Good Chromatographic Practices_GMP_Dr.A.Amsavel - PowerPoint PPT Presentation

About This Presentation
Title:

Good Chromatographic Practices_GMP_Dr.A.Amsavel

Description:

Introduction Solvents & Reagent Quality Mobile phase preparation & usage Buffer, Filtration, Degassing Storage & usage Preparation for analysis Sample preparation & System suitability Resolution, peak symmetry, tailing, capacity factor, Adjustment allowed as per USP Column management Washing and change of mobile phase Analysis and integration Audit trail, data backup, Review and Documentation Problems and reporting Qualification and Calibration – PowerPoint PPT presentation

Number of Views:584

less

Transcript and Presenter's Notes

Title: Good Chromatographic Practices_GMP_Dr.A.Amsavel


1
Good Chromatographic Practices
  • Dr. A. Amsavel

2
Content
  • Introduction
  • Solvents Reagent Quality
  • Mobile phase preparation usage
  • Buffer, Filtration, Degassing
  • Storage usage
  • Preparation for analysis
  • Sample preparation System suitability
  • Adjustment allowed as per USP
  • Column management
  • Washing and change of mobile phase
  • Analysis and integration
  • Documentation
  • Problems and reporting
  • Qualification and Calibration

3
Introduction
  • Chromatography analysis (HPLC, GC TLC etc ) of
    material in pharmaceuticals QC Lab is inevitable
  • Chromatographic technique is used for
    Identification, determination of assay, purity,
    related substances and residual solvents etc.
  • Purpose of Chromatography analysis is for
    final product , intermediate, reaction
    monitoring, process validation, cleaning
    validation, cleaning verification, stability
    testing etc
  • Chromatographic analysis has to be performed and
    documented as per GMP and to meet the regulatory
    requirements
  • There are challenges, while performing analysis
    and compliance

4
Qualification And Validation
  • HPLC GC instruments shall be qualified (URS,
    DQ, IQ, OQ and PQ)
  • Software used shall meet the 21CFR Part-11
    compliance
  • Hardware Software shall be qualified as per
    Computer system validation (CSV)requirement (
    GAMP-5)
  • Instruments shall be calibrated appropriately at
    defined interval
  • Test method shall be validated, compendia method
    shall be verified, and or method transfer as
    applicable
  • Analyst shall be qualified and trained
  • Ref Qualification Of Equipment Annex 1
    Qualification Of Liquid Chromatography Equipment

5
  • What are the challenges faced while
    Chromatographic analysis?
  • What are the possible errors
  • occur during analysis?

6
(No Transcript)
7
  • Problems or challenges !!!
  • How can you prevent the problems?
  • How can it be controlled?
  • How errors can be eliminated or minimized?

8
  • Follow Good Chromatographic practices
  • SOP for Good Chromatographic practices
  • Use Qualified the Instruments
  • Validated / verified and appropriate test method
  • Perform periodic Calibration Preventive
    maintenance
  • Right Quality solvents, proper sample
    preparation, system suitability
  • Train and Qualify the analyst
  • Let us discuss on HPLC analysis

9
(No Transcript)
10
HPLC Analysis Flow
11
Water for HPLC analysis
  • Always use only Ultra pure / Milli-Q water for
    HPLC analysis
  • Ultra pure HPLC water of 18M? resistivity
  •  Do not use RO water/de-ionised water for HPLC
    analysis. It will have organic and in-organics
    impurities
  • If water contains impurities, it will have higher
    absorption and lead to poor baseline , drift ,
    ghost peak and less accurate
  • Note
  • Use fresh water if possible or store properly
  • Store HPLC grade water in glass containers.
  • Plastic containers may leach the additives in
    the plastic into the water

12
Solvents Reagents
  • All reagents and solvents should be highest
    quality.
  • HPLC grade reagents solvents are high purity
    will have low UV absorbance
  • Low grade solvent contain impurities to produce
    spurious peaks, poor peak , high baseline etc
  • Reagent / acid for buffer preparation should be
    highest purity.
  • Use specific quality of reagent as stated in
    Pharmacopiea / test method as required
  • Eg. Different grade of Ammonium acetate based on
    the purity
  • TEA TFA are expected to be contaminated from
    the lab environment and should be refrigerated
    during storage.

13
Impact of Quality of Solvent
14
Buffers
  • Preferably all buffers should be prepared freshly
    on the day of use .
  • Buffer pH will affected by prolonged storage and
    prone to microbial growth both will affect badly
    in chromatography.
  • If buffer solutions are used for long or
    stored establish the shelf life and use
  • Where possible use reagents that without
    stabilizer. Eg Sodium metabisulphite is used as
    stabilizing agent
  • These stabilizing agents often may affect the
    optical and chromatographic behavior.
  • Avoid the repeated / multiple opening use of
    solid reagent since may easily contaminate
  • Based on requirement, purchase in low pack size
    to reduce the exposure.

15
Use of buffer
  • Do not leave the buffer solutions for long in
    system, to avoidcrystallization. This will
    affect the system.
  • Effect on pump - damage plunger and seal
  • Effect on column - creation of column voids
  • Effect on flow line - corrosion of stainless
    steel lines
  • Long storage may develop a bacterial growth eg
    phosphate buffers it is good medium for
    bacterial and fungus growth. Hence, prepared
    freshly and avoid storage.
  • All the solutions should be clear, homogenous
    free from particulate matter.
  • Buffer solutions must be filtered through 0.45µm
    filter
  • Note small amount of sodium azide to the aqueous
    solvent to inhibit microbial growth.

16
Filtration
  • All HPLC solvents or mobile phase should be
    filtered through a 0.45 µm filter before use.
  • Removes particulate matter that may cause
    blockages of column and HPLC system.
  • Do not keep mobile phase under the suction for
    more than 5 minutes. composition of mobile phase
    change if use highly volatile solvents
  • Filtration prevents the wear tear of parts.
  • Pump plungers, seals and check valves will
    perform better and lifetimes will be maximized.
  • Also column performance and life will be
    maximised
  • Filtered solvents should be stored in a covered
    reservoir to prevent evaporation dust
    contamination etc.

17
Changing Over of Mobile Phase- RP
  • Changing Over of Mobile Phase one analysis to
    another analysis is important for smooth
    chromatographic analysis.
  • Before changing to totally organic phase in the
    RP system, the whole system should be flushed
    with water organic (11) mixture to remove the
    buffers used.
  • Prevent the crystallisation when the organics
    come into contact with leftover buffers in
    system.
  • Eg 100 Acetonitrile use will lead to precipitate
    the salt in the buffer solution

18
Changing Over of Mobile Phase .
  • To convert a normal phase system/column to a
    reversed phase, flush with a solvent that is
    miscible with both the current normal phase
    solvents
  • If the final reversed phase solvents include a
    buffer, then it is advisable to move from the
    100 methanol flush to a 50 aqueous methanol
    flush. For example
  • Normal phase Hexane/Ethyl acetate
  • Flush IPA (isopropenyl acetate) then
    Methanol
    Finally (5050) Methanol/Water
  • Reversed phase Buffered aqueous methanol

19
Changing Over of Mobile Phase.
  • To convert a reversed phase system/column to a
    normal phase follow a similar way but in
    reverse,
  • For example,Reversed Phase Buffered Aqueous
    MethanolFlush 5050 Methanol/Water
    finally Methanol then IPANormal Phase
    Hexane/Ethyl Acetate

20
Mobile phase preparation- Isocratic
  • Different solvents has to be premixed for
    isocratic systems
  • Preparation tips Measure the solvents
    separately for accuracy and mix. E.g. To prepare
    1000ml of 5050 mixture of water/methanol, Measur
    e 500ml of milliQ water, transfer into a bottle
    transfer 500ml of Methanol and mix sonicate
    for about 10-15 minutes.
  • Protection of mobile phase storage
  • Use glass bottle
  • Use Plastic bottle for high sensitivity inorganic
    analysis.
  • Always cover the container to prevent evaporation
    of solvents dust freeNote Do not cover the
    mobile phase bottle with parafilm. Theleachable
    from the parafilm may contaminate the analysis.

21
Mobile phase preservation usage
  • Do not mix old mobile phase with new mobile phase
  • Do not expose the bottle to direct sunlight or
    wind.
  • Use always suction filter
  • Filter and sonicate the mobile phase.
  • Keep reservoir above solvent delivery system
  • Keep suction filter 2 cm above the bottom to
    prevent particulates from reaching pumps
  • Periodically clean the suction filter by
    sonicating in isopropanol or 1N nitric acid and
    water
  • Use mobile phase after reach the room temperature

22
Degassing
  • Mobile phase should be thoroughly degassed to
    remove all dissolved gasses.
  • Dissolved gas can be removed from solution by
    Bubbling with helium Sonication / Vacuum
    filtrationIf the mobile phase is not degassed,
    air bubbles can form in the high-pressure system
    resulting in problems with system instability,
    spurious baseline peaks etc.
  • Do not use Online degasser while THF as mobile
    phase due to degradation of Teflon in vacuum
    chamber.

23
Advantage of Degassing
24
Suction filter
  • All solvent line shall be fitted with an inlet
    filter to prevent the particles contamination
  • The filters should be kept clean to prevent cross
    contamination.
  • When system is not being used, store the
    filters in a solution of 50 Acetonitrile /50
    water and in closed condition.
  • This is to prevent the microbial growth, dust and
    dirt to clog the filter pores.
  • The solvent lines should be clean, clog free, no
    air bubble and should not have sharp bends or
    creases in them.
  • The solvent lines and filters should be of
    sufficient length to reach the bottom of the
    solvent reservoir
  • Solvent reservoirs should be placed higher than
    the pump inlet manifold.

25
Equilibration of Column
  • Column volume
  • This column volume is more correctly called as
    void volume (Vm).
  • The amount of mobile phase required to fill the
    column.
  • Otherway the volume of mobile phase which should
    be flushed through a column before it is ready to
    use
  • When a column has just installed on a reverse
    phase HPLC system then it will typically require
    10 to 20 column volumes.(exceptions use of
    pairing agents chiral methods)
  • This is typically approximately 70of the total
    column tube volume.

26
Equilibration of Column
27
HPLC Column Management
  • Do not store HPLC columns in buffers. A buffer
    may precipitate inside the column, it will clog
    and affects the packing material.
  • Mobile phases with 100 or close to 100 buffer
    may lead to bacterial growth, which can block the
    column frit and packing material.
  • Bacteria may also affect the analyte, and organic
    products from the dead bacteria may cause "ghost
    peaks" in chromatograms.
  • Do not store HPLC columns in solvents that
    degrade easily tetrahydrofuran (THF),
    triethylamine (TEA), trifluoroacetic acid (TFA).
  • Unstabilized THF can form peroxides which may
    degrade the column
  • All buffers should be washed out of the column
    before flushing with Acetonitrile.
  • Gradually start the column washing Eg Starts
    the flow with 0.2 ml/min and increases gradually
    to 2.0 ml/min and continues for 20min to 30 min
    or as per procedure
  • Have Dedicated Columns for each Method / each
    product

28
Washing of HPLC System Columns
  • Flush the system thoroughly with strong
    solventBecause switching to pure organic
    solvent, especially acetonitrile, can cause
    buffers or salts to precipitate.
  • Recommend switching to non-buffered mobile phase
    first. Just replace the buffer bottle with water
    and run the gradient again.
  • Run the gradient to 100 strong solvent (usually
    methanol or acetonitrile) and hold it at 100 for
    1015 min to thoroughlyflush the column and
    equipment
  • Do not shut off a system that contains buffers or
    salts

29
Washing Storage of HPLC Columns
  • Contamination may affects the chromatography and
    column performance.
  • Short term storage, i.e. over night, columns can
    be stored in the eluent used in last analysis.
  • Middle term storage, i.e. 2 days or over the
    weekend, columns should be flushed with pure
    water to prevent microbial growth.
  • Long term storage, silica based columns should be
    stored in an aprotic solvent.
  • The best storing solvent is Acetonitrile water
    content should not be higher than 50.
  • Buffer salts will precipitate in Acetonitrile and
    can block the capillaries and the column.

30
Column cleaning Regeneration
  • Column cleaning has to be used 40 60 column
    volumes appropriate solvent or as per procedure
    .
  • Check the column efficiency, capacity factor
    etc. Before and after clean-up procedure to
    confirm the improvement of column performance
  • Solvent can be used as follows for regenration
  • Normal phase media
  • flush with tetrahydrofuran
  • flush with methanol
  • flush with tetrahydrofuran
  • flush with methylene chloride
  • flush with benzene-free n-hexane
  • Reversed phase media
  • flush with HPLC water inject 4 aliquots of 200
    µL DMSO during this flush
  • flush with methanol
  • flush with chloroform
  • flush with methanol

31
System preparedness
  • Needle and plunger seals shall be washed before
    starting the analysis.
  • Use wash solutions, which are miscible with the
    mobile phase to wash and plunger seal
  • Use separate solutions and containers for plunger
    seal wash and needle wash.
  • Use of one solution for both functions may
    compromise the effectiveness because the
    functions of these solutions differ.

32
Seal wash - Composition
  • Over a period of time small amounts of mobile
    phase solvents seep through the seal to the back
    of the pump head.
  • If these solvents contain buffers then the salts
    may precipitate out forming deposits which can
    shorten the life of the seal.
  • Seal wash can start with aqueous solvent to
    dissolve buffers.
  • A small amount of organic solvent is added to
    prevent bacteria growth and also to reduce the
    surface tension of the water (this helps the wash
    solvent cling to surfaces).
  • Typical seal wash composition is 80 water and
    20 organic solvent.
  • The organic solvent may be methanol, acetonitrile
    or isopropyl alcohol (IPA)

33
Needle wash
  • The needle-wash required by flushing to prevent
    carryover of sample between injections.
  • To extends the life of the injector seals by
    removing buffer sample
  • Use Extended needle wash if the carry over is
    expected, especially analysis of Related
    substances.
  • The composition of the needle wash needs to be
    matched to the sample.
  • Proportions of aqueous and organic solvents in
    the mobile phase will be appropriate.

34
Injector Maintenance
  • Purge the flow line at least once a day.
  • Purge all the flow lines with mobile phase to
    ensure the solvent tubing is completely filled
    with mobile phase
  • Change purge liquid frequently.
  • Never use salt solutions as purge liquid.
  • Use good quality septum's for sample vial and
    do not reuse .
  • Do not fill vials to the brim.
  • Filter sample with 0.2µm or 0.45µm filters before
    injection.
  • If the sample contains a lot of solid, centrifuge
    it, and the filter the solution.

35
Fitting plumbing
  • Excess or short tube fitting will affect peak
    shape.
  • Estimate the length required and fix properly .
  • Allow extra length if the tubing is to go around
    corners as sharp bends in the tubing will distort
    the inside bore
  • Note Care should s be taken when tightening
    fittings.
  • Over-tightening can cause damage to the fitting
    threads, ferrules etc,
  • Lead to leak or break off in the housing

36
Guard Columns and or cartridges
  • Protect the analytical column from sample and
    system debris and contaminants for better column
    performance and efficiency.
  • Guard columns or cartridges are one of the most
    cost effective and efficient ways of trapping
    these unwanted system components.

Note Guards are designed to be disposable.
Should be replaced once contaminated Recommend
that the guard is replaced every 50 100
injections or based on user
37
Precaution
  • Avoid pressure shocks on the column.
  • Pressure shocks lead to channeling in the column,
    whichresults in peak splitting in corresponding
    chromatogram.
  • Always keep both ends of the column closed, after
    usage.
  • Keep the columns in the designated column
    cabinets after use.
  • Use the pH range of 2 to 8 or follow the
    manufacturersinstructions

38
Sample Preparation
  • Sample preparation is not only dissolution of a
    solid in a liquid.
  • If Samples require use filtration, extraction or
    derivative etc
  • Accurate weighing and/or dissolution or make up
    volume for assay.
  • Filtrate the sample -By syringe filter or using
    on-line pre-column filter
  • Sample solvents should match the mobile phase to
    avoid baseline errors and spurious peaks
  • The sample should be clean for long life of
    column and reliable
  • If sample contaminates the column, we may spend
    more time money
  • Important If using a membrane filter make sure
    that it is compatible with all the solvent used

39
Definition
  • Dwell volume (D) Also known as gradient delay
    volume, is the volume between the point at which
    the eluents meetand the top of the column.
  • Hold-up volume (VM) The volume of mobile phase
    required for elution of an unretained component.
  • Resolution (RS) The resolution is the separation
    of two components in a mixture,
  • Retention volume (VR) The volume of mobile phase
    required for elution of a component (Vm
    retention time (TR) X flow rate (F)
  • Separation factor (a) The relative retention
    calculated for two adjacent peaks (k2/k1)
  • Theoretical plates (N) A measure of column
    efficiency based on RT PW

40
System Suitability
  • Perform system suitability (SS) test to verify
    the chromatographic system is adequate for the
    intended analysis.
  • SS is based the performance equipment,
    electronics, analytical operations, and sample
    etc .
  • The following factors may affect chromatographic
    behavior
  • Composition, ionic strength, temperature, and pH
    of the mobile phase
  • Flow rate, column dimensions, column temperature,
    and pressure
  • Stationary phase characteristics- type of column
    (particle-based or monolithic), particle or pore
    size, porosity, and specific surface area
  • Reverse-phase and other surface modification of
    the stationary phases, the extent of chemical
    modification (as expressed by end-capping, carbon
    loading, and others)

41
System Suitability Parameters
  • Resolution between specified peaks ( gt2)
  • Theoretical plate (Ngt2000)
  • Precision / injection repeatability RSD of
    replicate injection unless otherwise specified
  • RSD for five replicate injections of the
    analyte- 2.0
  • RSD for five replicate injections of the
    analyte- gt 2.0
  • (Can vary based on the purpose assay , RC, etc
    from 0.7 to 10)
  • Tailing factor (T lt2)
  • Capacity factor ( k' gt 2 )

42
Resolution
  • Resolution is defined as the distance between two
    adjacent peak apexes, divided by the average base
    width of both peaks.
  • Resolution is dependant on three other variables,
    the column efficiency (N), capacity factor (k)
    selectivity (a).
  • Increasing N increases resolution because peak
    width decreases. Decreasing N decreases the
    resolution due to broad peak width
  • Decreasing k sharpens the peaks but decreases
    resolution. Increasing k broadens the peaks but
    improves resolution.
  • Increasing a increases resolution. One peak moves
    relative to the other. Likewise, decreasing a
    decreases resolution.

43
Capacity factor
  • Capacity factor (N) is affected by changes in
    mobile phase, operating temperature, analyte
    retention characteristics and changes to the
    surface chemistry of the column.
  • Changes in N that occur both with standard and
    sample mixes are likely to be due to changes in
    the column, temperature or mobile phase
    composition.
  • Changes in N that occur only in the sample mix
    and not the standard mix are most likely to be
    due to the composition of the sample.
  • Note Capacity factor will change by up to 10
    for a 5C rise in column temperature.
  • Selectivity is a measure of the relative
    retention of two adjacent peaks in a chromatogram

44
Band spreading
  • Broad peaks with change in retention time,
    indicate band spreading.
  • HPLC column or incorrect system plumbing will
    affects band spreading
  • Measure the band spreading due to the HPLC
    system.Remove the HPLC column from the system
    and replace with a zero dead volume union. Inject
    column efficiency solution and calculate
  • Band Spread (µL) Peak Width x (1/20) x 1 x 1000
    Confirm column effects or plumbing .

45
Solvent property
  • Understand the property of solvent, which may
    affect the HPLC analysis
  • Properties like viscosity, Polarity, RI,
    UV-cutoff, miscibility number etc
  • Know the miscibility numbers, to predict the
    miscibility of solvents.
  • If the smaller miscibility number is subtracted
    from the larger and the difference is 15 units or
    less, then the two liquids are soluble in all
    proportions at 15C.
  • Difference in Smaller miscibility number is lt 15
    mixer of solvents lead to poor chromatographic
    analysis and less accurate.

46
Solvent properties
47
Impact of PH
  • Reversed phase HPLC The retention of analytes is
    related to their hydrophobicity. The more
    hydrophobic the analyte are retained longer .
    When an analyte is ionized, it becomes less
    hydrophobic and retention decreases.
  • When separating mixtures containing acids and/or
    bases it is necessary to control the pH of the
    mobile phase of buffer in order to achieve
    reproducible results.
  • Eg Resolution changes to 1.4 from 3.0 due to less
    of 0.1pH

48
Adjustment allowed for HPLC condition
Property USP General Chapter 621 Ph.Eur.  Gen. Chapter 2.2.46
Column length 70 70
Particle size Reduction by 50 Reduction by 50
Particle size No increase No increase
Internal diameter Can be adapted as long as the linear flow velocity remains the same 25
Flow rate 50 or more, provided the linear flow velocity remains the same 50
Column temp. 10 C 10 C, maximum 60 C
pH -mobile phase 0.2 units 0.2 units (1 for neutral subs)
49
Adjustment allowed for HPLC condition
Property USP General Chapter 621 Ph.Eur.  Gen. Chapter 2.2.46
Injection volume Reduction allowed as far as precision and detection limit acceptable. No increase. Reduction allowed as far as precision and detection limit acceptable. No increase.
Salt conc. of the buffer 10, as far as the allowed change in pH value 10
Composition of mobile phase Minor components 30, if not more than 10 absolute Minor components 30,if not more than 2 absolute (greater value accepted)
Wavelength Not permitted , can be Max 3nm based on the validation
For gradient separation, a change of the mobile phase is not recommended For gradient separation, a change of the mobile phase is not recommended For gradient separation, a change of the mobile phase is not recommended
50
Starting HPLC analysis
  • Test method shall be created and verified
  • Sequence can be as below or as per test method
  • Blank, system suitability, reference solution ,
    test, and bracketing standard . Bracketing
    standard at defined number of sample injections
  • Data file shall be continuous (do not repeat the
    file no/name)
  • Load the sample and run the system suitability
    and once SS passes run the test
  • After completion, process the chromatograms
  • Print the method, sequence and data
  • Perform audit trail before batch release
  • Store the data and keep back up

51
Integration
  • Do not integrate any peak by manually.
  • Integrate all the sample sets by batch wise.
  • Always use same processing method for processing
    ofblank, standard sample chromatograms in case
    of Assay related substances, etc.
  • Verify the processing parameters like
  • Threshold,
  • Width,
  • System suitability,
  • Peak names etc.
  • Save the processing method

52
Integration
  • Integrate all the injections including entire
    set of failures.
  • For System suitability injection for which
    Tailing factor,Resolution or Theoretical plates
    are to be checked,consider only the first
    injection of the sequence.
  • Note Do not consider this criteria for average
    of all the standard injections unless otherwise
    specified.
  • Re-integration
  • Do not re-integrate the chromatograms
    withoutdocumenting.
  • Document reason for reintegration.

53
Common problem in HPLC analysis
  • System failures may occur during analysis due to
  • System over pressure
  • Communication error
  • Failure of system suitability
  • Peak splitting/ negative peak
  • Bracketing standard failure

54
Common problem in HPLC analysis
  • How to handle the failure ?
  • SOP shall be available and shall address the
    handling of Lab deviation/ incidents.
  • SOP shall define clearly the deviation/ incident
    , reporting investigation, CAPA and
    documentation
  • Record all the deviation/ incident happened in
    chromatographic analysis
  • Process all the injections including the invalid
    injection and report and store the data along
    with Raw data.

55
Common problem in HPLC analysis
  • How to handle the failure ?
  • Do not omit any injection
  • Investigate the deviation/ incident
  • Find the root cause for the sample set failure.
  • Rectify the problem, take appropriate CAPA and
    document
  • Repeat complete sample set of injections in case
    of sample injection failure

56
How to handle the problem if any
  • In case of interruption, due to power failure,
    computer interruption, time gap due to sample
    preparation or due to injection for 4 hours and
    if system is in continuous state of equilibrium
  • Inject bracketing standard and proceed otherwise
    restart
  • Deviation in RT for sample or std, is gt 15 of
    specified RT
  • Inhibit the peak upto void volume
  • Use the same integration parameter for entire set
  • Reprocess shall be at the same time for entire
    sequence

57
Documentation
  • Ensure are followed contemporaneously
  • Instrument Use Log
  • Routine Maintenance Log
  • Problem Log
  • Column History Log

58
PQ of HPLC
Ensure the below as parameter during PQ as min,
but not limited to
Component Parameter Acceptance Criteria
Pump Pump Flow Accuracy 0.5ml (0.475 to 0.525) 5.0ml (4.75 to 5.25)
Pump Pump flow precision RT RSDNMT0.50
Pump Gradient composition in 20, 40, 60, 80 2.0
Column oven Column Temperature Accuracy Column Oven lt 2.0
Column oven Column Temperature Stability Column Oven lt 1.0
Sample oven Sample temperature Accuracy Set temperature 4C gt -2.00 and lt 5.00
UV Detector Wavelength Accuracy (201nm to 209nm) 205 nm lt 2
UV Detector Wavelength Accuracy (241nm to 249nm) 245 nm lt 2
UV Detector Wavelength Accuracy (269nm to 277nm) 273 nm lt 2
UV Detector Noise and Drift Noise lt 0.040 mAU Drift lt 0.500 mAU/Hr
UV Detector Signal to Noise gt 3000
UV Detector Response Linearity (Resp.Factors) Correlation coefficient gt 0.99900
Lamp Intensity 1000
Sampler / Injectors Injector Precision Volume Delivery - Linearity RSD for Area lt 1.0 RSD for Height lt 2.00
Sampler / Injectors Injector Precision Volume Delivery - Linearity Correlation coefficient NLT0.99
Sampler / Injectors Injection Carryover Carryover for Area lt 0.20
Sampler / Injectors Injection Carryover Carryover for Area lt 0.40
59
Calibration of HPLC
Component Calibration test Acceptance Criteria
Pump Leak Test  No leak
Pump Flow rate Accuracy 0.5ml (0.49 to 0.51) 1.0ml (0.98 to 1.02) 2.0ml (1.96 to 2.04)
Pump Flow stability RT RSDNMT1.0
Pump Gradient Delivery Accuracy in 20, 40, 60, 80 2.0
Column oven Sampler Temperature Accuracy by Calibration of Thermocouple and Air temperature  Column Oven 25C/40C/60C 2.0 
UV Detector Wavelength Accuracy(266nm to 276nm) 271 to 273nm
UV Detector Dynamic Short-term Noise (Single to Noise Ratio) Noise0.04 mAU or less Drift5.0mAU/hr or less
UV Detector Response Linearity Correlation coefficient NLT0.99
UV Detector Lamp energy Low intensity (gt 200) Average intensity (gt 5000) Highest intensity (gt 10000)
Sampler / Injectors Volume Precision RSD NMT 1.0
Sampler / Injectors Volume Delivery - Linearity Correlation coefficientNLT0.99
Sampler / Injectors Injector Carryover NMT 0.1
60
Reference
  • WHO 791 Guideline for GOOD CHROMOTOGRAPHY
    PRACTICES (February 2019)
  • Troubleshooting LC Systems, John W Dolan Lloyd
    R Snyder, Humana Press
  • Qualification Of Equipment Annex 1 Qualification
    Of Liquid Chromatography Equipment

61
  • Thank you
Write a Comment
User Comments (0)
About PowerShow.com