Title: HPLC and UHPLC Part 2


1
HPLC and UHPLC Separations at Mid to High pH
Part 2
2
Introduction
  • LabNote continues the discussion of the use of
    higher pH for method development and analysis
  • There are many advantages to working at higher
    pH, but also some items to be aware of

3
Advantages
  • Orthogonal selectivity (different elution order)
    can be obtained for many analytes
  • Basic analytes are retained longer and have
    better loading capacity, linearity, and peak shape

4
Advantages
  • Signal and signal-to-noise ratios are usually
    improved for basic analytes
  • Chromatographic separations at higher pH allow
    compounds that are unstable at lower pH to be
    analyzed without on-column degradation

5
What to Watch Out For
  • Analytes that are unstable at higher pH could
    degrade on column
  • With LC-UV detection the UV spectra and lmax 
    values of ionizable analytes often change
    dramatically
  • Can make peak tracking more challenging

6
What to Watch Out For
  • Column lifetimes for all silica-based columns
    will be shorter using buffers
  • On-column reaction or decomposition can occur for
    metal-sensitive analytes

7
Method Development
  • Parameters that affect selectivity (relative
    retention) include
  • Column bonded phase
  • Organic modifier choice
  • Mobile phase pH
  • organic modifier or gradient steepness
  • Column temperature
  • Buffer identity and concentration
  • Newest addition to the ACE column family the
    ACE SuperC18 phase

8
ACE SuperC18 Features and Benefits
  • Unique bonding chemistry provides increased
    resistance to degradation
  • Can be used at pHs ranging from 1.511 with all
    analytes types

9
ACE SuperC18 Features and Benefits
  • Designed for use with LC-MS compatible mobile
    phase buffers
  • Does not show hysteresis effects (memory effects)
    when changing between low and high pH mobile
    phases

10
ACE SuperC18 Features and Benefits
  • The ACE SuperC18 column produces highly
    symmetrical peaks for bases, acids, and neutral
    analytes
  • Baselines were observed to be better when NH4OH
    or NH4OAc were premixed with the organic modifier
    in the B solvent

http//www.mac-mod.com