Title: FieldSpec
1FieldSpec HH UV/VNIR
- HandHeld Spectroradiometer
2Analytical Spectral Devices, Inc.
- 5335 Stearling Drive, Suite A
- Boulder, CO 80301
- (303) 444-6522 fax(303) 444-6825
- www.asdi.com
- Asd_support_at_asdi.com
3Device Specifications
- Range 325-1075 nanometers
- 512 channel silicon photodiode array overlaid
with an order separation filter - Each channel will receive light in a 1.6
nanometer nominal band - Spectral resolution of approximately three
nanometers at 700 nanometers
4Optics
- Light is collected through fiber optic head
called a bare head - The bare head has a field of view (FOV) of 25
degrees. Foreoptics may be added to restrict this
FOV. - 10 and 1 degree FOV restricting foreoptics are
included.
5Pelican case is not waterproof
6Thinking Inside the Box
Shore Power
Serial to USB Cable
Battery
FOV 1 and 10 deg
Spectroradiometer
Help
Battery Charger
Small White Reference
7The Spectroradiometer
8Serial to USB Cable
9Charger
RED Light Charging
GREEN Light Charged
FLASHING RED Check connection
10Mounting Device
11Secure to Tripod
- Remove quick release head from tripod
- Insert only into bottom hole on device
- Do not over tighten only use fingers
- Check tripod tight-points before trusting
12Back of Device
Battery and Power Indicators
Power
Battery Port
Shore Power Port
Battery Port
Serial Port
13FOV Device
14FOV Limiting Foreoptics
- Place threaded end on bare head
- Do not over tighten
- Use extreme care
- Threads are located on end with label
15Insert Battery and Turn On
16Review
- Dont over tighten anything
- 1.) connect cables and FOV limiting foreoptics
- 2.) add battery and turn on by holding down power
- 3.) now you are ready to attach to the computer!!
17IMPORTANT
- Do not plug the GPS into the serial port until
you have turned on the computer. If the computer
freezes you may need to turn off the ballpoint
mouse in the device manager - Dont let the spectroradiometer get too hot! If
the spectra looks odd and it is hot out, first do
a DC. If this will not help you will need to let
it cool down.
18Tough Book
19Convert to Tablet (if needed)
20Tough Book Overview
Connections
Controls
Power on
21Tough Book Controls
On-Screen Keyboard
Lock
Rotate Display
22Tough Book Connections
Power Port
USB for FieldSpec HH
23Once FieldSpec HH is connected and on
For high-glare environments
For normal lighting
24View Mode Buttons
Raw DN and reflectance selector
White Reference and DC Buttons
Status Bars
Spectral View Area
Save Status Bar
GPS Display
25Detail of Top Bar
Pan mode allows for navigation while zoomed
Freezes current spectra to allow examination
Zoom mode lets you examine spectra closer
XY is normal view mode
Closes shutter
26Set the Foreoptic First
27Set the Integration Time
28Integration Time
- Adjust to improve signal strength
- Avoid saturation (right), that will appear as
off-graph and be displayed in red above sample
status bar
29CDA
- Configurable Default Averaging will help to
reduce noise - Higher number will take more time, but improve
quality of signal - Control gt Adjust Configuration
30CDA Window
10 is default, but no less than 20 should be
used.
31- Always open Spectrum save before the first save
- In this window set file name and other save
settings - Control gt Spectrum Save
- Spacebar will save spectra
32This information is encoded in every saved
spectrum
Save Location
Amount / Frequency of save
Save Name/Start Number
Comments
33Good Luck