Title: Michigan Xray Rules for Suppliers and Installers


1
Michigan X-ray Rules for Suppliers and Installers
  • Michigan Radiation Safety Section
  • Tod Van Wieren, Health Physicist
  • Talk about some current rules affecting suppliers
    and installers
  • Identify possible rule updates of interest to
    suppliers and installers

2
Topics of Discussion
  • Public Health Code
  • Ionizing Radiation Rules
  • Suggested State Regulations

3
Public Health Code
  • Act 368 of 1978 Part 135
  • The law giving RSS authority
  • Establishes registration, fees, and provisions
    for administrative rules
  • Draft updates planned
  • X-ray operator training
  • Allow future non-ionizing device rules (laser,
    tanning, etc.)

4
Ionizing Radiation Rules
  • Part 4 Registration
  • Register, pay fee, apply tags, post certificate
  • Vendor Obligation Rule 195. (3) A person shall
    not make, sell, lease, transfer, lend, assemble,
    or install radiation machines.unless the
    equipment.meets the requirements of these
    rules.

5
Ionizing Radiation Rules
  • Part 5 Standards for Protection
  • Dose Limits 1250 mR/quarter
  • Postings Certificate, Notice to Employees,
    Rules or reference
  • Monitoring All occupationally exposed
    personnel, Dual badges when lead aprons are
    routinely worn.

6
Ionizing Radiation Rules
  • X-ray room Shielding
  • The degree of protection required for an
    enclosureshall be subject to design approval by
    the department.
  • Mobile equipment used routinely in one location
    is considered fixed and rooms must be shielded.
  • More this afternoon

7
Ionizing Radiation Rules
  • Manufacturers guidance or other state
    regulations do not always comply with Michigan
    rules.
  • If there are ANY questions about rule
    requirements please contact us.

8
Suggested State Regs
  • X-ray room design
  • Plan reviews required for all new installations
    or modifications prior to construction
  • Facility must have a scale drawing of the room
    and shielding installed available for inspection
  • Operators booth design criteria

9
Suggested State Regs
  • X-ray services registration
  • X-ray installation and machine services shall be
    registered
  • Machine performance
  • X-ray machine output accuracy, reproducibility,
    and linearity limits

10
Suggested State Regs
  • Film processing
  • Darkroom fog testing
  • Method to prevent accidental entry
  • Film storage
  • Regular screen inspection and cleaning
  • No outdated film unless it passes sensitometry
    testing

11
Suggested State Regs
  • Quality assurance testing
  • States starting to add QA for non-mammo
    facilities
  • Testing may be similar to mammography, but no
    details yet
  • Suggested State Regs do not address this yet

12
Questions?