linearity gt0.99 coefficient of correlation
- Carryover lt0.1 from previous injection
- Cross Contamination lt0.1 from previous sample
- Injection Range 0.1 µL - 2000 µL (in 0.1 µL
increments)
28Agilent Standard Autosampler
- 0.1 - 100 µl
- Typically lt0.5 RSD from 5-100 µl,
- typically lt1 RSD from 1 - 5 µl
- Up to 1500 µl
- Typically lt0.5 (injection volume lt1000 µl),0.5
(injection volume gt1000 µl) - 0.1 - 900 µl l
- Typically gt0.5 (injection volume lt5 µl),
- lt0.5 (injection volume 5 - 900 µl)
- Up to 1800 µl
- lt0.5 up to 1800 µl
- Up to 5000 µ
- lt0.5 (injection volume lt 2000 µl),
- 1 (injection volume gt 2000 µl)
29Uncertainty in HPLC Analysis
- It is impossible to run all autosamplers at a
rsd better than 0.5. In fact, it is difficult
to keep them in good enough shape to provide an
overall rsd of nmt 1.0. Experience shows a
reasonable uncertainty of 2 or more from the
autosampler. - Autosamplers-A Major Uncertainty Factor in HPLC
Analysis Precision, S. Kuppers, B. Renger, and
V.R. Meyer, LC-GC Europe, February, 2000.
30Waters 1500 SERIES HPLC PUMPS
- Programmable Flow Rate Range 0.00 - 10.00 mL/min,
in 0.01 mL increments - Flow Precision lt 0.1 RSD or 2 seconds SD, six
replicates, based on retention time or volumetric
measures, at 1 mL/min, 1000 - 2000 psi
backpressure - Flow Accuracy 1.0 or 30 uL/min, whichever is
greater, at 1 mL/min, degassed methanol, 1000 -
2000 psi backpressure (each pump individually) - Maximum Operating Pressure 6000 psi (41,370 kPa,
401 bar) - Delay Volume (1525 pump)lt200 uL (without manual
injector or column heater and with one GM 150
mixer) - Gradient Compositional Accuracy (1525) 0.5 of
setting at 1 mL/min and 1000 psi backpressure
(MethanolMethanol with propylparaben) with one
GM 150 mixer
31Pump Specifications 2
- Gradient Compositional Precision (1525) lt0.5 RSD
of setting at 1 mL/min and 1000 psi backpressure
(Methanol Methanol with propylparaben) with one
GM 150 mixer (based on 6 replicates of
compositional accuracy) - Acoustic Noise lt70 dB (A) at operator position
- Operating Temperature Range 4 C - 40 C
- Operating Humidity Range 20 - 80,
non-condensing - Physical Size (W x H x D) (without bottle holder)
- 12 inches x 17 inches x 24 inches
- 30.5 cm x 43 cm x 61 cm
- Weight
- 1515 pump 45 lbs. or 20.4 kg
- 1525 pump 60 lbs. or 27.2 kg
32Agilent 1100 HPLC Pump
- Pump Flow
- Precision lt0.3 RSD (typically lt0.15 RSD)
- Range Setpoints from 0.001 to 5.0 ml/min, in
0.001 ml/min increments - Pressure
- Operating range 0 - 40 MPa (0 - 400 bar, 0 - 5880
psi) - Pulsation lt2 amplitude (typically lt1 ) at 1
ml/min isopropanol at all pressures gt1 MPa (gt 10
bar, gt147 psi) - Gradient
- Delay volume 180 - 480 µl (600-900 µl with mixer)
dependent on back pressure - Composition precision lt0.20 SD, at 0.1 and 1
ml/min - Recommended pH range 1.0 - 12.5
- Solvents with pH lt2.3 should not contain acids
which attack steel
33Influence of Poor Peak Shape on Integration This
slide courtesy of Waters, Inc.
34BMW Z Decisions
35BMW Z3
36(No Transcript)
37Invicta S1
- Engine
- 4.6 Liter V-8
- DOHC Per Bank
- 320 bhp at 5,900 rpm
- 300 lb-ft at 4,800 rpm
- Performance (claimed)
- Top Speed 180 mph
- 0-60 mph in 5 sec.
- 0-100 mph in 11 sec.
38Bottom Lines
- The World Is Infinitely Complex.
- It Is Impossible To Validate The Intricate
Minutiae In Complex Systems. - Appropriate Performance Parameters Must Be
Established In SOPs To Demonstrate Fitness For
Use. - Control Samples Which Challenge The Typical
Analytical Method Validation Dimensions Should Be
Used.
39What Do We Want From Our Instrument?The
Performance Required For Method Validation
- Accuracy
- Precision
- Specificity
- Detection limit
- Quantitation limit
- Linearity
- Range
- Robustness
- Typical Validation Characteristics ICH and, USP
lt1225gt