Title: Continuous Glucose Monitoring System CGMS
1Continuous Glucose Monitoring System (CGMS)
- Chien-Wen Chou MD
- Division of Endocrinology Metabolism
- Chi-Mei Medical Center
- 30 June 2006
2Continuous Glucose Monitoring System
- A continuous glucose monitoring system (CGMS) is
an FDA-approved device that records glucose
levels throughout the day and night. - only approved device -- Medtronic's MiniMed
device-can provide up to 288 glucose measurements
every 24 hours. - The system is used to measure an average blood
glucose for up to 3 days, while the person with
diabetes continues daily activities at home
3How Does the Device Work? (1)
- First, a tiny glucose-sensing device called a
"sensor" is inserted just under the skin of your
abdomen. - Tape is used to hold it in place.
- The sensor measures the level of glucose in the
tissue every 10 seconds and sends the information
via a wire to a pager-sized device called a
"monitor" that you attach to a belt or the
waistline of your pants. - The system automatically records an average
glucose value every 5 minutes for up to 72 hours.
4How Does the Device Work? (2)
- Results of at least four finger stick blood
glucose readings taken with a standard glucose
meter and taken at different times each day are
entered into the monitor for calibration. - Any insulin taken, exercise engaged in, and meals
or snacks consumed are both entered into a
paper-based "diary" and recorded into the monitor
(by pushing a button to mark the time of the
meals, medication, exercise, and other special
event you wish to record). - After 3 days, the sensor is removed at the
doctor's office and the information stored in the
CGMS is downloaded into a computer. - The information will be presented as graphs or
charts that can help reveal patterns of glucose
fluctuations.
5When Is the Device Used?
- The CGMS is not intended for day-to-day
monitoring or long-term self-care and it is not a
replacement for standard blood glucose
monitoring. - The main advantage of continuous glucose
monitoring is that it can help identify
fluctuations and trends that would otherwise go
unnoticed with standard HbA1c tests and
intermittent finger stick measurements. - For example, the device can capture dangerously
low overnight blood glucose levels which often go
undetected, reveal high blood sugar levels
between meals, show early morning spikes in blood
sugar, evaluate how diet and exercise affect
blood sugars, or provide up to a 72-hour complete
review of the effects of changes - Continuous monitoring is reimbursed by Medicare
and covered by many private insurance plans
6Continuous Glucose Monitoring Roadmap for 21st
century diabetes therapy
- David C. Klonoff, MD, FACP
- Diabetes Care 281231-1239, 2005
7Purposes
- provides information about the direction,
magnitude, duration, frequency, and causes of
fluctuations in blood glucose levels. - provides much greater insight into glucose levels
throughout the day. - help identify and prevent unwanted periods of
hypo- and hyperglycemia.
8Technologies (1)
- Five CGMs have been approved by the U.S. Food and
Drug Administration (FDA) for use in the U.S. or
carry CE marking for use in Europe. - Continuous Glucose Monitoring System Gold (CGMS
Gold Medtronic MiniMed, Northridge, CA) - GlucoWatch G2 Biographer (GW2B Cygnus, Redwood
City, CA) - Guardian Telemetered Glucose Monitoring System
(Medtronic MiniMed) - GlucoDay (A. Menarini Diagnostics, Florence,
Italy) - Pendra (Pendragon Medical, Zurich, Switzerland)
- FreeStyle Navigator Continuous Glucose Monitor
(Abbott Laboratories, Alameda, CA) -- premarket
approval application has been submitted to the FDA
9Technologies (2)
- minimal invasiveness through continuous
measurement of interstitial fluid (ISF) or with
the - noninvasive method of applying electromagnetic
radiation through the skin to blood vessels in
the body. - bringing a sensor into contact with ISF include
inserting an indwelling sensor subcutaneously
(into the abdominal wall or arm) to measure ISF
in situ or harvesting this fluid by various
mechanisms that compromise the skin barrier and
delivering the fluid to an external sensor - After a warm-up period of up to 2 h and a
device-specific calibration process, each
devices sensor will provide a blood glucose
reading every 110 min for up to 72 h with the
minimally invasive technology and up to 3 months
with the noninvasive technology. - Results are available to the patient in real time
or retrospectively. - Every manufacturer of a CGM produces at least one
model that sounds an alarm if the glucose level
falls outside of a preset euglycemic range.
10(No Transcript)
11(No Transcript)
12Target Populations
- The ideal time to calibrate is either after
fasting or at least 3 h postprandially, but not
right after exercise or when the blood glucose
level is likely to be rising or falling. - Without such calibration, continuous readings may
be inaccurate. - Currently available CGMs that provide real-time
readings should not be used to make therapeutic
decisions, such as whether to dose insulin or
eat, because they are not sufficiently accurate. - Instead, an abnormal reading should prompt a
finger-stick blood glucose measurement whose
value can be acted upon. - Patients require a thorough training program to
calibrate and operate a CGM.
13 Accuracy (1)
- A real-time CGM can be programmed to sound an
alarm for readings below or above a target range - The most important use of an alarm is to detect
unsuspected hypoglycemia (such as during sleep)
so that glucose can be administered to prevent
brain damage. - There is a trade-off between an alarms
sensitivity and specificity. - In general, if the alarm is set to sound at a
lower level than the hypoglycemic threshold, then
the specificity will be good but the sensitivity
may be poor. - If the alarm is set to sound at a glucose level
higher than the hypoglycemic threshold, then the
sensitivity will be good but the specificity may
be poor. - The greater accuracy a continuous monitor can
provide, the less of a trade-off is necessary
14Accuracy (2)
- The Diabetes Research in Children Network
(DirecNet) is a U.S. network of five clinical
centers and a coordinating center dedicated to
researching glucose monitoring technology in
children with type 1 diabetes - The networks investigators, the DirecNet Study
Group, assessed the accuracy of the first- and
second-generation CGMS and the GW2B in children
with type 1 diabetes in concurrently published
studies - The second-generation CGMS Gold, compared with
the first-generation CGMS, had a lower median
relative absolute difference (RAD) between CGMS
glucose and reference serum glucose paired values
(11 and 19, respectively) - For the GW2B, the median RAD between GW2B glucose
and reference serum glucose paired values was 16
- Similar RAD values of 21 have been reported for
the first-generation CGMS by Kubiak et al. - RAD values of 12.8 and 12.815.7 have been
reported for the second-generation CGMS Gold
system by Goldberg et al. and Guerci et al.
respectively.
15Accuracy (3)
- The DirecNet Study Group found the CGMS Gold
system, which is the second generation of CGMS
technology, as well as the GW2B, which is the
second generation of GlucoWatch technology, to
have inversely proportional sensitivity and
specificity rates during hypoglycemia in children
and adolescents with type 1 diabetes. - A series of alarm settings were selected for a
reference blood glucose of 60 mg/dl. - For CGMS Gold, the settings with sensitivity and
specificity were 60 mg/dl, 49 and 42 80 mg/dl,
84 and 36 100 mg/dl, 100 and 25 and 120
mg/dl, 100 and 16. - With the GW2B, the settings were 60 mg/dl, 23 and
49 80 mg/dl, 59 and 33 100 mg/dl, 84 and 20
and 120 mg/dl, 92 and 15. - The authors concluded, "These data show that the
GW2B and the CGMS do not reliably detect
hypoglycemia. - Both of these devices perform better at higher
glucose levels, suggesting that they may be more
useful in reducing HbA1c levels than in detecting
hypoglycemia"
16Accuracy (4)
- The International Organization for
Standardization (ISO) standards for accuracy of
point blood glucose tests require that a sensor
blood glucose value be within 15 mg/dl of
reference for a reference value 75 mg/dl and
within 20 of reference for a reference value gt75
mg/dl. - Sensor accuracy by this definition is expressed
as the percentage of data pairs meeting these
requirements. - The DirecNet group found that for hypoglycemic
blood glucose levels (determined by a reference
blood glucose monitor, the OneTouch Ultra), the
CGMS Gold met the ISO standards in only 48 of
readings and the GW2B met these standards in only
32 of readings - The percentage of data points attaining ISO
accuracy standards climbed as the blood glucose
level rose, topping out for the highest segment
of reference blood glucose levels (i.e., blood
glucose values 240 mg/dl). - In this glycemic category, the CGMS Gold and
GW2B, respectively, met ISO accuracy for 81 and
67 of data points. - In a separate series of 15 healthy nondiabetic
children undergoing continuous glucose monitoring
over 24 h, the DirecNet Group reported that the
median absolute difference in concentrations for
the GW2B was 13 mg/dl and for the CGMS was 17
mg/dl. - Furthermore, 30 of the values from the GW2B and
42 of the values from the CGMS deviated by gt20
mg/dl from the reference value
17(No Transcript)
18(No Transcript)
19(No Transcript)
20Clinical Indications (1)
- when adjusting therapy
- quantifying the response in a trial of a diabetes
therapy - assessing the impact of lifestyle modifications
on glycemic control - monitoring conditions where tighter control
without hypoglycemia is sought (e.g., gestational
diabetes, pediatric diabetes, in the intensive
care unit) - diagnosing and then preventing hypoglycemia
(e.g., during sleep, with hypoglycemia
unawareness) - diagnosing and preventing postprandial
hypoglycemia. - facilitate adjustments in therapy to improve
control (most important use)
21Clinical Indications (2)
- Specific therapeutic adjustments include changing
from regular to a synthetic ultrashort-acting
insulin analog at mealtime, changing from NPH to
a synthetic ultralong-acting insulin once or
twice per day, increasing or decreasing the
mealtime insulin bolus dosage, increasing or
decreasing the basal insulin rate, altering the
treatment of intermittent hypoglycemia or
hyperglycemia, changing the insulin-to-glucose
correction algorithm for premeal hyperglycemia,
changing the insulin-to-carbohydrate ratio at
mealtime, changing the method for counting
carbohydrates, changing the carbohydrate
composition of the diet, changing the discount in
short-acting insulin dosage for exercise,
changing the nighttime regimen because of the
dawn phenomenon, changing the target preprandial
or postprandial blood glucose values, or before
referring a patient for psychological counseling
to improve adherence to the treatment regimen. - The most frequent therapy adjustment by Sabbah et
al.(out of eight adjustments) was to increase the
mealtime bolus dosage. - The most frequent therapy adjustment by Kaufman
et al.(out of nine adjustments) was to modify the
type of basal long-acting insulin.
22Advances in Glucose Monitoring
- As recently as July 2003, the FDA approved the
first wireless combination system, consisting of
a glucose monitor and an "intelligent" insulin
pump (co-developed by Medtronic MiniMed and
Becton, Dickinson and Company). - The next phase of advances will allow insulin
pumps to not only simply recommend proper insulin
dosages, but actually automatically deliver them. - More recently, in August 2005, Medtronic has
expanded its CGMS line and announced FDA approval
of its newest device, called the Guardian RT. - This system works just like the MiniMed device
but instead, displays the "real-time" glucose
levels every five minutes. This information
alerts the patient immediately to glucose levels
that are too high or low, allowing for
adjustments in therapy
23Light Waves Instead of Finger Pricks
- Several companies are developing non-invasive
glucose monitoring devices that rely on light
waves. - The monitors shine infrared (or near-infrared)
light onto the skin of the patient's forearm and
analyse the light that is reflected back to
determine the concentration of glucose in the
tissue. - The company Sensys has made progress in
developing a commercial, portable version of an
infrared monitor. - Their latest model, the Sensys Medical Glucose
Tracking System is about the size of a computer
tower and is over 90 accurate. - CME Telemetrix has developed a product called
GlucoNIR, a non-invasive infrared system which
may also be able to non-invasively measure HbA1c.
- Animas is developing an implantable version of an
infrared optical sensor, intended to be implanted
in the body for up to 5 years, but the device is
not expected to be available until 2005.
24A Glucose-Monitoring Skin Patch
- SpectRx and Abbott Laboratories are developing a
continuous glucose monitor which will be worn as
a skin patch. - This monitor would measure glucose levels in
interstitial fluid, collected through microscopic
holes created by a laser in the dead outer layer
of skin, and measured through glucose oxidase
reaction/electrical current generation in a patch
containing a glucose sensor. - This device is currently being assessed in human
trials in adults and children.
25Glucose-Sensing Contact Lens
- Researchers at the University of Texas and Ciba
Vision are developing glucose sensing contact
lenses which are designed to be used in
conjunction with a palm-sized light source. - The contact lens is made using a meshwork that
traps fluorescent molecules inside the lens. - The patient inserts the contact lenses in the
usual way, holds the light device up to the eye
and activates it, sending a small burst of
glowing light into the contact lens. - The fluorescent molecules in the lens bind to and
react with the glucose in the users tears. - The device reads the wavelength of the
fluorescence reflected from the contact lens and
translates the reading into a measure of the
glucose. - Higher levels of fluorescence mean higher levels
of glucose. - There is a seven-minute delay before a hand-held
device stores the glucose data. - This device also includes an alarm that signals a
patient if the glucose levels rise too quickly. - Clinical trials are underway.
26A Smart Tattoo
- A collaboration between Texas AM University and
Penn State University is developing a "smart
tattoo" that could provide accurate blood glucose
readings. - Polyethylene glycol beads coated with fluorescent
molecules are injected beneath the skin surface
and interact with the interstitial fluid. - In low glucose, the tattoo is highly fluorescent.
- In high glucose, the fluorescence beads are
displaced by glucose binding and the overall
fluorescence of the tattoo decreases. - Fluorescence can be read by a detection light.
- Preliminary data in tattooed rats have yielded
promising results.