Title: MidPoint Review Report
1- Mid-Point Review Report
- 900 MHz NMR Probe Cryogenic Gas Supply System
- Team 2
- Rebecca Altman
- Jason Kitchen
- Zachary Stevenson
- Jessica Vanterpool
- February 26th, 2009
2Presentation Outline
- Subsystem Testing
- Temperature Variance
- Mass Flow Controller
- Pneumatic Actuator vs. Solenoid Valve
- Future Testing
- Future Plans
Team with Dr. Brey in front of 900 MHz magnet
3Project Overview
4Assembly Progress
- Refill controller
- All wires have been sauntered and connected
- Both alarms have been added
- Two pin connector for the solenoid valve has been
added
Senior Design Group 2 900 MHz NMR Probe
Cryogenic Gas Supply System
5Heat Exchanger Coil Testing
- Increased the flow rate in increments of 5 scfh
- 10 Minute Intervals
- Ideal Flow Rate is 60 scfh
- Very high pressure readings before 40 scfh
(debris) - Droplets at 65 scfh, temperature at 78.1 K
- Results
- Experimentally T 84.7 K (With possible
debris) - Theoretically T 79.3 K (Did not take debris
into consideration) - Within required range, perform experiment again
for verification
Senior Design Group 2 900 MHz NMR Probe
Cryogenic Gas Supply System
6Heat Exchanger Coil Testing Cont.
- Flow Rate and Pressure vs. Temperature
Senior Design Group 2 900 MHz NMR Probe
Cryogenic Gas Supply System
7Heat Exchanger Coil Testing Cont.
- Equipment and Schematic Video
Senior Design Group 2 900 MHz NMR Probe
Cryogenic Gas Supply System
8N2 Gas Flow Regulation Improvement
- Current valve block in spectrometer has 15 valve
setting combinations for N2 gas flow from 0 to
2000 L/hr - Problem gas flow fluctuates with supply
pressure changes, which leads to temperature
fluctuation - Solution purchased a mass flow controller from
Alicat Scientific, cost 1396.00
- Features
- Flow range is 0 to 50 L/min
- Self adjusts for pressure changes using solenoid
valve - Typical 100 ms response time
- Accuracy 0.8 of reading
- Repeatability 0.2
- 30 pre-defined gases, including N2
9Testing Progress
- Nitrogen Dewar
- Temperature variance test
- Several refills were carried out at three
different cryostat pressures (6,10,22 psig) - First the FTS system was set at -75 C
- The N2 gas flow was set to 25 scfh during each
refill - Temperature was also allowed to reach a steady
state before each refill began - Digital thermometer used to record temperature
reading at transfer line entrance
Liquid valve
Gas Valve
Senior Design Group 2 900 MHz NMR Probe
Cryogenic Gas Supply System
10Testing Progress Cont.
- Results
- Less boil off at 6 psig, however there was a
significant decrease in the refill rate and twice
had to open the supply valve further to continue
the refill - Pressure will have to be maintained to ensure a
good refill rate - lower supply cryostat pressures will lower the
effect of a refill on the output temp to the
probe - Significant decrease in the amount of N2 vapor
venting out of the HX each time the supply
pressure was lowered
Senior Design Group 2 900 MHz NMR Probe
Cryogenic Gas Supply System
11Solution Pneumatically Actuated Cryogenic Valve
- Solenoid valve moved 15 m from magnet axis,
outside of magnetic field - Solenoid now controls a pilot gas connected to
pneumatic valve actuator - Pneumatic valve can be located inside magnetic
field - Pneumatic actuation opens LN2 fill line to refill
from supply dewar to HX cryostat
12Purchase Summary
- Waiting for quotation from Burkert Fluid Controls
for pneumatically actuated cryogenic valve - Remaining purchases will be plumbing fittings for
hookup of valves to supply Dewar and HX cryostat
13Design Problem Refill Solenoid Valve Proximity
to Magnet
- LN2 supply dewar must be within 3 m of the 900
MHz magnets axis - Refill solenoid located here could eventually
fail as it becomes permanently magnetized
14Future Testing
- March 3rd Refill Controller Test 1
- Check Wire Diagrams
- Refill Controller will be connected to the supply
system where each switch will be test multiple
times - March 5th Hx Coil Test 2
- Mass Flow Controller instead of Wall Pressure
Regulator - Spring Break Dr. Fus Experiment
- Dr. Fu will be using our setup minus the
controller. - Will be given data from him after Spring Break
15Future Plans
- Nitrogen Dewar
- Come up with a test that will be able to bypass
or insulate the HX pre-cooling coils to see what
effect they have on the refill temp stability - Temperature Fluctuation Issues
- Talk with Dr. Van Sciver
- Temperature Fluctuation from the Supply Dewar
- Temperature readings taken at the outlet, better
place? - Things that can be done to prevent further
temperature flunctuation
Senior Design Group 2 900 MHz NMR Probe
Cryogenic Gas Supply System
16Questions?