Title: Sam Ju Cho


1
Sam Ju Cho
A Study on the characteristic of normoxic polymer
gel dosimeter according to its composition
  • Dept. of Radiation Oncology
  • Dongguk University International Hospital

2
Introduction
3
Introduction
To evaluate complex radiation therapy such as
IMRT, IGRT, SRS and brachytherapy, The polymer
gel has been developed as a 3D dose verification
dosimeter Need for the polymer gel to use as a
dosimeter Have to know the dose response
characteristic for different dosimetry
requirement Because the polymer gel is a type of
chemical dosimeter Component type and
composition ratio of polymer gel affect the dose
response characteristic
4
Introduction
5
Materials and Methods
6
Materials and Methods
Ø 1.6 cm, W 6 cm
Normoxic polymer gel fabrication and test vial
7
Materials and Methods
8
Materials and Methods
30x30x30 cm3 water phantom Medical linac 6 MV
Put the test vials in the 2 dose uniformity of
water phantom TRS 398 protocol for dose
calibration
9
Materials and Methods
MR image acquisition
Before imaging, put the test vials in the MR room
(24 h) Styrofoam case for the test vials 1.5T
and Head coil Multi spin-echo, MX 256, ST 5 mm,
FOV 180, BW, 130, NEX 1 TR 3000 ms, TE 30 240
ms equal space
10
Materials and Methods
Dose response delivered dose D and measured MR
T2 (1/R)time
relationship Dose sensitivity dose response
divide by delivered
dose D Dose reproducibility deviation of
T2 variation among
measured gel dosimter - 8 gelatin, 8 MAA,
10 mM THPC, 0.05 mM HQ - 2, 4 Gy to each 5 vials
11
Materials and Methods
Dosimeter analysis
  • Dose accuracy delivered dose Dd and measured
    dose
  • Dm relationship
  • - 8 gelatin, 8 MAA, 10 mM THPC, 0.05 mM HQ
  • 0, 2, 4, 6 Gy delivered to each 3 vials for
    fitting curve
  • using fitting curve, Absorbed doses were
    calculated
  • for 10 vials which used dose reproducibility
    study
  • Dose rate dependency dose rate and measured dose
    D
  • relationship
  • - 8 gelatin, 8 MAA, 10 mM THPC, 0.05 mM HQ
  • 200 MU/min, 600 MU/min and 2, 6 Gy delivered

12
Materials and Methods
Dose uncertainty used as measurement dose error
Combined standard uncertainty of the dose
For a 95 level of confidence
Dose resolution for a given level of
confidence, the minimal detectable dose
For a 95 level of confidence
Baldock C, Lepage M, Bäck SÅJ, et al. Dose
resolution in radiotherapy polymer gel dosimetry
effect of echo spacing in MRI pulse sequence.
Phys Med Biol 46449-4602001
13
Results
14
Results
T2 time calculation I
T2 times of each polymer gel were calculated by
using TR, TE and signal intensity
15
Results
T2 time calculation II
Signal intensity versus TE
16
Results
Dose response I
The concentration of the gelatin is high, the
threshold R2 value increase and the dose response
decreases.
17
Results
Dose response II
The concentration of the MAA is high, the
threshold R2 value decrease and also the dose
response decrease.
18
Results
Dose Sensitivity I
The concentration of the gelatin increase, the
dose sensitivity increase within some rage.
19
Results
Dose Sensitivity II
The concentration of the MAA increase, the dose
sensitivity increase within some rage.
20
Results
Dosimetric Characteristic I
21
Results
The solid line is a calibration curve and the
circles are measured data for dose accuracy
evaluation
22
Results
23
Results
Dose rate dependency is shown in the high dose
region of high dose rate
24
Conclusions
25
Conclusions
26
Thank you for your attention