Title: The Measurement of pH,
1Chapter 18
- The Measurement of pH,
- Selected Ions, and Conductivity
2The Importance and Definition of pH
- Chemistry of life based on water
- Cells contain 80-90 water
- pH
- Measure of the acidity of an aqueous solution
- Intrinsically related to water chemistry
- Essential for biological systems
- Food processing
- Sewage treatment
- Water purification
- Pharmaceutical production
- Must be monitored and maintained
3The Importance and Definition of pH
- Water naturally dissociated to form hydrogen (H)
and hydroxide (OH-) ions - H2O H OH-
- At 25OC, equilibrium is reached
- Concentration of H OH-
- 1 x 10-7 mole/L
- Neutral
4The Importance and Definition of pH
- Acids
- Release hydrogen ions into solution when added to
water - Causes increase of H to gt1x10-7
- Bases
- Cause hydrogen ions to be removed from solution
when added to water - Causes decrease of H to lt1x10-7
5The Importance and Definition of pH
- Strong acids and strong bases
- Completely dissociate in water
- Have strong effect on H concentration
- HCl H Cl-
- NaOH Na OH-
- Both are completely in ionic form in water
- Weak acids and weak bases
- Do not completely dissociate in water
- Have smaller effect on concentration of H in
solution
6The Importance and Definition of pH
- pH
- Convenient way to express H concentration in
solution - the negative log of the H concentration when
concentration is expressed in moles per liter - Practical range between 0 and 14
- Some acids and bases can fall between this
range..VERY DANGEROUS! - gt 7 bases
- lt 7 acids
7The Importance and Definition of pH
- Ultimate importance of H concentration
- Profoundly affects properties of aqueous
solutions - Concentrated sulfuric acid
- Dissolve iron nails
- Citric acid (weak acid)
- Contributes to taste of foods
- pH scale
- Logarithmic
- pH of 5.0 10x more H than pH 6.0
- Methods to measure pH
- Indicator dyes
- pH meter/electrode
8pH Indicators
- pH indicators
- Dyes whose color is pH dependent
- Change colors at certain pH values
- Directly dissolved in a solution
- Impregnated into strips
- Dipped into the solution to be tested
- Phenolphthalein
- Indicator dye (used to be found in some
laxatives) - Colorless to red between pH 8 and 10
- Litmus
- Extracted from lichens
- Oldest know pH indicator
- Pink in acidic solutions
- Blue in basic ones
9pH Indicators
- pH indicators, cont..
- Do not change colors sharply at a single pH
- Change over a range of 1-2 pH units
- Broad indicators only
- Some have range of dyes to give more accurate
readings - Good for 0.3 pH units
- Uses
- Biological systems
- Low probability of contamination to mixtures
- Solution can be placed into a cell culture system
to monitor cell wastes vs. environmental pHred
ok.yellowish tinge becoming acidic - Radioactive solutions
- Dip paper and dispose of it
- Inexpensive/simple to use
10The Design of pH Meter/Electrode Measuring Systems
- Overview
- pH meter/electrode systems
- Most common method of measuring pH in the
laboratory - Accurate, sensitive and flexible vs. indicators
- Measure to the nearest 0.1 pH unit or better
- Used with a variety of samples
- Consists of
- Voltmeter
- Measures voltage
- 2 electrodes
- Sample
- Whats being measured
- pH meter
- H concentration is a measure of the magnitude of
the electrical potential (of the sample) - Measured by the voltmeter
11The Design of pH Meter/Electrode Measuring Systems
- The Basic Design and Function of Electrodes
- Electrodes and the Measurement of pH
- Electrodes
- Heart of the pH system
- Actually two electrodes
- pH measuring electrode
- -thin, fragile glass bulbcalled glass electrode
(fig. 18.2) - - glass sensitive to H concentration
- - sets up electrical potential between inner and
outer surfaces of the glass - H concentration causes increase in magnitude of
potentialpH - - electode contains buffered chloride solution
12The Design of pH Meter/Electrode Measuring Systems
- The Basic Design and Function of Electrodes
- Electrodes and the Measurement of pH
- Reference electrode
- Stable, constant voltage
- Needed b/c not possible to measure potential of a
single electrode - Strip of metal and electrolyte solution in a
glass/plastic tube - electrolytes substances-acids, bases, salts-that
release ions when dissolved in water - --- metal and electrolyte react together to
develop constant voltage
13The Design of pH Meter/Electrode Measuring Systems
- The Basic Design and Function of Electrodes
- Electrodes and the Measurement of pH
- Reference electrode-pH events
- Electrical potential (voltage) of glass electrode
varies depending on the samples H concentration - Electrical potential of the reference electrode
is constant - The two electrodes are connected to one another
through the voltage meter - Meter measures the voltage difference between the
reference electrode and the measuring electrode.
This voltage is an electrical signal - The electrical signal is amplified and converted
to a display of pH
14Operation of a pH Meter System
- Calibration
- Also called standardization
- Critical step in operation
- Tells the meter how to translate the voltage
difference between the measuring and reference
electrodes into units of pH - Calibrate every day
- Response of electrodes declines over time
- Response between ph and voltage is linear
- Two buffers of known pH used for calibration
- 1st usually pH 7.00
- 2nd typically pH 4.00, 12.00, or 12.00
- If solution acidic, acid buffer used
- If basic, basic buffers used
15Operation of a pH Meter System
- Calibration-Process
- Immerse electrodes in pH 7.00 buffer
- Set display to read 7.00
- Meter internally adjusts self so that pH of 7.00
corresponds to 0 millivolts (mV) - Called set, zero offset, or standardized
- Meter sets to 0 mV
- Immerse electrodes in second pH buffer
- Set display to read the pH of that buffer
- Meter uses the 2nd reading to establish
calibration line with a particular slope - Standard curve
16Operation of a pH Meter System
- Calibration-Process
- Slope
- Measure of the response of the electrodes to pH
- Steeper the slope, more sensitive the electrodes
are to H - Should be close to -59.2 mV/pH units _at_ 25OC
- In older probes
- Less able to generate a potential
- Slope of the line declines (fig 18.9, pg 350)
- Can be used as indication of the condition of the
electrodes - Can be expressed as a percent of the theoretical
value - 97 slope -57.2 mV/pH unit
- Buffers
- Must be good quality
- Commonly purchased as ready to use solutions
- Often in colors
17Operation of a pH Meter System
- Temperature
- Two effects
- Effects the potential that develops with the
measuring electrodes response to pH - pH of the solution being measured can
increase/decrease with temperature changes
18Operation of a pH Meter System
- Operation of a pH Meter System
- Most pH meters can compensate for temperature
dependent changes in the electrode response - Has temperature adjustment knob or built in
thermometer - Called ATC-automatic temperature compensating
probes - Temperature effects on pH of solutions
- Measure pH when solution at temperature at which
it will be used - Calibration
- Use room temperature standards
19Operation of a pH Meter System
- Box 1, pg. 352 for operating a pH meter
- Allow meter to warm up as directed by the
manufacturer - Open the filling hole fill solution must be
nearly to the top - Set to pH mode to read the pH of a sample or
standard
20Operation of a pH Meter System
- Box 1, pg. 352 for operating a pH meter
- Calibrate system each day or before use
- Adjust meter temp setting to room temperature or
use ATC probe - Minimum of two standard buffers
- Rinse electrodes with DI water and BLOT dry
- Wiping creates static charge
- Immerse electrodes in RT buffer
- Junction is immersed
- Level of sample below level of filling solution
- Allow reading to stabilize
21Operation of a pH Meter System
- Box 1, pg. 352 for operating a pH meter
- Adjust meter to read 7.00 with knob, dial,
button, or other methods of adjustment - Remove electrodes, rinse with distilled water,
blot dry - Can rinse with next solution and NOT blot
- Place electrodes in 2nd standard buffer, allow
the reading to stabilize - Set meter to temperature of sample or use ATC
probe - Place electrodes in the sample
- Allow pH reading to stabilize
- Record relevant information
- Remove electrodes from sample, rinse them, and
store properly
22Operation of a pH Meter System
- Measuring the pH of Difficult Samples
- Sample is the key to the measurement system
- Must be homogeneous to avoid variation and drift
- Equilibrated to a particular temperature
- Chemical reactions can cause pH to change over
time - Among substances in the sample
- With CO2 from the air
- Some samples inherently difficult to pH-require
special techniques or types of electrodes - Nonaqueous
- High purity water, containing little salt
- High salt concentrations
- High protein concentrations
- S-2, Br-, or I-
- Tris buffers
- Viscous
- Turbid
23Trouble-Shooting pH Measurements
- General Considerations
- pH meter range response
- 100 to 10-14 moles/L
- To very low concentrations of H
- To very small changes to H concentration
- Detect H in the presence of other ionic species
- Meters
- Must have a wide range of response
- Must be VERY sensitive
- Must be VERY specific
24Trouble-Shooting pH Measurements
- General Considerations
- 1st step..recognizing that there is a problem
- Most can be quite subtle
- Sometimes overlooked.
- Symptoms of pH problems
- Reading drifts in one direction
- Will not stabilize
- Takes an unusually long time to stabilize
- Reading fluctuates
- Meter cannot be adjusted to both calibration
buffers - Apparent pH value for a buffer or sample seems to
be wrong - No reading at all
25Trouble-Shooting pH Measurements
- The Components of a pH Measuring System and their
Potential Problems - Problems can arise from
- Reference electrode
- Measuring electrode
- Meter
- Calibration buffers
- The sample
26Trouble-Shooting pH Measurements
- The Components of a pH Measuring System and their
Potential Problems - Reference electrode junction
- Most common source of problems
- Dirty
- Partially occluded
- Long stabilization time
- Reading drifts slowly toward the correct pH
- Slow equilibration can also be caused by
- -changes in temperature of the sample
- -reactions within the sample
- -incompatibility between the reference electrode
and the sample - -checking for occlusion in table 18.9, pg. 356
27Trouble-Shooting pH Measurements
- The Components of a pH Measuring System and their
Potential Problems - Measuring electrode bulb
- Must be clean
- Needed for contact with solution
- Sample residues
- High levels of protein
- Age of electrode
- Attack to the glass by aqueous solutions
- Form gel layer on bulb surface
- Thickens over time..becomes sluggish and slope
of calibration line decreases
28Trouble-Shooting pH Measurements
- The Components of a pH Measuring System and their
Potential Problems - Meter
- Complete lack of response
- Caused by meter itself
- Least likely component of the system to cause
problems - Buffers
- Can be checked by using solution of known pH
- Sample not homogenous/temperature not stable
- Ph readings will fluctuate or drift
29Trouble-Shooting pH Measurements
- Tips
- Simple mistakes
- Always look for and correct simple/embarrassing
mistakes first - Electrode measuring bulb and the junction not
immersed in the sample - Meter not turned on or plugged in or electrode
cables not connected to the meter - Reference electrode not filled with electrolyte
- Reference electrode filling hole is closed
- Sample not well-stirred
- Calibration buffers not fresh
- Electrode is cracked or broken and needs to be
replaced
30Trouble-Shooting pH Measurements
- Tips
- Trouble-Shooting based on symptoms
- Tables 18.9-18.12, pg 356
31Trouble-Shooting pH Measurements
- Procedures for Cleaning and Maintaining
Electrodes - Follow manufacturers instructions
- Commonly recommended suggestions
- Boxes 3-6, pg 357-58
32Other types of Selective Electrodes
- Indicator electrode
- Responds selectively to a specific molecule in
solution - Dissolved CO2
- Dissolved O2
- During fermentation
- Growth of microorganisms sensitive to oxygen
levels