Instrument Qualification DQ, IQ, OQ, PQ vs. Validation and Routine MaintenanceCalibration PowerPoint PPT Presentation

presentation player overlay
1 / 39
About This Presentation
Transcript and Presenter's Notes

Title: Instrument Qualification DQ, IQ, OQ, PQ vs. Validation and Routine MaintenanceCalibration


1
Instrument Qualification (DQ, IQ, OQ, PQ) vs.
Validation and Routine Maintenance/Calibration
  • Tom Layloff
  • Management Sciences for Health
  • www.msh.org
  • The views expressed here are those of the author
    and not necessarily those of MSH

2
Drivers
  • Society
  • Litigious
  • Risk-free
  • ICH

3
Regulators Want
  • Assurance Of Product Safety, Efficacy And Quality
  • Assurance Of Reliable Data And Information
  • Repeatable
  • Reproducible

4
Analytical Instrument Users Want
  • Versatile And Reliable Instruments Which Meet
    Their Requirements.
  • Make Them The Envy Of All Other Analysts.
  • Best Toy On The Block.
  • Their Budgeted Instrument Request Delivered
    Yesterday.
  • More
  • Sensitivity
  • Selectivity
  • Resolution
  • Accuracy
  • Precision

5
Instrument Manufacturers
  • Successful Analytical Instrument Manufacturers
    Want To Sell Lots Of Equipment To Satisfied
    Users.
  • Successful Analytical Instrument Manufacturers
    Constantly Monitor The Law, Regulation, And
    Guidances Seeking To Identify As Soon As Possible
    New Analytical Instrument Requirements.

6
Instrument Company Success
  • A Successful Instrument Manufacturer Must
    Anticipate The Availability Of Funds And Drivers
    For Additional Data.
  • The Societal Values Must Be Served.
  • Instrument Companies Which Do Not Adequately
    Address The Needs Of The Testing Community Do Not
    Survive.

7
USER REQUIREMENT SPECIFICATION (URS)
  • Sales And Marketing Personnel Visit With Users To
    Identify Needs And Bells And Whistles.
  • Minimum Performance Requirements Must Be
    Identified.
  • Focus Groups May Aid In Refining The URSs.

8
The Road Forward
  • The URSs Drive The Development Of Functional
    Requirement Specifications (FRS) Which Establish
    The Design Qualification (DQ).
  • The FRS Are Developed Into The Detailed Design
    Specification (DDS). The DDS Establishes The
    Basis For The Installation Qualification (IQ).
  • The DDS Provide The Basis For Construction.

9
(No Transcript)
10
FDA CMC Guidance
  • Methods validation is the process of
    demonstrating that analytical procedures are
    suitable for their intended use. The methods
    validation process for analytical procedures
    begins with the planned and systematic collection
    by the applicant of the validation data to
    support the analytical procedures.

11
Instrument ValidationToms Revision
  • Instrument validation is the process of
    demonstrating that analytical instruments are
    suitable for their intended use. The validation
    process for analytical instruments begins with
    the planned and systematic collection by the
    applicant of the validation data to support the
    analytical applications.

12
21 CFR 211.160 (b)(4)
  • The calibration of instruments, apparatus,
    gauges, and recording devices at suitable
    intervals in accordance with an established
    written program containing specific directions,
    schedules, limits for accuracy and precision, and
    provisions for remedial action in the event
    accuracy and/or precision limits are not met.
    Instruments, apparatus, gauges, and recording
    devices not meeting established specifications
    shall not be used.

13
GUIDELINE ON GENERAL PRINCIPLES OF PROCESS
VALIDATION, MAY, 1987
  • 1. Equipment and Process
  • a. Equipment Installation Qualification
  • b. Process Performance Qualification
  • Installation qualification - Establishing
    confidence that process equipment and ancillary
    systems are capable of consistently operating
    within established limits and tolerances.

14
Coup De Grâce
  • It is important that the manufacturer prepare a
    written validation protocol which specifies the
    procedures (and tests) to be conducted and the
    data to be collected. The purpose for which data
    are collected must be clear, the data must
    reflect facts and be collected carefully and
    accurately.
  • In assessing the suitability of a given piece of
    equipment, it is usually insufficient to rely
    solely upon the representations of the equipment
    supplier,
  • Information obtained from these studies should be
    used to establish written procedures covering
    equipment calibration, maintenance, monitoring,
    and control.

15
Continuing 1
  • Installation qualification studies establish
    confidence that the process equipment and
    ancillary systems are capable of consistently
    operating within established limits and
    tolerances. After process equipment is designed
    or selected, it should be evaluated and tested to
    verify that it is capable of operating
    satisfactorily within the operating limits
    required by the process.

16
Two Prongs of Equipment Purchasing/Receipt
  • Good Business Practices ISO Standards
  • Purchased Goods Must Conform to The Purchasers
    Requirements and Specifications
  • cGMP
  • Equipment Must Be Fit For Intended Use
  • Equipment Must Conform to Requesters
    Specifications
  • Performance Assessment

17
ISO 9001-2000
  • 7.4 Purchasing
  • 7.4.1 Purchasing process
  • The organization shall ensure that purchased
    product conforms to specified purchase
    requirements. The type and extent of control
    applied to the supplier and the purchased product
    shall be dependent upon the effect of the
    purchased product on subsequent product
    realization or the final product. The
    organization shall evaluate and select suppliers
    based on their ability to supply product in
    accordance with the organization's requirements.
    Criteria for selection, evaluation and
    re-evaluation shall be established. Records of
    the results of evaluations and any necessary
    actions arising from the evaluation shall be
    maintained (see 4.2.4).

18
  • 7.4.2 Purchasing information
  • Purchasing information shall describe the product
    to be purchased, including where appropriate
  • a) requirements for approval of product,
    procedures, processes and equipment,
  • b) requirements for qualification of personnel,
    and
  • c) quality management system requirements.
  • The organization shall ensure the adequacy of
    specified purchase requirements prior to their
    communication to the supplier.

19
  • 7.4.3 Verification of purchased product
  • The organization shall establish and implement
    the inspection or other activities necessary for
    ensuring that purchased product meets specified
    purchase requirements. Where the organization or
    its customer intends to perform verification at
    the supplier's premises, the organization shall
    state the intended verification arrangements and
    method of product release in the purchasing
    information.

20
ISO 17025 Section 4.6.2
  • (a) All incoming supplies and reagents that can
    affect the quality of test results shall be shown
    to be suitable for their intended use through
    acceptable control sample testing or other system
    suitability testing prior to their release for
    routine use or supplier certifications may be
    accepted after the reliability of the supplies
    has been established. A program to periodically
    monitor the supplies must be implemented to
    assure the continued adherence to those standards.

21
(No Transcript)
22
(No Transcript)
23
(No Transcript)
24
Fused Silica Capillary 1
  • The small improvement are overshadowed by
    the fact that most of the adverse quality issues
    of two decades ago remain. variations in
    capillary bore and outer dimensions, ovality,
    random brittleness, and surface activity.

25
Fused Silica Capillary 2
  • Although fused-silica capillary enables
    separation methods unimaginable prior to its
    introduction, most manufactured capillaries are
    highly variable in quality and less than optimal
    for any application.

26
Fused Silica Capillary 3
  • Not only have fused-silica capillary columns
    revolutionized GC, but the newer separation
    science technologies capillary electrophoresis
    (CE), capillary LC, and capillary
    electrochromatography (CEC)perhaps would not
    have been explored without the availability
  • Steve Griffin, LCGC North America, 2002, 20,
    928-938.

27
Waters 717 plus Autosampler
  • Precision   lt0.5 RSD (5 µL to 100 µL)
  • Linearity   gt0.99 coefficient of correlation
  • Carryover  lt0.1 from previous injection
  • Cross Contamination lt0.1 from previous sample
  • Injection Range 0.1 µL - 2000 µL (in 0.1 µL
    increments)

28
Agilent Standard Autosampler
  • 0.1 - 100 µl
  • Typically lt0.5 RSD from 5-100 µl,
  • typically lt1 RSD from 1 - 5 µl 
  • Up to 1500 µl
  • Typically lt0.5 (injection volume lt1000 µl),0.5
    (injection volume gt1000 µl) 
  • 0.1 - 900 µl  l
  • Typically gt0.5 (injection volume lt5 µl),
  • lt0.5 (injection volume 5 - 900 µl)
  • Up to 1800 µl
  • lt0.5 up to 1800 µl
  • Up to 5000 µ 
  • lt0.5 (injection volume lt 2000 µl),
  • 1 (injection volume gt 2000 µl)

29
Uncertainty in HPLC Analysis
  • It is impossible to run all autosamplers at a
    rsd better than 0.5. In fact, it is difficult
    to keep them in good enough shape to provide an
    overall rsd of nmt 1.0. Experience shows a
    reasonable uncertainty of 2 or more from the
    autosampler.
  • Autosamplers-A Major Uncertainty Factor in HPLC
    Analysis Precision, S. Kuppers, B. Renger, and
    V.R. Meyer, LC-GC Europe, February, 2000.

30
Waters 1500 SERIES HPLC PUMPS
  • Programmable Flow Rate Range 0.00 - 10.00 mL/min,
    in 0.01 mL increments
  • Flow Precision lt 0.1 RSD or 2 seconds SD, six
    replicates, based on retention time or volumetric
    measures, at 1 mL/min, 1000 - 2000 psi
    backpressure
  • Flow Accuracy 1.0 or 30 uL/min, whichever is
    greater, at 1 mL/min, degassed methanol, 1000 -
    2000 psi backpressure (each pump individually)
  • Maximum Operating Pressure 6000 psi (41,370 kPa,
    401 bar)
  • Delay Volume (1525 pump)lt200 uL (without manual
    injector or column heater and with one GM 150
    mixer)
  • Gradient Compositional Accuracy (1525) 0.5 of
    setting at 1 mL/min and 1000 psi backpressure
    (MethanolMethanol with propylparaben) with one
    GM 150 mixer

31
Pump Specifications 2
  • Gradient Compositional Precision (1525) lt0.5 RSD
    of setting at 1 mL/min and 1000 psi backpressure
    (Methanol Methanol with propylparaben) with one
    GM 150 mixer (based on 6 replicates of
    compositional accuracy)
  • Acoustic Noise lt70 dB (A) at operator position
  • Operating Temperature Range 4 C - 40 C
  • Operating Humidity Range 20 - 80,
    non-condensing
  • Physical Size (W x H x D) (without bottle holder)
  • 12 inches x 17 inches x 24 inches
  • 30.5 cm x 43 cm x 61 cm
  • Weight
  • 1515 pump 45 lbs. or 20.4 kg
  • 1525 pump 60 lbs. or 27.2 kg

32
Agilent 1100 HPLC Pump
  • Pump Flow
  • Precision lt0.3 RSD (typically lt0.15 RSD)
  • Range Setpoints from 0.001 to 5.0 ml/min, in
    0.001 ml/min increments
  • Pressure
  • Operating range 0 - 40 MPa (0 - 400 bar, 0 - 5880
    psi)
  • Pulsation lt2 amplitude (typically lt1 ) at 1
    ml/min isopropanol at all pressures gt1 MPa (gt 10
    bar, gt147 psi)
  • Gradient
  • Delay volume 180 - 480 µl (600-900 µl with mixer)
    dependent on back pressure
  • Composition precision lt0.20 SD, at 0.1 and 1
    ml/min
  • Recommended pH range 1.0 - 12.5
  • Solvents with pH lt2.3 should not contain acids
    which attack steel

33
Influence of Poor Peak Shape on Integration This
slide courtesy of Waters, Inc.

34
BMW Z Decisions
35
BMW Z3
36
(No Transcript)
37
Invicta S1
  • Engine
  • 4.6 Liter V-8
  • DOHC Per Bank
  • 320 bhp at 5,900 rpm
  • 300 lb-ft at 4,800 rpm
  • Performance (claimed)
  • Top Speed 180 mph
  • 0-60 mph in 5 sec.
  • 0-100 mph in 11 sec.

38
Bottom Lines
  • The World Is Infinitely Complex.
  • It Is Impossible To Validate The Intricate
    Minutiae In Complex Systems.
  • Appropriate Performance Parameters Must Be
    Established In SOPs To Demonstrate Fitness For
    Use.
  • Control Samples Which Challenge The Typical
    Analytical Method Validation Dimensions Should Be
    Used.

39
What Do We Want From Our Instrument?The
Performance Required For Method Validation
  • Accuracy
  • Precision
  • Specificity
  • Detection limit
  • Quantitation limit
  • Linearity
  • Range
  • Robustness
  • Typical Validation Characteristics ICH and, USP
    lt1225gt
Write a Comment
User Comments (0)
About PowerShow.com