Title: Claude B Anger, MS, Microbiology
1Claude B Anger, MS, (Microbiology)
- Current Retired, Consultant, CBA
MicroEnterprises - Past
- 18 years as Director of RD Microbiology
Cytotoxicity and the Corporate Microbiologist for
Allergan, Inc. - 10 years as Manager of RD Microbiology for
Allergan, Inc. - 5 years as microbiology supervisor for Baxters
Microbiological Reagents Division at Hyland
Laboratories (diagnostics mfg) - 3 years as senior research microbiologist for Max
Factor - 1 year as the chief microbiologist for Purex
Corporation - Honors
- 1985-1990 PMA, chairman, biological testing Div
of Biological Section - 1991-1995 USP, Expert advisor to the
Microbiology subcommittee - 1992-2001 ANSI microbiology expert to ISO for
Microbiology Cytotoxicity standards development
for contact lens care products
2Acanthamoeba Keratitis can be anOutbreak
diseaseHistory in the USA
3Disinfection is not SterilizationContact lens
disinfection is not hospital disinfection
- Disinfection selects for surviving resistant
microbes efficacy depends on temperature, agent
concentration contact time - Disinfection products are defined into three
efficacy categories, low, intermediate and high - Hydrophilic contact lens (a class II non-critical
device) disinfection was/is considered to be low
efficacy due to the potential toxicity to the
external eye from the hydrophilic contact lens
containing disinfectant. - Low efficacy disinfection is not required to be a
sporicidal, fungicidal or tuberculocidal. CL
disinfection is now asked to kill fungal spores
and perhaps amoebic cysts which will be
intermediate disinfection - can we change CL
disinfection stringency without increased ocular
toxicity from stronger formulations? - The CL disinfection rub step by the patient can
deposit 104 microbes from the finger/hand onto
the lens, thorough hand washing before removal of
lenses should be stressed as a major compliance
factor
4Contact Lens Disinfection History
- 1976-1981 FDA guidance declared CL disinfection
efficacy to be measured by (1) sterilization type
kill rate D-values for the active disinfectant
solution, (2) the 20 item sterility regimen test
and (3) contribution of elements lens tests
against the following organisms - Staphylococcus aureus
- Pseudomonas aeruginosa
- Serratia marcescens (the most highly pigmented
strain) - Candida albicans (yeast cell)
- Aspergillus fumigatus (mold spore lung infector,
high disinfection resistance) - Herpes simplex virus
- No instructions were given as to how the
disinfectant D-values were to be determined
(multi-sloped kill curves) nor any efficacy
criteria to meet except the 20 lens MIMC test
having no survivors. Contact lens soaking times
set at 2-3 x the largest D-value
5Contact Lens Disinfection History
- 1982-1990 FDA guidelines modified at least 6
times - Serratia marcescens strain changed to a
non-pigmented ocular infection isolate strain
(quaternary ammonium can be used) - Herpes simplex virus is dropped from the efficacy
indicator challenge organism list - very low
infection prevalence with CL - Aspergillus fumigatus is changed to Fusarium
solani as the mold spore efficacy indicator
organism because Fusarium was a more prevalent
ocular infector and it was less resistant to
contact lens disinfecting agents - D-value
criteria is set at fungistasis
6Contact Lens Disinfection History
- 1982-1990 FDA guidelines modified at least 6
times - 20 lens MIMC regimen test has the criteria
changed from zero recovery to 10 or less
survivors per lens/solution combination in the
lens case well - Acanthamoeba was not added to the efficacy
indicator list still working on AK root cause
and how to correct for the AK outbreak plus no
Acanthamoeba culture and challenge tests are
available
7Contact Lens Disinfection History
- 1990-2003
- Contact lens disinfection efficacy standard ISO
FDIS 14729 is developed by ANSI, FDA, EU and
Japan experts for world wide use and the FDA
adopts the ISO methods and criteria as their
guideline efficacy for contact lens disinfection
product approval. - The Acanthamoeba outbreak of 1984-1992 was ended
in the USA when salt tablets and the use of home
made solutions in contact lens care were removed
from the marketplace inadequate disinfection was
not the issue
8Contact Lens Disinfection History
- 1990-2003
- From 1993-2003 the AK incidence in the USA fell
to 19 cases/yr as compared to 115 cases/yr during
the 84-91 US outbreak and the only country with
an AK problem remaining was the UK. Consequently
the ISO and FDA did not consider it relevant for
CL disinfection to add Acanthamoeba to the list
of efficacy indicating challenge organisms
because CL disinfection efficacy was not the
apparent issue for the AK outbreak no
standardized/validated Acanthamoeba disinfection
challenge test methods were available and no
agents active against Acanthamoeba and compatible
with contact lens care were evident.
9Contact Lens Disinfection Efficacy Testing
Anomalies against Acanthamoeba
Borazjani RN, Kilvington S. Efficacy of
multipurpose solutions against Acanthamoeba
species. SCL Ant Eye 2005 28169-175
10Cyst strain production method vs Disinfection
A. polyphaga Log Reductions at 4 hr contact
Hughes R, Heaselgrave W, Kilvington S.
Antimicrob. Agents Chemother 2003 473080-3084
11What is the Root Cause of an AK outbreak
- The use of any water in the contact lens care
regimen was and still is the source of
Acanthamoeba for CL associated AK - ALWAYS FORBID
THIS PRACTICE. - The 1990s CL disinfection marketplace rejected
the highest efficacy systems, heat units and
2-step 3H2O2, for ease of use and comfort from
multi-purpose solution systems without increase
in the AK occurrence rate. Contact lens
disinfection will have resistant survivor
microorganisms in the lens case after multiple
procedures so the patient hygiene and care
regimen (water?) determines if Acanthamoeba will
be present. - In 1990 Dr Mary Beth Moore published in Cornea 9
Suppl 1S33-5 that for AK the patient was at
fault, and stated patients can not be relied upon
to follow routines true?
12What is the Root Cause of an AK outbreak
- The contact lens storage case is the culprit here
because it allows disinfection survivor microbes
to set up a biofilm after about 2 weeks of use
when it is not cleaned regularly. Rinsing the
lens case or lenses with water supplies the
Acanthamoeba and the biofilm is the food for
trophozoite growth. This is where AK begins what
is the infection dose for AK ,1 or 50? - What is the corrective action for AK root cause?
Absolutely stop patient use of water in lens care
and physically clean the lens case regularly
(also can put it in boiling water). - Will increasing contact lens disinfection
stringency really help reduce AK, especially if
the current products can pass the new criteria
which will mean there is no real change? Are
current products really inadequate? - An intensive program for the education of contact
lens wear patients to strictly follow the care
product regimen is warranted, with warnings that
failure to comply can lead to blindness.
13Contact Lens Storage Case Cleaning Disinfecting
- Practitioners and industry must stress NEVER USE
WATER IN THE CONTACT LENS CARE REGIMEN ANYWHERE!!
Not to rinse the lens case nor the lenses - Practitioners should advocate physical cleaning
of the contact lens storage case by a procedure
with a dedicated instrument (hard tooth brush)
and CL disinfectant or with boiling water every
week and never air-dry the lens case but keep it
filled with fresh disinfectant and sealed ready
for the next use (do not discard solution, use it
to disinfect lenses and discard after use) - To insure against lens case biofilm (estimate 2
weeks to set up) source infections practitioners
should advocate to patients a once a week
conventional home microwave treatment of CL
disinfectant filled lens cases with 50 power for
3 minutes and no lenses present.
Hiti K, Walochinik J, et al. Microwave
Treatment of Contact Lens Cases Contaminated with
Acanthamoeba. Cornea, 2001
14Issues for Standardizing CL Disinfection Efficacy
Testing against Acanthamoeba species challenges
- Product formulas tested for ability to cause
trophozoite to encyst - Survivor cell recovery must be quantitative and
not an estimate - The acceptable standard of error for these
methods is 0.5 log - The reasonable Acanthamoeba challenge level is
103/ mL - Will the new tests and criteria cause current
products to be off the market are current
products considered inadequate for CL
disinfection? - If the current CL disinfection products are
unacceptable what new anti-acanthamoeba agents
are available for CL disinfection use and will
there be a grace period (5 years) for research
and development of new products? - Any new test preamble should contain rationales
to address the above issues for testing personnel
to understand the methods, the acceptance
criteria and the expected standard error
15Contact Lens Disinfection Efficacy Challenge
TestsWhat Do They Really Measure
- STAND ALONE
- Ability of active disinfectant to kill an aqueous
challenge of various microbial cell types in
various resistance states. ID most resistant type - Can generate specific microbe kill kinetics and
the 3 log reduction times - Establishes contact lens soaking time for the
designated regimen solution compatibility with
different lenses and lens storage cases? - Selects for resistant organisms in the challenge
organism cell inoculum (weak, normal resistant
cells are present)
16Contact Lens Disinfection Efficacy Challenge
TestsWhat Do They Really Measure
- REGIMEN (test performed on sterile surgeon gloves
in a hood) - The inoculum contact time with the contact lens
is too short at 5 minutes, no time for microbes
to attach to test lenses. How long should this
time be? The rub step is a major source of
microbes, but physical rubbing with a surfactant
solution can dislodge attached microbes from the
lens surface for removal in the rinse step, so
lens rub is advised - Clean rinse with saline can remove nearly all
of the test lens inoculum, a high force stream
with a large volume (20 seconds) Data from
positive controls should be generated. - Organic soil presence has significance and should
be used - For Acanthamoeba challenges, the lens type makes
a big difference in recovery (very sticky cells
with high affinity for silicone hydrogel lenses)
17Straw-man proposed Acanthamoeba challenge Test
current systems have been demonstrated to pass
the regimen test
- Organism A. castellanii from the ATCC, a T-4
genotype from an ocular infection - Trophozoites Cultured axenically in a defined
PYG medium at 30ºC for 72 hours. - Cyst production Trophozoites transferred to
Neffs constant pH encystment medium to give
about 106 cells/mL and incubated at 30ºC for 7
days, centrifuge, then wash resuspend the
pellet to yield 5 3 x 105 cyst PFU /mL - Quantitation of Inocula The agar sandwich
bacterial plaque assay of serial dilutions
18Straw-man proposed Acanthamoeba challenge Test
current systems have been demonstrated to pass
the regimen test
- Product challenge 3 mL product in a contact lens
storage case is inoculated with 0.03mL of a
standardized cell suspensions in Pages saline to
yield 103 to 104 cells per mL of product and
incubated at 23ºC. (lens present? Which type) - Viable cell assay at designated contact time
intervals remove 0.5 mL of test solution and
serially dilute 10-fold in a neutralizer broth
and plate 0.5 mL of each dilution in duplicate by
the bacterial plaque assay method. - All steps must be properly validated for
precision, accuracy and reproducibility measures.
Standard error for plate count methods is 0.5
log. - CRITERIA 2 log reduction of trophozoite
viability and NLT 0.7 log reduction in cysts
during the manufacturer recommended minimum lens
soaking time after an n of 3