Title: Intubation and Anesthetic Machines
1Intubation and Anesthetic Machines
2Endotracheal Intubation
- Types of endotracheal tubes
- Plain
- Used in birds
- Occasionally used in cats
- A plain tube does not form an airtight seal
within the trachea - Will allow secretions, blood or gastro-intestinal
fluid to enter the lungs
3Endotracheal Intubation
- Types of endotracheal tubes
- Cuffed
- Used in all species
- Cuff is inflated with air to produce a leak proof
seal - A pilot balloon is used to estimate degree of
inflation - Alternatively the cuff is filled until no leak of
air can be heard when the lungs are inflated to
25 cm H2O pressure
4Endotracheal Intubation
- Types of endotracheal tubes
- Cole
- Originally designed for babies
- The tube has excellent pressure-flow
characteristics and provides less resistance to
breathing - Absence of cuff allows insertion of a larger tube
- A cole tube is commonly used for cats
- "Shoulders" of the tube form an airtight seal at
the entrance to the larynx
5Endotracheal Intubation
- Size of endotracheal tube
- Choose an endotracheal tube as large as possible
without forming a "push-fit - Check the tube length alongside the animal
- The tip of the tube must be beyond the larynx but
not extend past the thoracic inlet
6Endotracheal Intubation
- Size of endotracheal tube
- Tube sizes refer to the internal diameter of the
part of the tube residing in the trachea - A few manufacturers print both internal and
external diameters on their endotracheal tubes - Purchase tubes with thin walls especially when
they are to be used on cats and small dogs - A thin wall allows a tube with a larger lumen to
be used
7Endotracheal Intubation
- Technique of intubation
- Orotracheal Intubation
- Dogs and Cats
- Open the dog or cat's mouth and view the larynx
- Use a bright overhead light or a laryngoscope
- There is no excuse for doubt when inserting an
endotracheal tube - You can see where it is -- in the trachea or in
the esophagus
8Endotracheal Intubation
- Technique of intubation
- Orotracheal Intubation
- Dogs and Cats
- Pull the tongue forward gently in the dog and cat
to move the larynx more rostral . Do not pull
hard on the tongue or it will could lacerated or
have nerve damage - Do not put your fingers inside the mouth (jaw
tone and the ability to reflexively close the
jaws persists into light anesthesia) - A plastic or metal stilette can be used inside an
endotracheal tube to stiffen it - This is to improve control of the direction of
the tip of the tube, not so that you can force
the tube into the larynx - Make sure that the tip of the stilette is not
sharp and does not extend beyond the end of the
tube, where it might lacerate the trachea
9Endotracheal Intubation
- Technique of intubation
- Orotracheal Intubation
- Dogs and Cats
- If you use a laryngoscope
- Place the tip of the blade over the tongue and
under the epiglottis - Placing it on the epiglottis will distort the
laryngeal opening and make it harder to intubate - Tracheal intubation can be performed with the
animal on its side, sternum, or back - The position may be dictated by the animal's
condition (e.g., with GI obstruction always
intubate the animal in sternal position, with the
head up) or by personal preference of the surgeon
10(No Transcript)
11(No Transcript)
12Endotracheal Intubation
- Technique of intubation
- Orotracheal Intubation
- Pigs and small ruminants
- Technique used is the same for dogs and cats
- The anatomy of the pharynx and larynx of these
species differs - Adult Cattle
- Intubation can be performed using a laryngoscope
with an extra long blade - Intubation most frequently accomplished without
viewing the larynx - You introduce your hand and arm into the mouth to
palpate the laryngeal opening and to guide the
tip of the endotracheal tube into the trachea - Horses
- Intubation performed "blind", without viewing the
larynx and without palpation.
13Nasotracheal Intubation
- Dogs, cats, pigs and ruminants
- Not performed
- Horses
- Easy to insert a tube from the nares through the
ventral nasal meatus, into the trachea - Used in some foals to administer halothane or
isoflurane to induce anesthesia - Used in adult horses during recovery from
anesthesia to relieve nasal obstruction from
mucosal swelling
14Pharyngostomy Intubation
- Tracheal intubation through a pharyngostomy
- Dogs and cats
- May be used in dogs and cats requiring oral
surgery, e.g., repair of fractured jaw, cleft
palate repair, lacerations of hard palate - Allows the surgeon greater access to the surgical
field
15Exotics
- Reptiles Most reptiles can be intubated if
large enough. - Avian Endotracheal intubation sometimes used.
High incidence of mucous blockage in tubes due to
very small tracheal openings and high mucous
production. Most avian veterinarians prefer air
sac cannulization. This is a surgical procedure
when a tracheal tube is inserted directly into an
air sac near the leg.
16Endotracheal Intubation
- Complications of endotracheal intubation
- If the internal diameter of the endotracheal tube
is too small in relation to the animal - There will be increased resistance to breathing
- Hypoventilation will result
- If the cuff is over inflated, or if it is
inflated for too long - The tracheal mucosa, and sometimes the cartilage,
will die and slough - Later the animal will show signs of tracheitis
- Only inflate the tube cuff with enough air to
prevent a leak when the animal's lungs are
artificially inflated - If you have a device to measure intra-cuff
pressure, inflate the cuff to 25 cm H2O pressure - If the duration of surgery is long, provided that
regurgitation has not occurred, the cuff should
be deflated and the tube repositioned every 2
hours
17Endotracheal Intubation
- Complications of endotracheal intubation
- If the tube is too long
- A primary bronchus may be intubated
- One lung will collapse
- Cyanosis may develop
- Delivery of inhalation anesthetic is impaired
- The animal may wake up
- Endotracheal tubes can cause airway obstruction
- They become twisted or kinked
- The bevel of the tube lies against the tracheal
wall - The cuff is over inflated and squashes the tube
lumen - The cuff bulges over the end of the tube
- K-Y lubricant is allowed to dry inside the tube
- Mucus and secretions accumulate during anesthesia
(especially cats and small dogs) or are allowed
to dry inside the tube (improper cleaning after
anesthesia).
18Endotracheal Intubation
- Complications of endotracheal intubation
- Trauma of the larynx and trachea
- Can produce laryngitis/tracheitis which will last
for several days - Laryngeal trauma may also predispose the animal
to laryngospasm or to granuloma formation at a
later date - Mucosal swelling in the recovery period may cause
partial or complete airway obstruction - Preferably do not cause trauma, but if swelling
occurs, inject dexamethasone IV and provide
supportive treatment (oxygen, reintubate) until
the swelling is reduced - Irregular heart rhythm
- Laryngoscopy causes bradycardia
- Tracheal intubation causes tachycardia and,
sometimes, ventricular premature depolarizations - Atropine premedication should prevent bradycardia
- Lidocaine, 2 mg/kg, applied directly to the
larynx or injected intravenously before
intubation should modify or prevent tachycardia - Lidocaine is used when tachycardia is anticipated
to cause deterioration of the patient.
19Endotracheal Intubation
- Complications of endotracheal intubation
- Leak in the cuff preventing an airtight seal and
controlled ventilation - Always check the cuff for leaks before inducing
anesthesia - Transfer of infection
- Clean tubes thoroughly, inside and outside,
between patients - Use a bristle brush and betadine (or similar)
solution - If ethylene oxide (ETO) sterilization is used,
adequate aeration is needed to eliminate irritant
residues
20Endotracheal intubation
- Complications of endotracheal intubation
- Bracycephalic syndrome (short nose breeds)
- Stenotic nares
- Everted layrngeal saccules
- Elongated soft palate
- Hypoplastic trachea
- Airways can collapse easily after extubation
- Wait until the dog has a good gag reflex before
extubating - Preoxygenate for 5 minutes before induction in
case of a difficult intubation
21Endotracheal Intubation
- Complications of endotracheal intubation
- Cats
- Can be very difficult to intubate due to
laryngeal spasms - Small amount of lidocaine swabbed on larynx can
numb it long enough to intubate - Many need to use a stylet in endotracheal tube to
help pass it
22Endotracheal Intubation
- Complications of endotracheal intubation
- Collapsing trachea
- Genetically weak trachea
- Preoxygenate severe cases
- Intubate as gently as possible as not to irritate
the trachea
23Endotracheal Intubation
- Procedure for tracheal extubation in dogs and
cats - Leave tube in place until swallowing reflex
returns - Remove blood clots from nasopharynx
- Deflate the cuff before extubation
- Exceptions
- When bleeding into the nose and mouth has
occurred - Excessive salivation has occurred
- Regurgitation has occurred
- Prevent the animal biting the tube while it is
being removed - Cost prohibitive
- Animal (dog or cat) can easily inhale or swallow
pieces of tube
24Parts of an Anesthetic Machine
- Gas cylinders
- Oxygen and nitrous oxide are contained in
compressed gas cylinders - Found as E cylinders that are usually attached to
the machine via yokes that are equipped with a
specific pin system - Tanks also come in large G and H cylinders
25Parts of an Anesthetic Machine
- Gas cylinders
- Pressure gauge attached to the cylinder indicates
the pressure of the gas in the tank - Pressure in a full O2 cylinder is 2200 psi
- Oxygen tanks should be changed when pressure
drops below 500 to 600 psi - The volume of O2 in an E cylinder can be
calculated by multiplying the psi by 0.3 - A full tank of 2200 psi will contain 660 L of O2
26Parts of an Anesthetic Machine
- Gas cylinders
- Nitrous tanks are stored at lower pressures
- A full tank is 770 psi
- Nitrous tanks should be changed when pressure
gauge drops below 500 psi - Both liquid and gas states are present but the
gauge reads only the gas state - Liquid evaporates to gas as soon as the gas
leaves the tank so the pressure in the tank will
not change until all the liquid state has
evaporated
27O2 cylinder, pressure releasing valve, pressure
gauge
28Parts of an Anesthetic Machine
- Pressure releasing valve (regulator)
- Reduces the high pressure of the O2 or nitrous
leaving the tank to a low pressure of 50 psi
29Parts of an Anesthetic Machine
- Flow meter
- Measure O2 or nitrous in L/min
- Allows the anesthetist to set the O2 or nitrous
oxide flow rates that will be delivered to the
animal - As the gas passes through the flow meter gas
pressure is reduced further to 15 psi
30Flow meter
31Parts of an Anesthetic Machine
- Vaporizer
- Converts the liquid anesthetic into a gas state
- Controls the amount of vaporized amount of
vaporized anesthetic mixed with the carrier gas
32Sevoflurane vaporizer
33Parts of an Anesthetic Machine
- Check valves
- Inhalation/Exhalation flutter valves
- Insures a uni-directional flow of gas to and from
the patient when delivering a circle system - Y connector
- Connects the endotracheal tube to the inspiratory
and expiratory tubes of a circle system
34Parts of an Anesthetic Machine
- Rebreathing bag (reservoir bag)
- Allows the animal to breath easier from a
reservoir of gas - Can be used to deliver O2 (with or without
anesthetic gas) and manually assist respirations,
bagging - Bags should have a minimum volume of 60 ml/kg of
patient weight
35Parts of an Anesthetic Machine
- Carbon dioxide absorber
- Soda lime canister
- Used in rebreathing systems to remove CO2 from
the expired gases - Exhaust gases enter a canister containing soda
lime or barium hydroxide - Na, K, Ca2 and Ba2 hydroxide reacts with the
exhaled CO2 and water to form carbonate - Heat is liberated and the pH decreases
36Parts of an Anesthetic Machine
- Carbon dioxide absorber
- Soda lime canister
- A pH color indicator turns blue on consumption
- When the soda lime or barium hydroxide granules
turn color or the granules become hard instead of
crumbly, they are saturated with CO2 and should
be replaced - When in use, the granules will produce heat and
condensation within the canister - The color reaction is time limited
- Exhausted crystals should be removed immediately
and replaced with new granules
37Parts of an Anesthetic Machine
- Carbon dioxide absorber
- Soda lime canister
- Should be changed after 6 to 8 hours of use
depending on the size of the animal and the gas
flow rate - If machines are left standing for longer than 30
days, granules should be replaced before using
machine
38Parts of an Anesthetic Machine
- Exhaust valve
- Also called the pop-off valve, or pressure relief
valve - Exhaust gases leave the system via the exhaust
valve entering the scavenger system - Valve can be fully or partially open when a
patient is using the machine - Valve is closed for leak tests or when filling
the reservoir for assisted respirations
39Parts of an Anesthetic Machine
- Manometer
- Measures the pressure in the system in mm Hg or
cm H20 - Generally calibrated form -30 to 50 cm H20
- Gauge thus reflects the pressure of gas in the
animals airways and lungs - The pressure should be at 0 and never more than
15 cm H20 (11 mm Hg) - When providing positive assisted ventilation, the
pressure should not exceed 15 to 20 cm H20 (11 to
15 mm Hg)
40Parts of an Anesthetic Machine
- Oxygen flush valve
- O2 bypasses vaporizer, delivering 100 O2 to
breathing system - Enables the anesthetist to flush the system with
pure O2 - Fills the reservoir and system for leak test
- Also flushes the anesthetic gases out of the
circuit and replaces with pure oxygen - Never use O2 flush valve with a Bain circuit in a
small animal because it produces too much pressure
41Parts of an Anesthetic Machine
- Scavenger system
- Attached to the exhaust valve
- Consists of tubing that collects gases and
directs them outside the building or to a
charcoal canister - Can be active or passive
42Parts of an Anesthetic Machine
- Negative pressure relief valve
- Some newer machines have this safety feature
- Valve opens in response to a negative pressure
situation in the system - Allows room air into the circuit
- Negative pressure could be due to an active
scavenger system or a low oxygen supply
43Maintenance
- Oxygen tanks must be turned off to prevent excess
pressure on the regulator - Flush the remaining O2 to minimize damage to the
pressure gauge and reducing valves - Turn flowmeter off to prevent sudden rush of O2
into the flowmeter when O2 is turned back on - Dont over tighten because the knobs can be
easily twisted off
44Maintenance
- After each anesthesia induction, removable
machine parts and anesthetic equipment that come
in contact should be washed in a mild, soapy
solution, soaked in a cold disinfectant,
thoroughly rinsed and dried - The dome valves and absorbent canister should be
disassembled and wiped dry. - Flutter valves need periodic removal and cleaning
with a disinfectant to prevent adherence to the
machine housing
45Maintenance
- Vaporizers should be turned off when not in use
and periodically emptied to prevent buildup of
the preservative and other residue - Best to clean and recalibrate by authorized
personnel every 6 to 12 months - Isoflurane does not contain a preservative
46Maintenance
- Barium hydroxide or soda lime granules found in
the CO2 absorbers need replacing when the
granules have changed color or cannot be easily
crumbled - Do not tightly pack and leave about 1 cm (1/2
inch) of air space - Avoid having dust enter tubing or hoses of the
machine - Rubber items will likely need to be replaced
after prolonged use
47Environmental concerns
- Environmental pollution can be minimized through
proper equipment use and scavenging of the gases - Safe exposure limit for inhalant anesthetic
agents has been set at 2 p.p.m. in room air - Everyone, especially pregnant women, should avoid
high levels of waste anesthetic gases - Much of the anesthetic levels are because of
leaks in anesthetic machines
48Environmental concerns
- Vaporizers and CO2 absorbers should be filled
with minimal personnel in a well ventilated area
while wearing gloves and masks - Do not turn the vaporizer on and off until or
while, the patient is connected to the machine - During recovery, keep patient in a well
ventilated area and on the machine until expired
gases are scavenged
49Environmental concerns
- Use active scavenging systems whenever possible
to ensure waste gases are drawn out of the area - If passive scavenging is used, keep the hose as
short as possible and have it travel downward
toward the exhaust - If it is not possible to install scavengers in
all rooms where machines are used, either use an
activated charcoal cartridge that must be
replaced after 12 hours or substitute injectable
anesthesia
50Environmental concerns
- Before anesthesia, the machine should be checked
for both high and low pressure leaks - Leakage of nitrous is the major environmental
concern - A high pressure system test monitors NO2 and O2
leakage - A low pressure system leak is in the anesthetic
machine itself
51Environmental concerns
- Low pressure system test
- A low pressure system leak occurs between the
flowmeter and the patient - Turn the tank on, close the pop-off valve, and
occlude the end of the hose so the gas should
have nowhere to escape - Adjust the flowmeter to at least 2 L/min of O2
allowing the bag to fill gradually and then turn
off the flowmeter - If there is no escape of air when the bag is
gently squeezed, then there is no low pressure
system leakage
52Environmental concerns
- Low pressure system test
- System can also be checked by occluding as above
and using the flowmeter to allow the system to
pressure at 30 cm H2O - Turn off the flowmeter
- The pressure should be maintained for at least 10
seconds - One can also listen for the hiss of escaping air
or use a detergent solution as described earlier
53Breathing Systems
- Rebreathing systems
- Circle systems
- Rebreathing refers to breathing a mixture of
expired gases and fresh gases - The amount of CO2 in inhaled gases depends on
- Whether the rebreathing system has a CO2 absorber
- The flow rate of fresh gases (the higher the
fresh gas flow rate, the more expired gas is
pushed out the scavenger and not rebreathed - Depending on the flow rate of fresh gas, the
system is classified as a closed system (total
rebreathing of expired gases) or semi-closed
system (partial rebreathing of expired gases)
54Breathing Systems
- Closed rebreathing systems
- With closed systems the fresh gas flow rate is
does not exceed the patients metabolic O2
consumption of 5 to 10 mL/kg/min - The system may be used with a closed pop-off
valve and a fresh gas flow rate of 5 to 10
mL/kg/min - Expired gases are recirculated (after CO2
removal) with incoming fresh gases
55Breathing Systems
- Closed rebreathing systems
- Danger of increased CO2 accumulation if CO2
absorber not working efficiently - It is economical and there is minimal pollution
- It takes longer to change planes of anesthesia
- O2 depletion and N2O buildup are common, so do
not use N2O with this system - Requires constant monitoring to ensure pressures
do not build up in the system if the O2 flow
delivered exceeds the metabolic requirement
56Breathing Systems
- Closed rebreathing systems
- It can be dangerous to run a rebreathing system
with the pop-off valve closed, if the pop-off
valve does not have a safety release at high
pressures - It is recommended that the pop-off valve be left
partially open to prevent increases in pressure
in the system and to adjust the O2 flow rate
accordingly to prevent the rebreathing bag from
collapsing - If the bag does not collapse you can be confident
that sufficient O2 is being delivered to meet the
patients metabolic requirements
57Breathing Systems
- Semi-closed or partial rebreathing systems
- With semi-closed systems, the fresh gas is
delivered in excess of metabolic consumption at
25 to 50 mL/kg/min (suggested economical flow
rate) - The gas escapes through the pop-off valve to the
scavenger or after having the CO2 removed by the
soda lime and then recirculated with the fresh
gases - Higher flow rates can be used
- Less rebreathing will occur
- N2O buildup is less of a concern with higher flow
rates - Important to flush the system to prevent nitrogen
buildup from the expired gases
58Breathing Systems
- Non-rebreathing systems
- There is no mixing of inhaled and exhaled gases
and no rebreathing of expired gases all expired
gas goes to the scavenger - CO2 absorber not required
- Fresh gas flow rates required at 200 to 300
mL/kg/min - Fresh gas flow rates required at 130 to 200
mL/kg/min with the Bain system - May be some rebreathing of exhaled gases if a
reservoir bag and low flow rate
59Breathing Circuits
- Many kinds of breathing circuits available
- Circle system
- Universal F-circuit
- Bain system (Coaxial)
60Breathing Circuits
- Circle system
- CO2 absorber
- Inspiratory and expiratory unidirectional valves
(check valves or flutter valves) - Two breathing hoses connected with a Y-piece to
the patient - Rebreathing bag
- Pop-off valve (exhaust valve)
- scavenger
61Breathing Circuits
- Circle system
- Can be used as a non-rebreathing system (200
mL/kg/min) - Can be used as a partial rebreathing system (25
to 50 mL/kg/min) - Can be used as a total rebreathing system (5 to
10 mL/kg/min)
62Breathing Circuits
- Circle system
- An advantage is the mixture of expired gases with
incoming gases - Humidifies and warms the incoming gases
- Main disadvantages of the circle system occur
with smaller patients - Excess weight and bulk of the hoses
- Excess dead space
- Resistance to breathing caused by the
unidirectional valves
63Breathing Circuits
- Universal F-circuit
- Basically a modified circle system where the
inspiratory hose is placed within the expiratory
hose - Still requires a CO2 absorber, rebreathing bag,
unidirectional valves, pop-off valve and scavenger
64Breathing Circuits
- Universal F-circuit
- Incoming fresh gas is warmed also by expired
gases - The advantage is lighter weight and less bulk
- Disadvantage is that if the system is stretched,
the end of the inspiratory hose pulls away from
the end of the expiratory hose - Considered a safety feature so that the hoses
dont break - Increases the dead space within the circuit
- Even when not stretched, the dead space is
equivalent to the circle system
65Breathing Circuits
- Bain system
- Consists of one tube inside the other
- Fresh gases flow through the inner tube
- Unused and exhaled gases flow through the outer
tube - There also is a rebreathing bag with a clip on
the tubing between the reservoir bag and
scavenger connection but no CO2 absorber - Between breaths, the fresh gases flow through the
inner tube toward the patient and then back
through the outer tube toward the scavenger
66Breathing Circuits
- Bain system
- When the patient inspires, the gases are drawn
from the inner tube, which will be 100 fresh
gases or a mixture of fresh gases and expired
gases, depending on the fresh gas flow rate - This system can be used as a non-rebreathing
system with a fresh gas flow rate of 200 to 300
mL/kg/min - The high flow rate pushes exhaled gases away down
the outer tube so there is no rebreathing of
exhaled gases - By changing the flow rate to 130 to 200 mL/kg/min
the system acts as a partial rebreathing system - Most of the gases get pushed away but there is
partial rebreathing of some exhaled gases
67Breathing Circuits
- Bain system
- Ideal for small patients (lt7kg)
- Lightweight
- Minimal dead space
- Little resistance to breathing
- Good for all small animals in general but is not
economical when the patient weighs in excess of
10kg - Limiting factor is the size of the patient
- The O2 flowmeter must provide flow rates required
for a partial or non-rebreathing system (130 to
300 mL/kg/min) - Total volume of the Bain hose must be greater
than the tidal volume of respiration of the
patient to effectively prevent rebreathing
68Breathing Circuits
- Bain system
- Good for procedures involving the head (less
tubing in the way) - Good for procedures with much manipulation (i. e.
radiography, because there is less weight pulling
on the head) - Warming and humidification are minimal with
partial rebreathing - Requires a precision vaporizer
69Vaporizers
- Vapor pressure is characterized by the amount of
vapor related to its liquid in a closed container - The pressure exerted by the gas is called the
vapor pressure and will increase with increases
in temperature - Most anesthetics vaporize at a concentration
higher than necessary for clinical anesthesia - So a vaporizer is used to deliver diluted
anesthetics to patients
70Vaporizers
- Precision vaporizer
- Enables delivery of controlled concentrations of
anesthetic vapor independent of time, temperature
and fresh gas flow rate - Temperature and flow rate are compensated for by
the vaporizer or manually by the anesthetic
technician
71Vaporizers
- VOC (vaporizer out of circle)
- Vaporizer is added to the system between the O2
flowmeter and the circle - The circle consists of the inspiratory and
expiratory valves, breathing hoses,CO2 absorber,
pop-off valve, scavenger and rebreathing bag
72Vaporizers
- VIC ( vaporizer in circle)
- Vaporizer is placed inside the breathing system,
usually between the inspiratory valve and the
patient - VICs are always non-precision
- The carrier gas passes over the surface of the
anesthetic liquid or past a wick - Incoming gases mix with warm exhaled gases in the
system - Better vaporization of liquid is obtained when
low fresh gas flows are used - High flows cool liquid and reduce vaporization
- Are also safest when used with agents with low
vapor pressure (e. g. methoxyflurane)