Title: First, very preliminary Results from the Solid-State Electron EDM Experiment
1First, very preliminary Results from the
Solid-State Electron EDM Experiment
- nEDM collaboration meeting
- June 21, 2005
2(No Transcript)
3Sensitivity estimate
4Enhancement Factor of eEDM
- Buhmann, Dzuba, Sushkov, Phys. Rev. A 66, 042109
(2002). - Dzuba, Shushkov, Johnson, Safronava, Phys. Rev.
A 66, 032105 (2002). - Kuenzi, Sushkov, Dzuba, Cadogan, Phys. Rev. A
66, 032111 (2002). - Mukhamedjanov, Dzuba, Sushkov, Phys. Rev. A 68,
042103 (2003).
5HV Electrodes Magnetic Flux pick-up coils
(planar gradiometer)
0
_
2.5
5
- Common mode rejection of external
- uniform B field and fluctuations.
- CMRR238, (0.4 area mismath.)
- Enhancement of sample flux pick-up.
6Waveforms
HV monitor
Current In the ground plate
SQUID signal
7Data Processing
1
-3
3
-1
Get rid of the long time drifting function up to
the second order terms.
8Flux sensitivity estimate for the prototype
experiment
9EDM signal
4K, 2.8kVpp, 1.13Hz, 50 minutes
SQUID signal (-2.68-5.5)e-7 V
-6.6E-08 - 2.8085E-07 V (drift corrected)
Leakage Current (-4.6e-5 - 1e-6 ) x10 uA
10SQUID signal
Standard deviation
SQUID transfer function
B.V. Vasilev and E.V. Kolycheva, Sov. Phys.
JETP, 47 2 243 (1978) de(8.1 ?11.6)?10-23 e-cm
4K, 2.8kVpp, 1.13Hz, 50 minutes
11eEDM sensitivity that can be achieved with the
prototype
- With the same configuration, 4 days of data
averaging will give an eEDM limit of 1e-24 e-cm. - Increase HV from 2.8kVpp to 7kVpp will push the
eEDM limit to 4e-25 e-cm in 4 days data at 4K. - Using a 2K bath, eEDM limit 2e-25 e-cm.
- At 400mK, eEDM limit 4e-26 e-cm.
- At 40mK, eEDM limit4e-27 e-cm.
12Finite Leakage Current ???
13- Liquid Helium level runs low during the run.
14Zero-crossing Waveform
15EDM measurements vs time
16Standard deviation
17Spin Noise
18EDM signal
19Leakage Current