RAE/Honeywell in Partnership with BNC Scientific
  • GammaRAE II R Training

2
QPlacing the device into operation
  • A See manual pg. 9-12

3
QPlacing the device into operation
4
  • QSetting user adjustable features (what the
    setting features represent, i.e.,threshold
    setting and their effect on false alarm and
    sensitivity of device

5
ASee manual pg. 24 and 25
6
  • Q displays and alerts (monitoring system status
    and recognizing and understanding system
    notifications)
  • ASee manual pg. 8

7
  • QConducting basic operator maintenance (charging
    of batteries, etc)
  • A See manual pg. 6

8
  • QOverview of how alarm algorithm works and how
    the device sets and/or uses background including
    answers to following questions Is alarm
    threshold based on the detectors gross count rate
    or a derived quantity such as
  • AThere are two modes Search Mode and Safety
    Mode. Search Mode alarm threshold (Background
    radiation level) (Search Alarm) x (standard
    deviation of background radiation level). Safety
    Mode sets Low or High absolute alarm threshold
    based on dose and dose rate.

9
  • Q Is alarm threshold based on the detectors
    gross count rate or a derived quantity such as
    exposure rate or an energy-dependent count rate?
  • AThe alarm threshold is based on detector gross
    count rate and accumulated count.

10
  • QWhat time period (i.e. integration time) that
    is used to assess whether or not radiation has
    exceeded threshold?
  • A125ms

11
  • Q How frequently is the threshold condition
    assessed?
  • A1Hz

12
  • QIs the alarm fixed or does it vary with
    radiation background?
  • ASafety Mode alarms (Low and High) are fixed
    based on dose and dose rate including background.
    Search Mode alarm varies with the background
    level.

13
  • QIf the threshold is fixed what are the
    threshold values?
  • ASafety Mode alarms (Low and High) are
    programmable which can be set by users. Safety
    High default setting 200 µR/h range 0600R/h.
    Safety Low default setting 50 µR/h range
    0600 R/h. Dose High default setting 25 R
    range 0999.9R. Dose Low default setting 5R
    range 0999.9R.

14
  • QIf the threshold is variable, how high above
    background is the threshold set? If the
    difference is based on a formula should the
    formula be provided?
  • AThere are two modes Search Mode and Safety
    Mode. Search Mode alarm threshold (Background
    radiation level) (Search Alarm) x (standard
    deviation of background radiation level). Search
    Alarm number is how many sigmas (standard
    deviation of stored background reading which was
    recorded after initial 36-second background
    reading countdown). Safety Mode sets Low or High
    absolute alarm threshold based on dose and dose
    rate. Search Alarm default setting 5.8 range
    19.9. Safety LOW and HIGH settings 0600mR/h.

15
  • QIf the threshold varies with the background,
    how frequently and for what duration is the
    background measured?
  • AThe background was measured only when the
    instrument was initially turned on. However, at
    any time the background can be updated by users
    in Search Mode.

16
  • QIf the alarm decision is based on multiple
    variations, intermediate and multiple decisions
    (i.e., energy windowing, region of interest
    analysis, etc) how are the decisions processed
    and alarm decisions made?
  • AAlarm decision is only based on dose or dose
    rate.