Title: Getting Started with the EVAPro
1Getting Started with the EVAPro
1
After you remove your EVAPro from the box, check
to make sure you have all accessories shown below.
Click here for next screen
22
Power
ON
Flow Meter
Check Oil
Fail Pass
Flag
Power
On
Flow Meter Ball
Check Oil
Selector Switch
Test Smoke 0.040 0.020 Future
Quick Reference
Tester Hook Up Phase-One Phase-Two
Click here for next screen
33
Nitrogen / Shop Air Connector Fitting
Remote Smoke Button
12-Volt DC Power Cable
Smoke Supply Hose
Accessories Compartment
Click here for next screen
44
Connect nitrogen or shop air supply hose to the
quick-connector shown below. The EVAPro comes
with a built-in regulator that allows you to use
nitrogen for all EVAP testing or shop air for all
of your general purpose leak detection needs.
Click here for next screen
55
Choose the proper EVAP port adaptor for the
Vehicle from the accessory kit.
The Schrader removal tool is used for taking the
core out of the EVAP service port.
Click here for next screen
66
1. Remove the OBD-II service port adaptor cap.
2. With the Schrader removal tool turn clockwise
and remove the core from the service port . You
must remove the core before testing. After
testing,dont forget to reinstall the OBD-II
service port Schrader core and cap. If the
vehicle being tested does not have an EVAP
service port, then access the EVAP system at the
purge valve line that leads to the charcoal
canister.
Click here for next screen
77
Lets do a step by step EVAP test. Remember that
you must use nitrogen!
1. Connect the red lead to positive on battery.
2. Connect the black lead to chassis ground.
With power leads connected, the green light
should be ON.
A low battery will cause the Testers green Power
light to blink. See Trouble Codes in your
Operator Manual.
Click here for next screen
88
For the purpose of this presentation, we will use
the .040 standard.
- Position the Testers selector switch to the
.040 calibration. - Turn the Tester ON.
- Set the Pass / Fail mark by using your thumb to
align the red pointer flag with the ball. - Now the red flag is set to equal a .040
leak and it is calibrated for that vehicle.
Power ON Check Oil
Click here for next screen
Remember to calibrate for .040 or .020 before
each test.
9Have your scan tool ready. Go to Vent Solenoid
( On / Off ) as shown below.
9
If scantool doesnt close vent solenoid, refer to
Tech Tip Booklet
Click here for next screen
10Phase One
10
1. Set selector switch to TEST. 2. Close EVAP
vent solenoid. 3. Turn the Tester ON.
Power ON Check Oil
At first the ball will go to the top of the flow
meter. The ball will slowly drop until it no
longer goes down. 3. Once the ball stops moving
downward, check the alignment with the red flag.
If the ball is below the red flag, the leak(s)
are less than .040 PASS If the ball is above
the red flag, the leak(s) are greater than .040
FAIL Go to Phase Two.
Five minutes after turning the Tester ON, it will
turn off automatically.
Click here for next screen
11Phase Two
11
Power ON Check Oil
- Set selector switch to SMOKE.
- Turn the Tester ON.
- 3. Look for Smoke and/or dye exiting the leak(s).
Five minutes after turning the Tester ON, it will
turn off automatically.
Click here for next screen
12Look for both smoke and the ultraviolet-activated
fluorescent dye.
888-822-8832
www.vacutec.com
Click here to Exit PowerPoint