Title: SenoScan FullField Digital Mammography System
1SenoScan Full-Field Digital Mammography System
2SenoScan FFDM
Tuesday September 25, 1202 pm Eastern Time
Press Release SOURCE Fischer Imaging
Corporation Fischer Imaging Receives Final FDA
Approval for Its SenoScan Full Field Digital
Mammography System
3SenoScan System
4SenoScan Tech. Shield Acquisition Station
5SenoScan
6The 3 Key Characteristics
- Higher Resolution - 25 50 Micron Pixel size
- Accommodates more women - 21x29cm field of view
- Higher DQE, Lower dose - slot scanning means no
grid
7SCANNING MOTION OF DETECTOR
8Beam Quality
- High-output Tungsten-Rhenium target
- 8 kW power rating
- 0.3 mm focal spot
- 20 kVp to 45 kVp voltage range
- Pivoting arm
- Collimation to Detector Width
- No grid
9Typical technique
- SenoScan typical technique for the 4.2 cm
accreditation phantom - 29 kVp, 160 mA
- (or less than 35 mAs).
10Beam Quality
11Detector Design
- Mechanical parameters
- Scintillator
- Fibre-Optic plate
- CCD operating in TDI mode
- Correlated Double Sampling technology
12Cesium Iodide
221 mm
Fiber Optic Plate
4 CCDs
10 mm
COMPOSITION OF SENOSCAN DETECTOR
13Detector Characteristics
- 4 CCDs, 25 micron native resolution
- 50 micron standard resolution
- 2-D area detector
- Time delay and Integration (TDI)
- CsI for high DQE
- 10 lp/mm at 50 Microns
- 14 lp/mm at 25 Microns
14Detector Design
- Large charge well capacity 2 M electrons
- minimal dead space between CCDs
- Four CCDs butted for 220 mm length
- CCD matrix 400 x 2048 pixels or 200 x 1024
pixels - Fibre-optic plate shields CCD from X-rays for
minimum noise
15Key System Specifications
- 4096 x 5625 matrix, both resolution modes
- Field of view 21cm x 29 cm (standard),
11cm x 15 cm (high res) - Scan time five seconds
- Effective exposure time 200 milliseconds
- Two-monitor high resolution diagnostic
workstation - Dicom3 compatibility
16SenoScan
17Detector Sensitometric Response
18Detector performance - low exposure
19Spatial Resolution
- More than 5 lp/mm needed
- 70 of DCIS presents as calcifications
- 8 mm average size
- Dershaw 29th NCBC 4/00
- Prevalence of DCIS growing as screening
prevalence increases - SenoScan 5o and 25 Micron Pixel Size Valuable
Screening and Diagnostic Tool
20(No Transcript)
21SenoScan MTF
22Comparative dose, SenoScan vs. FSM
23Detective Quantum Efficiency
- Defined as
- DQE SNR2(image)/SNR2(incident)
- Can be shown to be proportional to
- MTF2(f)/NPS
- Often heuristically described as
- DQE ? (Image Quality) / (Patient Dose)
-
24(No Transcript)
25Effect of DQE
- Over an ensemble of images a large set of
mammograms, what is likely to matter is the
integral of the DQE curve - So that a measure of system effectiveness is
given by the area under the DQE curve
26From the DQE equations
- Recall that
-
- DQE ? (Image Quality) / (Patient Dose)
-
27(No Transcript)
28Patient Ergonomics
- 200 milliseconds effective exposure time
(velocity x pixels/column) - Breast support and paddle design cup the breast -
reducing chance of motion - Patient feedback indicates that the curved breast
support increases comfort for same amount of
compression
29(No Transcript)
30(No Transcript)
31SenoScan
32Unique Features of SenoScan
33(No Transcript)
34(No Transcript)
35(No Transcript)